Download - Frank`s Hospital Workshop

Transcript
Aisys
10.27.2 Electronic Vaporizer - Valve Block
9
1
5 (6) 7
8
4
a
3
b
**
2
c
Item
Description
Stock Number
Qty
1*
Valve Block Subassembly
1011-7000-000
2
Valve Connector Assembly
1011-7003-000
a - Spring, Wave
1011-3126-000
b - Seal, Spring energized
1011-3124-000
c - O-ring ID7.00 BCG OD11.00 Viton Duro 75
1011-3125-000
3
Insertion tool, spring energized seal
Refer to section 10.1
4
Valve, Solenoid 2-way BCG
1011-3115-000
5
Cap, Liquid Flow Prevention
1011-3129-000
6
Ball, 4.00 DIA BCG Polypropylene
1011-3127-000
7
O-ring, ID12.37 BCG OD17.61 Viton Duro 75
1011-3128-000
8*
O-ring, ID3.68 OD7.24 BCG Viton Duro 75
1011-3139-000
9
Screw, M4x45 Hex Cap Head
N122024
(2)
(3)
(4)
* A replacement valve block assembly (Item 1) includes a strip of tape over the o-ring (Item 8) to
hold it in place. Be sure to remove the tape before replacing the valve block assembly.
** Apply thin coat of silicone sealant to threads of mounting post before securing solenoid with
thumb nut. Install washer with dome facing outward (toward nut).
10-40
05/05 M1046983