Advanced Programming with Hospira MedNet TM Software
7
Select either YES> or NO> to
purge the system See Purging
the System on page 4- 28 for more
information.
MORPHINE PF
1 mg/mL
PURGE ?
PURGING
RECOMMENDED
YES
NO
8
If YES> is selected,
disconnect the set from the
patient, and press and hold
the PURGE> Softkey.
While purging is occurring,
the word PURGING is
displayed.
MORPHINE PF
1 mg/mL
DISCONNECT
SET FROM
PATIENT
PRESS AND HOLD
PURGE KEY
PURGE
PREVIOUS
NOTE: Purging is recommended
to remove system slack when a new vial is inserted. The
maximum volume delivered during a purge is 3 mL.
Upon release of the PURGE>
Softkey, the display asks if
the purge is complete.
9
Select YES> to continue, or
select NO> to purge again
until complete.
MORPHINE PF
1 mg/mL
PURGE
COMPLETE ?
YES
NO
10 After purging is complete,
reconnect set to patient.