Calculation of the critical test value:
The algorithm uses the phase angle and the amplitude of the first eight harmonics of
the response to calculate the test value. The statistical test is carried out in the following
way: The first test value will be calculated with a sample of the first 15 epochs. If the
test value is less than the critical test value the sample will be increased by one epoch
and tested again. This procedure will be continued until the test value is equal to or
greater than the critical test value. Then the testing is stopped with a “Pass” result. If
the critical test value is not reached after 180 epochs, the test will stop with a "refer"
result. This successive procedure is more efficient than using a fixed large sample size
and a single statistical testing. However, using such repeated testing one has to consider that the error probability of the test increases with each test step leading to a decrease of sensitivity. A so-called Bonfferoni correction of the critical test value is known
to compensate for this effect. But the Bonfferoni correction leads to an overcorrection
in the case of dependent samples. Therefore, a new method for the calculation of the
critical test value for repeated testing of dependent samples had been developed. The
new procedure for the calculation of the critical test values for repeated testing of dependent samples is described in detail in Stürzebecher et al. 2005 "Automated auditory
response detection: statistical problems with repeated testing" , Int J Audiol, 42(2):
110-117, 2005
With the given exact critical test value a sensitivity of nearly 100% (99.3%) is maintained. This is proved by testing noise data (no response) generated by extensive Monte
Carlo Simulations.
A detailed description of the used ASSR procedure is published by
Stürzebecher E, Cebulla M and Neumann K. "Click-evoked ABR at high stimulus repetition rates for neonatal hearing screening" Int J Audiol, 42(2):59-70, 2003