CIC Pro™ Clinical Information Center
Bedrock Hardware Platform
Service Manual
2001099-156
g
Revision C
GE Medical Systems
Information Technologies
gemedicalsystem.com
NOTE: Due to continuing product innovation, specifications in this manual are subject to change without
notice.
NOTE: This manual supports CIC Pro information centers with the product code JA1. The CIC Pro
information centers with the JA1 product code use the Bedrock hardware platform and are only compatible
with software versions 4.1 and later.
Listed below are GE Medical Systems Information Technologies trademarks used in this manual. All other
trademarks contained herein are the property of their respective owners.
UNITY NETWORK, ApexPro, CD TELEMETRY, and RSVP are trademarks of GE Medical Systems
Information Technologies registered in the United States Patent and Trademark Office.
CD TELEMETRY®–LAN, CENTRALSCOPE, CIC Pro, ICMMS, Prism, Octacomm, and Octanet are
trademarks of GE Medical Systems Information Technologies.
© 2004 General Electric Company. All rights reserved.
T-2
CIC Pro Clinical Information Center
2001099-156
Revision C
04 April 2005
Contents
1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
License Agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Manual Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Hazard Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
CIC Pro Specific Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Equipment Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
2
Equipment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
CIC Pro Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Basic Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Back Panel/Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Back Panel/Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
CIC Pro System Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Unity Network Description/Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Unity Network MC Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Unity Network IX Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Network Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
CIC Pro Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
How Trends are Calculated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
File/Data Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Full Disclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
CIC Pro Service Logons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
System Logons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Service Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Webmin Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Integrated Internet Explorer Browser Interface . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Revision C
CIC Pro Clinical Information Center
2001099-156
i
CIC Service Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Miscellaneous Service Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
CIC Pro Command-line Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
Run CIC Cmd (Webmin) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
Feedback Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
Error/Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
Audible Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
System Environment Monitor Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Pre-Installation Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Un-Interruptible Power Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Pre-Installation Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Physical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Connect Keyboard and Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Install the Speakers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Install the Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Connect/Configure the Video Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
CIC Pro System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Enhanced Write Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
IP Address Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Connect to Unity Networks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Computer Name Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Printer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
CIC Pro Application Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Access Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Calibrate the Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
Set the CIC Pro Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Set the Telemetry Unit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-43
Set the Full Disclosure Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-47
Set the Display Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-50
Set the Current Telemetry Listings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-52
Lock or Unlock Beds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-54
Change CIC Pro Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-54
Install Optional Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-56
Commit Changes to the Write Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-57
Time and Date Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-58
Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-60
Backup CIC Pro Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-60
Exit Service Mode and Proceed to Checkout Procedures . . . . . . . . . . . . . . . . . . 3-60
ii
CIC Pro Clinical Information Center
2001099-156
Revision C
4
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Manufacturer Recommendation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
PM Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Safe Shutdown/Restart Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Controlling Electrostatic Discharge Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Before Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
External Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Display Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Wall Receptacle Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Ground (Earth) Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Ground (Earth) Wire Leakage Current Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Enclosure (Chassis) Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Test Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Backup-Restore CIC Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Planning Backup/Restore Strategy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Data Module Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Checkout Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
About the Checkout Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Disk Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Audio Verification Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Verify Access to All Other Care Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Verify Full Disclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Verify Each CIC Pro Can Print a Full Disclosure Report . . . . . . . . . . . . . . . . . . . 4-31
CIC Service Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Maintenance Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
“Visual Inspection” on page 4-6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
“Cleaning” on page 4-7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
“Electrical Safety Tests” on page 4-10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
“Backup-Restore CIC Configuration” on page 4-16 . . . . . . . . . . . . . . . . . . . . . . 4-32
“Checkout Procedures” on page 4-28 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Repair Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Revision C
CIC Pro Clinical Information Center
2001099-156
iii
5
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Blank Display, Unit Doesn’t Boot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Hardware Component Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Webmin Hardware Monitoring Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Software Component Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Webmin Software Monitoring Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Incorrect CIC Pro Application Colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Browser Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
CIC Service Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
List Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Duplicate TTX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Open RTERM Session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Event Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
CIC Pro Log File Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Log File Command Prompt Procedures/Functions . . . . . . . . . . . . . . . . . . . . . . . 5-18
Canceling CIC Pro Print Jobs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
6
Parts Lists, Drawings and Replacement . . . . . . . . . . . . . 6-1
Ordering Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Field Replaceable Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
FRU Installation/Field Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Field Replacement Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Hard Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Internal Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
CIC Pro Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
Optional Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
Keyboard Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
A
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Technical Specifications–CIC Pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
iv
CIC Pro Clinical Information Center
2001099-156
Revision C
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Recorder (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Display Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
Monitor (Display) Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
B
Supported CIC Pro Parameters . . . . . . . . . . . . . . . . . . . . .B-1
Arterial Blood Gas (POC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-3
Arterial Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-3
BIS Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-4
Cardiac Calculations (Cardiac Calcs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-4
Cardiac Output (CO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-4
Central Venous Pressure (CVP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
Cerebral Perfusion Pressure (CPP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
Continuous Cardiac Output (CCO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
CO2 Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
ECG Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6
Episodic Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-7
Femoral Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-7
Inter-Cranial Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-7
Left-Arterial Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-7
Mass Spec . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-7
Non-Invasive Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-8
Non-Invasive Cardiac Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-8
Other Cardiac Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-9
Pulmonary Artery Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-9
Pulmonary Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-9
Right Arterial Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-10
Respiration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-10
Respiration Mechanics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-11
SPO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-11
Special Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-11
ST Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-12
SvO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-12
Transcutaneous CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-12
TramModule Transcutaneous CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-13
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-13
Umbilical Pressure (UAC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-13
Umbilical (UV, UVC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-13
Ventilator Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-14
Revision C
CIC Pro Clinical Information Center
2001099-156
v
C
Webmin Runtime Diagnostics Utilities . . . . . . . . . . . . . . .C-1
Runtime Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Closed-loop Audio Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-4
Ping Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-5
Hard Drive Integrity Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-6
Instruction Store Integrity Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-7
Temp/Voltage Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-8
Network Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-9
OS Hotfix Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-10
Process Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-11
OS Service Process Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-12
Device Driver Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-13
Timezone Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-14
OS Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-15
OS System Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-16
OS Application Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-17
OS Security Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-18
System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-19
D
CIC Pro Service Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-1
Service Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3
Audio Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-4
Drive Integrity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-6
User Asset Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-8
Environment Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-9
Network Integrity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-11
Printing Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-13
Video Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-15
Watchdog Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-18
E
Electromagnetic Compatibility . . . . . . . . . . . . . . . . . . . . . 1-1
Electromagnetic Compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions . . . . . . . 1-3
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity . . . . . . . . 1-4
Recommended Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Compliant Cables and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
vi
CIC Pro Clinical Information Center
2001099-156
Revision C
1
Revision C
Introduction
CIC Pro Clinical Information Center
2001099-156
1-1
For your notes
1-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Introduction: License Agreement
License Agreement
IT IS IMPORTANT THAT YOU CAREFULLY READ THE TERMS AND
CONDITIONS OF THIS LICENSE AGREEMENT BEFORE COMMENCING
THE USE OF THE CLINICAL INFORMATION CENTER WORKSTATION
(THE “WORKSTATION”) AND THE CLINICAL INFORMATION CENTER
PROGRAM RECORDED THEREIN AND ANY ACCOMPANYING USER
DOCUMENTATION (“PROGRAM”). THIS LICENSE REPRESENTS THE
ENTIRE LICENSE AGREEMENT CONCERNING THE PROGRAM BETWEEN
YOU AND GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
(“INFORMATION TECHNOLOGIES”) AND SUPERSEDES ALL OTHER
COMMUNICATIONS OR ADVERTISING RELATED TO THE PROGRAM
EXCEPT ANY TERMS AND CONDITIONS OF SALE OR WARRANTIES OR
WARRANTY LIMITATIONS RELATIVE TO THE PROGRAM AND/OR THE
WORKSTATION AS MAY BE EMBODIED IN ANY DOCUMENTATION
SUPPLIED WITH THE WORKSTATION. BY COMMENCING THE USE OF
THE WORKSTATION AND THE PROGRAM CONTAINED THEREIN, YOU
ARE ACCEPTING AND AGREEING TO BE BOUND BY ALL THE TERMS
AND CONDITIONS OF THIS LICENSE AGREEMENT. IF YOU ARE NOT
WILLING TO BE BOUND BY THE TERMS AND CONDITIONS OF THIS
LICENSE AGREEMENT, YOU SHOULD PROMPTLY RETURN THE
WORKSTATION TO GE Medical Systems Information Technologies AND YOU
WILL RECEIVE A REFUND OF THE PURCHASE PRICE.
I.
GRANT
The Program is capable of coupling one to sixteen patient monitoring units to
the Workstation. GE Medical Systems Information Technologies hereby grants
you a non-exclusive, non-transferable right and license to use the Program for
coupling the number of patient monitoring units to the Workstation for which a
per-unit royalty has been paid pursuant to Article II hereof.
II. ROYALTY
You have paid GE Medical Systems Information Technologies a one-time, perunit royalty equal to GE Medical Systems Information Technologies’ current
published price for the use of the Program. The per-unit royalty is based on the
actual number of patient monitors intended to be coupled by the Program to the
Workstation as stated in the Purchase Order for the Workstation and the
Program. If you use the Program to couple any patient monitoring units to the
Workstation in addition to the number for which a per unit royalty was
previously paid, you agree to pay GE Medical Systems Information
Technologies an additional per-unit royalty equal to GE Medical Systems
Information Technologies' then current published royalty for the Program for
each such additional patient monitoring unit so coupled. The additional per-unit
royalty shall be paid to GE Medical Systems Information Technologies within
30 days of the use of the Program to couple any such additional patient
monitoring units to the Workstation.
Revision C
CIC Pro Clinical Information Center
2001099-156
1-3
Introduction: License Agreement
III. LIMITATIONS
You hereby agree not to: (1) use the Program in any network or system other
than to couple patient monitoring units to the Workstation; (2) make any copy
of the Program for any reason, or allow or assist others to do so; (3) modify,
reverse engineer, de compile or disassemble the Program or merge any part of
the Program into any other program; (4) rent, sell, sublease, assign, transfer or
otherwise share the Program or any of your rights in the Program under this
Agreement with any third party; or, (5) remove or alter any copyright notice,
labels or trademarks from the Program or the Workstation.
IV. TITLE
This License is not a sale. Title and all copyrights to the Program and any copy
made by you remains the sole property of GE Medical Systems Information
Technologies.
V. TERM
This Agreement shall continue in force until terminated. This Agreement shall
terminate automatically when you cease using the Workstation and the Program
for their intended purpose. GE Medical Systems Information Technologies may
terminate this Agreement on 30 days written notice if you make any
unauthorized copies of the Program or fail to comply with any of the restrictions
on use of the Program as set forth herein.
VI. LIMITED WARRANTY, DISCLAIMER AND LIMITATION
OF LIABILITY
A. Licensor warrants that on the acceptance date the Program shall be free from
significant programming errors and shall operate and conform to the
published functional specifications applicable thereto, and that the Program
shall conform to the standards generally observed in the industry for similar
software.
B. This warranty shall be invalidated by your modification of the Program if
such modification or the interaction between such modification and the
Program as supplied by Information Technologies is the cause of the defect,
error or non-conformity.
C. Except as stated above, the warranty covering the Program and the
Workstation shall be either Information Technologies Standard Warranty or
Limited Extended Parts Warranty as published by Information Technologies
and hereby made a part hereof.
D. EXCEPT FOR THE EXPRESS WARRANTIES STATED HEREIN, GE
Medical Systems Information Technologies DISCLAIMS ALL
WARRANTIES WITH REGARD TO THE PROGRAM INCLUDING
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR
A PARTICULAR PURPOSE.
1-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Introduction: License Agreement
E. GE Medical Systems Information Technologies entire liability to you arising
out of or in connection with this Agreement shall not exceed the per-unit
royalty paid to GE Medical Systems Information Technologies for use of the
Program. You acknowledge that the amount paid to Information
Technologies for use of the Program is insufficient for Information
Technologies to undertake any greater risk. IN NO EVENT SHALL
INFORMATION TECHNOLOGIES BE LIABLE FOR ANY INDIRECT,
INCIDENTAL, CONSEQUENTIAL, SPECIAL OR EXEMPLARY
DAMAGES (INCLUDING WITHOUT LIMITATION, LOST PROFITS,
BUSINESS INTERRUPTION, LOSS OF BUSINESS INFORMATION,
PERSONAL INJURY OR ANY OTHER PECUNIARY LOSS) ARISING
FROM THE USE OF THE PROGRAM, EVEN IF INFORMATION
TECHNOLOGIES HAS BEEN ADVISED OF THE POSSIBILITY OF
SUCH DAMAGES.
VII. GOVERNING LAW
This Agreement shall be governed by the laws of the State of Wisconsin.
VIII.PARTIAL INVALIDITY
If any provision of this Agreement is held invalid or unenforceable, the
remaining portions of the Agreement shall continue in full force and effect.
Revision C
CIC Pro Clinical Information Center
2001099-156
1-5
Introduction: Manual Information
Manual Information
Revision History
The document part number and revision letter are listed at the bottom of each page
in this manual. The revision letter identifies the document’s update level. The
revision history of this document is summarized below.
Revision History
Revision
Comment
A
Initial Release
B
Various edits supporting additional features/feedback/clarifications
obtained after previous release.
C
Enhancements to the Write Filter procedure.
Conventions
Bold text
Indicates keys on the keyboard, text to be entered, or hardware items
such as buttons or switches on the equipment.
Italicized text
Indicates software terms that identify menu items, buttons, or options
in various windows.
Ctrl+Esc
Indicates a keyboard operation. A plus (+) sign between the names of
two keys indicates that you must press and hold the first key while
pressing the second key once.
For example, “Press Ctrl+Esc” means to press and hold down the
Ctrl key while pressing the Esc key.
<Space>
Indicates you must press the spacebar. When instructions are given
for typing a precise text string with one or more spaces, the point
where the spacebar must be pressed is indicated as <Space>.
Enter
Indicates you must press the Enter or Return key on the keyboard.
Do not type “enter”.
Manual Purpose
This manual contains technical information supporting equipment maintenance and
repair, considered field repeatable down to the assembly level.
Where necessary, this manual identifies additional sources of relevant information
and technical assistance.
See the operator’s manual for instructions necessary for safe equipment operation, in
accordance with proper function and intended use.
1-6
CIC Pro Clinical Information Center
2001099-156
Revision C
Introduction: Manual Information
Intended Audience
This manual is intended for use by service representatives and technical personnel
who maintain, troubleshoot, or repair the equipment.
Revision C
CIC Pro Clinical Information Center
2001099-156
1-7
Introduction: Safety Information
Safety Information
General
Responsibility of the Manufacturer
GE Medical Systems Information Technologies is responsible for the effects of
safety, reliability, and performance only if:
„
assembly operations, extensions, readjustments, modifications, or repairs are
carried out by persons authorized by GE Medical Systems Information
Technologies;
„
the electrical installation of the relevant room complies with the requirements of
the appropriate regulations; and
„
the device is used in accordance with the instructions for use.
„
These devices are intended for use under the direct supervision of a licensed
health care practitioner.
„
These devices are not intended for home use.
„
Federal law restricts these devices to be sold by or on the order of a physician.
„
Contact GE Medical Systems Information Technologies for information before
connecting any devices to the equipment that are not recommended in this
manual.
„
Parts and accessories used must meet the requirements of the applicable UL/
IEC/UL 60601 series safety standards.
„
Parts and accessories used must meet all local building and safety requirements.
„
Periodically, and whenever the integrity of the device is in doubt, test all
functions.
„
The use of ACCESSORY equipment not complying with the equivalent safety
requirements of the device may lead to a reduced level of safety of the resulting
system. Consideration relating to the choice shall include:
Intended Use
„
‹
use of the accessory in the PATIENT VICINITY; and
‹
evidence that the safety certification of the ACCESSORY has been
performed in accordance to the appropriate UL/IEC/UL 60601 national
standards.
If the installation of the equipment, in the USA, will use 240V rather than 120V,
the source must be a center-tapped, 240V, single-phase circuit.
Hazard Definitions
NOTE
A NOTE provides application tips or other useful information.
Warnings and Cautions are used throughout this manual and designate a degree or
level of hazardous situations. A hazard is defined as a source of potential injury to a
person.
1-8
CIC Pro Clinical Information Center
2001099-156
Revision C
Introduction: Safety Information
WARNING
A WARNING indicates a potential hazard or unsafe practice
which, if not avoided, could result in death or serious injury.
CAUTION
A CAUTION indicates a potential hazard or unsafe practice
which, if not avoided, could result in minor personal injury or
product/property damage.
CIC Pro Specific Hazards
WARNINGS
LOSS OF MONITORING — If monitoring at the CIC Pro is
temporarily interrupted, alternate monitoring devices or close
observation should be used until the monitoring function at the
CIC Pro is restored.
"Loss of monitoring function" indicators:
‹
RED SCREEN (RED/GRAY screen for second monitor in
dual-display configured systems) indicates the CIC Pro
application is restarting itself and patient monitoring at the
CIC Pro is NOT occurring. The monitoring function at the
CIC Pro will automatically resume. No user action is
required.
‹
BLUE SCREEN indicates the Windows operating system has
a functional error and patient monitoring at the CIC Pro is
NOT occurring. If the CIC Pro does not automatically restart
after 60 seconds, the monitoring function at the CIC Pro will
not resume until you turn off the power to the CIC Pro and
then turn the power back on. The monitoring function should
resume in approximately 3 to 4 minutes.
Once the monitoring function at the CIC Pro has been restored,
verify the correct monitoring state and alarm function.
Revision C
CIC Pro Clinical Information Center
2001099-156
1-9
Introduction: Safety Information
WARNINGS
CIC Pro V4.1.x IS in-unit compatible with CIC Pro V4.0.x, but IS
NOT with versions of CIC Pro V3.1 or earlier. Sharing of the
same care unit name across CIC Pros having incompatible
software versions can result in lost or corrupted telemetry alarm
defaults data and loss of audible alarms.
Both hardwire and telemetry beds are limited in the number of
remote view connections that can be supported.
Do not exceed a maximum of 15 CIC Pros in a single logical
care unit.
Attempting simultaneous displays of a patient monitor
(bedside or telemetry) at too many CIC Pros may cause lost
or intermittent communication between CIC Pros and the
patient monitor. This is evidenced by NO COMM or
intermittent communication conditions for the beds.
The maximum CIC Pros viewing a patient monitor can vary
depending on patient monitor capabilities and network design.
Do not load any software other than that specified by GE Medical
Systems Information Technologies onto the CIC Pro. Installation
of software not specified by GE Medical Systems Information
Technologies may cause damage to the CIC Pro or loss or
corruption of data.
1-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Introduction: Equipment Symbols
Equipment Symbols
The following symbols appear on the CIC Pro.
ATTENTION: Consult accompanying documents before using the equipment.
Provide electrostatic discharge damage Protection.
USB PORT ICON: USB port. Used primarily for mouse and keyboard connections
UNITY NETWORK MC NETWORK CONNECTION ICON: Ethernet port used for the
Unity Network MC network connection.
UNITY NETWORK IX NETWORK CONNECTION ICON: Ethernet port used for the
Unity Network IX network connection.
SPEAKER OUT ICON: Speaker/audio port.
VIDEO 1 ICON: Primary video port.
VIDEO 2 ICON: Secondary video port.
COM 1 ICON: Primary communication port.
COM 2 ICON: Secondary communication port.
Power On: Indicates power switch ON position.
Revision C
CIC Pro Clinical Information Center
2001099-156
1-11
Introduction: Equipment Symbols
Power Off: Indicates power switch OFF position.
Chassis ground connection.
1-12
CIC Pro Clinical Information Center
2001099-156
Revision C
Introduction: Service Requirements
Service Requirements
Follow the service requirements listed below, and in the “Maintenance” chapter of
this manual, starting on page 4-1.
„
Refer equipment servicing to GE Medical Systems Information Technologies’
authorized service personnel only.
„
Any unauthorized attempt to repair equipment under warranty voids that
warranty.
„
It is the user’s responsibility to report the need for service to GE Medical
Systems Information Technologies or to one of their authorized agents.
„
Failure on the part of the responsible individual, hospital, or institution using
this equipment to implement a satisfactory maintenance schedule may cause
undue equipment failure and possible health hazards.
„
Regular maintenance, irrespective of usage, is essential to ensure that the
equipment is always functional when required.
Equipment Identification
Every GE Medical Systems Information Technologies device has a unique
serial number for identification. A sample of the information found on a serial
number label is shown below.
### ## ## #### # #
A
B
C
D
E
F
Description
A
product code1
B
year manufactured
C
fiscal week manufactured
D
production sequence number
E
manufacturing site
F
miscellaneous characteristic
1. The product code for the CIC Pro is JA1
Revision C
CIC Pro Clinical Information Center
2001099-156
1-13
Introduction: Equipment Identification
1-14
CIC Pro Clinical Information Center
2001099-156
Revision C
2
Revision C
Equipment Overview
CIC Pro Clinical Information Center
2001099-156
2-1
For your notes
2-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: CIC Pro Description
CIC Pro Description
Basic Components
The CIC Pro supports a choice from 18-inch, 20-inch, or 21-inch color displays.
Touchscreen display is also supported. See “FRU Installation/Field Replacement
Procedures” on page 6-4.
Specifications
See Appendix A, “Technical Specifications” .
Back Panel/Connections
Unity Network MC
Network Connection
Chassis Ground
Connection
Secondary
Video
Connection
Fuse Drawer
Unity Network IX
Network Connection
COM 1 Port
COM 2
Port
Ventilation Ports
Power Receptacle
Primary
Video
Connection
Power Switch
External
Speaker
Connection
USB Connections
093A
Back Panel/Label
131A
Revision C
CIC Pro Clinical Information Center
2001099-156
2-3
Equipment Overview: CIC Pro Description
Speaker
Speaker
Network
Cable
COM 1
SERIAL PORT
COM 2
SERIAL PORT
VGA MONITOR
PORT
Unity IX
Network
ETHERNET
PORT
Laser Printer
(Optional)
Network
Cable
Touchscreen Serial Cable
(Optional)
DDW Cable
PRN-50, DDW
(Optional)
CIC Pro
Computer
SPEAKER/AUDIO
PORT
Unity MC
Network
VGA MONITOR
PORT
USB
PORT
Keyboard
ETHERNET
PORT
Mouse
USB
PORT
CIC Pro System Block Diagram
Monitor Cable
Monitor Cable
(Optional)
Monitor
Monitor
(Optional)
Uninterruptable
Power Supply
(Optional)
2-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Unity Network Description/Connection
Unity Network Description/Connection
Unity Network MC Network
Unity Network MC (Mission Critical) network functions:
„
Connects CIC Pro systems to other GE patient monitoring devices
„
Connects CIC Pro systems to telemetry systems
„
The Unity Network MC passes the following data:
‹
Waveforms
‹
Parameters
‹
Alarms
‹
Other time-sensitive information
Unity Network IX Network
The Unity Network IX (Information eXchange) network functions:
„
Network laser printing
„
Connects CIC Pros together within a care unit for license sharing.
„
Full Disclosure
„
Remote Serviceability
NOTE
This network is used if a connection to outside your facility is required (a
browser internet connection, for example).
Network Connections
Connections to the Unity Network MC and IX networks are via Ethernet. For
connector locations see “Back Panel/Connections” on page 2-3.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-5
Equipment Overview: Theory of Operation
Theory of Operation
CIC Pro Functional Description
The CIC Pro application is designed to provide real-time patient data and alarms for
central nurses’ stations in hospitals. It can display real-time waveforms and vital
sign data, with visual and audible alarms, for up to 16 patients simultaneously. CIC
Pro supports both hardwired and telemetry data.
CIC Pro software allows users to select any bed on the Unity Network MC network
and display an expanded view of that bed’s real-time parameters and waveforms.
This expanded view also allows users to view and modify settings within the care
unit, and view other patient data (including alarm histories, graphic trends and
tabular trends).
The user can configure the number of patients displayed by the system, and the
number of displayed waveforms per patient. Waveform colors are configurable.
All configuration data is stored, and is restored after a system power cycle or
software restart.
The CIC Pro is connected to the Unity Network MC network via Ethernet.
How Trends are Calculated
CIC Pros trend two types of physiological data: periodic and episodic.
Periodic Data
Periodic data is constantly updated. Data is sampled every two seconds to yield 30
samples per minute. The displayed value is the median of the 30 samples. It is
always the median of a one-minute time frame, regardless of the interval selected.
The interval is the time BETWEEN values, not the time of the value itself. For
example, a five minute interval means one minute samples spaced five minutes
apart, NOT five minute samples, and NOT a median of the five one minute samples.
Odd number values are rounded down to the nearest even number.
If the calibration of the system clock changes (for example, daylight savings time),
the “time” for periodic data “slides” into the revised time. However, episodic data is
time-stamped and retains its original time.
Examples of periodic data include heart rate (HR) and blood pressure (BP).
Episodic Data
Episodic data are events that are user, or system, generated. Examples of episodic
data include temperature (Temp) and non-invasive blood pressure (NiBP).
2-6
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Theory of Operation
File/Data Handling
Log Files
Log files generated by the CIC Pro application, other associated applications and the
CIC Pro operating system are used during system analysis, problem diagnosis and
troubleshooting. See “Event Logs” on page 5-17 for more information about log
files.
RWHAT Packets
All monitoring devices on the GE Unity Network periodically broadcast information
about themselves in “RWHAT” packets. Among other things, RWHAT packets
contain IP address, port number, name, and offered services information about each
device.
All monitoring devices listen for RWHAT packets, and maintain a database of
information about other devices on the network. When devices need to
communicate, the appropriate IP address information is obtained from the database,
Unity Network-protocol messages are created, and operating system services are
used to transmit the message on the network.
For example, when a CIC Pro computer communicates with a telemetry device, the
telemetry device’s IP address is retrieved from the CIC Pro computer RWHAT
database, the Unity Network messages are created, and the CIC Pro Windows
operating system sends the messages to the telemetry device.
Storage
„
The FD Data interface uses the Unity Network IX network to retrieve data from
the database.
„
Each bed is stored in a separate directory, containing one index file and many
record files.
„
The Report printout is a configurable, long-time-period overview of waveform
activity. It is activated by pressing the Report Print button on the FD display.
„
The Strip printout is a quick snapshot of what is currently on the screen. It is
activated by pressing the Print button while the FD tab page is being viewed.
„
There is a separate FD printer selection from the main laser printer selection on
the setup page. Specifying a printer here does not advertise this CIC Pro as a
print server on RWHAT (like the main laser selection). It is used exclusively by
the local FD system. A printer must be “Added” to the operating system before
it will show up in the drop-down list.
Printing
Revision C
CIC Pro Clinical Information Center
2001099-156
2-7
Equipment Overview: Theory of Operation
Full Disclosure
The CIC Pro Full Disclosure (FD) system stores all waveform and parametric data
from a patient for up to 76 hours. This data can be randomly accessed later in a static
display that looks similar to the real-time display window. To accomplish this, every
Unity Network waveform packet (4/sec) and every parameter packet (1 every 2
seconds) is stored on the CIC Pro. In addition, one RWHAT packet and one admit
packet is stored every minute to help recreate the patient’s history.
Unlike an alarm history event stored at the bedside, which only stores a 10 second
snap shot surrounding the event, Full Disclosure allows the user to scroll back in
time (prior to the event and leading up to the event).
Behaviors/Rules
Installation
„
Incompatible with old FD system (prior to CIC Pro v3.x).
„
Certain Unit Defaults are incompatible with CIC Pro v2.x systems.
Start-up modes
„
Auto For All – Automatically detects admitted beds that are not currently being
FD’d anywhere and attempts to start FD on one of the CIC Pros within the unit.
This works regardless of whether the bed was admitted directly at the bedside or
the CIC Pro. FD cannot be stopped on a particular bed until the bed is
discharged.
„
Auto If Listed – Same as Auto For All, but only if the bed in question is entered
into the list on the FD Unit Defaults setup screen. All other beds are not Full
Disclosed.
„
Manual Mode – Beds are NOT automatically full disclosed upon admission.
User can manually start and stop FD for a particular bed by using a button
located on the Full Disclosure tab sheet. All FD data is deleted when Full
Disclosure is stopped for a bed. See the CIC Pro operator’s manual for more
information.
