Download Mars IVF
Transcript
Mars IVF User Manual 7.1 Transport The cabinet can be moved through a standard 800-mm wide door. 7 2 Preparation The installations site for the unit must be draught-free and should be selected so that frequent passing of people in front of the work opening is avoided. 1. The tabletops of stainless steel are mounted over the trough. 2. Adjust the levelling screws to assure that the tabletop is in horizontal position and levelled. 3. Valves for gases or vacuum are installed in the side windows. A qualified technician must make the connections for the supply. 4. For connection of the exhaust air to the exhaust air systems, special precautions which must be discussed with the cabinet supplier must be taken. When the cabinet has been installed 1. Check that the front window is in the correct position, i.e. parallel to the front shield. 2. Never lift the front window manually; , this will cause window to mall function. 3. Check that the window slides from top to bottom without any irregular sound. 4. If the window does not move when the UP or DOWN button is pressed, authorized service personnel must be called to correct the fault. 5. Always keep hands and arms away from the working chamber when activating the front window. Code No. 9.900.900.752 \\Server1\data\QA\Manualer\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanLaf\Origio\Mars IVF manual-Origiorev.5.doc Page 13 af 32