Download Fluke Corporation - Electrocomponents
Transcript
1735 Users Manual Note If you replace the battery please use original spare parts only – see section “Standard and Optional Accessories”. Maintenance If the Logger is used appropriately it does not require special maintenance or repair. If the Logger gets dirty wipe it off carefully with a damp cloth (without cleaning agents). Maintenance work may be executed only by trained and qualified personnel. This work may only be done at a company related service center within the guarantee period. See www.fluke.com for locations and contact information of Fluke Service Centers worldwide. Calibration As an additional service we offer the regular examination and calibration of your Logger. Storage If the Logger is stored for longer time or is not in use for longer time, you should charge the battery at least once every six months. Measurement Theory The following formulas are the basics of the measuring values: Voltage and current measurement VRMS = 1 2 v dt T∫ RMS value of voltages I RMS = 1 2 i dt T∫ RMS value of currents I N = I1 + I 2 + I 3 RMS value of neutral-conductor current Neutral is calculated when not measured, i.e., no 4phase Flexi set is connected. 48