„
Support for up to 16 patients per CIC Pro
„
Stores 76 hours of data per patient regardless of license type. Licensing controls
amount of data that is viewable.
„
Unit Licensing Mode sets which type of license is requested. Supported types:
None (1 hour), 24, 48, 72 hours.
„
Failure to obtain a license results in no FD (0 hours).
Licensing
2-8
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Theory of Operation
„
Once a license is successfully obtained, if the license server is removed, FD
access is restricted to 1 hour, until the server is returned.
„
Licensing information is stored with data, so restrictions follow data regardless
of where it is viewed, even if viewed in a different unit with a different Unit
License Mode.
Refer to the GE “Unity Network® Information Suite (IS) Server License
Configuration Guide” for more complete licensing information.
Control
CIC Pro Master and Full Disclosure
NOTE
There is only one “master” CIC Pro per care unit (with respect to FD). The CIC
Pro with the lowest IP address becomes the master.
CIC Pro master wakes up every minute to perform the following activities:
„
Detects admitted (but not FD’d) beds within the care unit, for Auto modes.
„
Assigns beds (to be FD’d) to CIC Pros within the care unit.
„
Applies Multi-FD and Twin-Bed Rules.
„
Uses the FDSvr process to find the "FULL DISCLO" service, identifying CIC
Pros with FD ability. The "FULL DISCLO" service is used by the CIC Pro
master while searching for other CIC Pros to query, control & assign beds.
NOTE
The CIC Pro master checks the versions of the other CIC Pros before trying
to identify CIC Pros with FD ability. Versions earlier than 2.5 are ignored.
„
Uses the LocateFD service (provided by FDSvr) to guarantee that the right CIC
Pro is contacted if retrieving data while a multi-FD condition is occurring.
NOTE
A CIC Pro will NOT act as master during first minute after it starts up. Instead,
the latest data is first collected from the other CIC Pros on the network, and this
data determines which CIC Pro becomes master. Multiple masters on a network
might occur, but only briefly (this will self-correct within a couple of minutes).
Minute Rule/Offline Storage Setting
(see “Offline Storage” on page 3-49):
The Offline Storage Setting determines the length of time the CIC Pro will maintain
FD data for a bed from which it has stopped receiving data (a NO COMM
condition).
„
Revision C
If the NO COMM condition ends WITHIN the Offline Storage setting time
frame, FD for the bed continues.
CIC Pro Clinical Information Center
2001099-156
2-9
Equipment Overview: Theory of Operation
„
If a NO COMM condition exceeds the Offline Storage Setting time frame, all
data for the bed is deleted. If (while in Auto mode) the bed comes back online
(the NO COMM condition ends) after this point, the bed is reassigned as a new
and different bed.
„
When FD starts up on a CIC Pro, it determines when data was last received
from the assigned beds.
If the latest data for a bed:
‹
IS NOT older than the Offline Storage Setting, then FD data collection
continues for the bed.
‹
IS older than the Offline Storage Setting, existing data associated with the
bed is deleted, and NEW data collection is started for the bed.
Multi-FD Rule
If multiple CIC Pros are running in a unit, the Multi-FD Rule allows switching of
FD data collection from one CIC Pro to another, if a CIC Pro goes offline (reboot,
shutdown, etc.).
The following examples describe how the Multi-FD Rule works when using 2 CIC
Pros (CICA and CICB).
NOTE
In the examples, start up mode is “Auto”, and only one bed (BED1) is used.
Example 1:
1.
CICA goes offline and stops FD'ing BED1. This causes CICB to begin FD'ing
BED1.
NOTE
The master CIC Pro detects CICA going offline, and switches FD over to
CICB, within a minute of the offline event occurring.
2.
CICA comes back online within the Offline Storage Setting time frame, and
continues FD'ing BED1 again. At this point, CICA and CICB are BOTH FD'ing
BED1.
3.
The Multi-FD Rule does not allow more than one CIC Pro to FD the same bed
at the same time--only the data that goes the farthest back in time (CICA) is
kept. Because of this, CICA continues FD'ing BED1, and CICB stops.
Example 2:
2-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Theory of Operation
1.
CICA goes offline and stops FD'ing BED1. This causes CICB to begin FD'ing
BED1.
NOTE
The master CIC Pro detects CICA going offline, and switches FD over to
CICB, within a minute of the offline event.
2.
CICA comes back online after exceeding the Offline Storage Setting time
frame. Because of this, existing BED1 data is deleted from CICA, and NEW
BED1 data collection begins on CICA. At this point, CICA and CICB are
BOTH FD'ing BED1.
3.
The Multi-FD Rule does not allow more than one CIC Pro to FD the same bed
at the same time (only the data that goes the farthest back in time (CICB) is
kept). Because of this, CICB continues FD'ing BED1, and CICA stops.
NOTE
At every wake-up cycle (1 minute interval), the CIC Pro master scans all CIC
Pros within the unit to determine if more than one CIC Pro is FD’ing the same
bed (this can happen as part of normal operation). If the master detects this
condition, all CIC Pros are instructed to stop FD’ing the bed except for the one
CIC Pro with the oldest data for the bed.
Combo mode (Twin-Bed Rule)
In combo mode:
1.
Two beds are on the network: one is a hardwired bed, and the other is telemetry
bed. Both beds share the same name (e.g. BED), but '*' is added to the telemetry
bed name (e.g. BED and BED*). Both beds represent the same patient. NOTE:
If NOT in combo mode, these two beds would be treated as distinct and
separate.
2.
Only the oldest bed data is kept. Data for the other bed is deleted.
3.
No matter which data is kept, FD continues, and the data is stored under the bed
name, but without the '*'.
4.
Combo mode is stopped (breaking combo) by discharging either bed. If this
happens, FD continues on the UNDISCHARGED bed, and the data is stored
under the UNDISCHARGED bed name.
NOTE
If the hardwired bed is discharged, '*' would again be included in the
name.
To follow are combo mode examples:
Combo Mode Example 1:
a.
Revision C
A hardwired bed BED is admitted and is FD'ed.
CIC Pro Clinical Information Center
2001099-156
2-11
Equipment Overview: Theory of Operation
b.
A telemetry bed BED* is admitted and is FD'ed. Two separate FD data
stores exist, one for each bed (possibly on a different CIC Pros).
c.
The two beds are put into combo mode. When this happens, BED data is
kept because it is older than BED* data, and BED* data is deleted. FD
continues, and the data is stored under the name BED.
d.
Combo is broken by discharging BED*. FD continues, with the data is still
stored under the name BED.
Combo Mode Example 2:
a.
A hardwired bed BED is admitted and is FD'ed.
b.
A telemetry bed BED* is admitted and is FD'ed. Two separate FD data
stores exist, one for each bed (possibly on a different CIC Pros).
c.
The two beds are put into combo mode. When this happens, BED data is
kept because it is older than BED* data, and BED* data is deleted. FD
continues, and the data is stored under the name BED.
d.
Combo is broken by discharging BED. FD continues, with the data is NOW
stored under the name BED*.
Combo Mode Example 3:
2-12
a.
A telemetry bed BED* is admitted and is FD'ed.
b.
A hardwire bed BED is admitted and is FD'ed. Two separate FD data stores
exist, one for each bed (possibly on a different CIC Pros).
c.
The two beds are put into combo mode. When this happens, BED* data is
kept because it is older than BED data, and BED data is deleted. FD
continues, and the data is stored under the name BED*.
d.
Combo is broken by discharging BED*. FD continues, with the data is still
stored under the name BED.
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Theory of Operation
Combo Mode Example 4:
a.
A telemetry bed BED* is admitted and is FD'ed.
b.
A hardwired bed BED is admitted and is FD'ed. Two separate FD data
stores exist, one for each bed (possibly on a different CIC Pros).
c.
The two beds are put into combo mode. When this happens, BED* data is
kept because it is older than BED data, and BED data is deleted. FD
continues, and the data is stored under the name BED.
d.
Combo is broken by discharging BED. FD continues, with the data is NOW
stored under the name BED*.
Bed Name/IP Address Changes
The CIC Pro tracks FD data for individual beds using a combination of bed name
and IP address.
If a bed name changes during FD, information within the data reflects the new name
at the point the name change occurred.
If a bed IP address changes during FD, but the name does not, it is interpreted as a
new and different bed, and a new FD data store is created and utilized for the new
bed.
If a bed goes offline, and another bed with the same name but a different IP address
comes online, it is interpreted as a new and different bed and a FD data store is
created and utilized for it. However, the following conditions apply:
„
If the bed that went offline (with the original IP address) comes back online
within the Offline Storage Setting time frame, the two beds continue to be
treated as separate, since the different IP addresses positively identify them as
distinct and separate.
„
If the bed that went offline (with the original IP address) DOES NOT come
back online within the Offline Storage Setting time frame, the bed with the new
IP address is interpreted as a REPLACEMENT for the bed with the original IP
address. In this case, the FD data for the bed with the original IP address is
deleted, and FD continues, storing data for the bed with the new IP address
ONLY.
„
Each unit operates independently with respect to Full Disclosure. The only
interaction is when displaying data across units.
„
License Mode – All beds within a unit will have access to the same amount of
data storage (based on time) unless there is a failure to get a license (0 hours).
Can be different across different units (i.e. UnitA-72hours, UnitB-None
(1hour).
„
Start-up Mode – The same start-up rules apply to all beds within a Unit.
Unit Boundary
Revision C
CIC Pro Clinical Information Center
2001099-156
2-13
Equipment Overview: Theory of Operation
2-14
„
Default – The same FD defaults are used on all CIC Pros within a unit.
„
Master – One per unit.
„
Acquisition/Storage – The master CIC Pro will only assign beds within the unit
to CIC Pros within that Unit. At no time will data for a bed be stored on a CIC
Pro in a different unit.
„
Bed Unit Changes – If a bed changes its unit name while being FD’d, all data
for the bed is deleted. The unit where the bed moved becomes responsible for
the bed.
„
CIC Pro Unit Changes – If a CIC Pro changes its unit name while FD’ing beds,
all data for all beds being FD’d by that CIC Pro is deleted. The unit with the
original name is responsible for the beds.
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: CIC Pro Service Logons
CIC Pro Service Logons
The following table lists all logons associated with CIC Pro service.
CIC Pro Service Logons
Additional
Information
Logon Description
User Name
Password
Webmin
service
salsa1,3,5,7
“Webmin Login”
on page 2-18
CIC Pro Service Tool
admin
tango1,3,5,7
“Logon” on
page 2-24
CIC Pro Administrator
administrator
admin1,3,5,7
“Administrator
Logon” on page 216
CIC Pro Service Mode
N/A
mms_cic
Accessed from the
Service Password
tab within the CIC
Pro application
Command-Line
Utilities
N/A
mms_com
“CIC Pro
Command-line
Utilities” on
page 2-29
System Logons
The following procedures provide CIC Pro software access at the operating system
level. Each logon provides a different level of accessibility at startup. Follow these
steps to logon as appropriate.
CAUTION
If the beds admitted at this CIC Pro are not transferred to a
different monitoring station, they are unmonitored while this CIC
Pro is shut down. Beds admitted on this CIC Pro display “NO
COMM” at other locations if they are not admitted elsewhere.
To shut down the CIC Pro application and log on as a different user:
Revision C
1.
Select Setup CIC.
2.
Select the Service Password tab.
3.
Type mms_com as the Password and press Enter.
4.
In the Windows command prompt that opens, type stop and then press Enter.
5.
Click on Start then select Shut Down and Log off as CIC and press Enter.
CIC Pro Clinical Information Center
2001099-156
2-15
Equipment Overview: CIC Pro Service Logons
6.
Hold down the LEFT Shift key and select Yes. Continue holding until prompted
for the password.
7.
Choose from the following logons and continue as instructed.
Administrator Logon
User name: type administrator and press Tab.
Password: type admin1,3,5,7 and press Enter.
Run-time Logon
User name: type cic and press Tab.
Password: type cic and press Enter.
CIC Pro Service Mode
The CIC Pro service mode is accessed from within the CIC Pro application by
clicking the Setup CIC button and clicking the Service Password tab.
Password: type mms_cic and press Enter.
The service mode is required to configure a number of CIC Pro application settings.
2-16
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Service Utilities
Service Utilities
The Service Utilities provide several advanced and specialized functions for
configuring and troubleshooting the CIC Pro.
WARNING
CIC Pro service utilities are intended for use ONLY by properly
trained, qualified personnel. Do NOT “experiment” with the
service utilities, or use them in any way other than shown in this
manual. Consequences of misuse include loss of patient data,
corruption of CIC Pro or operating system software, or disruption
of the entire Unity Network MC network.
Webmin Overview
Webmin, an internet-based web application, is used for CIC Pro administration and
service.
Local Webmin Access from the CIC Pro Computer
1.
Click the Browse button in the upper right corner of the CIC Pro multiple
patient viewer window.
NOTE
If there is no Browse button, or if the button is disabled, see “Browser
Configuration” on page 3-38.
2.
Depending upon browser configuration, select either the Favorites menu or the
star icon. Select LocalWebmin.
Remote Webmin Access from an Intranet Computer
Intranet Domain Configuration
The intranet computer (usually an FE laptop) must be TCP/IP configured for
contacting the CIC Pro before you can connect to Webmin:
„
The intranet computer IP address scheme must be the same as the CIC Pro.
„
Optional subnet mask and default gateway settings on the intranet computer
must match the CIC Pro.
If necessary, refer to the CIC Pro IP address labels or the site survey documentation
for CIC Pro intranet domain settings.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-17
Equipment Overview: Service Utilities
Intranet Computer LAN Configuration
It may be necessary to change the Internet Explorer LAN settings on the intranet
computer to connect to Webmin. To change Internet Explorer LAN settings:
1.
In Internet Explorer, select Tools > Internet Options.
2.
Click the Connections tab.
3.
Click LAN Settings.
4.
Disable the Automatic Configuration and/or Proxy Server selections, as
required.
5.
Click OK.
Connect to Webmin
1.
Connect the intranet computer to a Unity Network IX network switch, or
connect the intranet computer directly to the CIC Pro Unity Network IX
network connection port using a crossover cable.
2.
Start the Internet Explorer application.
3.
Type http://[CIC Pro server IP address]:10000 into the Address field and
press Enter.
NOTE
[CIC Pro server IP address] is the Unity Network IX network IP address for
the CIC Pro server.
Webmin Login
‹
Username: service
‹
Password: salsa1,3,5,7
This password uses the number 1, not the letter l.
Webmin functionality is divided across a number of different tab screens. Refer to
the following locations for more information about functionality associated with
each screen:
2-18
„
“Webmin Tab” on page 2-19.
„
“System Tab” on page 2-20.
„
“Applications Tab”, starting on page 2-21.
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Service Utilities
Webmin Tab
013A
The following links are found on the Webmin tab.
Webmin Tab Links and Descriptions
Link Name
Description
Webmin Actions Log
Includes a utility for viewing CIC Pro Webmin activity over a configurable period of
time. See “Webmin Actions Log” on page 5-9.
Logout
Logs user out of Webmin application and displays the login window.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-19
Equipment Overview: Service Utilities
System Tab
014A
The following links are found on the System tab.
System Tab Links and Descriptions
Link Name
Description
Asset Management
Includes utilities for viewing CIC Pro part number/serial number information and listing
Unity Network devices connected to the Unity Network MC network. See “CIC Pro
Part/Serial Number Utility” on page 5-3 and “Unity Devices” on page 5-6.
Computer Name
Used to configure CIC Pro computer name and reset licensing for a reconfigured
computer name. See “Computer Name Configuration” on page 3-23.
Diagnostics Logs
Used to download CIC Pro diagnostic logs. See “Diagnostic Logs” on page 5-4.
Runtime Diagnostics
Includes a wide variety of system runtime diagnostic utilities. See “Runtime
Diagnostics” on page 5-5. Also see “Webmin Runtime Diagnostics Utilities” on page C1. for details about specific runtime diagnostics utilities.
Set IP Address
Used to configure CIC Pro IP addresses. See “IP Address Configuration” on page 319.
Statistics
Includes utilities for viewing operating system runtime and fan runtime statistics. See
“CIC Pro Operating System Runtime Statistics” on page 5-10 and “Fan Runtime
Statistics” on page 5-6.
Write Filter
Includes CIC Pro write filter utilities. See “IP Address Configuration” on page 3-19.
Logout
Logs user out of Webmin application and displays the login window.
2-20
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Service Utilities
Applications Tab
015A
The following links are found on the Applications tab screen.
Application Tab Links and Descriptions
Link Name
Description
Availability
Used to view the uptime availability of the CIC Pro over a configurable period of
time. See “Availability (CIC Pro Uptime Report)” on page 5-11.
Backup-Restore CIC Configuration
Utility which allows you to backup and restore certain tools, data files and list files.
See “Backup-Restore CIC Configuration” on page 4-16.
CIC Configuration
Used to add/delete laser printers, view printer configuration information, and to
configure the CIC Pro workstation language settings. See “Network Laser Printer
Installation/Deletion” on page 3-26, “Printer Configuration Information” on page 57 and “Change CIC Pro Language” on page 3-54.
Create Favorite
Used to add favorites to the CIC Pro server browser Favorites menu. See “Create
Webmin Favorite” on page 3-41.
Download Logs
Used to download CIC Pro log files. See “Download Logs (Webmin)” on page 519.
Run CIC Cmd
Used to run CIC Pro commands. Similar in function to “CIC Pro Command-line
Utilities” on page 2-29.
Update CIC
Used to install CIC Pro software upgrades. See the CIC Pro Clinical Information
Center Upgrade Instructions.
View Logs
Used to view CIC Pro log files. See “View Logs (Webmin)” on page 5-18.
Logout
Logs user out of Webmin application and displays the login window.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-21
Equipment Overview: Service Utilities
Integrated Internet Explorer Browser Interface
When CIC Pro Browser Configuration is set to Integrated mode (see “Browser
Configuration” on page 3-38), icons are used for navigation instead of menus.
148A
The following table describes integrated browser icon functionality:
Icon
2-22
Description
Function
Back navigation button
Click to return to previous
webpage (previous webpage
must be cached).
Forward navigation button
Click to advance to next
webpage (next webpage must be
cached).
Stop button
Click to stop the webpage
loading process.
Reload button
Click to reload the current
webpage.
Home button
Click to go to the home webpage.
The home webpage is set by
configuring internet options in
Internet Explorer. See “Browser
Setup” on page 3-39.
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Service Utilities
Icon
Revision C
Description
Function
Print button
Click to print the current
webpage.
Print preview button
Disabled in this release of the
CIC Pro software.
Favorites button
Click to view a list of webpage
shortcuts (favorites). List displays
in the right frame in the browser.
See “Create Webmin Favorite”
on page 3-41 to add to this list of
shortcuts.
CIC Pro Clinical Information Center
2001099-156
2-23
Equipment Overview: Service Utilities
CIC Service Tool
The CIC Service Tool is available from the Windows desktop. It consolidates access
to several service utilities and functions in one area.
NOTE
The Windows desktop is available only if using a full-access logon when
starting up the CIC Pro. See “Administrator Logon” on page 2-16.
Logon
1.
If necessary, shut down the CIC Pro application and log on using a full-access
logon according to “Administrator Logon” on page 2-16.
CAUTION
If the beds admitted at this CIC Pro are not transferred to a
different monitoring station, they are unmonitored while this CIC
Pro is shut down. Beds admitted on this CIC Pro display “NO
COMM” at other locations if they are not admitted elsewhere.
2.
Double-click the CIC Service Tool icon on the Windows desktop. The Service
Tool - Login window displays.
101A
3.
Type admin into the User Name field, and tango1,3,5,7 into the Password
field. Click OK. The CIC - Service Tool window displays.
NOTE
To close the CIC Service Tool at any time, click the Close button at the
bottom left corner of the window.
2-24
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Service Utilities
Service Tools
Click the Service Tools button to display the available service tool buttons.
102A
Double-click the icon to open the associated tool/utility.
NOTE
See “CIC Pro Service Tools” on page D-1 for procedures using the Service
Tools.
Service Tools Icons and Descriptions
Icon Name
Description
Audio Test.
Tests the CIC Pro computer audio equipment function. See “Audio Test” on
page D-4.
Drive Integrity Test
Tests the read/write integrity of CIC Pro computer electronic storage media (see
“Drive Integrity Test” on page D-6). Similar to the DOS chkdsk utility (see “Disk
Check” on page 4-29).
User Asset Management
Displays CIC Pro asset management information. See “User Asset Management”
on page D-8.
Environmental Monitoring
Tests the status of the CIC Pro computer hardware environment, and
environment-effecting components, such as voltage, fans, and temperature. See
“Environment Monitoring” on page D-9.
Network Integrity Test
Tests connectivity to network devices. Similar to DOS ping utility. (ApexPro
Telemetry Server will not return ping over the Unity Network IX network). See
“Network Integrity Test” on page D-11.
Printing Test
Tests CIC Pro printing function (printer installation, configuration, connectivity,
etc.). See “Printing Test” on page D-13.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-25
Equipment Overview: Service Utilities
Service Tools Icons and Descriptions
Icon Name
Description
Video Test.
Tests CIC Pro video function. See “Video Test” on page D-15.
Watchdog Test.
Tests CIC Pro watchdog countdown function. See “Watchdog Test” on page D18.
Log Files
Click the Log Files button to display CIC Pro log files.
104A
2-26
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Service Utilities
System Tools
Click the System Tools button to display the available service tool buttons.
103A
Double-click the icons to open the associated tool/utility. For more information
about Windows tools, refer to documentation for the specific Windows operating
system supplied with the CIC Pro.
Service Tools Icons and Descriptions
Icon Name
Description
Event Viewer.
Opens the Windows Event Viewer window.
Command Prompt.
Opens a Windows command prompt window.
Dr. Watson.
Opens the Dr. Watson for Windows utility. Dr. Watson is used to help detect,
decode and log errors that are encountered while windows or windows programs
are running.
Performance Viewer.
Opens the Windows Performance window.
Registry Editor.
Opens the Windows Registry Editor window.
Task Manager.
Opens the Windows Task Manager window.
System Information.
Displays system information.
Computer Management
Opens the Windows Computer Management window.
Date and Time
Opens the Windows Date and Time Properties window. See “You must restart the
CIC Pro for the new IP Address settings to take effect.” on page 3-22.
Display
Opens the Windows Display Properties window. See “Incorrect CIC Pro
Application Colors” on page 5-12.
Internet Options
Opens the Windows Internet Properties window. See “Browser Setup” on page 339.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-27
Equipment Overview: Service Utilities
Service Tools Icons and Descriptions
Icon Name
Description
Network Connections.
Opens the Windows Network Connections window.
Add a printer
Opens the Windows Add Printer Wizard utility. See “Printer Installation” on
page 3-25.
Regional and Languages
Opens the Windows Regional and Language Options window.
NOTE: See “Completion” on page 3-60 for the preferred procedure for changing
languages.
Sounds and Audio Devices
2-28
Opens the Windows Sounds and Audio Devices Properties window.
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Miscellaneous Service Utilities
Miscellaneous Service Utilities
CIC Pro Command-line Utilities
The command-line utilities use the Windows DOS command prompt for configuring
and troubleshooting the CIC Pro software.
WARNING
Command-line utilities are intended for use ONLY by qualified
personnel with training and experience with their use. Do NOT
“experiment” with any commands other than those shown in this
manual. The consequences of misuse include loss of patient data,
corruption of CIC Pro or operating system software, or disruption
of the entire Unity Network MC network.
To run the command-line utilities:
1.
Use the mouse to select Setup CIC at the bottom of the screen.
2.
From the CIC Setup screen, select the Service Password tab.
3.
At the service password prompt type mms_com and press Enter. This enables
and displays a DOS command prompt window.
Run CIC Cmd (Webmin)
You can also run command-line utilities using Webmin. Webmin provides a
command prompt interface.
To run command-line utilities, using the Webmin interface:
1.
If necessary, login to Webmin as described in “Webmin Overview” on page 217.
2.
Click the Applications tab.
3.
Click the Run CIC Cmd link to access the Run CIC Cmd window.
The Run CIC Cmd window allows you to execute single line Windows Command
Prompts specific to CIC Pro.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-29
Equipment Overview: Miscellaneous Service Utilities
2-30
1.
Type the CIC Pro command code into the Command field.
2.
Click Run Cmd.
CIC Pro Clinical Information Center
2001099-156
Revision C
Equipment Overview: Feedback Indicators
Feedback Indicators
Error/Warning Messages
WARNING
LOSS OF MONITORING — If monitoring at the CIC Pro is
temporarily interrupted, alternate monitoring devices or close
observation should be used until the monitoring function at the
CIC Pro is restored.
"Loss of monitoring function" indicators:
‹
RED SCREEN (RED/GRAY screen for second monitor in
dual-display configured systems) indicates the CIC Pro
application is restarting itself and patient monitoring at the
CIC Pro is NOT occurring. The monitoring function at the
CIC Pro will automatically resume. No user action is
required.
‹
BLUE SCREEN indicates the Windows XPe operating
system has a functional error and patient monitoring at the
CIC Pro is NOT occurring. If the CIC Pro does not
automatically restart after 60 seconds, the monitoring
function at the CIC Pro will not resume until you turn off the
power to the CIC Pro and then turn the power back on. The
monitoring function should resume in approximately 3 to 4
minutes.
Once the monitoring function at the CIC Pro has been restored,
you should verify the correct monitoring state and alarm function.
Audible Alarms
The CIC Pro is equipped with two internal speakers, and an external speaker
connection.
On a mirrored CIC, it may be desirable to reduce the alarm volume to OFF. See “Set
the Alarm Volume” on page 3-37.
NOTE
External speakers are normally connected during operation. An alert message
displays if this connection is broken.
System Environment Monitor Alert Messages
Alert messages display at the bottom of the screen if the CIC Pro detects an error
condition.
Revision C
CIC Pro Clinical Information Center
2001099-156
2-31
Equipment Overview: Feedback Indicators
For your notes
2-32
CIC Pro Clinical Information Center
2001099-156
Revision C
3
Revision C
Installation
CIC Pro Clinical Information Center
2001099-156
3-1
For your notes
3-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Pre-Installation Requirements
Pre-Installation Requirements
Read and fully understand the following pre-installation requirements. Verify all
requirements are met before starting CIC Pro system installation.
Site Requirements
The CIC Pro is capable of producing heat. It must be physically located to provide
proper cooling ventilation.
WARNINGS
SERVER LOCATION — The CIC Pro is UL/IEC/EN 60601-1
certified equipment which may not meet the leakage requirements
for patient care area equipment. The CIC Pro must not be located
in the patient vicinity as a shock hazard may exist, unless the
medical electrical system standard, UL/IEC/EN 60601-1 is
followed.
Failure to adequately ventilate the CIC Pro may cause equipment
failure or improper operating conditions which may endanger the
patients being monitored.
CAUTION
The CIC Pro uses an internal forced-air cooling system, but most
displays do not. The user must determine the heat dissipation
requirements of the selected display and provide for any required
cooling ventilation.
Environmental Limits
The CIC Pro operates reliably within normal office environmental limits. Select a
site which meets the following criteria:
Revision C
„
Clean and reasonably free of excess dust. Dust accelerates system wear.
„
Well ventilated and away from sources of heat.
„
Away from sources of vibration or physical shock.
CIC Pro Clinical Information Center
2001099-156
3-3
Installation: Pre-Installation Requirements
Electromagnetic Fields and Electrical Noise
The CIC Pro should be isolated from strong electromagnetic fields and electrical
noise produced by electrical devices such as:
„
Elevators
„
Copy machines
„
Air conditioners, large fans
„
Large electric motors
„
Radio and TV transmitters
„
High frequency security devices
„
High-load medical devices (e.g. imaging, defibrillators, etc.)
Clearance and Airflow
Keep the CIC Pro ventilation openings free of obstructions. The CIC Pro physical
location should provide at least the following minimal ventilation clearances:
Minimum Ventilation Clearance
Front
20.5 centimeters (8 inches)
Back
20.5 centimeters (8 inches)
1
1 centimeter (3/8 inches)
Bottom
1. The vent holes in the bottom of the unit must not be obstructed. A minimum
airspace of 1 centimeter (3/8") must be provided between the vent holes and the
supporting surface. When installed with the bottom vent holes down, the unit
must be on a hard flat surface with the bottom feet intact.
