white interfaces. Criterion for pass: bright point number is 0; number of
continuous dark point pair is ≤3, and no continuous dark point appears in the
image area; there are no three or more than three continuous dark point;
dark point defects number is ≤7, and number of dark points in the image area
is ≤2; point flash defects is 0; space between fail points is ≥5mm.
4. Click [Exit] or press [ESC] to return to the main screen interface.
7.1.4
NOTE:
Monitor Parameter Setting
After changing the main monitor, parameter loading should be performed to match the
monitor and the system.
The parameters of the main monitor include color temp, brightness/contrast, Gamma, and
Default, information of loading Default parameters are introduced here:
1. Press <1> and <4> key on the monitor as shown in the figure for more than 3 seconds to enter
the screen.
2. Press <1> or <2> to select Default, the cursor is on the Default position, press <4> to enter the
parameter automated loading, and it takes about 3 seconds to finish.
3. Color temperature is default to be 11000K, and adjustment is not recommended.
4. After all settings are finished, move the cursor to Exit or press <3> to exit the menu.