Fluke 125 Users Manual Multimeter disturbance: • • VDC, VAC, and VAC+DC with STL120 and short ground lead. OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM. Table 3 Disturbance less than 1% of full scale E= 3 V/m E= 10 V/m Frequency range 10 kHz to 27 MHz VDC, VAC, VAC+DC OHM, CONT, DIODE CAP 500 mV to 1250V 50 Ω to 30 MΩ 50 nF to 500 μF 500 mV to 1250V 50 Ω to 30 MΩ 50 nF to 500 μF Frequency range 27 MHz to 1 GHz VDC, VAC, VAC+DC OHM, CONT, DIODE CAP 500 mV to 1250V 50 Ω to 30 MΩ 50 nF to 500 μF 500 mV to 1250V 50 Ω to 30 MΩ 50 nF to 500 μF Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale. 10-14