Do not remove the bottom feet or operate with the bottom of the unit on a
carpeted surface.
If installed with the vent holes in the bottom surface to the side (standing on a
side, or "vertical" installation), a minimum of 1 centimeter (3/8") clearance
must be provided between the vent holes and the adjacent surface.
Access Space
Provide ample access space for CIC Pro maintenance, especially at the back, to
allow access to cable and power connections.
WARNING
Ample access for AC power cord disconnect (from the wall outlet,
or from the back of the unit) is vital to provide positive AC power
disconnection for service or in the event of emergency.
Physical Specifications
Refer to “Physical Specifications” on page A-4
3-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Pre-Installation Requirements
Power Requirements
Duplex Wall Receptacles
WARNING
If power to the CIC Pro is lost, patient monitoring information
will no longer be displayed or stored.
A properly grounded duplex wall receptacle is required for each CIC Pro. Additional
receptacles may be required to accommodate connected peripheral equipment. The
wall receptacle must be installed in an approved junction box. Use only a threeprong, polarized, hospital-grade wall receptacle to accept the three-prong polarized
CIC Pro power plug.
WARNING
DO NOT use plug adapters that defeat the grounding capability of
the three-prong power plug. An ungrounded electrical device
presents a potentially severe and dangerous shock hazard.
Equipment Grounding
WARNING
SHOCK HAZARD — The CIC Pro and all peripheral equipment
must be adequately grounded or a shock hazard may exist.
DO NOT use plug adapters that defeat the grounding capability of
the three-prong power plug. An ungrounded electrical device
presents a potentially severe and dangerous shock hazard.
The grounding pin of all wall receptacles and all exposed metal parts (beds,
radiators, water pipes, etc.) in any patient area should be electrically connected
together. This common ground point should be connected to the nearest
equipotential ground through a bonded grounding system, or with a 10 AWG
stranded copper grounding cable.
The equipotential ground point should be as close to earth ground potential as
possible. If a bonded grounding system is not available, the ground pin of each wall
receptacle must be individually connected to a central grounding point. Do NOT
jumper from ground pin-to-ground pin of the receptacles.
The grounding system must not carry current, such as a grounded neutral, since the
current flow will produce potential differences along the ground path. These
potential differences are a shock hazard source for equipment users and patients.
Do NOT use conduit as a ground conductor. Plastic (PVC) piping or fittings used in
the conduit runs can break the electrical connection to ground, resulting in potential
shock hazards.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-5
Installation: Pre-Installation Requirements
The electrical grounding system should be connected to an earth ground. If this is
not possible, then a good ground reference, such as a metal water pipe, or an
electrically-conductive building component, should be used. IT IS MORE
IMPORTANT THAT ALL GROUNDED OBJECTS IN THE PATIENT AREA
ARE AT THE SAME GROUND POTENTIAL RATHER THAN AT EARTH
GROUND POTENTIAL.
Un-Interruptible Power Supplies
UPS Installations
See “Optional Components” on page 6-25 for a listing of UPS units available for the
CIC Pro.
GE Medical Systems Information Technologies recommends using an uninterruptible power supply (UPS) with the CIC Pro system.
Without a UPS, power line outages may result in:
‹
Improper shut down of the CIC Pro system, causing lengthy disk scan
procedures on reboot
‹
Data loss.
Follow the manufacturer’s recommendations for installing the UPS.
UPS Run-times and Options
CAUTION
Connect the UPS to the CIC Pro (s) and display monitor(s) only.
Do NOT connect printers or other devices to a UPS, as such
devices may shorten estimated run-times. If AC line power is not
restored before UPS run-time is exceeded, the CIC Pro shuts
down and patients will not be monitored.
UPS run-times are affected by:
3-6
„
Battery age,
„
Ambient temperature,
„
Site specific UPS usage patterns, and
„
Load characteristics.
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Pre-Installation Requirements
Your actual run times may be different..
UPS Description
Runtime Estimates
Power Supply UPS 600VA 120 Volts
One CIC Pro with one 18” or 20” display for
approximately 60 minutes
(Approximate total load = 90 watts)
One CIC Pro with two 18” or 20” displays for
approximately 38 minutes
(Approximate total load = 140 watts)
Two CIC Pros with two 18” or 20” displays for
approximately 30 minutes
(Approximate total load = 180 watts)
As a point of reference, the CIC Pro V4.1 CPU typically draws about 40 watts, and
the 20" and the 18" displays each typically draw about 50 watts.
Refer to “Optional Components” on page 6-25 for UPS part number information.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-7
Installation: Pre-Installation Check
Pre-Installation Check
1.
Complete “Pre-Installation Requirements” on page 3-3.
2.
Prepare and install the UPS device according to the manufacturer’s instructions.
3.
Remove all equipment, including all peripheral devices, from the shipping
cartons. Inspect for damage. If any damage is found, contact GE Medical
Systems Information Technologies Technical Support.
4.
Identify all required cables and prepare them for installation.
Installation
WARNINGS
CIC Pro V4.1.x IS in-unit compatible with CIC Pro V4.0.x, but IS
NOT with versions of CIC Pro V3.1 or earlier. Sharing of the
same care unit name across central stations having incompatible
software versions can result in lost or corrupted telemetry alarm
defaults data and loss of audible alarms.
Both hardwire and telemetry beds are limited in the number of
remote view connections that can be supported.
Do not exceed a maximum of 15 CIC Pros in a single logical
care unit.
Attempting simultaneous displays of a patient monitor
(bedside or telemetry) at too many CIC Pros may cause lost
or intermittent communication between CIC Pros and the
patient monitor. This is evidenced by NO COMM or
intermittent communication conditions for the beds.
The maximum CIC Pros viewing a patient bedside monitor can
vary depending on bedside monitor capabilities and network
design.
3-8
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Installation
WARNINGS
Only external devices specifically designed to be connected to the
CIC Pro, or approved by GE Medical Systems Information
Technologies for use with the CIC Pro system, should be
connected, as specified in this manual or as otherwise specified by
the manufacturer.
A shock hazard may exist if external devices are connected
differently from described in this manual, or as directed by the
manufacturer.
External equipment must be connected to the CIC Pro only by
qualified biomedical engineering personnel.
When using keyboard/video/mouse (KVM) switching devices
with the CIC Pro, only the keyboard and mouse should be
switched through the KVM. The video for each CIC Pro should be
displayed on its own monitor. Failure to do so may result in
delayed recognition of alarm events.
CAUTIONS
All external cabling used with the CIC Pro must be routed so it
does not interfere with access to, or operation of, the CIC Pro
system. Install cabling to guard against tripping and accidental
cable disconnection.
Do NOT apply power until all equipment is installed and ready for
use.
This assembly is static sensitive and should be handled using
precautions to prevent electrostatic discharge damage.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-9
Installation: Installation
Physical Installation
The CIC Pro may be physically mounted using one of the following strategies:
„
Desktop — The CIC Pro may be placed on a desktop, under the monitor stand/
base, or on a shelf or floor without the use of any mounting hardware.
„
Wall Mounting — The CIC Pro may be physically mounted to a wall or the
bottom side of a desk or shelf using standard GCX mounting hardware.
Connect Keyboard and Mouse
Connect the keyboard and mouse cables to the USB ports on the back of the CIC
Pro.
NOTE
Any available USB port can be used for keyboard/mouse connections.
USB Ports
090A
CAUTION
If using a Keyboard/Video/Mouse switching device that accepts
either a USB or a PS/2 keyboard, use a PS/2 keyboard only. USB
keyboards may not provide complete functionality in all
languages.
3-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Installation
Install the Speakers
In addition to two internal speakers, the CIC Pro is equipped with an 1/8” external
speaker connection port.
WARNING
CIC Pro audible alarms will not sound for patients with bedside
monitoring devices configured to “Operating Room” mode.
External
Speaker
Port
090A
1.
Identify the sound output jack on the back of the PC tower. The speaker output
receptacle is labeled SPKR.
2.
Plug the speaker into the sound output receptacle.
NOTE
External speakers are connected during normal operation. A alert message
displays when no external speaker connection is present. This is true even if the
CIC Pro is used as a mirrored CIC Pro and its audible alarm volume is set to
OFF.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-11
Installation: Installation
Install the Power Cord
1.
Plug the AC power cord into the receptacle on the back of the CIC Pro.
AC Power Cord
Receptacle
AC Power
Cord Clamp
AC Power
Cord
2.
Install the provided cable clamp around the AC power cord.
3.
Tighten the cable clamp screw to secure the clamp to the chassis.
088A
Connect/Configure the Video Display
Connecting a Single Monitor
Plug the video display monitor into the primary video connection at the back of the
CIC Pro.
Primary Video
Connection
092A
3-12
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Installation
Connecting Dual Monitors
Dual monitors allow a desktop display area twice as wide as a single monitor.
1.
If necessary, shut down the CIC Pro according to “Safe Shutdown/Restart
Procedure” on page 4-4.
2.
Plug the second monitor into the secondary video connection on the back of the
CIC Pro.
3.
Start the CIC Pro.
NOTE
The video connection supports DVI-A analog/digital connections. Analog VGA
monitors require either a VGA to DVI-A connection cable, or a VGA to DVI-A
plug adapter.
Secondary Video
Connection
Primary Video
Connection
092A
Dual Display Configuration
NOTE
You should commit (save) the following configuration changes to the CIC Pro
write filter:
‹ Dual display
‹
Touch screen display
‹
Printer installation
If a hard drive failure occurs, the CIC Pro does not save these uncommitted
(unsaved) configuration changes. After the CIC Pro reboots, you must
reconfigure these settings.
See “Commit Changes to the Write Filter” on page 3-57.
Revision C
1.
Type mms_com and press the Enter key. The Windows command prompt
opens with:
C:\Program Files\Marquette\CIC\4.1.x>
2.
Type stop and press the Enter key.
CIC Pro Clinical Information Center
2001099-156
3-13
Installation: Installation
3.
Select Start > Shut Down, then select Log off [username].
4.
Hold down the left Shift key and click OK. Continue holding the Shift key and
wait for the log on screen.
5.
Type administrator, press the Tab key and type admin1,3,5,7 (with no
spaces). Press the Enter key.
6.
Right-click the ATI icon in the right corner of the toolbar at the bottom of the
screen.
121A
7.
Select Settings > ATI Display Settings.
094A
8.
Click the Settings tab in the Display Properties window.
9.
In the Display Properties window, click the Settings tab. Select the 2 monitor
icon.
096A
3-14
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Installation
10. Select the Extend my Windows desktop onto the monitor checkbox. Click OK.
The desktop expands onto the secondary monitor.
NOTE
The secondary monitor displays a mirror image of the primary monitor
display if the Extend my Windows desktop onto the monitor checkbox is not
selected.
11. If desired, change the Screen resolution setting for the 2 monitor. Note the
setting and select the 1 monitor. Make sure the Screen resolution setting is the
same for both monitors.
12. Click the Apply, and then the Advanced, buttons.
13. Select the General tab in the (Multiple Monitors) and MOBILITY RADEON
9000 Properties window.
095A
14. Select the Apply the new display settings without restarting radio button. Click
Apply.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-15
Installation: Installation
15. Select the Displays tab.
Monitor
Activation
Button
099A
16. Click the monitor activation button for the newly connected secondary monitor.
NOTE
The monitor activation button is located in the upper left corner of the
Monitor title bar. This button is red when the associated monitor is
disabled, and changes to green when the monitor is activated.
17. Click Apply. Click OK. The display appears on the secondary monitor.
18. Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
3-16
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: Installation
Connecting/Configuring Touchscreen Monitor
NOTE
You should commit (save) the following configuration changes to the CIC Pro
write filter:
‹ Dual display
‹
Touch screen display
‹
Printer installation
If a hard drive failure occurss, the CIC Pro does not save these uncommitted
(unsaved) configuration changes. After the CIC Pro reboots, you must
reconfigure these settings.
See “Commit Changes to the Write Filter” on page 3-57.
1.
Plug the video cable for the touchscreen monitor into the primary video
connection on the back of the CIC Pro, and configure the monitor settings
according to “Connect/Configure the Video Display” on page 3-12.
2.
If you have not already done so, complete “Calibrate the Screen” on page 3-32
and “Browser Configuration” on page 3-38.
3.
Plug the data cable for the touchscreen monitor into the COM 1 connection port
on the back of the CIC Pro.
COM 1
Connection
Port
090A
Revision C
4.
Stop the CIC Pro application, shut down the CIC Pro and log back in as
“administrator” according to “Feedback Indicators” on page 2-31.
5.
Select Start > Settings > Control Panel.
6.
Double-click the elo Touchscreen
CIC Pro Clinical Information Center
2001099-156
icon.
3-17
Installation: Installation
7.
If necessary, select the General tab in the Elo Touchscreen Properties window.
123A
8.
Click Align.
9.
As directed, touch the screen at the displayed target
icons.
10. Touch/click OK to close the Elo Touchscreen Properties window.
11. Select Start > Shut Down > Log off administrator.
12. Restart the CIC Pro. It will start up in normal CIC Pro user mode.
13. Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
3-18
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro System Configuration
CIC Pro System Configuration
Enhanced Write Filter
The write filter is used by the CIC Pro to prevent compact flash drive failure or data
corruption. The write filter operates by strictly limiting the write/program operations
occurring to the compact flash drive.
WARNING
LOSS OF MONITORING — You must commit (save) the
following configuration changes to the CIC Pro write filter:
‹ IP address
‹ Computer name
‹ Optional licenses
‹ Language
If a hard drive failure occurs, the CIC Pro discards uncommitted
(unsaved) configuration changes. During the reboot process, the
CIC Pro uses the last committed (saved) configuration settings,
which may be different from the settings the CIC Pro was using
before the hard disk failed.
See “Commit Changes to the Write Filter” on page 3-57.
IP Address Configuration
WARNING
Duplicate IP addresses will cause erratic system communication
and data loss! Be extremely careful not to assign the same IP
address to two different devices.
WARNING
LOSS OF MONITORING — You must commit (save) the
following configuration changes to the CIC Pro write filter:
‹ IP address
‹ Computer name
‹ Optional licenses
‹ Language
If a hard drive failure occurs, the CIC Pro discards uncommitted
(unsaved) configuration changes. During the reboot process, the
CIC Pro uses the last committed (saved) configuration settings,
which may be different from the settings the CIC Pro was using
before the hard disk failed.
See “Commit Changes to the Write Filter” on page 3-57.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-19
Installation: CIC Pro System Configuration
1.
If the Internet Explorer browser is not already running: with the CIC Pro
running, click the Browse button at the top right corner of the multiple patient
viewer window to start the Internet Explorer browser.
2.
Access the Webmin application according to “Local Webmin Access from the
CIC Pro Computer” on page 2-17.
3.
Log in to the Webmin application according to “Webmin Login” on page 2-18.
The Webmin tab window displays.
013A
4.
Click the System tab.
014A
3-20
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro System Configuration
5.
Click the Set IP Address link at the left side of the System tab window. The Set
IP Address window displays.
133A
WARNING
Duplicate IP addresses will cause erratic system communication
and data loss! Be extremely careful not to assign the same IP
address to two different devices.
6.
If the site uses a custom Unity Network MC network addressing scheme (ie not
126.x.x.x scheme), change the Unity Network MC network IP address so that it
is unique on the network.
NOTE
Default IP address settings are recommended. Refer to the site survey for
IP address values.
NOTE
Typically the Unity MC Subnet Mask setting should remain the default
setting 255.0.0.0. See the site survey.
7.
Enter the Unity IX IP Address, Subnet Mask and Default Gateway settings
according to the site survey.
NOTE
Typically the Unity IX Subnet Mask setting should remain 255.255.255.0.
See the site survey.
NOTE
It is necessary to open the Unity IX IP address to the internet before the
browser can access internet web addresses (external to the hospital
network). Contact your hospital IT Administrator.
8.
Revision C
If the Unity Network IX network IP address/subnet mask has been changed,
replace the label on the CIC Pro cover with a new label displaying the IP
address/subnet mask value(s).
CIC Pro Clinical Information Center
2001099-156
3-21
Installation: CIC Pro System Configuration
9.
If desired, select the checkbox to update the CIC Pro computer name using the
last 9 digits of the Unity Network MC network IP address.
140A
CAUTION
LOSS OF DATA — Select the license reset feature only if you are
certain you want to delete the existing license. Previous license
file configuration is lost if this feature is enabled.
10. If desired, select the checkbox to reset the license so it is based on the IP
address (enabled only if the other checkbox selected).
11. Click Save.
12. Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
13. Select the Start > Shut Down
14. From the Shut Down Windows dialog, select Restart. Click OK.
NOTE
You must restart the CIC Pro for the new IP Address settings to take effect.
Connect to Unity Networks
CAUTION
Perform “IP Address Configuration” on page 3-19 before
connecting to the Unity networks.
Do not adjust time/date settings while connected to the network.
Changes to the time/date data affects all units on the network. This
can result in other monitors altering the time and date parameter of
some patient data.
3-22
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro System Configuration
Connect the Unity Network MC and IX network drop cables to the back of the CIC
Pro as indicated in the following figure.
Unity Network
MC Network
Connection
Unity Network
IX Network
Connection
090A
Computer Name Configuration
WARNING
LOSS OF MONITORING — You must commit (save) the
following configuration changes to the CIC Pro write filter:
‹ IP address
‹ Computer name
‹ Optional licenses
‹ Language
If a hard drive failure occurs, the CIC Pro discards uncommitted
(unsaved) configuration changes. During the reboot process, the
CIC Pro uses the last committed (saved) configuration settings,
which may be different from the settings the CIC Pro was using
before the hard disk failed.
See “Commit Changes to the Write Filter” on page 3-57.
Revision C
1.
If the Internet Explorer browser is not already running: with the CIC Pro
running, click the Browse button at the top right corner of the multiple patient
viewer window to start the Internet Explorer browser.
2.
Access the Webmin application according to “Local Webmin Access from the
CIC Pro Computer” on page 2-17.
CIC Pro Clinical Information Center
2001099-156
3-23
Installation: CIC Pro System Configuration
3.
Log in to the Webmin application according to “Webmin Login” on page 2-18.
The Webmin tab window displays.
013A
4.
Click the System tab.
014A
5.
Click the Computer Name link at the left side of the System tab window.
6.
Type a computer name for the CIC Pro into the New name field.
CAUTION
LOSS OF DATA — Select the license reset feature only if you are
certain you want to delete the existing license. Previous license
file configuration is lost if this feature is enabled.
3-24
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro System Configuration
7.
If desired, select the checkbox to reset the license so it is based on the new
computer name.
126A
8.
Click Submit.
127A
9.
Select the Start > Shut Down.
10. From the Shut Down Windows dialog, select Restart. Click OK.
NOTE
You must restart the CIC Pro for the new computer name setting to take
effect.
11. Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
12. Replace the label on the CIC Pro cover with a new label displaying the new
Computer Name.
Printer Installation
NOTE
You should commit (save) the following configuration changes to the CIC Pro
write filter:
‹ Dual display
‹
Touch screen display
‹
Printer installation
If a hard drive failure occurs, the CIC Pro does not save these uncommitted
(unsaved) configuration changes. After the CIC Pro reboots, you must
reconfigure these settings.
See “Commit Changes to the Write Filter” on page 3-57.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-25
Installation: CIC Pro System Configuration
Digital Writer Printer Connection
NOTE
A DDW cannot be used for the following (Use a laser printer instead. See
“Network Laser Printer Installation/Deletion” on page 3-26):
1.
‹
to print from the Graphic Trends or Alarm Histories tabs if “PDS” is
selected for those tabs.
‹
to print Full disclosure and Flow Volume Loop information.
Connect the PRN 50 or PRN 50-m to the COM 2 port on the back of the CIC
Pro. Connect an AC power cord from the power receptacle on the PRN 50 or
PRN 50-M to a power outlet.
COM 2
Connection
Port
090A
NOTE
The PRN 50 or PRN 50-M must use software V2B or later to operate
properly with the CIC Pro and telemetry systems.
2.
Load the printer with paper.
3.
In CIC Setup, on the CIC Defaults tab, select COM2 from the DDW selection
list. See “Printer/Writer Selections” on page 3-34.
Network Laser Printer Installation/Deletion
NOTE
The following installation procedure requires a Unity Network IX network
connection. See “You must restart the CIC Pro for the new IP Address settings
to take effect.” on page 3-22.
NOTE
The following only applies when the printer TCP/IP scheme follows Unity
Network IX network IP addressing factory defaults (192.168.x.x), and the
printer is physically located on the Unity Network IX network. When the
printer uses a custom TCP/IP addressing scheme, refer to the site survey for
resolving TCP/IP values.
3-26
1.
If necessary, shut down the CIC Pro application and log on using a full-access
logon according to “System Logons” on page 2-15.
2.
Ensure the printer is on.
3.
Using the Internet Explorer application, click the Browse button at the top right
corner of the multiple patient viewer window to start the Internet Explorer
browser.
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro System Configuration
4.
Access the Webmin application according to “Local Webmin Access from the
CIC Pro Computer” on page 2-17.
5.
Log in to the Webmin application according to “Webmin Login” on page 2-18.
The Webmin tab window displays.
6.
Click the Applications tab.
7.
Click the CIC Configuration link at the left side of the Applications tab
window.
015A
Laser Printer Installation
a.
Click Install Laser Printer.
028A
b.
Type the IP address assigned to the printer (e.g. 192.168.x.x) into the
Printer’s IP Address field.
NOTE
If the addressing scheme has been changed from factory defaults, the
assigned addresses must be tracked and maintained at the site. This
includes printers. The above scheme only applies when the Unity Network
IX network IP addressing has not changed from the factory defaults and/or
the printer is not connected to your hospital’s network.
c.
Select the printer type from the Printer Type pull-down menu. Supported
HP LaserJet printers include: 4100 Series PS, 4300 PS, 4050 Series PS,
4200 PS and 4000 Series PS.
d.
If desired, enter a name for the printer into the Printer Name field (29
characters max.).
NOTE
If a printer name isn’t specified, the printer type will be used as the
printer’s name.
e.
Revision C
If desired, enter a comment about the printer into the Comment field. (29
characters max.).
CIC Pro Clinical Information Center
2001099-156
3-27
Installation: CIC Pro System Configuration
f.
Select Yes or No from the Test Page pull-down menu for printing a test page
as part of the printer installation (recommended).
g.
Click Submit.
134A
h.
The results of the printer installation operation display. If the printer was
successfully installed, it will be displayed in the printer list.
135A
i.
Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
Delete Laser Printer
a.
Click Delete Printer.
028A
b.
Select the printer you want to delete from the Printer Name pull-down
menu in the Select the printer to delete section of the window.
c.
Click Submit.
136A
3-28
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro System Configuration
d.
The results of the printer delete operation display. If the printer was
successfully deleted, it will be removed from the printer list.
137A
e.
Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
Setting Default Paper Size
CIC Pro default paper size can be configured.
NOTE
Default paper size is set to 8 x 11 inches (letter) during manufacturing.
1.
If necessary, shut down the CIC Pro application and log on using a full-access
logon according to “System Logons” on page 2-15
2.
Select Start > Settings > Printers and Faxes.
3.
Right-click the printer you want to configure.
4.
Select Properties.
5.
Click the Advanced tab.
138A
6.
Revision C
Click Printing Defaults.
CIC Pro Clinical Information Center
2001099-156
3-29
Installation: CIC Pro System Configuration
7.
Click Advanced.
139A
8.
Select default paper size (Paper Size).
9.
Click OK (Advanced Options).
10. Click OK (Printing Defaults).
11. Click OK (Properties).
12. Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
3-30
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
CIC Pro Application Configuration
Access Service Mode
WARNING
The service mode is intended for use only by qualified personnel
with training and experience in its use. The consequences of
misuse include loss of alarm configuration, loss of patient data,
corruption of the CIC Pro operating system software, or disruption
of the entire network.
1.
Use the mouse to select Setup CIC at the bottom of the screen.
2.
From the CIC Setup screen, select the Service Password tab.
3.
When prompted for the service password, type mms_cic and press Enter.
NOTE
When you are in service mode, a cross-shaped (+) cursor replaces the
normal arrow cursor.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-31
Installation: CIC Pro Application Configuration
Calibrate the Screen
CAUTION
When using a video splitter with the CIC Pro, screen calibration
may be possible with only one of the monitors connected to the
splitter. This is because changing calibration for one monitor will
effect the calibration of all other monitors connected to that same
splitter.
When using monitors connected to a splitter, only the last monitor
calibrated will have proper calibration. Manual measurements
should be made from that monitor only.
NOTE
A flexible clear plastic ruler calibrated in inches and centimeters is
recommended for this procedure.
3-32
1.
Click the Screen Calibration tab. The Screen Calibration window displays.
2.
The illustration shows the Screen Calibration screen. In the Units box, use the
mouse to choose Inches or Centimeters, as desired.
3.
Click the Default button to set the display to the default resolution.
4.
Hold your ruler horizontally on the computer screen, aligning the zero mark of
your ruler with the horizontal zero mark of the screen ruler.
5.
Position the arrow cursor on the screen ruler. Hold down the left mouse button
while you drag (or pull) the screen ruler as required until the calibration marks
of the screen ruler match the calibration marks of your ruler, then release the
mouse button.
6.
Repeat this procedure to calibrate the vertical screen ruler.
7.
When you finish, click Apply at the lower right corner of the calibration screen
to save the calibration settings.
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
NOTE
The Maintain Aspect Ratio button on the calibration screen allows you to
adjust both screen rulers simultaneously. However, because of differences
in monitors and screen resolution, it is recommended that each ruler be
adjusted separately. (A check mark in the box indicates when this option is
selected. Click on the check box to deselect it.)
Set the CIC Pro Defaults
Open the CIC Defaults Screen
Click on the CIC Defaults tab to open the CIC Defaults screen shown.
Set the Central Name
NOTE
All other systems on the GE Medical Systems Information Technologies’ Unity
Network MC network use the Central name to identify this CIC Pro system.
The central name should be relevant to the location, such as CS-2 or SDU3.
Make sure no other central stations have the same central name.
Press Tab to enter the Name fields. The maximum length of the central name is
four characters.
Click the text cursor in the Central field.
Type the Central name. When finished, click Apply at the lower right corner of the
screen to save the setting.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-33
Installation: CIC Pro Application Configuration
Set the Unit Name
NOTE
All other systems on the Unity Network MC network use the care Unit name to
identify this CIC Pro system. The care unit name should be relevant to the
location, such as CCU or ICU-1.
It is very important to enter the correct care unit name. Be especially careful of
the name’s spelling.
If any other CIC Pro systems are intended to have the same care unit name,
make sure the names match exactly.
CAUTION
Once the care unit name is programmed, and the CIC Pro system
is placed in use, avoid changing the care unit name. Changing the
care unit name deletes all Full Disclosure data and the list of
transmitters stored on the CIC Pro. You must re-enter that data
after you change the care unit name.
Click on the arrowhead next to the Unit name box to open the pull-down list of unit
names. Click to select the desired name, or type a name into the Unit name box and
press Enter.
NOTE
If the care unit name is not listed in the pull-down list, you must type it in. The
name cannot exceed 7 characters.
When finished, click Apply at the lower right corner of the screen to save the setting.
Set the Waveform Parameters
Select the waveforms you wish to display (up to three in addition to the ECG).
„
Graph Waveform ECG 1: Default from ECG Source.
„
Graph Waveform 2 – 4:Allows you to enable/disable graphing for subsequent
ECG leads. Choices are Off, I, II, III, V, aVR, aVL, and aVF.
When finished, click Apply at the lower right corner of the screen to save the setting.
Printer/Writer Selections
NOTE
A DDW cannot be used for the following. Use a laser printer instead:
‹
if the CIC Pro is used with Unity PDS.
‹
to print Full disclosure and Flow Volume Loop information.
Full disclosure reports can potentially be very large and take a very long time to
print. This can block alarm graphs sent to that printer. Because of this, the print
setup for full disclosure reports may be different from other patient printing.
It is important to note that the printer selection list for full disclosure is different
from the printer selection list for graphing patient data. The following printer list
may not include printers defined on the GE Unity Network IX network.
