Download Agilent 1260 Infinity Micro Degasser
Transcript
Maintenance Introduction to Maintenance and Repair 7 3 Press down the clip of the fuse holder and pull out of the power socket. 8a^e ;jhZ]daYZg Figure 13 Fuse Holder and Clip 4 Remove the fuses from the fuse holders. 5 Ensure the fuse wires inside the fuses are not broken. If a test meter is available, check the resistance of each fuse. A good fuse shows a low resistance (approximately 0 Ohm). 6 If a fuse is defective (wire broken or high resistance), insert a new fuse. 7 Reinsert the fuse holders and the power cable. 8 Switch on the power switch. 1260 Infinity Micro Degasser User Manual 55