5.2 PERFORMANCE VERIFICATION TEST (1.5 SERIES)
DOOR LATCH (OPEN POSITION)
DOOR RELEASE TAB
BUBBLE DETECTORS
(1.6 SERIES ONLY)
PINS (7)
96F07001
Figure 5-1.
Mechanical Elements Behind Cassette Door
5.1.3
SANITIZING THE INFUSION SYSTEM
Sanitize the external surfaces of the infusion system using a cleaning solution listed in
Table 5-1, Cleaning Solutions.
Note: Not all cleaning solutions are sanitizers. Check product labeling.
CAUTION: Do not sterilize the infusion system using heat, steam, ethylene oxide
(ETO), or radiation. These methods may cause the instrument to malfunction.
5.2
PERFORMANCE VERIFICATION TEST
(1.5 SERIES)
The performance verification test (PVT) consists of the tests described in the following
sections. The PVT can be used for diagnostic purposes during the troubleshooting of a
malfunctioning infusion system. The PVT should be used for performance verification
before an infusion system is placed back in service after repair. If any malfunction is
detected as a result of the PVT, refer to Table 6-3, Troubleshooting with the PVT (1.5 Series).
Note: The PVT must be performed exactly as described in this manual to assure effective
and reliable product evaluation information.
This section consists of the PVT for 1.5 series infusion systems. For performance testing
of 1.6 series infusion systems, use the PVT in Section 5.3, Performance Verification Test
(1.6 Series).
LifeCare 5000 Series
5-3
430-03714-004 (Rev. 8/96)