3-34
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Laser: – Select your network or installed parallel laser printer from the drop down
list.
DDW: – If a Digital Data Writer is connected to the CIC Pro COM2 port, select
COM2 from the drop-down list.
Full Disclosure: – Select your network.
Click Apply to save the new settings.
Setup a Mirrored CIC Pro
The Mirror Central Display option allows double-monitoring of patients from
remote, or secondary CIC Pros. The display on the remote CIC Pro is configured to
mirror the originating CIC Pro (i.e. the same patients are shown in the same display
slots).
If the remote CIC Pro is located in the same care unit as the originating CIC Pro, it
can be used to perform the Admit/Discharge, or any other display-type operation(s).
The changes affect both CIC Pros. For example, if the patient in a display slot is
changed at one CIC Pro, the change is also automatically performed on the
mirroring CIC Pro.
CAUTION
Use caution when configuring mirror CIC Pro displays. Since
changes at one CIC Pro can affect the other display, there may be
an inadvertent loss of patient monitoring at the primary CIC Pro.
You can set up a mirrored CIC Pro for remote monitoring. This is set in the service
mode. Changes made on one CIC Pro are NOT always made at the other CIC Pro.
The main and mirrored CIC Pro share bed lists. When a mirrored CIC Pro is set up,
these rules are in place:
Revision C
„
If the user selects new parameters or colors to view on one display, that view is
NOT “mirrored” on the other display.
„
Auto Display is disabled at the mirrored CIC Pro. However, it is still active on
the main CIC Pro. You must select Disable Auto Display on the Display Format
tab of the main CIC Pro.
„
When using the mirror feature, configure both the main and mirrored CIC Pro
to the same Display Format tab settings. Whenever these settings are changed
on one CIC Pro, they are NOT reflected on the other.
„
The title bar of the mirror CIC Pro displays mirror of [CIC SELECTED].
CIC Pro Clinical Information Center
2001099-156
3-35
Installation: CIC Pro Application Configuration
To set up a mirrored CIC Pro:
3-36
1.
At the remote CIC Pro (mirroring the main CIC Pro), click on Setup CIC.
2.
Auto Display is disabled on the remote CIC Pro. However, it is still active on
the main CIC Pro. Select the Display Format tab and then click the Disable
Auto Display button.
3.
Configure both the main and remote CIC Pro to the same Display Format tab
settings for the mirror feature to work correctly. Set the Columns and Rows in
the Display Format tab of the remote CIC Pro to match the main CIC Pro.
4.
Select the CIC Defaults tab.
5.
At Mirror Central Display there is a drop-down box. All CIC Pros on the
network are listed in this box. Select the main CIC Pro you want mirrored on
the remote CIC Pro.
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
6.
Check the title bar on the remote CIC Pro to assure that it is mirroring the
correct CIC Pro.
7.
When complete, click Apply at the lower right corner of the screen to save the
new settings.
Set the Alarm Volume
This box allows you to set the alarm volume for the CIC Pro. The minimum alarm
volume is set in service mode. Actual alarm volumes can be set to any level above
this minimum while either in user or service mode.
To adjust the minimum alarm volume:
1.
If not already in service mode, enter service mode according to “Access Service
Mode” on page 3-31.
2.
Click the Minimum drop-down menu in the Alarm Volume control area. The
Alarm Volume drop down menu opens.
063A
3.
Click on the desired minimum volume level. The minimum volume can be set
between OFF and 100% in the user mode.
4.
Values above the Minimum volume setting can be selected from the Current
drop-down menu while either in service or user mode. The value selected in the
Current list determines the actual alarm volume
Adjust the Color Set
This option allows you to set a color scheme for waveforms. Preset choices are
Clinical, Transducer, or Custom. You must be in the service mode to make
modifications to the color set. See “Access Service Mode” on page 3-31.
Revision C
„
Clinical—This sets the colors for single-parameter or double-parameter patient
monitoring. ECG waveforms display in orange; ART, PA, FEM, CVP, RA, LA,
ICP, SP, UAC, and UVC display in green; and RESP, SPO2, and CO2 display in
blue.
„
Transducer—This sets the colors for multi-parameter patient monitoring.
Display colors are: ECG in brown, ART in red, PA in yellow, FEM in red, CVP
and RA in blue, LA and ICP in white, SP in green, UAC in red, UVC in blue,
RESP and SPO2 in green, and CO2 in white.
„
Custom—This selection allows you to set each waveform color individually.
CIC Pro Clinical Information Center
2001099-156
3-37
Installation: CIC Pro Application Configuration
NOTE
When you select Custom, you can select colors for each of the fourteen
waveform types individually.
When you select either Clinical or Transducer, access to the color palettes for
individual waveforms is disabled.
To access and change the Color Set:
1.
Click on the arrow next to the waveform name box to open the pull-down list of
colors.
2.
Click on the desired color to select it.
3.
When color selections are complete, click Apply at the lower right corner of the
screen to save the new color settings.
Browser Configuration
This control allows you to disable/enable and configure the web browser. When
enabled, the Browser Configuration control includes the following possibilities.
„
Single Monitor—Browser is Integrated into Single Patient Viewer,
„
Single Monitor—Browser is Free Floating,
„
Dual Monitor—Browser displays on second monitor.
You must be in the service mode to make modifications to browser configuration.
See “Access Service Mode” on page 3-31.
NOTE
The browser interface differs slightly depending upon the selected mode:
Free Floating option — Displays text menus; favorites list displays in left
frame.
Integrated option — Displays icons instead of menus; favorites list displays in
right frame. See “Integrated Internet Explorer Browser Interface” on page 2-22.
To exit the web browser, select the Close button.
3-38
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Browser Setup
WARNING
LOSS OF MONITORING — If the browser function is
inappropriately used, loss of monitoring function may result. Use
alternate monitoring devices or close patient observation until the
monitoring function at the CIC Pro is restored.
When using the browser, follow these restrictions:
‹
Do NOT attempt to access the file systems of the CIC Pro
through the use of the browser.
‹
Do NOT attempt to download files of any type. This includes,
but is not limited to, audio or video files.
1.
If not already in service mode, enter service mode according to “Access Service
Mode” on page 3-31.
2.
Click the Enable checkbox in the Browser Configuration control area. Select
the Free Floating radio button.
3.
With the CIC Pro running, click the Browse button at the top right corner of the
multiple patient viewer window to start the Internet Explorer browser. Select
Tools > Internet Options. The Internet Options window appears.
4.
In the Address field, in the Home Page section enter the internet address you
want for your browser home page, or select Use Blank. Click OK.
NOTE
When Internet Explorer starts, the browser displays the specified Home
Page.
5.
Revision C
Click Settings in the Temporary Internet files control.
CIC Pro Clinical Information Center
2001099-156
3-39
Installation: CIC Pro Application Configuration
6.
Ensure the Temporary Internet files folder disk space does not exceed 1 MB and
click OK.
157A
7.
If you use Automatic configuration or a Proxy server, then select the
Connections tab and click LAN Settings...
8.
Enter the Automatic configuration script or Proxy server address and port
according to the hospital IT Administrator. Click OK
NOTE
Before the browser can access internet web addresses (external to the
hospital network), a connection for the Unity Network IX network to the
3-40
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
internet is required. Contact the hospital IT Administrator.
9.
Proceed to “IP Address Configuration” on page 3-19.
Create Webmin Favorite
The Create Webmin Favorite feature is useful for adding favorites to the CIC Pro
browser. It is a simpler, alternative method to the procedure described in “MUSE
Web Frame/No Frame Connection Errors” on page 5-12.
Revision C
1.
If the Internet Explorer browser is not already running: with the CIC Pro
running, click the Browse button at the top right corner of the multiple patient
viewer window to start the Internet Explorer browser.
2.
Access the Webmin application according to “Local Webmin Access from the
CIC Pro Computer” on page 2-17.
3.
Log in to the Webmin application according to “Webmin Login” on page 2-18.
4.
Click the Applications tab.
5.
Click the Create Favorite link to display the following window.
6.
Type a name into the Please enter a name for this favorite field. The name will
identify the favorite in the browsers Favorites menu.
CIC Pro Clinical Information Center
2001099-156
3-41
Installation: CIC Pro Application Configuration
7.
Type the internet address or IP address into the Please enter an IP address or
URL (webpage address) field.
8.
Click to select a radio button associated with the link for the favorite.
9.
If necessary, click Reset Form to clear the fields and selections on the page.
10. Click Create Favorite.
11. Click the link to test the URL for the new favorite.
Delete Favorites
3-42
1.
If not already in service mode, enter service mode according to “Access Service
Mode” on page 3-31.
2.
Click the Enable checkbox in the Browser Configuration control area. Select
the Free Floating radio button.
3.
With the CIC Pro running, click the Browse button at the top right corner of the
multiple patient viewer window to start the Internet Explorer browser. Select
Organize Favorites. The Organize Favorites window displays.
4.
Select the favorite you want to delete from the list of favorites.
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
5.
Click Delete.
6.
Click Yes in the Confirm File Delete window to delete the selected favorite.
Set the Telemetry Unit Defaults
Open the Telemetry Unit Defaults Screen
Click on the Telemetry Unit Defaults tab as shown.
This option allows you access to the telemetry default settings. In user mode, all of
the controls on the Telemetry Unit Defaults tab sheet are view-only. You must be in
the service mode to review and make modifications to the Telemetry Unit Defaults
tab sheet at the CIC Pro.
NOTE
For more information on setting Telemetry Unit Defaults, refer to the telemetry
system’s operator manual.
Graph Setup
NOTE
When changing the graph setup for admitted patients, the changes will not take
effect until the patients are discharged or the settings are changed in the Graph
Setup settings for each patient.
Locations – The Default locations for this CIC are only used for telemetry beds.
These defaults determine where patient data prints for either manual or alarm
Revision C
CIC Pro Clinical Information Center
2001099-156
3-43
Installation: CIC Pro Application Configuration
conditions. Since a telemetry patient is not linked to a monitor, such as a Solar 8000
monitor, these defaults are necessary to specify the destination for alarm and manual
graph printouts.
The Location options are:
„
Manual Graph Location —allows you to designate the default manual graph
location for telemetry patients.
„
Alarm Graph Location —allows you to designate the default alarm graph
location for telemetry patients.
„
Print Window Location —allows you to designate the default print window
location for telemetry patients.
Waveform – Under the Graph Setup Waveform option you can designate the
primary ECG lead for printing and enable/disable graphing for subsequent ECG
leads.
The Waveform options are:
„
Graph Waveform ECG 1:—allows you to designate the primary ECG lead for
printing.
„
Graph Waveform 2 – 4—allow you to enable/disable graphing for subsequent
ECG leads. Choices are Off, I, II, III, V, aVR, aVL, and aVF.
Transmitter Graph – This option allows you to turn on/off transmitter graph
printing.
Alarm Graph On/Off – This option allows you to turn on/off alarm graph printing.
Event Marker Graph On/Off – This option allows you to turn on/off event marker
graph printing.
ECG
Display Lead – This option allows you to set the primary ECG lead for display in
the patient’s waveform window. Choices are I, II, III, V, aVR, aVL, and aVF.
Arrhythmia – This option allows you to enable/disable an arrhythmia analysis
program. Choices are Full, Lethal, and Off.
Lead Analysis – This option allows you to designate Single-Lead or Multi-Lead
analysis for ECG and arrhythmia analysis.
ST Analysis – This option allows you to enable/disable ST analysis. Choices are On
or Off.
Va Lead/Vb Lead – This option allows you to set the default for the V leads that
will be monitored in these positions. A 6-lead cable is required for multiple V-lead
monitoring. Choices for Va are V1, V2, V3, V4, V5, and V6. Choices for Vb are V2,
V3, V4, V5, and V6.
Detect Pace – This option allows you to enable/disable pacer detection. Choices are
Pace 1, Pace 2, and Off.
3-44
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
PDS
This option allows you to enable/disable use of the Patient Data Server (PDS).
The PDS is a rack-mounted server that collects graphic trend and alarm history data
from data acquisition devices on the Unity Network MC network. Data may be
collected from both hard wired and wireless devices and returns the data to a CIC
Pro for viewing. This data is stored in the PDS for 72 hours, providing temporary
patient data history. Additionally, the data is stored under a patient’s PID so that as
the patient is moved from one data acquisition device or care unit to another, the
new data is appended to the patient’s record, creating one continuous 72-hour string
of data.
Patient Age, Transmitter Alarm Pause and Alarm Pause Breakthrough
Patient Age – This option allows you to set patient age. Choices are 0–2 Years, 3–
11 Years, 11–13 Years, and Adult.
Transmitter Alarm Pause – This option allows you to turn on/off transmitter
alarm pausing. Choices are Enabled, Disabled, and Off.
Alarm Pause Breakthrough – This option allows you to turn on/off transmitter
pause breakthrough. Choices are Always on and Always off.
When finished, click Apply at the lower right corner of the screen to save the
settings.
Event Marker – This option allows you to turn on/off event marker alert.
Set the Telemetry Alarm Control Defaults
Open the Telemetry Alarm Control Defaults Screen
Click on the Telemetry Alarm Control Defaults tab as shown.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-45
Installation: CIC Pro Application Configuration
In user mode, all of the controls on the Telemetry Alarm Control Defaults tab sheet
are view-only. You must be in the service mode to set default limits and alarm
settings for telemetry patients.
NOTE
For more information on setting Telemetry Alarm Control Defaults, refer to the
telemetry system’s operator manual.
Parameter Limits and Alarm Levels
This option allows you to set default parameter alarms limits and alarm levels for the
following parameters: HR, NBP-S, NBP-D, NBP-M, SPO2, SPO2-R, ST-I, ST-II,
ST-III, ST-V, ST-V2, ST-V3, ST-V4, ST-V5, ST-V6, ST-aVR, ST-aVL, ST-aVF,
and PVC.
Arrhythmia Alarm Levels
This option allows you to set default alarm levels for the following arrhythmia calls:
ASYSTOLE, VFIB/VTAC, V TACH, V BRADY, VT>2, ACC VENT, PAUSE,
TACHY, BRADY, R ON T, COUPLET, BIGEMINY, TRIGEMINY, PVC,
IRREGULAR and ATRIAL FIB.
NOTE
The default alarm levels for ASYSTOLE and VFIB/VTACH cannot be moved
from the Crisis level.
System Alarm Levels
This option allows you to set default alarm levels for the following system alarms:
CHANGE BATTERY, OFF NETWORK, ARR SUSPEND, LEADS FAIL, PROBE
OFF.
When finished, click Apply at the lower right corner of the screen to save the
Telemetry Alarm Control Defaults setting.
3-46
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Set the Full Disclosure Defaults
Open the Full Disclosure Defaults Screen
Click on the Full Disclosure Defaults tab as shown.
Report Duration
This option allows you to designate how much data is included in the report. The
maximum report duration is 72 hours, depending upon licensing.
To set the report duration, place the cursor on the scroll bar below the Report
Duration display field. Move the scroll bar to the left for shorter duration or to the
right for longer duration.
Report Hole Location
This option allows space for binding printed reports. Options are none, top, bottom,
left, and right.
Report Include
This option allows you to enable print characteristics. Options include Graybar,
Arrhythmia Annotations, and Heart Rate. You may select all or none of these
options.
Revision C
„
In Graybar, every other line of the report appears on a shaded background to
provide visual differentiation from other lines.
„
In Arrhythmia Annotations, the name of an applicable arrhythmia call appears
underneath its occurrence in the report.
„
In Heart Rate, the last active heart rate included in the report appears at the end
of the report line.
CIC Pro Clinical Information Center
2001099-156
3-47
Installation: CIC Pro Application Configuration
Report Line Time
This option allows you to designate how much data shows on an individual report
line. Choices are 15sec, 30sec, and 1min.
Strip Duration
This option allows you to designate how much data is included in the strip. The
maximum strip duration is 60 minutes.
To set the strip duration, place the cursor on the scroll bar below the Strip Duration
display field. Move the scroll bar to the left for shorter duration or to the right for
longer duration.
Strip Hole Location
This option allows space for binding printed report strips. Options are none, top,
bottom, left, and right.
Start Data Storage
NOTE
You must be in the service mode to modify this setting.
This option allows you to designate how full disclosure is enabled for patients at the
time of admission. Choices are automatically for all beds, automatically if listed,
and manually.
Bed List
NOTE
You must be in the service mode to modify this setting. See “Access Service Mode”
on page 3-31.
This option shows a listing of beds for which full disclosure data is automatically
stored.
Unit License Default: Full Disclosure License Type
NOTE
You must be in the service mode to modify this setting. See “Access Service Mode”
on page 3-31.
This option shows a list of Full Disclosure license options. Choices are none, 24
hours, 48 hours, and 72 hours.
NOTE
If the default does NOT match the actual license, Full Disclosure does NOT
work.
3-48
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Offline Storage
Offline storage exists as a precautionary measure when contact with a monitor has
been lost. The feature is applied most often when the monitor is being used in the
Rover application, since the monitor “roves” from room to room, on and off the
network.
If a discharge/admit cycle occurs off the network for a bed that is being full
disclosed, the full disclosure sub-system is unaware that the patient has changed.
When the bed returns to the network, full disclosure will present the data from the
discharged and admitted patients as part of the more recent patient’s data. Offline
storage helps to prevent this by assuming that the patient has been discharged if the
monitor is off the network for greater than the number of minutes selected with the
Offline Storage control.
WARNING
POTENTIAL DATA LOSS — Do not allow a NO COMM
(patient offline) event to exceed the time limit selected in the
Offline Storage setting. The patient’s full disclosure data is
deleted if the time limit is exceeded.
NOTE
You must be in the service mode to modify the Offline Storage setting. See “Access
Service Mode” on page 3-31.
To modify an offline storage time period, follow these steps.
1.
Select the Full Disclosure Defaults tab.
2.
Click the Offline Storage field. A down arrow icon is displayed next to the
current offline time period.
3.
Click on the down arrow to display a list of time periods and select the desired
time period. Choices are 30 mins, 1, 2, 4, 8, and 12 hours.
4.
A prompt appears. Select Yes if you are sure you want to change the offline
storage time. The offline storage menu closes, with the selected time remaining
visible.
Restore
This button allows you to clear any modifications made to the Full Disclosure tab
sheet and revert to the previous settings.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-49
Installation: CIC Pro Application Configuration
Set the Display Format
Open the Display Format Screen
“Map” of new
display format
Click on the Display Format tab as shown.
The Display Format tab sheet allows you to format the clinical information center’s
multiple patient viewer with the required number of patient windows.
NOTE
Modifications to display format are subject to licensing restrictions. Licensing
determines the number of beds displayed on the CIC Pro screen. Modifying the
display to show more than the licensed number of beds will result in blank slots
in the display.
Columns
This option allows you to designate the number of columns of patient windows in
the multiple patient viewer.
Rows
This option allows you to designate the number of rows of patient windows in the
multiple patient viewer.
Map of new display
This option reflects the current Columns/Rows selections graphically. When the
Columns or Rows selections change, this map is immediately updated.
3-50
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Auto Display Button
This option allows you to set the function of the Auto Display button in the Main
Menu, or to disable it. Choices are Maximize Waveform Length, Maximize Number
of Waveforms, and Disable Auto Display Button.
Changing the Display Format
To change the display format, follow these steps.
NOTE
You must remove admitted beds from the display before you can select a
display format that would eliminate those patient slots from the display.
To do this, right-click the mouse pointer in the appropriate slot, and choose
None from the Select Care Unit then Bed Number menu.
NOTE
An alarm will sound when removing admitted beds from the display if the beds
are not viewed on another CIC Pro.
1.
Click on the selection corresponding to the number of columns you want in the
multiple patient viewer.
2.
Click on the selection corresponding to the number of rows you want in the
multiple patient viewer.
3.
Visually verify in the map of the new display that the new setting is what you
want.
When finished, click Apply at the lower right corner of the screen to save the
settings.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-51
Installation: CIC Pro Application Configuration
Set the Current Telemetry Listings
NOTE
Current telemetry listings should only be done after the CIC Pro name and
computer name have been configured.
Open the Current Telemetry Listings Screen
Click on the Current Telemetry Listings tab as shown.
In user mode, the Current Telemetry Listings tab sheet is view-only. You must be in
the service mode to make modifications to the telemetry settings. See “Access
Service Mode” on page 3-31.
3-52
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Admitted Telemetry Patients
The Admitted Telemetry Patients control is for display only.
NOTE
This overview covers the controls in a row. Each row contains information for
one telemetry patient (sorted by TTX number).
The second line of an entry shows the current software level for the patient bed
in question.
„
Tower—indicates which telemetry receiver cabinet (tower) this telemetry
patient is communicating with.
„
Receiver—indicates which receiver assembly inside the telemetry receiver
cabinet this telemetry patient is communicating with.
NOTE
The following information only applies to CDT-LAN with software
versions prior to 6D, it does not apply to CDT-LAN software version 6D or
ApexPro.
„
Unit|Bed—indicates the unit and bed to which this patient has been assigned.
„
Type —indicates the type of patient this is: Tele Bed or Tele Combo.
„
Transmitter—indicates the identification number assigned to this patient’s
transmitter.
When finished, click Apply at the lower right corner of the screen to save the setting.
Bed and Transmitters
The Bed and Transmitters controls can only be modified from within service mode.
„
Telemetry Beds—allows you to add, modify, or delete a telemetry bed name.
„
Hardwire Beds—allows you to add, modify, or delete a hardwire bed name.
„
Transmitters—allows you to add, modify, or delete a telemetry transmitter.
Enter a New Item
Enter a new item in a list as follows:
1.
Position the cursor in the blank box at the top of the list.
2.
Type in the desired information and press Enter to add the entry to the list.
3.
When selections are complete, click Apply at the lower right corner of the
screen to save the new settings.
Editing Entries
Edit existing entries as follows:
Revision C
1.
Locate the item in the scroll list.
2.
When you click on the entry, it appears in the editing box.
3.
Edit the information and press Enter to update the entry in the list.
CIC Pro Clinical Information Center
2001099-156
3-53
Installation: CIC Pro Application Configuration
4.
When selections are complete, click Apply at the lower right corner of the
screen to save the new settings.
NOTE
Telemetry beds are distinguished from monitoring beds by an asterisk at the
end of the bed number. When entering or editing information in the
Telemetry Beds list, be sure to add an asterisk to the end of the bed name.
Lock or Unlock Beds
In the Service Mode, the clinical information center can be configured with the bed
names either in locked or unlocked mode. When locked, the bed name is
permanently allocated to a particular slot on the CIC Pro and users are unable to
move the bed to another slot.
NOTE
It is possible to admit a patient to a window with a bed name that is locked to
NONE. To avoid duplication of patient waveforms, a window locked as NONE
should not be used to admit a patient.
To lock or unlock a bed:
1.
Position the mouse pointer in the patient’s waveform window.
2.
Press the right mouse button to open the right-click menu.
3.
Check either LOCK or UNLOCK. The right-click menu closes, and the change
takes effect immediately.
Change CIC Pro Language
WARNING
LOSS OF MONITORING — You must commit (save) the
following configuration changes to the CIC Pro write filter:
‹ IP address
‹ Computer name
‹ Optional licenses
‹ Language
If a hard drive failure occurs, the CIC Pro discards uncommitted
(unsaved) configuration changes. During the reboot process, the
CIC Pro uses the last committed (saved) configuration settings,
which may be different from the settings the CIC Pro was using
before the hard disk failed.
See “Commit Changes to the Write Filter” on page 3-57.
3-54
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
The language configuration utility allows you to configure the display language for
the CIC Pro application.
NOTE
You must log in to the CIC Pro as “administrator” before you can change the
CIC Pro language. See “System Logons” on page 2-15.
To change the CIC Pro language:
1.
If necessary, connect and login to Webmin locally as described in “Local
Webmin Access from the CIC Pro Computer” on page 2-17 and “Webmin
Login” on page 2-18.
2.
Click the Applications tab.
3.
Click the CIC Configuration link. The CIC Configuration window displays.
028A
4.
Click the Languages button. The Languages window displays.
030A
5.
Select the language for the installation location from the drop-down list.
6.
Click Submit.
7.
Select Start > Shut Down, then select Restart.
8.
Click OK. Wait for the log on screen.
9.
Allow the CIC Pro to restart normally in user mode.
10. Complete any other pending CIC Pro configuration procedures and then
complete “Commit Changes to the Write Filter” on page 3-57.
Revision C
CIC Pro Clinical Information Center
2001099-156
3-55
Installation: CIC Pro Application Configuration
Install Optional Licenses
WARNING
LOSS OF MONITORING — You must commit (save) the
following configuration changes to the CIC Pro write filter:
‹ IP address
‹ Computer name
‹ Optional licenses
‹ Language
If a hard drive failure occurs, the CIC Pro discards uncommitted
(unsaved) configuration changes. During the reboot process, the
CIC Pro uses the last committed (saved) configuration settings,
which may be different from the settings the CIC Pro was using
before the hard disk failed.
See “Commit Changes to the Write Filter” on page 3-57.
Install any optional application licenses at this time. Refer to the GE “Unity
Network® Information Suite (IS) Server License Configuration Guide”.
CIC Pro Full Disclosure License Management Setup
Every CIC Pro with version 4.1.x software has the ability to store 16 beds of Full
Disclosure (FD) data for one hour without a license. A CIC Pro with the Full
Disclosure option and version 4.1.x software has the ability to store up to 16 beds for
72 hours with the appropriate licenses and license settings.
When operating two CIC Pros within the same care unit, with one having the Full
Disclosure option, and the other without: it is necessary to turn full disclosure OFF
on the CIC Pro without the full disclosure option, to prevent a conflict between the
two different storage capacities of the CIC Pros.
To set turn full disclosure OFF, set the full disclosure License type to none. See
“Unit License Default: Full Disclosure License Type” on page 3-48.
Commit License Changes to the Write Filter
Complete any other pending CIC Pro configuration procedures and then complete
“Commit Changes to the Write Filter” on page 3-57.
3-56
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
Commit Changes to the Write Filter
After all CIC Pro configuration is complete, commit (save) all the configuration
changes to the write filter.
WARNING
LOSS OF MONITORING — You must commit (save) the
following configuration changes to the CIC Pro write filter:
‹ IP address
‹ Computer name
‹ Optional licenses
‹ Language
If a hard drive failure occurs, the CIC Pro discards uncommitted
(unsaved) configuration changes. During the reboot process, the
CIC Pro uses the last committed (saved) configuration settings,
which may be different from the settings the CIC Pro was using
before the hard disk failed.
See “Commit Changes to the Write Filter” on page 3-57.
NOTE
You should commit (save) the following configuration changes to the CIC Pro
write filter:
‹ Dual display
‹
Touch screen display
‹
Printer installation
If a hard drive failure occurs, the CIC Pro does not save these uncommitted
(unsaved) configuration changes. After the CIC Pro reboots, you must
reconfigure these settings.
See “Commit Changes to the Write Filter” on page 3-57.
Commit changes to the write filter as follows:
Revision C
1.
If you are not already logged into the CIC Pro as ”administrator”, stop the CIC
Pro, shut down the CIC Pro and log back in as “administrator” according to
“Access Service Mode” on page 3-31
2.
Double-click the Internet Explorer icon on the Windows desktop. Access and
login to Webmin locally as described in “Webmin Overview” on page 2-17.
3.
Click the System tab.
4.
Click the Write Filter link. The Write Filter Commands window displays.
CIC Pro Clinical Information Center
2001099-156
3-57
Installation: CIC Pro Application Configuration
5.
Click the Commit Changes button.
130A
6.
Restart the CIC Pro according to “Safe Shutdown/Restart Procedure” on
page 4-4.
Time and Date Configuration
NOTE
Print out all pertinent time-sensitive admitted patient information before
adjusting the time and date settings.
CAUTION
Disconnect the CIC Pro from the Unity MC network before
setting the time/date.
Changes to the time/date data affects all units on the network. This
can result in other monitors altering the time and date parameter of
some patient data.
3-58
1.
Disconnect the Unity Network IX network connection from the back of the CIC
Pro. See “You must restart the CIC Pro for the new IP Address settings to take
effect.” on page 3-22.
2.
Select Setup CIC at the bottom of the screen.
3.
From the CIC Setup screen, select the Service Password tab.
4.
Type mms_com at the password prompt. Press Enter. Verify the Windows
command prompt window opens and the title bar is blue, indicating it is active.
If not, place your cursor inside the prompt window and click to activate it.
CIC Pro Clinical Information Center
2001099-156
Revision C
Installation: CIC Pro Application Configuration
NOTE
Software version 4.1.x shown in the following steps is used for this
example only. The version number displayed in the command prompt
should match the version found in the lower left corner of the CIC Setup
screen.
5.
Verify C:\Program Files\Marquette\CIC\4.1.x> appears, indicating the current
working directory. If not, type the following:
cd\Program Files\Marquette\CIC\<your software version> and press Enter.
6.
At the command prompt type su<space>cicadm<space>cicadm1,3,5,7 and
press Enter.
7.
Type timedate and press Enter. The Date and Time Properties window
displays.
8.
Select the Date & Time tab. Set the correct date and time. Click Apply.
9.
Select the Time Zone tab, and verify the correct time zone for the installation
site is selected.
10. Verify the Automatically adjust clock for daylight saving changes box is NOT
checked.
NOTE
DO NOT select the Automatically adjust clock for daylight saving changes
checkbox. You must manually make Daylight Savings Time changes in the
monitoring system. This change may affect patient data as noted above
Revision C
CIC Pro Clinical Information Center
2001099-156
3-59
Installation: Completion
11. Click Apply.
NOTE
DO NOT make any changes on the Internet Time tab.
12. Click OK to close all windows when finished.
13. Reconnect the Unity Network IX network connection cable to the back of the
CIC Pro. See.“You must restart the CIC Pro for the new IP Address settings to
take effect.” on page 3-22
Completion
Backup CIC Pro Configuration
To simplify system restore if it should become necessary: once the CIC Pro is
installed and configured, backup the CIC Pro configuration to a remote location
according to “Backup-Restore CIC Configuration” on page 4-16.
Exit Service Mode and Proceed to Checkout Procedures
Click on Setup CIC at the bottom of the screen and respond Yes to the exit Service
Mode message. Proceed with the “Checkout Procedures” on page 4-28.
3-60
CIC Pro Clinical Information Center
2001099-156
Revision C
4
Revision C
Maintenance
CIC Pro Clinical Information Center
2001099-156
4-1
For your notes
4-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Schedule
Schedule
Manufacturer Recommendation
WARNING
Failure to implement a satisfactory maintenance schedule may
result in equipment failure and present health hazards. The sole
responsibility for performing the recommended maintenance
schedule rests with the individuals, hospitals, or institutions
utilizing the device. The manufacturer does not in any manner
assume the responsibility for performing the recommended
maintenance schedule, unless an Equipment Maintenance
Agreement exists. Information Technologies service personnel
may, at their discretion, follow the procedures provided in this
manual as a guide during visits to the equipment site.
To make sure the CIC Pro hardware remains in proper operational and functional
order, a proper maintenance schedule must be observed. The following schedule is
recommended:
„
Visual Inspection: This should be performed by service personnel upon receipt
of the equipment, one time per year thereafter, and prior to servicing the unit.
„
Cleaning: This should be performed by service personnel upon receipt of the
equipment, one time per year thereafter, and each time the unit is serviced.
„
Calibration: Screen calibration should be performed by service personnel upon
receipt of the equipment, as directed in “Calibrate the Screen” on page 3-32.
„
Electrical Safety Checks: These should be performed by service personnel
upon receipt of the equipment, one time per year thereafter, and each time the
unit is serviced.
„
Checkout Procedure: This should be performed by qualified service personnel
upon receipt of the equipment, one time per year thereafter, and each time the
unit is serviced.
„
Replacing the Hard Drive: You should replace the hard drive every three
years. Refer to “Hard Drive” on page 6-10 for hard drive replacement
information.
PM Form
For the latest PM form associated with this product, contact GE Medical Systems
Information Technologies Service. Make a copy of the GE Unity Network CIC Pro
PM form and use the copy as a guide as you go through this section of the manual.
The PM form can then be archived for reference after you have finished all the steps
required to completely test the equipment.
If, for any reason, any of the procedures or tests results are not within the indicated
standards, contact GE Medical Systems Information Technologies Technical
Support (see “How to Reach Us...”).
Revision C
CIC Pro Clinical Information Center
2001099-156
4-3
Maintenance: Safe Shutdown/Restart Procedure
Safe Shutdown/Restart Procedure
WARNING
Beds displayed on the CIC Pro will not be monitored while the
unit is shut down. Additionally, beds admitted on a shutdown CIC
Pro will display “NO COMM”, when displayed at other stations.
Proper system shutdown is important. Errors can occur if the system is shut down
incorrectly.
To properly shut down the system:
1.
Select Setup CIC at the bottom of the screen.
2.
Select the Service Password tab.
3.
Type mms_com at the password prompt. Press Enter.
4.
At the Windows command prompt, type stop and press Enter.
5.
Open the Start menu located in the lower left corner of the Windows main
screen.
6.
Select Shut Down from the Start menu.
7.
From the Shut Down Windows dialog you can log off the current user, restart
the CIC Pro or shut it down. Select either Log off, Shut down, Restart, as
required, and click OK.
8.
If you intend to log back on immediately after shutting down, simultaneously
hold down the left Shift key while clicking OK. Continue holding the Shift key
until prompted for logon.
9.
If you intend to shut the CIC Pro down, wait until the message “It is now safe to
turn off your computer”, and turn power off at the switch on the back panel.
CAUTION
EQUIPMENT DAMAGE/DATA LOSS — Turn the CIC Pro
power switch off only when the message “It is now safe to turn off
your computer” is display. Equipment damage or data loss can
occur if this instruction is not followed.
4-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Controlling Electrostatic Discharge Damage
Controlling Electrostatic Discharge Damage
All external connector inputs and outputs of the device are protected from
ElectroStatic Discharge damage (ESD) damage. However, if the interior of the
device needs to be accessed for any reason, internal components and assemblies are
susceptible to ESD damage. This includes human hands, non-ESD protected work
stations and/or improperly grounded test equipment.
The following guidelines help make a service workstation more resistant to ESD
damage:
„
Discharge any static charge you may have built up before handling
semiconductors or assemblies containing semiconductors. This can be done by
touching any bare metal on the CIC Pro chassis, the cable connector jacks or the
ground post on the back of the unit. Do this frequently and repeatedly while
working on the unit.
„
Wear a grounded, antistatic wristband (3M part number 2046 or equivalent) or
heel strap at all times while handling or repairing assemblies containing
semiconductors.
„
Use properly grounded soldering and test equipment.
„
Use a static-free work surface (3M part number 8210 or equivalent) while
handling or working on assemblies containing semiconductors.
„
Keep the work surface free of nonconducting materials such as ordinary plastic
assembly aids and foam packing.
„
Do NOT remove semiconductors or assemblies containing semiconductors
from antistatic containers (Velo-stat bags) until absolutely necessary.
„
Make sure power to an assembly is turned off before removing or inserting a
semiconductor.
„
Do NOT slide semiconductors or electrical/electronic assemblies across any
surface.
„
Do NOT touch semiconductor leads unless absolutely necessary.
„
Semiconductors and electrical/electronic assemblies should be stored only in
antistatic bags or boxes.
These guidelines cannot guaranty a 100% static-free workstation, but greatly reduce
the potential for failure of any electrical/electronic assemblies due to electrostatic
discharge.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-5
Maintenance: Visual Inspection
Visual Inspection
Carefully inspect the CIC Pro and all peripheral components prior to installation,
one time each year thereafter, and each time the equipment is serviced.
Follow these steps to perform a complete visual inspection:
4-6
1.
GENERAL INSPECTION: Inspect the external surfaces of the CIC Pro
components and display monitor for obvious physical damage. Do NOT use the
device if physical damage exists. Refer damaged equipment to qualified service
personnel for repair.
2.
FANS: Inspect all cooling fans in the CIC Pro to make sure they are in good
working order and are not clogged with dust. These fans draw outside air into
the unit to cool internal components.
3.
CONNECTORS: Inspect all external connectors for damaged pins, prongs and
connector housings. Refer damaged equipment to qualified service personnel
for repair. Ensure that all connections are properly seated and secure.
4.
CABLE INSULATION: Inspect all cables, insulation, strain-reliefs and
connectors for damage, cracks or degradation. Refer damaged equipment to
qualified service personnel for repair.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Cleaning
Cleaning
Before Cleaning
1.
Complete the “Safe Shutdown/Restart Procedure” on page 4-4.
2.
Turn off, and unplug, all equipment.
NOTE
When unplugging, pull on the plug, not the cord.
WARNING
In order to remove power from the CIC Pro, disconnect the power
cord from the wall outlet. The on/off button (a convex button) on
the front panel does not disconnect the CIC Pro from AC power.
Additionally, wait 40 seconds after disconnecting the CIC Pro
power cord from the wall outlet. The unit remains energized for a
period of time after shutdown.
External Surfaces
Use a soft, lint-free cloth dampened with clean water to wipe down the external
surfaces of the CIC Pro. Do NOT use aerosol liquid sprays, solvents, or abrasives.
CAUTION
Make sure water does not run into any equipment openings. An
electrical shock hazard exists and/or the equipment can be
damaged if water is allowed to get inside.
Do NOT use aerosol liquid sprays, solvents, or abrasives.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-7
Maintenance: Cleaning
Internal Components
1.
Verify all equipment is turned off and unplugged.
2.
Disconnect all peripheral devices from the CIC Pro.
3.
Remove the CIC Pro cover. Refer to “Complete “Field Replacement
Preparation” on page 6-5.” on page 6-8.
4.
Using a compressed air duster recommended for computers and electronics,
carefully blow out any dust that may have accumulated on internal components.
For component level disassembly/reassembly information, refer to the “Parts Lists,
Drawings and Replacement” Chapter, starting on page 6-1.
Fans
Remove all of the dust from the fans and the fan intakes. The fans and fan intakes
usually accumulate the most dust. Accumulated dust can clog the fans and fan
intakes and prevent proper cooling.
Keyboard
Lightly wipe the keyboard with a clean, soft cloth dampened with water.
Mouse
The mouse may require cleaning if the on-screen cursor no longer smoothly follows
mouse movements.
Optical Mouse
Clean an optical-style mouse with a clean, soft cloth dampened with water.
NOTE
The optical mouse has no field-serviceable moving parts. If cleaning does not
restore function, the mouse may need to be replaced.
Mechanical Mouse
NOTE
Using the mouse on a clean, lint-free surface will help reduce the amount of
cleaning the mouse requires.
4-8
1.
Turn the mouse upside-down and turn the plastic ring on the bottom
counterclockwise to disengage it.
2.
On some mouse devices, you may need to press the plastic ring (rather than turn
it) to disengage it.
3.
Turn the mouse right-side up with one hand and catch the ring and the ball with
your other hand.
4.
Clean the three small rollers inside the mouse with a cotton swab moistened
with water. Rotate the rollers to clean all around them.
5.
Wipe the mouse ball with a clean, soft, dry, and lint-free cloth.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Cleaning
6.
If necessary, wash the mouse ball with warm, soapy water (use a mild soap such
as a dish washing liquid) and then dry the mouse ball thoroughly.
7.
Using a compressed air duster recommended for computers and electronics,
blow into the mouse case to remove any dust that has collected there.
8.
Put the ball and ring back in place.
9.
Test the mouse. If it does not track smoothly, repeat this procedure.
NOTE
It is possible for the mouse to become worn past the point where cleaning can
restore proper tracking function. When this occurs, the mouse should be
replaced.
Display Screen
To clean the display screen, put household glass cleaner on a soft cloth and wipe the
screen.
CAUTION
DO NOT spray the cleaner directly on the display screen. Liquid
getting inside the monitor or CIC Pro could cause damage or
present an electrical shock hazard.
Follow the display monitor’s manufacturer’s recommendations for cleaning the
display screen.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-9
Maintenance: Electrical Safety Tests
Electrical Safety Tests
General
Electrical safety tests provide a method of determining if potential electrical health
hazards to the operator of the device exist.
These tests apply to the server, all peripherals, and their three-prong plugs.
Recommendations
To help you establish a systematic maintenance routine, GE Medical Systems
Information Technologies recommends that you perform all safety tests presented in
this chapter:
„
Upon receipt of the device,
„
Every twelve months thereafter,
„
Each time the main enclosure is disassembled or a circuit board is removed,
tested, repaired, or replaced, and
„
Record the date and results on the “Maintenance Checklist” on page 4-32.
WARNING
Failure to implement a satisfactory maintenance schedule may
cause undue equipment failure and possible health hazards.
Unless you have an Equipment Maintenance Contract, GE
Medical Systems Information Technologies does not in any
manner assume the responsibility for performing the
recommended maintenance procedures. The sole responsibility
rests with the individual or institution using the equipment.
Information Technologies service personnel may, at their
discretion, follow the procedures provided in this manual as a
guide during visits to the equipment site.
Test Conditions
Electrical safety tests may be performed under normal ambient conditions of
temperature, humidity, and pressure.
4-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Electrical Safety Tests
Test Equipment
The manufacturer recommended test equipment required to perform electrical safety
tests is listed below. Equivalent equipment may be substituted as necessary.
Required Tools/Equipment
Item
Specifications
Leakage Current Tester
See figures below
Digital Multimeter (DMM)
resistance, AC voltage
Wall Receptacle Test
Before starting the tests, the wall receptacle from which the device will get electrical
power must be checked. This test checks the condition of the wall receptacle to
ensure correct results from leakage tests.
For international wall receptacles, refer to the internal standards agencies of that
particular country. Use a digital multimeter to ensure the wall receptacle is wired
properly.
If other than normal polarity and ground is indicated, corrective action must be taken
before proceeding. The results of the following tests will be meaningless unless a
properly wired wall receptacle is used.
Ground (Earth) Integrity
Listed below are two methods for checking the ground (earth) integrity, “Ground
Continuity Test” and “Electrical Leakage Tests”. These tests determine whether the
device’s exposed metal and power inlet’s earth (ground) connection has a power
ground fault condition.
Perform the test method below that is required by your Country/Local governing
safety organization.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-11
Maintenance: Electrical Safety Tests
Ground Continuity Test
Completion of this test is checked by the following steps:
1.
Disconnect the DUT (device under test) from the wall receptacle.
2.
Connect the negative (–) lead of the ohm meter to the protective earth terminal
(ground pin in power inlet connector) or the protective earth pin in the MAINS
PLUG (ground pin in power cord). Refer to the US 120VAC power cord figure
below.
Ground
Pin
3.
Set the DMM to measure Ohms.
4.
Connect the positive (+) lead of the Ohm meter to all exposed metal surfaces on
the DUT. If the metal surfaces are anodized or painted scrape off a small area in
a inconspicuous area for the probe to make contact with the metal.
5.
Resistance should read to pass:
‹
0.1 ohm or less to power inlet ground pin
‹
0.2 ohms or less to power cord ground pin
Electrical Leakage Tests
Impedance of Protective Earth Connection
This test, unlike a ground continuity test, will also stress the ground system by using
special ground bond testers e.g. Kikusui (model 872 or TOS 6100) or Associated
Research model HYAMP® Jr. Model 3030D.
This test normally is only required as a manufacturing production test to receive
safety agency compliance UL 609050, CSA 609050-00 (C22.2) and EN60950.
Some country agency’s DO require this test after field equipment repairs e.g.
Germany’s DIN VDE 0751 standards.
Consult your country/local safety agency if there is a question.
Compliance is checked by the following steps:
4-12
1.
A current not less than 10A and not exceeding 25A from a current source with a
frequency of 50 or 60 Hz with a no-load voltage not exceeding 6 V is passed for
at least 5 seconds through the PROTECTIVE EARTH TERMINAL or the
protective earth pin in the MAINS PLUG and each ACCESSIBLE METAL
PART which could become LIVE in case of failure in BASIC INSULATION.
2.
The voltage drop between the parts described is measured and the impedance
determined from the current and voltage drop. It shall not exceed the values
indicated.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Electrical Safety Tests
For EQUIPMENT without a POWER SUPPLY CORD the impedance between the
PROTECTIVE EARTH TERMINAL and any ACCESSIBLE METAL PART
which is PROTECTIVELY EARTHED shall not exceed 0.1 ohms.
For EQUIPMENT with a POWER SUPPLY CORD the impedance between the
protective earth pin in the MAINS PLUG and any ACCESSIBLE METAL PART
which is PROTECTIVELY EARTHED shall not exceed 0.2 ohms.
When taking this measurement, move the customer’s power cord around, no
fluctuations in resistance should be observed.
Ground (Earth) Wire Leakage Current Tests
Perform this test to measure current leakage through the ground (earth) wire of the
equipment during normal operation.
1.
Set the leakage tester switches as follows:
‹
Set the GND switch to GND OPEN.
‹
Set the polarity switch to NORM/FORWARD.
‹
Set the power switch to OFF.
2.
Connect a digital multimeter (DMM) to the METER jacks on the leakage tester.
Set the DMM to measure AC millivolts.
3.
Connect the power cord of the device under test to the power receptacle on the
rear of the leakage tester.
NOTE:
The device under test is to be tested at its normal operating voltage.
Revision C
4.
Set the leakage tester power switch to ON.
5.
Set the power switch of the device under test to ON.
6.
Read the current leakage indicated on DMM. If the reading is greater than 3.5
mA, the device under test fails and should be repaired and tested again.
7.
Set the polarity switch on the leakage tester to RVS (reverse).
8.
Read the current leakage indicated on DMM. If the reading is greater than 3.5
mA, the device under test fails and should be repaired and tested again.
CIC Pro Clinical Information Center
2001099-156
4-13
Maintenance: Electrical Safety Tests
9.
Set the leakage tester power switch to OFF.
NOTE
The MD (measuring device) is the circuitry defined by the appropriate
standard for measuring leakage current.
The measuring devices, defined by various standard organizations (IEC,
UL, etc.), produce almost identical test measurement results.
Enclosure (Chassis) Leakage Current Test
Perform this test to measure current leakage through exposed conductive surfaces on
the device under test during normal operation.
1.
4-14
Set the leakage tester switches as follows:
‹
GND switch - OPEN,
‹
Polarity switch - NORM.
2.
Connect a meter lead between the CHAS connector on the rear of the leakage
tester and an unpainted, non-anodized chassis ground on the unit under test.
3.
Set the leakage tester power switch to ON.
4.
Read the current leakage indicated on DMM. If the reading is greater than 3.5
mA the device under test fails and should be repaired and tested again.
5.
Set the polarity switch to RVS and observe the same meter readings as in the
previous step.
6.
Set the GND switch on the leakage tester to CLOSED.
7.
Read the current leakage indicated on DMM. If the reading is greater than 3.5
mA the device under test fails and should be repaired and tested again.
8.
Set the polarity switch to RVS and observe the same meter readings as in the
previous step.
9.
Set the polarity switch to NORMAL and observe the same meter readings as in
the previous step.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Electrical Safety Tests
10. Set the leakage tester power switch to OFF and remove the meter lead
connected in step 2.
Test Completion
Disconnect all test equipment from the device. Disconnect the device power cord
plug from the leakage tester power receptacle. Disconnect the leakage tester from
the wall receptacle.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-15
Maintenance: Backup-Restore CIC Configuration
Backup-Restore CIC Configuration
NOTE
The backup portion of this section should be performed at regular intervals and/
or whenever changes are made to CIC Pro configuration settings, to ensure
complete system restore if it should become necessary.
Click the Backup-Restore CIC Configuration link to display the following window:
NOTE
See “Data Module Detail” on page 4-24 for a summary of the data contained in
the individual modules displayed in the Available Tools list.
Planning Backup/Restore Strategy
When planning backup/restore strategy in a care unit running more than one CIC
Pro, it’s important to know that when a CIC Pro starts up, it will update some of it’s
configuration data to match any other CIC Pros it finds on the care unit network. The
exact data that is automatically updated in this fashion is shown in “Data Module
Detail” on page 4-24. In general, this would be data common to the care unit that
should be consistent between all the CIC Pros within that care unit. All other
configuration data is specific to the local CIC Pro only, and is not shared between
the CIC Pros in the care unit.
To follow are examples of CIC Pro use cases, demonstrating how this automatic
update impacts the backup/restoration process:
Restore in Care Unit Running One CIC Pro
Automatic update does not occur, and is not a factor in this situation: all
configuration data written to the CIC Pro during the restoration process remains
intact, since no other CIC Pros exist to trigger an automatic update.
4-16
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Backup-Restore CIC Configuration
Adding/Upgrading a CIC Pro in Care Unit Running Multiple CIC Pros
When a CIC Pro is added to, or upgraded, in a care unit already running other CIC
Pros, configuration data common to the care unit is changed in the new/upgraded
CIC Pro, when it starts up, to match the other CIC Pros found in the care unit
network. Configuration data specific only to the new/upgraded CIC Pro does not
change.
Restore Previous Configuration Data in Care Unit Running Multiple CIC Pros
In certain cases, it may be necessary to revert the CIC Pros in a care unit to back to a
past configuration.
In order to do this, it is necessary to:
1.
Shut down all but one of the CIC Pros in the care unit. This prevents the
restore from being overwritten by associated data from other CIC Pros
running on the care unit network.
NOTE
The CIC Pro application must be stopped in order to perform a restore.
Because configuration data is automatically updated to match other
CIC Pros found running in the care unit. If other CIC Pros are left
running, when the restored CIC Pro application starts back up, some
configuration data would be overwritten by the automatic update.
Revision C
2.
Stop the CIC Pro application on the one CIC Pro left running, and perform
the restore procedure. Restart the CIC Pro application.
3.
Start up the other CIC Pros, one at a time. Common configuration data
associated with the care unit is updated with data from the restored CIC
Pro as the other CIC Pros start back up.
CIC Pro Clinical Information Center
2001099-156
4-17
Maintenance: Backup-Restore CIC Configuration
Backup
1.
Backup must be done from a remote computer. Access the Webmin application
remotely according to “Remote Webmin Access from an Intranet Computer” on
page 2-17.
2.
Log in to the Webmin application according to “Webmin Login” on page 2-18.
3.
Click the Applications tab.
4.
Click the Backup-Restore CIC Config link.
5.
Select the modules from the Available Tools list you want to backup. Click
Unselect All to deselect selected files.
NOTE
Ctrl + click to select multiple modules. Click Unselect All to deselect
modules in the list.
6.
4-18
Click Backup. The selected files are temporarily written to the CIC Pro and are
prepared for downloading to your local computer.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Backup-Restore CIC Configuration
7.
Click Show Log to view a summary of the backup configuration file building
process.
8.
Click the cic.cfd link to begin the download process. A file download window
displays.
NOTE
Depending upon the operating system of your local computer, the
following download windows may be vary slightly in appearance.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-19
Maintenance: Backup-Restore CIC Configuration
9.
Click Save. The Save As window displays. Navigate to a directory on your local
computer where you want to save the downloaded configuration file.
10. Click Save. After the file is downloaded, the Download complete window
displays. Click Close to close the Download complete window.
Restore
1.
Shut down all of the CIC Pros in the care unit.
NOTE
The CIC Pro application must be stopped in order to perform a restore.
Because configuration data is automatically updated to match other CIC
Pros found running on the care unit, if other CIC Pros are left running,
when the restored CIC Pro application starts back up, some configuration
data would be overwritten by the automatic update.
2.
Access the Webmin application remotely according to “Remote Webmin
Access from an Intranet Computer” on page 2-17.
NOTE
Access the Webmin on the CIC Pro you want to restore.
4-20
3.
Log in to the Webmin application according to “Webmin Login” on page 2-18.
4.
Click the Applications tab.
5.
Click the Backup-Restore CIC Config link.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Backup-Restore CIC Configuration
6.
Select the modules from the Available Tools list you want to restore.
NOTE
Modules selected from the list should match the modules that were selected
when the configuration was backed up during “Backup” on page 4-18.
Modules selected for restore, not selected during backup, will not contain
any data.
For accurate information about the results of a restore procedure, view the
restore log, as directed in step 12.
NOTE
Ctrl + click to select multiple items from the lists. Click Unselect All to
deselect items in the lists.
Revision C
7.
Click Restore. The Upload CFD file window displays.
8.
Click Browse. The Choose file window displays.
CIC Pro Clinical Information Center
2001099-156
4-21
Maintenance: Backup-Restore CIC Configuration
NOTE
Depending upon the operating system used to run the browser, the
following download windows may be slightly different.
9.
Navigate to the previously downloaded configuration file. Select the file and
click Open. The path to the file displays in the Upload CFD file window.
10. Click Continue. The configuration file is uploaded to the ConfigTool directory
on the CIC Pro.
11. Click Restore to complete the configuration restore process. Progress of the
restoration process is displayed.
4-22
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Backup-Restore CIC Configuration
NOTE
The CIC Pro application must be stopped before the configuration settings can
be restored.
12. Click Show Log to view a summary of the completed configuration restore
process.
13. Restart the CIC Pro application on the CIC Pro just restored.
14. Start up the other CIC Pros, one at a time. Common configuration data
associated with the care unit is updated with data from the restored CIC Pro as
the individual CIC Pros start back up.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-23
Maintenance: Backup-Restore CIC Configuration
Data Module Detail
The following table lists summary detail information associated with data modules
listed as Available Tools in the Backup-Restore CIC Configuration window.
NOTE
Items marked with * are automatically updated by other CIC Pros running in a
care unit.
Backup-Restore CIC Configuration Data Module Detail
Data backed up/Restored
Data Source
Telemetry Alarm Control Defaults
Parameter Limits*
Limits.def/Limits.mei
Parameter Alarm Levels*
Param.def/Param.mei
Arrhythmia Alarm Levels*
Arrhy.def/Arrhy.mei
System Alarm Levels*
System.def/System.mei
Full Disclosure Defaults
Report Duration*
NewTelem.def
Report Hole Location*
NewTelem.def
Include Graybar*
NewTelem.def
Include Arrhythmia Annotations*
NewTelem.def
Include Heart rate*
NewTelem.def
Line Time*
NewTelem.def
Strip Duration*
NewTelem.def
Strip Hole Location*
NewTelem.def
Full Disclosure License*
NewTelem.def
Full Disclosure Offline Storage*
NewTelem.def
Start Data Storage*
NewTelem.def
Bedlist*
Fd.all
Display Format
4-24
Columns
Config.dat
Rows
Config.dat
Disable Auto Display button
Config.dat
Maximize Waveform length
Config.dat
Maximize No. of Waveforms
Config.dat
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Backup-Restore CIC Configuration
Backup-Restore CIC Configuration Data Module Detail
Data backed up/Restored
Data Source
Current Telemetry Listings
Telemetry Bed List*
Telebeds.all
Hardwire Bed List*
Hardwire.all
Transmitter List*
Ttx.all
CIC Defaults
Revision C
Central Name
Config.dat
Unit Name
Config.dat
ECG1WaveForm
Config.dat
Waveform2
Config.dat
Waveform3
Config.dat
Waveform4
Config.dat
Laser Printer
Config.dat
DDW Printer
Config.dat
FullDisclosure Printer
Config.dat
Mirror Central Display
Config.dat
Alarm Volume
Config.dat
Browser Enabled
Config.dat
Browser Status
Config.dat
ColorSetECG0
Config.dat
ColorSetECG1
Config.dat
ColorSetECG2
Config.dat
ColorSetECG3
Config.dat
ColorSetECG4
Config.dat
ColorSetECG5
Config.dat
ColorSetECG6
Config.dat
ColorSetART
Config.dat
ColorSetPA
Config.dat
ColorSetFEM
Config.dat
ColorSetCVP
Config.dat
ColorSetRA
Config.dat
ColorSetLA
Config.dat
CIC Pro Clinical Information Center
2001099-156
4-25
Maintenance: Backup-Restore CIC Configuration
Backup-Restore CIC Configuration Data Module Detail
Data backed up/Restored
Data Source
ColorSetICP
Config.dat
ColorSetSP
Config.dat
ColorSetUAC
Config.dat
ColorSetUVC
Config.dat
ColorSetRESP
Config.dat
ColorSetSPO2
Config.dat
ColorSetCO2
Config.dat
ColorOption
Config.dat
Central Name
Config.data
Telemetry Unit Defaults
4-26
Graph Set Manual
Config.dat
Graph Setup Alarm
Config.dat
Graph Setup Print Window
Config.dat
Graph Setup ECG1*
Telem.def
Graph Setup Waveform2 *
Telem.def
Graph Setup Waveform3*
Telem.def
Graph Setup Waveform4*
Telem.def
Transmitter Graph*
Telem.def
Alarm Graph*
NewTelem.def
Nurse Call Graph*
NewTelem.def
Display Lead*
Telem.def
Arrhythymia*
Telem.def
Lead Analysis*
Telem.def
ST Analysis*
Telem.def
Va Lead*
NewTelem.def
Vb Lead*
NewTelem.def
Detect pace*
Telem.def
Patient Age*
Telem.def
Transmitter Alarm pause*
Telem.def
Alarm Pause Breakthrough*
NewTelem.def
PDT*
NewTelem.def
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Backup-Restore CIC Configuration
Backup-Restore CIC Configuration Data Module Detail
Data backed up/Restored
Transmitter Nurse Call*
Data Source
NewTelem.def
Bed Slot Assignment
Display Format
Config.dat
Beds Configured
Config.dat
Slot Information (From Slot 0 – Slot 32)
Revision C
Row, Column and Bed Name
Config.dat
Lead_or_site, position, scroll_speed
Config.dat
Colors
Config.dat
Bed Assignment
Config.dat
Lock Status
Config.dat
CIC Pro Clinical Information Center
2001099-156
4-27
Maintenance: Checkout Procedures
Checkout Procedures
About the Checkout Procedures
The checkout procedure process provides a method of verifying operational and
functional performance without having to disassemble the unit. Failure to attain the
prescribed results indicates a need for calibration, configuration, or repair of the
equipment.
Perform the checkout procedures upon receipt of the equipment, every year
thereafter, and each time a component is removed or replaced.
The checkout procedure process is based on the assumption that the equipment
being tested is using known, good cables. It also requires that the user be familiar
with the operation of the system. For more information concerning the operation of
these components, refer to the CIC Pro operator manual.
Perform all of the checkout procedures included in this section to completely test the
system for proper function and operation.
NOTE
Close all files before starting the checkout procedures.
General
1.
Check the keyboard to make sure all of the keys work properly and do not stick
when pressed. If any keys stick, or are otherwise non-functional, replace the
keyboard.
2.
Check the mouse to make sure the cursor follows mouse movements smoothly
and the mouse buttons function properly. Refer to “Cleaning” on page 4-7 if
necessary.
NOTE
Refer to the operating instructions packaged with the display monitor to
complete the following step.
3.
Check the monitor display screen, and if necessary, adjust for:
‹
‹
4-28
proper contrast and color (adjust for best viewing in the environment where
the monitor is used)
proper screen focus
‹
correct horizontal and vertical linearity
‹
correct horizontal and vertical position
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Checkout Procedures
Disk Check
NOTE
The CIC Pro application must be shut down before running disk check. If the
disk drive is in use, a message asks if you want to reschedule the disk check for
the next time the system is restarted. Selecting this option will cause disk check
to occur during the next restart.
Your disk drive will not be available to run other tasks while disk check runs.
WARNING
Beds displayed on the CIC Pro will not be monitored while the
CIC Pro application is shut down. Additionally, beds admitted on
a shutdown CIC Pro will display “NO COMM”, when displayed
at other stations.
To run disk check:
1.
Log off the current CIC Pro user and log back on as administrator, according to
“System Logons” on page 2-15.
2.
Using My Computer or the Windows Explorer application, select the disk drive
you want to check.
3.
On the File menu, click Properties.
4.
Select the Tools tab.
5.
Under Error-checking, click Check Now.
6.
Select both disk check options and click Start.
Audio Verification Test
WARNING
CIC Pro audible alarms will not sound for patients with bedside
monitoring devices configured to “Operating Room” mode.
NOTE
Active alarms override this test.
To test audio output:
Revision C
1.
Select the Patient View window.
2.
Select the Alarm Control tab.
3.
Select the Alarm Help button.
4.
Select the Advisory Alarm button.
5.
Verify the audio tone plays through both speakers.
6.
Select the Close button in the lower right corner of the screen.
CIC Pro Clinical Information Center
2001099-156
4-29
Maintenance: Checkout Procedures
Verify Access to All Other Care Units
1.
Select the Patient View window.
2.
Select the View Other button.
3.
Verify that the current list is complete.
Verify Full Disclosure
Verify that all admitted beds in the care unit are full disclosed with the correct
license type.
1.
Select Setup CIC > Service Password and type mms_com.
2.
At the Windows command prompt, type fdcmd -lu and press the Enter key.
The utility may take several seconds to execute (allow up to two minutes after
boot up or after changing the Full Disclosure License type).
3.
If the Full Disclosure Licenses type has been changed, verify that all the beds
are being full disclosed, and that the correct license type is used for the admitted
beds. Refer to the figure below for an explanation of the fdcmd -lu command.
r
n
o
p
q
t
s
4-30
n
Up to 16 FD slots per CIC Pro.
o
Shows Unit|Bed name of the bed being Full Disclosed.
p
Identifies that a license was pulled from FlexLM and matches the Full
Disclosure license type. 1hr is displayed if no licenses and Full Disclosure
License type is set to NONE.
q
Displays the maximum amount of time that data is available in the
database.
r
Identifies who is the FD Master within the care unit.
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Checkout Procedures
s
Shows earliest and latest time of data collected. If the date is shown as 01/
01/70, then data is not being collected. Run fdcmd –lu command again to
view changes.
t
Shows the amount of data collected thus far. In this example, only 3 hours
of data is collected and it will max out at 76 hours.
Verify Each CIC Pro Can Print a Full Disclosure Report
1.
View an admitted patient from the single viewer screen.
2.
Select the Full Disclosure tab and verify waveforms appear.
NOTE
Full Disclosure data will not appear for approximately 2-minutes for newly
admitted patients.
3.
Select Print Report and modify the report for a one-page report, and select print.
4.
Verify the Full Disclosure report printed out at the printer
5.
Repeat the previous step for every CIC Pro within the unit.
CIC Service Tool
Perform all the service functions included in the CIC Service Tool. See “Service
Tools” on page D-3.
Revision C
CIC Pro Clinical Information Center
2001099-156
4-31
Maintenance: Maintenance Checklist
Maintenance Checklist
Use the following checklist to ensure maintenance completion.
“Visual Inspection” on page 4-6
1. ___ General
2. ___ Fans
3. ___ Connectors
4. ___ Cable Insulation
“Cleaning” on page 4-7
1.
___ External Surfaces
2.
___ Internal Components
3. ___ Fans
4. ___ Keyboard
5.
___ Mouse
6. ___ Display Screen
“Electrical Safety Tests” on page 4-10
1. ___ Wall Receptacle Test
2. ___ Ground Continuity Test
3.
___ Ground (Earth) Wire Leakage Current Tests
4.
___ Enclosure (Chassis) Leakage Current Tests
“Backup-Restore CIC Configuration” on page 4-16
1.
___ Backup
“Checkout Procedures” on page 4-28
4-32
1.
___ General
2.
___ Disk Check
3.
___ Audio Verification Test
4.
___ Verify Access to All Other Care Units
5.
___ Verify Full Disclosure
6.
___ General
7.
___ CIC Service Tool
CIC Pro Clinical Information Center
2001099-156
Revision C
Maintenance: Repair Log
Repair Log
A repair log is included for your convenience to record the repair history of the
product.
Unit Serial Number:
Institution Name:
Date
Revision C
Maintenance/Repair
CIC Pro Clinical Information Center
2001099-156
Technician
4-33
Maintenance: Repair Log
Unit Serial Number:
Institution Name:
Date
4-34
Maintenance/Repair
CIC Pro Clinical Information Center
2001099-156
Technician
Revision C
5
Revision C
Troubleshooting
CIC Pro Clinical Information Center
2001099-156
5-1
Troubleshooting:
For your notes
5-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Blank Display, Unit Doesn’t Boot
Blank Display, Unit Doesn’t Boot
1.
Verify the electrical outlet the CIC Pro is plugged into is operating properly.
2.
Verify that all components are connected correctly.
3.
Ensure that all cables are properly seated.
4.
Shutdown, turn off and unplug the CIC Pro, follow all antistatic precautions
according to “Controlling Electrostatic Discharge Damage” on page 4-5,
remove the cover and ensure all internal components are properly seated.
Hardware Component Troubleshooting
NOTE
Also consult any additional documentation, supplied with each component of
the system, for additional troubleshooting information.
Webmin Hardware Monitoring Utilities
CIC Pro Part/Serial Number Utility
The Webmin part/serial number utility displays CIC Pro system ID, serial and part
numbers, and asset information.
To use the part/serial number utility:
1.
Login to Webmin as described in “Webmin Overview” on page 2-17.
2.
Click the System tab.
3.
Click the Asset Management link.
4.
Click the Part No. and Serial Info button.
019A
Revision C
CIC Pro Clinical Information Center
2001099-156
5-3
Troubleshooting: Hardware Component Troubleshooting
5.
CIC Pro part/serial number, system ID and asset information displays.
020A
Diagnostic Logs
The Diagnostics Logs page will have content only if diagnostic utilities were run
previously.
To download diagnostic logs:
1.
Login to Webmin as described in “Webmin Overview” on page 2-17.
2.
Click the System tab.
3.
Click the Diagnostic Logs link.
4.
Right-click the diagnostic log you want to download in the list.
021A
5-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Hardware Component Troubleshooting
5.
Select Save Target As from the pop-up menu. The Save As window displays.
022A
6.
Navigate to a location to save the log file, and type a name for the log, or accept
the default. Click Save
023A
NOTE
Saved logs are written in text file format.
Runtime Diagnostics
Runtime diagnostics utilities provide several tools for viewing/troubleshooting a
variety of system components. See “Webmin Runtime Diagnostics Utilities” on
page C-1 for details about runtime diagnostics utilities.
To access the runtime diagnostics utilities:
Revision C
1.
Login to Webmin as described in “Webmin Overview” on page 2-17.
2.
Click the System tab.
3.
Click the Runtime Diagnostics link.
CIC Pro Clinical Information Center
2001099-156
5-5
Troubleshooting: Hardware Component Troubleshooting
4.
See “Webmin Runtime Diagnostics Utilities” on page C-1 for details about
specific runtime diagnostics utilities.
024A
Fan Runtime Statistics
1.
If necessary, login to Webmin as described in “Webmin Overview” on page 217.
2.
Click the System tab.
3.
Click the Statistics link.
025A
Unity Devices
View a list of Unity devices connected to the Unity Network MC network using the
following procedure:
5-6
1.
If necessary, login to Webmin as described in “Webmin Overview” on page 217.
2.
Click the System tab.
3.
Click the Asset Management link.
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Hardware Component Troubleshooting
4.
Click the Unity Devices button. The Unity Devices window displays.
100A
5.
Select/enter items, as desired, to configure the search.
NOTE
If no selections/entries are made, the search will display a list of all Unity
devices found on the network.
6.
Click the Run Query button.
The information displayed includes a list and summary of all devices
communicating over the Unity Network MC network.
Also see “List Network” on page 5-13.
Printer Configuration Information
The printer configuration information utility allows you to view configuration of
printers connected to the CIC Pro.
To use the printer configuration information utility:
Revision C
1.
If necessary, login to Webmin as described in “Webmin Overview” on page 217.
2.
Click the Applications tab.
3.
Click the CIC Configuration link.
CIC Pro Clinical Information Center
2001099-156
5-7
Troubleshooting: Hardware Component Troubleshooting
4.
Click the Printer Configuration Info button.
028A
5.
The Printer Configuration Info window displays.
029A
5-8
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Software Component Troubleshooting
Software Component Troubleshooting
Webmin Software Monitoring Utilities
Webmin Actions Log
The Webmin Actions Log provides details about Webmin application access and use.
The information can be filtered by user, module and date range.
To access the Webmin Actions Log:
1.
Login to Webmin as described in “Webmin Overview” on page 2-17.
2.
If necessary, click the Webmin tab.
3.
Click the Webmin Actions Log link.
The Webmin Actions Log search settings default to return all records from the
actions log.
017A
To filter/start a search:
Revision C
1.
Select one of the three user filter radio buttons, as desired, to either include or
exclude specific users from the search results. Select from the drop-down lists,
as appropriate, to limit the search to a specific user, or to exclude a specific user
from the search.
2.
Select one of the two module filter radio buttons, as desired, to either include all
the modules in the search, or to limit the search to a specific module. If
appropriate, select the module where you want to limit the search from the
drop-down list.
3.
Select one of the three date filter radio buttons, as desired, include all webmin
access data in the search, to limit the search to the current day only, or to limit
the search to a specific range of days. If appropriate, select the dates from the
drop-down lists to specify the date range for the search.
CIC Pro Clinical Information Center
2001099-156
5-9
Troubleshooting: Software Component Troubleshooting
4.
Once the search filter selections/configurations are made, click Search. The
associated search results display showing the logged action and the associated
module, user, client IP address, and date/time of the action.
018A
CIC Pro Operating System Runtime Statistics
The CIC Pro operating system runtime statistics utility displays the last time the
operating system, and the CIC Pro program, were started, as well as real-time
information about operating system and CIC Pro application status (running / not
running)
1.
If necessary, login to Webmin as described in “Webmin Overview” on page 217.
2.
Click the System tab.
3.
Click the Statistics link.
4.
Click the CIC OS Runtime Stats button.
025A
5-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Software Component Troubleshooting
5.
The CIC OS Runtime Stats window displays.
026A
Availability (CIC Pro Uptime Report)
View the CIC Pro uptime availability and shutdown event history for a given period
of time using the following procedure:
1.
If necessary, login to Webmin as described in “Webmin Overview” on page 217.
2.
Click the Applications tab.
3.
Click the Availability link. The Availability window displays.
4.
Define the time limits for the uptime report data. Select the starting date from
the Start Date drop-down list, and the ending date from the End Date list.
027A
5.
Select either the Yes or No startups and shutdowns radio button:
‹
Yes will include CIC Pro startup/shutdown information in the report
‹
No will exclude this information.
NOTE
If the startups and shutdowns No radio button is selected, daily report, GUI
events, CIC messages information will not be included in the report,
regardless of the Yes/No radio buttons selected for these items.
Revision C
CIC Pro Clinical Information Center
2001099-156
5-11
Troubleshooting: Software Component Troubleshooting
6.
If you selected the Yes radio button in the previous step, select either the Yes or
No radio buttons as desired to either include or exclude the remaining
availability information components (daily report, GUI events, CIC messages).
7.
Click Submit.
Incorrect CIC Pro Application Colors
If the colors in the CIC Pro application are not correct, verify High Color (16-bit) is
the Color Pallet selection in the Settings tab in the Display Properties Control Panel.
For dual display, see “Dual Display Configuration” on page 3-13.
Browser Troubleshooting
A parsing defect in the Microsoft Internet Explorer browser can, in some cases,
prevent access to certain sites. Also, the level of security present on the CIC Pro
server prevents running applications from the Windows desktop taskbar. The
combination of these two factors contribute to loss of browser function under certain
circumstances.
Work-arounds for this are contained in the following sections.
Address Field Entries
In some cases, adding a forward slash (/) to the end of an internet address in the
address field can allow access to a web page.
Specific Web Applications
MUSE Web Frame/No Frame Connection Errors
Button on the MUSE Web server introduction page allow you to select a display
with, or without, frames. These buttons attempt to access a function disallowed by
the CIC Pro.
Add the following suffixes to the MUSE Web address to bypass this problem:
Description
5-12
Suffix
Example
Frame (with
borders)
Museweb.dll?InitializeDefault
Page?Frames=1
http://museserver1/
Museweb.dll?InitializeDefaultPage?Fr
ames=1
No Frame
(without
borders)
Museweb.dll?InitializeDefault
Page?Frames=0
http://museserver1/
Museweb.dll?InitializeDefaultPage?Fr
ames=0
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Software Component Troubleshooting
Statview Administration Configuration Utility Connection
Errors
No known CIC Pro browser connection problems are known to exist associated with
the Statview Administration Configuration Utility. If problems are encountered,
contact GE Medical Systems Information Technologies Technical Support.
Patient Viewer Version 1
The level of security present on the CIC Pro server currently prevents browser
connection to Patient Viewer.
Other Connection Errors
If problems are encountered which are not included in this section, contact GE
Medical Systems Information Technologies Technical Support. Reference
MDOC724.
CIC Service Tool
The CIC Tool has functions useful for the troubleshooting process. See “Service
Tools” on page D-3.
List Network
The list network function provides a list and summary of all devices that are
communicating over the network.
To run the list network function:
1.
Select Setup CIC at the bottom of the screen.
2.
From the CIC Setup screen, select the Service Password tab.
3.
At the service password prompt type mms_com and press Enter. A Windows
command prompt window displays.
4.
Type lw, and the following modifiers, as required:
Example: lw<Space>-d and press Enter.
lw [-d] [-s] [-u] [-a] [-l] [-c] [-q] [-h], where:
-d <device> limits the search to <device> device types
-s <service> limits the search to devices offering <service>
-u <unit> limits the search to devices in <unit>
-a <IP:port> limits the search to devices which have <IP:port>
A “?” within any search string matches any character
-l lists resource matches
-c print a summary count of search results
-q suppress printing of search criteria summary
-h presents help screen
Revision C
CIC Pro Clinical Information Center
2001099-156
5-13
Troubleshooting: Software Component Troubleshooting
A typical listing would include the following data:
IP Address
Device Type
126.0.17.206
TELE TOWER
List Network
Secondary
Address
(2000)
126.0.17.206
126.0.6.77
126.0.6.61
126.0.177.21
TELE BED
TRAMSCOPE
TRAMSCOPE
CENTRAL
(2001)
(2000)
(2000)
(2000)
Care Unit:
Device Name
“SICU:
TOWR7e0011
“SICU:214-A”
“ICU:BED1”
“ICU:BED4”
“ICU:CS4”
Patient Name
“JACKSON”
“HUNTER”
“BLAND”
IP Address
IP addresses uniquely identifies each device on the network, whether the
devices are manufactured by GE Medical Systems Information Technologies,
or are part of the hospital information system.
Device Type
Includes:
CENTRAL: CIC Pro, Centralscope central station
MONITOR: Eagle, Solar, Tramscope monitors
MRT2: Monitoring Review Terminal 2
MUSE: MUSE system
TELE BED: Telemetry bed, CD, or APEX
TELE TOWER: Telemetry cabinet
Secondary Address
Identifies telemetry receivers in the cabinet assembly. Ignoring the base 2000
number, the last digit indicates which receiver is related to each telemetry bed.
Identifies the receivers by numbers in the cabinet assembly.
Care Unit: Device Name
Identifies the device in terms more understandable to the user.
Patient Name
Identifies the currently monitored patient.
5-14
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Software Component Troubleshooting
Duplicate TTX
During normal operation, the CIC Pro software does not allow you to enter duplicate
transmitter TTX numbers when you admit a patient. Many problems could arise if
two transmitters operating at the same frequencies are used in the telemetry system
at the same time.
But there are circumstances where duplicate transmitters are useful, such as during
telemetry system troubleshooting. This situation can occur when multiple CD
Telemetry-LAN cabinets or a telemetry system has a failure. The simplest way to
continue coverage of the patients lost during the failure is to move them to open
receivers elsewhere on the telemetry system.
But if duplicate TTX numbers are not allowed in the system, it is impossible to
admit the transferred patients. The duplicate TTX feature, however, overrides this
limitation, allowing for duplicate TTX numbers. With duplicate TTX enabled, you
can then admit the patients to the open receivers.
Once the option is enabled, the system allows the entry of duplicate TTX numbers
for five minutes. When the five minute period expires, the software again prevents
duplicate TTX number entry.
Admitted patients with duplicated TTX numbers are identified on the CIC Pro
display. When the telemetry malfunction is repaired, the duplicated/transferred
patients are then discharged and restored to the repaired receivers.
The normal state for this option is NOT ALLOWED.
To enable/disable the duplicate TTX function:
1.
Select Setup CIC at the bottom of the screen.
2.
From the CIC Setup screen, select the Service Password tab.
3.
At the service password prompt type mms_com and press Enter.
4.
At the command prompt, type:
setflags<Space>-dup<Space>on and press Enter to allow duplicate
TTX).
setflags<Space>-dup<Space>off and press Enter to turn off duplicate
TTX).
NOTE
Once you select this option, you have five minutes to enter the duplicate
TTX numbers. If more time is necessary, select this option again.
During the five-minute period you can admit duplicate TTX numbers. Admission
information can also be changed during this time.
Revision C
CIC Pro Clinical Information Center
2001099-156
5-15
Troubleshooting: Software Component Troubleshooting
Open RTERM Session
The remote terminal (RTERM) feature is similar to connecting a terminal directly to
a remote device, except that the CIC Pro acts as the terminal and it is connected to
the device through the network.
NOTE
You cannot use this function to log into another CIC Pro. It only allows you to
log into CD Telemetry-LAN Cabinets or Centralscope Central Stations.
WARNING
Using the RTERM session option can cause other central stations
to lose monitoring functions, patient data, or even the operating
software. Only trained service personnel should use this option.
To run the RTERM function:
1.
Select Setup CIC at the bottom of the screen.
2.
From the CIC Setup screen, select the Service Password tab.
3.
At the service password prompt type mms_com and press Enter. A Windows
command prompt window displays.
4.
Type the following command:
rterm “UNIT|CENTRAL”
where UNIT|CENTRAL is the fully specified UNITY Network device name.
5.
Press Enter.
NOTE
Description of this option is beyond the scope of this manual. Specialized
training in service monitor-level troubleshooting is necessary before you use
this option. Consult with GE Medical Systems Information Technologies
Service for information about the advanced training.
5-16
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Event Logs
Event Logs
For more information about CIC Pro logs, and retrieving log files, refer to the
following:
„
“Download Logs (Webmin)” on page 5-19.
„
“View Logs (Webmin)” on page 5-18.
„
“CIC Pro Log File Description” on page 5-17.
„
“getlog Function” on page 5-20.
„
“MUSE Web Frame/No Frame Connection Errors” on page 5-12.
For information about other types of log files, refer to the following:
„
“Diagnostic Logs” on page 5-4.
„
“OS Event Log” on page C-15.
„
“OS System Log” on page C-16.
„
“OS Application Log” on page C-17.
„
“OS Security Log” on page C-18.
CIC Pro Log File Description
Error Logs
CIC Pro records the occurrence of fatal and nonfatal errors in a user-readable error
log.
Connection Logs
CIC Pro records the usage of each available communication channel by logging
when a connection is established and which beds are viewed.
CIC Pro Log File Location/Naming
The CIC Pro log files are located in the following directory:
\Program Files\Marquette\CIC\Log
The naming convention for the log files is as follows:
yyyymmdd.log
For example: 20020422.log is the log file for April 22, 2002. There is one log file
per day.
Revision C
CIC Pro Clinical Information Center
2001099-156
5-17
Troubleshooting: Event Logs
Log File Command Prompt Procedures/Functions
View Logs (Webmin)
1.
Login to Webmin as described in “Webmin Overview” on page 2-17.
2.
Click the Applications tab.
3.
Click the View Logs link.
4.
Select the log files you want to view from the CIC Logs and/or Dr. Watson lists.
Click Unselect All to deselect selected files.
NOTE
Ctrl + click to select multiple items from the lists.
5.
5-18
Click View Selected Files. The selected log files display.
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Event Logs
Download Logs (Webmin)
NOTE
Download Logs only works when connecting from a networked, remote
computer. It will not work locally at the CIC Pro.
1.
Login to Webmin as described in “Webmin Overview” on page 2-17.
2.
Click the Applications tab.
3.
Click the Download Logs link.
NOTE
At the end of each day, log files are compressed into.bfp files. Because of this,
current day log files are uncompressed. Uncompressed files cannot be
downloaded using the Download Logs window. To view and download current
day log files, see “View Logs (Webmin)” on page 5-18.
4.
Revision C
Click a log you want to download. The File Download window displays.
CIC Pro Clinical Information Center
2001099-156
5-19
Troubleshooting: Event Logs
5.
Click to select the Save this file to disk radio button. Click OK. The Save As
window displays.
6.
Navigate to a location to save the log, and type a name for the logs, or accept the
default. Click Save.
NOTE
Saved logs are compressed into a *.zip formatted file.
getlog Function
Once you locate the desired error log, you can copy the log over the network to
another drive. Since error logs are text files, you can copy and read them into other
computers using a text editor or a word processing application.
To retrieve the desired log files, use the getlog function as follows:
Usage: getlog<Space>[-n] [-s] [-d] [-t] [-p]
Function Descriptions
Function
5-20
Description
getlog<Space>-n [UNIT|]BED
Fetches logs from BED (‘UNIT’ is optional, ‘BED’ is mandatory).
getlog<Space>-s YYYYMMDD
Starts on date yyyymmdd as in ‘19940428’ (default is today’s date).
getlog<Space>-d #days
Number of days prior to start date (default is ALL available days).
getlog<Space>-t type
Formats the output to a specific type (default is determined by
device type):
1 - Tramscope Version 6
2 - Tramscope Version 7
3 - Tramscope Version 17
4 - Eagle 3000
5 - Eagle 4000
6 - Eagle Tmp
7 - Type 1TTX:Dropout
getlog<Space>-p [dev:]/path
Saves logs to dev:/path (default outputs logs to this device).
CIC Pro Clinical Information Center
2001099-156
Revision C
Troubleshooting: Canceling CIC Pro Print Jobs
Canceling CIC Pro Print Jobs
Use the following procedure to cancel a print job from a CIC Pro:
1.
Select CIC Setup at the bottom of the screen.
2.
Click on the CIC Defaults tab.
3.
Click the appropriate Cancel Print Jobs button.
NOTE
This only cancels print jobs originating locally from the CIC Pro. You
cannot cancel print jobs originating from other CIC Pros on the network.
Revision C
CIC Pro Clinical Information Center
2001099-156
5-21
Troubleshooting: Canceling CIC Pro Print Jobs
For your notes
5-22
CIC Pro Clinical Information Center
2001099-156
Revision C
6
Revision C
Parts Lists, Drawings
and Replacement
CIC Pro Clinical Information Center
2001099-156
6-1
For your notes
6-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Ordering Parts
Ordering Parts
The parts lists and drawings in this chapter supply enough detail for you to order
parts considered field replaceable. If you require additional information, schematics,
or troubleshooting assistance, contact Tech Support.
To order parts, contact Service Parts at the address or telephone number listed on the
"How to Reach Us...," page found in the front of this manual.
Field Replaceable Units
Item Number
Revision C
Item Description
2020183-002
FRU CIC Pro CPU
2020183-003
FRU CIC Pro Fan
2020183-004
FRU CIC Pro Dual Speakers
2020183-005
FRU CIC Pro Power Supply
2020183-006
CIC Pro 3AG 4A SB Fuses
2020183-007
FRU CIC Pro Hardware
2020183-008
FRU CIC Pro Front Bezel
2020183-009
FRU CIC Pro Light Pipe
2020183-010
FRU CIC Pro Hard Drive
80274-006
CORD PWR 125V 6FT STR
401855-001
PWR SPLY CRD RA CONT Euro 10A 250V 2.5M
401855-002
PWR SPLY CRD RA British 10A 250V 2.5M
401855-003
PWR SPLY CRD RA Italian 10A 250V 2.5M
401855-004
PWR SPLY CRD RA Israeli 10A 250V 2.5M
401855-005
PWR SPLY CRD ST Harness 10A 125V 2M
401855-006
PWR SPLY CRD ST Harness 10A 250V 2M
401855-007
PWR SPLY CRD RA Swiss 10A 250V 2.5M
401855-008
PWR SPLY CRD RA Indian 10A 250V 2.5M
401855-009
PWR SPLY CRD RA Danish 10A 250V 2.5M
401855-010
PWR SPLY CRD RA Australian 10A 250V 2.5M
CIC Pro Clinical Information Center
2001099-156
6-3
Parts Lists, Drawings and Replacement: FRU Installation/Field Replacement Procedures
FRU Installation/Field Replacement Procedures
WARNING
PERSONAL INJURY/EQUIPMENT DAMAGE — FIRST
complete “Field Replacement Preparation” on page 6-5 before
performing any other procedure in this chapter. Failure to follow
this instruction could result in serious injury or product/property
damage.
Field replacement information for the following CIC Pro components are included
in this chapter:
6-4
„
“Fuses” on page 6-6
„
“Cover” on page 6-8
„
“Battery” on page 6-9
„
“Hard Drive” on page 6-10
„
“Fans” on page 6-12
„
“CPU Board” on page 6-13
„
“Power Supply” on page 6-16
„
“Internal Speaker” on page 6-19
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
Orientation
The following figure illustrates front, back, left and right orientation referred to
throughout the field replacement procedures.
Top
Left
Back
Front
Right
Bottom
Field Replacement Preparation
Only a qualified service technician should perform the procedures in this chapter.
Observe these safety guidelines before opening the CIC Pro case.
1.
Turn the CIC Pro off, and unplug the AC power cord from the back of the CIC
Pro and from the wall outlet.
WARNING
The on/off button (a convex button) on the back panel does not
turn off the system AC power. To remove power from the server,
you must unplug the AC power cord from the wall outlet or power
cord connector on the unit.
The CIC Pro remains energized for a period of time after
shutdown. Wait 30 seconds before proceeding.
2.
Turn off all connected peripheral devices.
3.
Label and disconnect all peripheral cables and telecommunication lines
connected to the I/O connectors or ports at back of the server.
4.
Always provide electrostatic discharge (ESD) protection by wearing an
antistatic wrist strap attached to the chassis ground of the system (any unpainted
metal surface) whenever handling CIC Pro internal components.
CAUTION
The CIC Pro is extremely static sensitive and should be handled as
directed in “Controlling Electrostatic Discharge Damage” on
page 4-5.
Revision C
CIC Pro Clinical Information Center
2001099-156
6-5
Parts Lists, Drawings and Replacement: Orientation
Fuses
1.
Turn the CIC Pro off and remove the power cord from the unit.
2.
Insert a flat-bladed screwdriver into the latch at the top of the fuse door and pry
the door open.
047A
3.
Insert a flat-bladed screwdriver under the lip in the fuse holder, pry the fuse
holder loose and slide it out of the unit.
132A
6-6
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
4.
Using gloves, a tissue, or a clean rag between your fingers and the fuses,
remove the fuses from the fuse holder.
NOTE
Do not touch fuses with your bare hands. Oils/acids on your skin can
greatly diminish fuse life.
056A
Revision C
5.
Examine both fuses. Replace any blown or damaged fuses.
6.
Reverse the steps in this procedure to install the fan fuses.
CIC Pro Clinical Information Center
2001099-156
6-7
Parts Lists, Drawings and Replacement: Orientation
Cover
WARNING
SHOCK HAZARD — Perform the “Electrical Safety Tests” on
page 4-10 whenever replacing the cover. Whenever the cover is
removed, the possibility of disrupting internal components is
present, which presents the risk of disruption of proper power/
ground connections. Potential for electrical shock exists if the
proper tests are not conducted after the cover is replaced.
1.
Complete “Field Replacement Preparation” on page 6-5.
2.
Using a phillips screwdriver, remove the eight screws securing the cover on
both sides and lift off the cover.
038A
3.
6-8
To install the cover, reverse the steps in this procedure and complete the
“Electrical Safety Tests” on page 4-10.
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
Battery
1.
Complete “Field Replacement Preparation” on page 6-5
2.
Remove the cover according to “Cover” on page 6-8.
3.
Push the battery latch away from the battery until the battery pops loose and
remove the battery.
052A
4.
Revision C
Reverse the steps in this procedure to install the battery.
CIC Pro Clinical Information Center
2001099-156
6-9
Parts Lists, Drawings and Replacement: Orientation
Hard Drive
1.
If possible, complete the backup procedure in “Backup-Restore CIC
Configuration” on page 4-16 before continuing with this procedure.
2.
Complete “Field Replacement Preparation” on page 6-5
3.
Remove the cover according to “Cover” on page 6-8.
4.
Disconnect the hard drive ribbon cable from the hard drive.
045A
5.
Using a phillips screwdriver, completely loosen the 3 screws securing the hard
drive bracket to the chassis and remove the hard drive and bracket.
032A
6-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
6.
7.
Using a phillips screwdriver, loosen the 4 screws securing the hard drive to the
bracket.
NOTE
Do not remove the screws. It is necessary only to loosen the screws enough to
allow the hard drive to slip out of the bracket, as described in the next step.
050A
8.
Slide the hard drive until the screws clear the holes in the bracket and remove
the hard drive. For reassembly, note the position of any washers on the screws.
053A
9.
Reverse the steps in this procedure to install the hard drive.
10. Perform the CIC Pro V4.1.x reload procedure according to the CIC Pro Clinical
Information Center Software Reload Instructions.
CAUTION
EQUPMENT MALFUNCTION/DATA LOSS — Always perform
both the compact flash and hard drive ghost procedures. Never
perform one without the other. Equipment malfunction and/or data
loss can occur if this instruction is not followed.
11. Complete “CIC Pro Application Configuration” on page 3-31.
Revision C
CIC Pro Clinical Information Center
2001099-156
6-11
Parts Lists, Drawings and Replacement: Orientation
Fans
NOTE
The CIC has 2 fans. References in this procedure to the "appropriate" fan is the
fan to which the procedure is applied.
1.
Complete “Field Replacement Preparation” on page 6-5
2.
Remove the cover according to “Cover” on page 6-8.
3.
Disconnect the appropriate fan power connector from the CPU board.
060A
4.
Using a phillips screwdriver and an adjustable wrench, remove the appropriate
2 screws securing the fan to the back of the chassis and remove the fan.
044A
5.
6-12
Reverse the steps in this procedure to install the fan.
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
CPU Board
1.
Complete “Field Replacement Preparation” on page 6-5
2.
Remove all plugs from the CIC Pro back panel.
3.
Remove the cover according to “Cover” on page 6-8
4.
Disconnect the speaker connector from the back left corner of the CPU board.
035A
5.
Disconnect the light pipe from the front left corner of the CPU board.
062A
Revision C
CIC Pro Clinical Information Center
2001099-156
6-13
Parts Lists, Drawings and Replacement: Orientation
6.
Disconnect the hard drive ribbon cable from the connector at the front left
corner of the CPU board.
054A
7.
Disconnect the power supply cable from the connector at the front right corner
of the CPU board.
NOTE
Pry back the retaining tab to free the connector. It may be necessary to use
a flat-bladed screwdriver to pry back the tab.
064A
8.
6-14
Remove the hard drive according to “Hard Drive” on page 6-10.
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
9.
Using a phillips screwdriver, remove the 10 screws securing the CPU board to
the bottom of the chassis.
034A
10. Using a phillips screwdriver, remove the 2 screws securing the CPU board to
the back of the chassis.
031A
11. Remove the CPU board.
12. Reverse the steps in this procedure to install the CPU board.
13. Perform the CIC Pro V4.1.x reload procedure according to the CIC Pro Clinical
Information Center Software Reload Instructions.
CAUTION
EQUPMENT MALFUNCTION/DATA LOSS — Always perform
both the compact flash and hard drive ghost procedures. Never
perform one without the other. Equipment malfunction and/or data
loss can occur if this instruction is not followed.
14. Complete “Computer Name Configuration” on page 3-23.
Revision C
CIC Pro Clinical Information Center
2001099-156
6-15
Parts Lists, Drawings and Replacement: Orientation
Power Supply
WARNING
SHOCK HAZARD — Never touch components inside the power
supply. There are no field serviceable components inside.
Capacitors in the power supply present a shock hazard even with
the power switched off.
1.
Complete “Field Replacement Preparation” on page 6-5
2.
Remove the cover according to “Cover” on page 6-8.
CAUTION
During CIC Pro operation, the power supply gets hot enough to
cause severe burns. Allow the power supply and insulation shield
to thoroughly cool after shutting down the CIC Pro, before
touching the power supply.
3.
Disconnect the connector at the front of the power supply at the right front of
the chassis.
NOTE
Pry back the retaining tab on the back of the socket to free the connector. It
may be necessary to use a flat-bladed screwdriver to pry back the tab.
051A
6-16
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
4.
Disconnect the connector at the back of the power supply at the right back of the
chassis.
NOTE
Pry back the retaining tab on the back of the socket to free the connector. It
may be necessary to use a flat-bladed screwdriver to pry back the tab.
048A
5.
Using a phillips screwdriver, remove the 4 screws (2 in back and 2 in front)
securing the power supply to the chassis.
057A, 058A
Revision C
CIC Pro Clinical Information Center
2001099-156
6-17
Parts Lists, Drawings and Replacement: Orientation
6.
Remove the power supply.
055A
7.
6-18
Reverse the steps in this procedure to install the power supply.
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: Orientation
Internal Speaker
1.
Complete “Field Replacement Preparation” on page 6-5
2.
Remove the cover according to “Cover” on page 6-8.
3.
Disconnect the speaker connector from the socket at the back left corner of the
CPU board.
035A
4.
Using a phillips screwdriver, completely loosen the 2 screws securing the
speakers to the chassis and remove the speakers.
040A
5.
Revision C
Reverse the steps in this procedure to install the speakers.
CIC Pro Clinical Information Center
2001099-156
6-19
Parts Lists, Drawings and Replacement: CIC Pro Parts List
CIC Pro Parts List
38
15
7
15
22
43
26
42
6-20
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: CIC Pro Parts List
28
27
16
31
32
PART OF ITEM 32
30
Revision C
CIC Pro Clinical Information Center
2001099-156
6-21
Parts Lists, Drawings and Replacement: CIC Pro Parts List
15
9
20
37
2
3
24
10
5
1
11
33
14
15
6
23
6-22
CIC Pro Clinical Information Center
2001099-156
Revision C
Parts Lists, Drawings and Replacement: CIC Pro Parts List
10
25
17
Revision C
CIC Pro Clinical Information Center
2001099-156
6-23
Parts Lists, Drawings and Replacement: CIC Pro Parts List
Item
Number
1
Part Number
2020183-006
Qty
CIC Pro 3AG 4A SB Fuses
N/A
2
PLUG M6 EQUIPOTENTIAL
1
3
WASHER LOCK SERRATED F/M-6
1
5
NUT HEX KEPS M3 - .5 CLASS 8 ZP
1
6
WASHER LOCK M3 EXT TOOTH
2
7
JACKSCREW 4-40X.18 W/VIBRATITE
8
9
CLAMP CABLE
2
10
SCR SL BDGH 6-32 X 3/16 NYLON
4
11
AC INLET MDL CONN/SW/DUAL FZ
1
14
CBL ASM 9500 PWR INLET CHOKE
1
15
SCR MACH PNHD M3X6LG SST W/THD LOCK
19
16
SCR MACH PNHD M3X4LG SST W/THD LOCK
4
17
SCR MACH FLHD M3 X 6MM SS N8.25 GRAY
4
20
CLIP SNAP-LOK 13X4.8 SADDLE
2
22
2020183-002
FRU CIC Pro CPU
N/A
23
2020183-005
FRU CIC Pro Power Supply
N/A
24
CHASSIS CIC PRO
1
25
COVER CIC V4.1
1
26
2020183-004
27
28
FRU CIC Pro Dual Speakers
BRACKET CIC PRO HARD DRIVE MNT
2020183-010
FRU CIC Pro Hard Drive
N/A
1
N/A
30
CABLE ASSY CIC PRO POWER
1
31
CABLE ASSY CIC V4.1 HARD DRIVE
1
32
FRU CIC Pro Light Pipe
N/A
33
CABLE ASSY CIC PRO AC POWER
N/A
34
CABLE ASSY CIC PRO FAN
35
2020183-009
2020183-008
FRU CIC Pro Front Bezel
2
N/A
38
PANEL ASM CIC PRO CPU I/O
1
42
SCR FAN SELF-TAPPING FLHD
4
43
6-24
Description
2020183-003
FRU CIC Pro Fan
2001099-114
MNL OPR CIC V4 ENG
CIC Pro Clinical Information Center
2001099-156
N/A
1
Revision C
Parts Lists, Drawings and Replacement: CIC Pro Parts List
Item
Number
Part Number
Description
Qty
2001099-157
CD SVCE CIC PRO V4.1 ENG
1
2001323-001
KOSS HDM/5 COMPUTER SPEAKERS
1
Optional Components
Item Number
Revision C
Item Description
2022144-002
USB Mouse Optical scroll lighted
XXXXXX-XXX
USB KEYBOARD KIT (see “Keyboard Kits” on page 6-26)
422310-001
MOUSE PAD GE
2006550-001
ADAPTER DB9F TO RJ-45
420650-009
ADAPTER VGA TO DVI-A
2011097-083
KIT SERVICE CICV4.1 SOFTWARE RELOAD
2011097-084
KIT UPGRADE CIC PROV4.1ENG
2011097-085
KIT UPGRADE CIC PROV4.1SWE
901136-(001-005)
4 to 1 keyboard/mouse switch box w/ extension cables ranging from 10 ft. – 100 ft
CIC18CFP
18” NEC color flat panel display
CIC20CFP
20” NEC color flat panel display
FPD18ADT-PS
18” color touchscreen display
901125-001
Video splitter drives up to 3 slaves with either an additional primary display or an
additional VDA
2001099-114
MNL OPR CIC V4 ENG
2001099-157
CD SVCE CIC PRO V4.1 ENG
2001323-001
KOSS HDM/5 COMPUTER SPEAKERS
2022038-001
PWR SPLY UPS 600VA 120 VOLTS
2022038-002
PWR SPLY UPS SMK 600VA 220 VOLTS
2022038-003
PWR SPLY UPS SMK 2000VA 120 VOLTS
CIC Pro Clinical Information Center
2001099-156
6-25
Parts Lists, Drawings and Replacement: CIC Pro Parts List
Keyboard Kits
Item Number
Item Description
Item Number
Item Description
2012217-021
KYBD KIT CIC USB ENG US
2012217-034
KYBD KIT CIC USB WIN95-SWED/FIN
2012217-022
KYBD KIT CIC USB WIN95-BELGIAN
2012217-030
KYBD KIT CIC USB WIN95-SWISS
2012217-023
KYBD KIT CIC USB WIN95-DANISH
2012217-031
KYBD KIT CIC USB WIN95-CZECH
2012217-024
KYBD KIT CIC USB WIN95-ENG EU
2012217-032
KYBD KIT CIC USB WIN95- DUTCH
2012217-025
KYBD KIT CIC USB WIN95-ENG UK
2012217-033
KYBD KIT CIC USB WIN95-RUSSIAN
2012217-026
KYBD KIT CIC USB WIN95-FRENCH
*
KYBD KIT CIC USB WIN95-SLOVENIA
2012217-027
KYBD KIT CIC USB WIN95-GERMAN
*
KYBD KIT CIC USB WIN95-CHINESE
2012217-035
KYBD KIT CIC USB WIN95-ITALIAN
*
KYBD KIT CIC USB WIN95-JAPANESE
2012217-028
KYBD KIT CIC USB WIN95-NORWEGIAN
*
KYBD KIT CIC USB WIN95-POLISH
2012217-036
KYBD KIT CIC USB WIN95-PORTUGUESE
*
KYBD KIT CIC USB WIN95-HUNGARIAN
2012217-029
KYBD KIT CIC USB WIN95-SPANISH
*Contact GE Medical Systems Information Technologies for part number
information.
6-26
CIC Pro Clinical Information Center
2001099-156
Revision C
A
Revision C
Technical Specifications
CIC Pro Clinical Information Center
2001099-156
A-1
For your notes
A-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Technical Specifications: Technical Specifications–CIC Pro
Technical Specifications–CIC Pro
Due to continual product innovation, GE Medical Systems Information
Technologies designs and specifications are subject to change without notice.
Performance Specifications
Performance Specifications
Number of Patients
Display Format
Displayed Wavelengths
Display Information
Full Disclosure Display
1 to 16 in standard display format, 1 to 17 in “view patient” format
User selectable for up to 16 patients per display
3.5 – 11 seconds, dependent on configuration choice
Access to complete patient information including: real-time patient
waveforms & monitored parameters, arrhythmia histories, graphic
trends, tabular vital signs and full disclosure waveforms & parameter
data.
9 waveforms, all parameters. 1 hour standard, licensed storage
includes 24, 48 and 72 hour storage.
Controls
„
USB Compatible Keyboard
„
USB Compatible Mouse
Processing
Main Processor: Intel Plll ULV 933 MHz ULV w/512 kB L2 cache
Printers
„
2” External Digital Writer (PRN 50/PRN 50-M)
„
Laser Printer
Alarms
Alarms
Patient Status Alarms
System Alarms
3 Levels: Warning, Advisory and Message
Notification
Audible, visual, graph runs, and history storage (dependent on
level), alarm display button to view alarming beds
Display of Alarm Information
Alarm information located in waveform window, parameter box
and alarm display buttons
Silencing
Current patient alarms silenced for 1 minute via alarm silence
button or alarm silence key on keyboard. Alarm silence
notification icon in parameter box.
Display of Limits
Revision C
4 Levels: Crisis, Warning, Advisory and Message
All parameter limits available
CIC Pro Clinical Information Center
2001099-156
A-3
Technical Specifications: Technical Specifications–CIC Pro
Recorder (optional)
„
2-inch external Direct Digital Writer (PRN 50/PRN 50-M)
„
Laser printer
General
512 MB RAM; CD-ROM drive, 40 GB hard drive (with or without Full
Disclosure); 1 GB EIDE Compact Flash
Storage
Software Updates
Via network connection
Serial Outputs
RS232 9 pin DB9 serial ports (two)
Parallel Output
IEEE 1284 Standard
USB
USB Type A connectors (four - two reserved for keyboard/mouse)
Network Connectivity
Unity Network MC network--real-time monitoring; Unity Network IX network-Unity information exchange network
Video Output
Onboard high resolution dual video ATI chip video board, 1280 x 1024 @ 75
Hz; 16-bits per pixel, supporting 16.7 million or true 24-bit color. DVI-A
connection.
Audio Output
Integral audio controller; GE Medical Systems Information Technologies
supplied speakers required
Environmental Specifications
Power Requirements
Input Range: 100-240V.
Frequency: 50-60 Hz
Input Current: 4.0 A
Power Consumption
50 watts (max)
5V±5% supply (10 Amps maximum)
+12V +/- 5% (1 Amp maximum)
Fuse Rating
3AG 4A SB/250V~
Cooling
Forced air 28CFM
Operating Environment Specifications
+0°C (+32°F) to +50°C (+122°F)
Rate of change not to exceed 10°C/Hr
Maximum Altitude: 3,000 meters (10,000 ft)
Relative Humidity: 10% -90% RH non-condensing
Non-operating Storage Environment
Specifications
+70°C (+158°F) for 84 hours
-20°C (-4°F) for 4 hours
Maximum Altitude: 12,000 meters (40,000 ft);
Relative Humidity: 10% - 90% RH non-condensing
Storage Pressure
500 kPA to 1060 kPA
Physical Specifications
Physical Specifications
A-4
Height
42 cm (16.6 inches)
Depth
20 cm (7.9 inches)
CIC Pro Clinical Information Center
2001099-156
Revision C
Technical Specifications: Technical Specifications–CIC Pro
Physical Specifications
Width
48 cm (18.9 inches)
Weight
11.35 (25 lbs.)
Display Specifications
Display Specifications
Display Size
Dot Pitch
Resolution
17 to 21 inch diagonal, color CRT
18 or 20 inch color flat panel
.028 mm maximum
1280 x 1024 @ 75 Hz
Electrical
Input Voltage
Range
Controls
Auto-ranging for either
85 – 130 VAC/60 Hz, 1.7 Amps, or
190 – 270 VAC/50 Hz, 1.2 Amps
On/Off, Brightness, Contrast
Environmental
Operating
Temperature
Humidity
10 to 40°C
10 to 90% (non-condensing)
Monitor (Display) Connections
Variety of interconnect video cables available.
Certification
UL/CSA/CE certification as required per the original equipment manufacturer
(OEM).
Revision C
CIC Pro Clinical Information Center
2001099-156
A-5
Technical Specifications: Technical Specifications–CIC Pro
For your notes
A-6
CIC Pro Clinical Information Center
2001099-156
Revision C
B
Revision C
Supported CIC Pro
Parameters
CIC Pro Clinical Information Center
2001099-156
B-1
For your notes
B-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Supported CIC Pro Parameters:
Arterial Blood Gas (POC)
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
PH
--
--
--
--
Yes
Partial Pressure CO2
PCO2
--
--
--
--
Yes
Partial Pressure O2
PO2
--
--
--
--
Yes
Bicarbonate
HC03
--
--
--
--
Yes
Temperature
Temp
--
--
--
--
Yes
Beb
--
--
--
--
Yes
Transcutaneous CO2
TCO2
--
--
--
--
Yes
Base Excess Extracellular Blood
Beeb
--
--
--
--
Yes
O2 Staturation
O2Sat
--
--
--
--
Yes
Hematacrit
HCT
--
--
--
--
Yes
Hemoglobin
Hb
--
--
--
--
Yes
Potassium
Ka
--
--
--
--
Yes
BUN
--
--
--
--
Yes
Sodium
Na
--
--
--
--
Yes
Chloride
Cl
--
--
--
--
Yes
Calcium
Ca
--
--
--
--
Yes
Barometric Pressure
BP
--
--
--
--
Yes
Glucose
Gluc
--
--
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Systolic Pressure
Sys
Yes*
Yes
Yes
Yes
Yes
Diastolic Pressure
Dias
Yes*
Yes
Yes
Yes
Yes
Mean Pressure
Mean
Yes*
Yes
Yes
Yes
Yes
Peripheral Pulse Rate
PPR
--
Yes
Yes
Yes
Yes
Parameter
pH
Base Excess Blood
Blood Urea Nitrogen
Arterial Pressure
Parameter
* - As 1 waveform
Revision C
CIC Pro Clinical Information Center
2001099-156
B-3
Supported CIC Pro Parameters:
BIS Data
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Suppression Ratio
SR
--
Yes
--
--
Yes
Signal Quality Index
SQI
--
Yes
--
--
Yes
Bi-spectral Index
BIS
--
Yes
--
--
Yes
Electromyograph
EMG
--
Yes
--
--
Yes
Parameter
Cardiac Calculations (Cardiac Calcs)
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Cardiac Output
CO
Yes
Yes
--
--
Yes
Heart Rate
HR
Yes
Yes
Yes
Yes
Yes
Mean Arterial Pressure
MAP
--
--
--
--
Yes
Central Venous Pressure
CVP
--
--
--
--
Yes
Pulmonary Artery Mean
PAM
--
--
--
--
Yes
Pulmonary Artery Wedge
PAW
--
--
--
--
Yes
Body Surface Area
BSA
--
--
--
--
Yes
Cardiac Index
CI
--
--
--
--
Yes
Stroke Volume
SV
--
--
--
--
Yes
Systemic Vascular Resistance
SVR
--
--
--
--
Yes
Systemic Vascular Resistance Index
SVRi
--
--
--
--
Yes
Pulmonary Vascular Resistance
PVR
--
--
--
--
Yes
Pulmonary Vascular Resistance Index
PVRi
--
--
--
--
Yes
Left Ventricular Stroke Work Index
LVSWi
--
--
--
--
Yes
Right Ventricular Stroke Work Index
RVSWi
--
--
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Cardiac Output
CO
--
Yes
--
--
Yes
Blood Temperature
BT
--
Yes
Yes
Yes
Yes
Parameter
Cardiac Output (CO)
Parameter
B-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Supported CIC Pro Parameters:
Central Venous Pressure (CVP)
Parameter
Mean Blood Pressure
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Mean
Yes
Yes
Yes
Yes
Yes
Cerebral Perfusion Pressure (CPP)
Parameter
Mean Blood Pressure
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Mean
--
Yes
--
--
Yes
Continuous Cardiac Output (CCO)
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
SVO2
--
Yes
Yes
--
Yes
Blood Temperature
BT
--
Yes
--
--
Yes
Continuous Cardiac Index
CCI
--
Yes
--
--
Yes
Cardiac Output
CO
--
Yes
--
--
--
Cardiac Index
CI
--
Yes
--
--
--
Systemic Vascular Resistance
SVR
--
Yes
--
--
Yes
Systemic Vascular Resistance Index
SVRi
--
Yes
--
--
Yes
Continuous Cardiac Output
CCO
--
Yes
--
--
Yes
Signal
Strength
--
Yes
--
--
--
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
CO2e/CO2i
--
Yes
Yes
Yes
Yes
O2e/O2i
--
Yes
Yes
Yes
Yes
Parameter
Systemic Venous Oxygen
Signal Strength
CO2 Parameter
Parameter
Carbon Dioxide (Expired/Inspired)
Oxygen (Expired/Inspired)
Revision C
CIC Pro Clinical Information Center
2001099-156
B-5
Supported CIC Pro Parameters:
ECG Parameters
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
I
Yes
--
Yes*
--
--
II
Yes
--
Yes*
--
--
III
Yes
--
Yes*
--
--
AVR
Yes
--
Yes*
--
--
AVL
Yes
--
Yes*
--
--
AVF
Yes
--
Yes*
--
--
V1
Yes
--
Yes*
--
--
V2
Yes
--
Yes*
--
--
V3
Yes
--
Yes*
--
--
V4
Yes
--
Yes*
--
--
V5
Yes
--
Yes*
--
--
V6
Yes
--
Yes*
--
--
V
Yes
--
Yes*
--
--
HR
Yes
Yes
--
Yes
Yes
Parameter
* - Excluded are specific alarm types: a) Asystole and b) VFIB/VTAC
B-6
„
Arrhythmia analysis is done at the bedside device.
„
The following are recognized alarm types at the CIC:
‹
Asystole
‹
VFIB/VTAC
‹
Vtach
‹
V2>2
‹
V Brady
‹
Brady
‹
Acc Vent
‹
R on T
‹
Couple T
‹
Bigeminy
‹
Trigeminy
‹
PVC
‹
Irregular (including AFIB)
CIC Pro Clinical Information Center
2001099-156
Revision C
Supported CIC Pro Parameters:
Episodic Temperature
Parameter
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Temp
--
Yes
--
--
Yes*
Temperature
* - Data presented only as periodic
Femoral Pressure
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Systolic Pressure
Sys
Yes*
Yes
Yes
Yes
Yes
Diastolic Pressure
Dias
Yes*
Yes
Yes
Yes
Yes
Mean Pressure
Mean
Yes*
Yes
Yes
Yes
Yes
Peripheral Pulse Rate
PPR
--
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Mean
Yes
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Mean
Yes
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
CO2-I/CO2-E
--
Yes
Yes
Yes
Yes
O2-I/O2-E
--
Yes
Yes
Yes
Yes
NO2-I/NO2-E
--
Yes
Yes
Yes
Yes
Parameter
* - As 1 waveform
Inter-Cranial Pressure
Parameter
Mean Pressure
Left-Arterial Pressure
Parameter
Mean Pressure
Mass Spec
Parameter
Carbon Dioxide (Inspired/Expired)
Oxygen (Inspired/Expired)
Nitrous Oxide (Inspired/Expired)
Revision C
CIC Pro Clinical Information Center
2001099-156
B-7
Supported CIC Pro Parameters:
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
N2-I/N2-E
--
Yes
Yes
Yes
Yes
Desflorane (Inspired/Expired)
DES-I/DES-E
--
Yes
Yes
Yes
Yes
Enflurane (Inspired/ Expired)
ENF-I/ENF-E*
--
Yes
Yes
Yes
Yes
Halothane (Inspired/Expired)
HAL-I/HAL-E
--
Yes
Yes
Yes
Yes
Helium (Inspired/Expired)
HEL-I/HEL-E
--
Yes
Yes
Yes
Yes
Isoflorane (Inspired/Expired)
ISO-I/ISO-E
--
Yes
Yes
Yes
Yes
ARG-I/ARG-E
--
Yes
Yes
Yes
Yes
SEV
--
Yes
Yes
Yes
Yes
--
--
--
Yes
Yes
--
Parameter
Nitrogen (Inspired/Expired)
Argon (Inspired/Expired)
Sevane
No Breath
* - Enflurane can also be shown as ETH
Non-Invasive Pressure
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Systolic Pressure
Sys
--
Yes
Yes
Yes
Yes
Diastolic Pressure
Dias
--
Yes
Yes
Yes
Yes
Mean Pressure
Mean
--
Yes
Yes
Yes
Yes
---
--
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
NCO
--
Yes
Yes
Yes
Yes
NPCBF
--
Yes
--
--
Yes
NonInvasive Cardiac Index
NCI
--
Yes
--
--
Yes
NonInvasive Cardiac Stroke Volume
NSV
--
Yes
--
--
Yes
Parameter
Rate
Non-Invasive Cardiac Output
Parameter
NonInvasive Cardiac Output
NonInvasive Pulmonary Capillary Blood
Flow
B-8
CIC Pro Clinical Information Center
2001099-156
Revision C
Supported CIC Pro Parameters:
Other Cardiac Parameters
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Heart Rate
HR
--
Yes
Yes
Yes
Yes
Premature Ventricular Contraction
PVC
--
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Systolic Pressure
Sys
Yes*
Yes
Yes
Yes
Yes
Diastolic Pressure
Dias
Yes*
Yes
Yes
Yes
Yes
Mean Pressure
Mean
Yes*
Yes
Yes
Yes
Yes
Pulmonary Arterial Wedge
PAW
--
Yes
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Fractional Inspired Oxygen
FIO2
--
--
--
--
Yes
Positive end expiratory pressure
PEEP
--
--
--
--
Yes
Respiration Rate
RR
--
--
--
--
Yes
Tidal Volume
TV
--
--
--
--
Yes
Peak Inspiratory Pressure
PIP
--
--
--
--
Yes
Cardiac Output
CO
--
--
--
--
Yes
PBAR
--
--
--
--
Yes
Hb
--
--
--
--
Yes
PaCO2
--
--
--
--
Yes
Partial Pressure of O2 in Arterial Blood
Arterial Oxygen Saturation
PaO2
--
--
--
--
Yes
Partial Pressure of O2 in Mixed Venous
SaO2
--
--
--
--
Yes
Systemic Venous Oxygen
SvO2
--
--
--
--
Yes
Parameter
Pulmonary Artery Pressure
Parameter
* - As 1 waveform
Pulmonary Calculations
Parameter
Barometric Pressure
Hemaglobin
Partial Pressure of CO2 in Arterial
Blood,
Revision C
CIC Pro Clinical Information Center
2001099-156
B-9
Supported CIC Pro Parameters:
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
BSA
--
--
--
--
Yes
Cdyn(Ced)
--
--
--
--
Yes
Minute Volume
MV
--
--
--
--
Yes
Cardiac Index
CI
--
--
--
--
Yes
Alveolar Arterial Oxygen Gradient
AaDO2
--
--
--
--
Yes
Arterial Oxygen Content
CaO2
--
--
--
--
Yes
Oxygen Delivery Index
DO2I(O2DI)
--
--
--
--
Yes
Mixed Venous Oxygen Content
CvO2
--
--
--
--
Yes
Arterial Venous Oxygen content
difference
a-vO2
--
--
--
--
Yes
Oxygen Consumption Index
VO2I(O2CI)
--
--
--
--
Yes
Oxygen Extraction Ratio
O2ER(O2R)
--
--
--
--
Yes
Pa/FiO2
--
--
--
--
Yes
Qs/Qt
--
--
--
--
Yes
FICKCO
--
--
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
BP
Yes
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
RR
Yes
Yes
Yes
Yes
Yes
RR-APNEA
--
--
Yes
Yes
--
Parameter
Body Surface Area
Compliance Dynamic
Oxygenation Ratio
Shunt Fraction
Fick Cardiac Output
Right Arterial Pressure
Parameter
Mean Pressure
Respiration Data
Parameter
Respiration Rate
Respiration Rate-Apnea
B-10
CIC Pro Clinical Information Center
2001099-156
Revision C
Supported CIC Pro Parameters:
Respiration Mechanics
Parameter
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Minute Volume (Total, Spontaneous,
and Mechanical),
rMV, rMVs,
rMVm
--
Yes
--
--
Yes
Tidal Volume (Total, Spontaneous, and
Mechanical)
rTV, rTVs,
rTVm
--
Yes
--
--
Yes
rPIP
--
Yes
--
--
Yes
Mean airway pressure
Rmawp
--
Yes
--
--
Yes
Positive end expiratory pressure
rPEEP
--
Yes
--
--
Yes
Intrinsic peak end expiratory pressure
rPEEPi
--
Yes
--
--
Yes
Respiration rate (Total, Spontaneous,
and Mechanical)
rRR, rRRs,
rRRm
--
Yes
--
--
Yes
I:E ratio denominator
rI:E
--
Yes
--
--
Yes
Compliance (dynamic)
rCDYN
--
Yes
--
--
Yes
Resistance (expiratory)
rRAWe
--
Yes
--
--
Yes
Work of breathing (ventilator)
rWOVm
--
Yes
--
--
Yes
Rpef
--
Yes
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Pulmonary Oxygen Saturation Mean
SPO2
Yes
Yes
Yes
Yes
Yes
Pulmonary Oxygen Saturation Rate
SPO2-R
--
Yes
Yes
Yes
Yes
Pulmonary Oxygen Saturation Mean
(from Tram module)
SPO2-M
Yes
Yes
Yes
Yes
Yes
Pulmonary Oxygen Saturation Rate
(from Tram module)
SPO2-M-R
--
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Mean
Yes
Yes
Yes
Yes
Yes
Peak inspiratory pressure
Peak Expiratory Flow
SPO2
Parameter
Special Pressure
Parameter
Mean Pressure
Revision C
CIC Pro Clinical Information Center
2001099-156
B-11
Supported CIC Pro Parameters:
ST Parameter
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
I
--
--
Yes
Yes
Yes
Yes
II
--
--
Yes
Yes
Yes
Yes
III
--
--
Yes
Yes
Yes
Yes
AVR
--
--
Yes
Yes
Yes
Yes
AVL
--
--
Yes
Yes
Yes
Yes
AVF
--
--
Yes
Yes
Yes
Yes
V1
--
--
Yes
Yes
Yes
Yes
V2
--
--
Yes
Yes
Yes
Yes
V3
--
--
Yes
Yes
Yes
Yes
V4
--
--
Yes
Yes
Yes
Yes
V5
--
--
Yes
Yes
Yes
Yes
V6
--
--
Yes
Yes
Yes
Yes
V
--
--
--
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
SVO2
--
Yes
Yes
Yes
Yes
“*” notation
--
Yes
--
--
--
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Transcutaneous Carbon Dioxide
tcCO2
--
Yes
--
--
--
Transcutaneous Oxygen
tcO2
--
Yes
--
--
--
Temperature
temp
--
Yes
--
--
--
Power
power
--
Yes
--
--
--
Timer
timer
--
Yes
--
--
--
Parameter
SvO2
Parameter
Oxygen Saturation
Signal Strength
Transcutaneous CO2
Parameter
B-12
CIC Pro Clinical Information Center
2001099-156
Revision C
Supported CIC Pro Parameters:
TramModule Transcutaneous CO2
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Transcutaneous Carbon Dioxide
tcmCO2
--
--
--
--
Yes
Transcutaneous Oxygen
tcmO2
--
--
--
--
Yes
Temperature
mtemp
--
--
--
--
Yes
Power
mpower
--
--
--
--
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Temperature 1
Temp1
--
Yes
Yes
Yes
Yes
Temperature 2
Temp2
--
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Systolic Pressure
Sys
Yes*
Yes
Yes
Yes
Yes
Diastolic Pressure
Dias
Yes*
Yes
Yes
Yes
Yes
Mean Pressure
Mean
Yes*
Yes
Yes
Yes
Yes
Rate
Rate
--
Yes
Yes
Yes
Yes
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
BP
Yes
Yes
Yes
Yes
Yes
Parameter
Temperature
Parameter
Umbilical Pressure (UAC)
Parameter
* - As 1 waveform
Umbilical (UV, UVC)
Parameter
Mean Pressure
Revision C
CIC Pro Clinical Information Center
2001099-156
B-13
Supported CIC Pro Parameters:
Ventilator Data
Abbreviation
RT
Waveforms
RT
Numerics
Settable
Alarming
Settable
Limits
Historical
Trending
Positive End Expiratory Pressure
PEEP
--
Yes
Yes*
--
Yes
Fraction of Inspired Oxygen
FiO2
--
Yes
Yes*
--
Yes
Peak Inspiratory Pressure
PIP
--
Yes
Yes*
--
Yes
Mean Airway Pressure
MAWP
--
Yes
Yes*
--
Yes
Ventilator Rate
VNT-RR
--
--
Yes*
--
Yes
Flow Rate
FLW-RT
--
--
Yes*
--
Yes
INSP
--
--
Yes*
--
Yes
I:E
--
--
Yes*
--
Yes
Sensitivity
SENS
--
Yes
Yes*
--
Yes
Respiratory Rate
PT-RR
--
Yes
Yes*
--
Yes
Plateau Pressure
PPLAT
--
Yes
Yes*
--
Yes
Inspiration Hold Time
IN-HLD
--
--
Yes*
--
Yes
PPS
--
--
Yes*
--
Yes
Exhaled Tidal Volume
TV
--
Yes
Yes*
--
Yes
Minute Volume
MV
--
Yes
Yes*
--
Yes
Parameter
Inspiratory Time
Inspiratory: Expiration Ratio
Denominator
Positive Pressure Support
* - Single ‘Vent’ Alarm for all parameters
B-14
CIC Pro Clinical Information Center
2001099-156
Revision C
C
Revision C
Webmin Runtime
Diagnostics Utilities
CIC Pro Clinical Information Center
2001099-156
C-1
For your notes
C-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Runtime Diagnostics
Webmin runtime diagnostics utilities provide several tools for viewing/
troubleshooting a variety of system components. See “Webmin Runtime
Diagnostics Utilities” on page C-1 for details about runtime diagnostics utilities.
To access the runtime diagnostics utilities:
1.
Login to Webmin according to “Webmin Overview” on page 2-17.
2.
Click the System tab.
3.
Click the Runtime Diagnostics link.
024A
Runtime diagnostics functionality is divided across a number of different screens.
Refer to the following locations for more information about functionality associated
with each screen:
Revision C
„
“Closed-loop Audio Info” on page C-4
„
“Ping Command” on page C-5
„
“Hard Drive Integrity Info” on page C-6
„
“Instruction Store Integrity Info” on page C-7
„
“Temp/Voltage Info” on page C-8
„
“Network Info” on page C-9
„
“OS Hotfix Info” on page C-10
„
“Process Info” on page C-11
„
“OS Service Process Info” on page C-12
„
“Device Driver Info” on page C-13
„
“Timezone Info” on page C-14
„
“OS Event Log” on page C-15
„
“OS System Log” on page C-16
„
“OS Application Log” on page C-17
„
“OS Security Log” on page C-18
„
“System Information” on page C-19
CIC Pro Clinical Information Center
2001099-156
C-3
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Closed-loop Audio Info
Closed-loop Audio Info displays real-time internal/external speaker wiring integrity
information. Click the Closed-loop Audio Info button to display the information.
067A
The Closed-loop Audio Info window displays the current status of the speaker
wiring:
C-4
Plugged ([number of detected connections])
Speaker plug properly connected.
Unplugged ([number of detected connections
- usually “0”])
Speaker plug NOT properly connected.
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Ping Command
1.
Click the Ping Command button.
068A
2.
Type the IP address of the device you want to ping into the Address to Ping
field. Click Ping.
084A
Revision C
CIC Pro Clinical Information Center
2001099-156
C-5
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Hard Drive Integrity Info
Hard Drive Integrity displays hard drive model/serial number, firmware, operating
temperature (current/lowest/highest) and logical information. Click the Hard Drive
Integrity button to display the information.
069A
C-6
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Instruction Store Integrity Info
Instruction Store Integrity Info displays instruction store information. Click the
Instruction Store Integrity Info button to display the information.
070A
Revision C
CIC Pro Clinical Information Center
2001099-156
C-7
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Temp/Voltage Info
Temp/Voltage Info displays current CIC Pro internal temperature, system voltage,
fan and speaker status. Click the Temp/Voltage Info button to display the
information.
071A
The values should fall within the following limits:
System Temperature
CPU
5 — 90° C
Enclosure
5 — 70° C
System Voltage
5V
4850 — 5250 mV
12V
11400 — 12600 mV
Fan Status
CPU Fan
2700 — 4000 rpm
Chassis Fan 1700 — 4000 rpm
Speaker Status
C-8
External
Plugged
Internal
Plugged
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Network Info
Network Info displays network information associated with the CIC Pro. Click the
Network Info button to display the information.
072A
Displayed information includes:
Revision C
‹
Media State
‹
Description
‹
Physical Address
‹
DHCP Enabled
‹
IP Address
‹
Subnet Mask
‹
Default Gateway
‹
DHCP Server
‹
Primary WINS Server
‹
Secondary WINS Server
CIC Pro Clinical Information Center
2001099-156
C-9
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
OS Hotfix Info
OS Hotfix Info displays information about installed operating system security
patches. Check this information to determine if specific security patches are already
installed.
Click the OS Hotfix Info button to display the information.
073A
Displayed information includes:
C-10
‹
Hotfix ID
‹
Hotfix Service Pack
‹
Hotfix Description
‹
Comments
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Process Info
Process Info displays logged process information. Click the Process Info button to
display the information.
074A
Revision C
‹
Priority
‹
Virtual Size
‹
Working Set Size
CIC Pro Clinical Information Center
2001099-156
C-11
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
OS Service Process Info
OS Service Process Info displays logged service process information. Click the OS
Service Process Info button to display the information.
075A
Displayed information includes the following Service Process details:
C-12
‹
Name
‹
Path
‹
Description
‹
State (Running/Stopped)
‹
Start Mode (Auto/Manual)
‹
Logon
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Device Driver Info
Device Driver Info displays driver information, including current state. Click the
Device Driver Info button to display the information.
076A
Displayed information includes the following Driver details:
Revision C
‹
Name
‹
Path
‹
Description
‹
Error Control
‹
Service Type
‹
Start Mode
‹
Started (True/False)
‹
State (Running/Stopped)
‹
Status
CIC Pro Clinical Information Center
2001099-156
C-13
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
Timezone Info
Timezone Info displays detailed timezone information for the CIC Pro. Click the
Timezone Info button to display the information.
077A
C-14
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
OS Event Log
OS Event Log displays logged events for the CIC Pro. Click the OS Event Log
button to display the information.
078A
Event logs can indicate corruption of CIC Pro data. Items indicating data corruption
include FileSize values less than 1 or greater than 524288, or Path values either null
or something other than \winnt\system32\config\
Revision C
CIC Pro Clinical Information Center
2001099-156
C-15
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
OS System Log
OS System Log displays logged system events for the CIC Pro. Click the OS System
Log button to display the information.
079A
Displayed information includes the following system event details:
C-16
‹
Type
‹
Date/Time
‹
Source
‹
Category
‹
Event
‹
Message
‹
User
‹
Computer
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
OS Application Log
OS Application Log displays logged application event information. Click the OS
Application Log button to display the information.
080A
Displayed information includes the following application event details:
Revision C
‹
Type
‹
Date/Time
‹
Source
‹
Category
‹
Event
‹
Message
‹
User
‹
Computer
CIC Pro Clinical Information Center
2001099-156
C-17
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
OS Security Log
OS Security Log displays logged security event information. Click the OS Security
Log button to display the information.
081A
Displayed information includes the following security event details:
C-18
‹
Type
‹
Date/Time
‹
Source
‹
Category
‹
Event
‹
Message
‹
User
‹
Computer
CIC Pro Clinical Information Center
2001099-156
Revision C
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
System Information
System Information displays system information. Click the System Information
button to display the information.
083A
Revision C
CIC Pro Clinical Information Center
2001099-156
C-19
Webmin Runtime Diagnostics Utilities: Runtime Diagnostics
For your notes
C-20
CIC Pro Clinical Information Center
2001099-156
Revision C
D
Revision C
CIC Pro Service Tools
CIC Pro Clinical Information Center
2001099-156
D-1
For your notes
D-2
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
Service Tools
The CIC - Service Tool provides several utilities for viewing/troubleshooting a
variety of system components.
To access the runtime diagnostics utilities:
1.
Login to the CIC - Service Tool according to “CIC Service Tool” on page 2-24.
2.
Click the System Tools button.
102A
System Tools functionality is divided between a number of different utilities. Refer
to the following locations for more information about functionality associated with
each utility:
Revision C
„
“Audio Test” on page D-4
„
“Drive Integrity Test” on page D-6
„
“User Asset Management” on page D-8
„
“Environment Monitoring” on page D-9
„
“Network Integrity Test” on page D-11
„
“Printing Test” on page D-13
„
“Video Test” on page D-15
„
“Watchdog Test” on page D-18
CIC Pro Clinical Information Center
2001099-156
D-3
CIC Pro Service Tools: Service Tools
Audio Test
1.
Double-click the Audio Test icon. The Service Tool - Audio Test window
displays.
2.
Click the Audio Test tab.
105A
3.
Click the Play button. A sound should be heard from the CIC speakers.
NOTE
If necessary, adjust the sound volume with the right/left arrow buttons, or
drag the sound percentage slider using the mouse.
4.
The Audio Test pop-up window displays.
106A
D-4
5.
Click Yes if you heard a sound from the CIC Pro.
6.
If no sound was heard, adjust the audio volume according to the NOTE in step
3, and repeat this procedure. If a sound is still not heard, troubleshooting of the
sound card and/or other troubleshooting may be required.
7.
Click the Audio Information tab. Information about the CIC Pro audio system
displays. Information on this tab is used for audio troubleshooting.
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
NOTE
Click Update to refresh the displayed values.
107A
8.
Revision C
When finished, click Close to close the Service Tool - Audio Test window.
CIC Pro Clinical Information Center
2001099-156
D-5
CIC Pro Service Tools: Service Tools
Drive Integrity Test
1.
Double-click the Drive Integrity Test icon. The Service Tool - Drive Integrity
Test window displays.
2.
Click the Disk Drive Test tab.
108A
3.
Click the Hard Disk Drive button. The integrity of the hard drive is checked,
and the results are displayed in the Result field.
4.
Click the Compact Flash button. The integrity of the compact flash storage card
is checked, and the results are displayed in the Result field.
141A
D-6
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
5.
Click the Disk Drive Information tab. Information about the CIC Pro storage
media displays.
NOTE
Click Update to refresh the displayed values.
109A
6.
Click the S.M.A.R.T Information tab. SMART (Self Monitoring Analysis and
Reporting Technology) information about the CIC Pro storage media displays.
NOTE
Click Update to refresh the displayed values.
110A
7.
Revision C
When finished, click Close to close the Service Tool - Drive Integrity Test
window.
CIC Pro Clinical Information Center
2001099-156
D-7
CIC Pro Service Tools: Service Tools
User Asset Management
1.
Double-click the User Asset Management icon. The Service Tool - User Asset
window displays.
111A
D-8
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
Environment Monitoring
1.
Double-click the Environment Monitoring icon. The Service Tool Environment Monitor window displays.
2.
Click the Voltage/Temperature tab. Information about CIC Pro current internal
component voltage/temperature displays.
NOTE
Click Update to refresh the displayed values.
112A
Properties
Revision C
Value
System Voltage(5V)
Actual real-time voltage of the CIC Pro
5-volt internal power sub-system
(acceptable limits: 4850 to 5250 mV)
System Voltage(12V)
Actual real-time voltage of the CIC Pro
12-volt internal power sub-system
(acceptable limits: 11400 to 12600 mV)
CPU Temperature
Real-time CIC Pro internal temperature
at the CPU (acceptable limits: 5 —
90°C)
Enclosure Temperature
Real-time CIC Pro internal temperature.
(acceptable limits: 5 — 70°C)
CIC Pro Clinical Information Center
2001099-156
D-9
CIC Pro Service Tools: Service Tools
CPU FAN Speed
Real-time CIC Pro CPU fan speed.
(acceptable limits: 2700 — 4000 RPM)
CHASSIS FAN Speed
Real-time CIC Pro chassis fan speed.
(acceptable limits: 1700 — 4000 RPM)
External Speaker Status
Plugged (1) indicates connector plugged
into the external speaker socket;
Unplugged (0) indicates nothing plugged
into the external speaker socket.
Internal Speaker Status
Plugged (1) indicates connector plugged
into the internal speaker socket;
Unplugged (0) indicates nothing plugged
into the internal speaker socket.
3.
Click the Fan Control tab.
113A
D-10
4.
Click the OFF button. The results of the fan control test display in the Result
field.
5.
Repeat the previous step for the LOW, MID and HIGH buttons.
6.
When finished, click Close to close the Service Tool - Environment Monitor
window.
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
Network Integrity Test
1.
Double-click the Network Integrity Test icon. The Service Tool - Network
Integrity Test window displays.
2.
Click the Network Test tab.
114A
3.
Enter the IP address of a known device on the network into the field to the left
of the Pinging button.
4.
Click Pinging. The results of the ping display in the Result field.
142A
Revision C
CIC Pro Clinical Information Center
2001099-156
D-11
CIC Pro Service Tools: Service Tools
5.
Click the Network Information tab. Information about the CIC Pro network
connection devices displays.
NOTE
Click Update to refresh the displayed values.
115A
6.
D-12
When finished, click Close to close the Service Tool - Network Integrity Test
window.
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
Printing Test
1.
Double-click the Printing Test icon. The Service Tool - Printing Test window
displays.
2.
Click the Printing Test tab.
3.
Click the DDW Test or Laser Printer Test button.
4.
If DDW Test was clicked, click the Start - Test Pattern button.
5.
If laser Laser Printer Test was clicked:
a.
Select the laser printer from the pull down menu. Click Printing.
143A
b.
A test page is printed. Click Yes in the pop-up window when the test page
prints.
143A
Revision C
CIC Pro Clinical Information Center
2001099-156
D-13
CIC Pro Service Tools: Service Tools
6.
Click the Installed Printer Information tab. Information about installed printers
displays.
144A
NOTE
Click Update to refresh the displayed values.
117A
7.
D-14
When finished, click Close to close the Service Tool - Printing Test window.
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
Video Test
1.
Double-click the Video Test icon. The Service Tool - Video Test window
displays.
2.
Click the Video Test tab.
118A
3.
Revision C
To run a “one-step” video test, click the Start Video Test button. The monitor
will cycle through and display all five RGB color modes. The results of the test
display in the Result field.
CIC Pro Clinical Information Center
2001099-156
D-15
CIC Pro Service Tools: Service Tools
4.
To test individual RGB color modes, click the button corresponding to the color
mode you want to test. The associated color displays, and the results of the test
display in the Result field. Click Yes when the colors display correctly.
1458A
5.
D-16
Click the Video Information tab. Information about the CIC Pro video displays.
Information on this tab is used for video troubleshooting.
CIC Pro Clinical Information Center
2001099-156
Revision C
CIC Pro Service Tools: Service Tools
NOTE
Click Update to refresh the displayed values.
119A
6.
Revision C
When finished, click Close to close the Service Tool - Video Test window.
CIC Pro Clinical Information Center
2001099-156
D-17
CIC Pro Service Tools: Service Tools
Watchdog Test
NOTE
This procedure will cause the CIC Pro to reboot.
1.
Double-click the Watchdog Test icon. The Service Tool - Watchdog Test
window displays.
2.
Click the Start button. The watchdog test counts down until the Stop button is
clicked. The CIC Pro reboots when the count down is completed.
120A
3.
D-18
When finished, click Close to close the Service Tool - Watchdog Test window.
CIC Pro Clinical Information Center
2001099-156
Revision C
E
Revision C
Electromagnetic
Compatibility
CIC Pro Clinical Information Center
2001099-156
E-1
For your notes
E-2
CIC Pro Clinical Information Center
2001099-156
Revision C
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Electromagnetic Compatibility (EMC)
Changes or modifications to this system not expressly approved by GE
Medical Systems can cause EMC issues with this or other equipment.
This system is designed and tested to comply with applicable regulation
regarding EMC and must be installed and put into service according to
the EMC information stated in this appendix.
WARNING
Use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause
unexpected or adverse operation.
WARNING
The equipment or system should not be used adjacent to,
or stacked with, other equipment. If adjacent or stacked
use is necessary, the equipment or system should be
tested to verify normal operation in the configuration in
which it is being used.
Guidance and Manufacturer’s Declaration – Electromagnetic
Emissions
The CIC Pro is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure
that the CIC Pro is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
EN 55011
Group 1
The equipment uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
EN 55011
Class A
Harmonic Emissions
EN 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Complies
Revision C
The equipment is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
CIC Pro Clinical Information Center
2001099-156
E-3
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
The CIC Pro is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure
that the CIC Pro is used in such an environment.
Immunity Test
EN 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Electrostatic
Discharge (ESD)
EN 61000-4-2
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient/Burst
EN 61000-4-4
± 2 kV for power supply lines
± 2 kV for power supply
lines
Mains power should be that of a typical commercial
or hospital environment.
±1 kV for input/output lines
±1 kV for input/output
lines
Surge
EN 61000-4-5
± 1 kV differential mode
± 1 kV differential mode
± 2 kV common mode
± 2 kV common mode
<5% Ut (>95% dip in Ut)
for 0.5 cycles
<5% Ut (>95% dip in Ut)
for 0.5 cycles
<40% Ut (>60% dip in Ut)
for 5 cycles
<40% Ut (>60% dip in
Ut) for 5 cycles
<70% Ut (>30% dip in Ut)
for 25 cycles
<70% Ut (>30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 s
<5% Ut (>95% dip in Ut)
for 5 s
3 A/m
3 A/m
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN 61000-4-11
Power Frequency
(50/60 Hz)
Magnetic Field
EN 61000-4-8
Mains power should be that of a typical commercial
or hospital environment.
Mains power should be that of a typical commercial
or hospital environment. If the user of the equipment
requires continued operation during power mains
interruptions, it is recommended that the equipment
be powered from an uninterruptable power supply or
a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE:
Ut is the AC mains voltage prior to application of the test level.
E-4
CIC Pro Clinical Information Center
2001099-156
Revision C
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
The CIC Pro is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure
that the CIC Pro is used in such an environment.
Immunity Test
EN 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should not
be used closer to any part of the equipment, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
EN 61000-4-6
3 Vrms
150 KHz to 80 MHz
3 V rms
Radiated RF
EN 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2
P
d = 1.2
P 80 MHz to 800 MHz
d = 2.3
P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer, and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the
compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects,
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio,
AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above,
the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the equipment.
bOver the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
Revision C
CIC Pro Clinical Information Center
2001099-156
E-5
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Recommended Separation Distances
The table below provides the recommended separation distances (in
meters) between portable and mobile RF communications equipment and
the CIC Pro.
The CIC Pro is intended for use in the electromagnetic environment on
which radiated RF disturbances are controlled. The customer or the user
of the CIC Pro can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CIC Pro as
recommended below, according to the maximum output power of the
communications equipment.
Separation Distance in Meters (m) According to Frequency of Transmitter
Rated Maximum Output
Power of Transmitter in
Watts
150 kHz to 80 MHz a
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
a
P
d = 1.2
80 MHz to 800 MHz a
d = 1.2
P
800 MHz to 2.5 GHz a
d = 2.3
P
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated
using the equitation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE:
These guidelines may not apply in all instances. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
E-6
CIC Pro Clinical Information Center
2001099-156
Revision C
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Compliant Cables and Accessories
WARNING
The use of accessories, transducers and cables other than
those specified may result in increased emissions or
decreased immunity performance of the equipment or
system.
The table below lists cables, transducers, and other applicable
accessories with which GE Medical Systems claims EMC compliance.
NOTE: Any supplied accessories that do not affect EMC compliance are
not included.
Revision C
Part No
Description
Maximum Lengths
201792-001
DVI-D to DVI-D Cable
1.8 m / 6 ft
418335-00x
RJ45 Category 5 Cable
N/A
2016193-001
USB Keyboard
N/A
TBA
USB Mouse
N/A
2019795-001
USB Cable
1.8 m / 6 ft
2006733-001
Serial Cable
1.8 m / 6 ft
2001323-001
Speakers
N/A
80274-006
AC Power Cord
1.8 m / 6 ft
2020737-001
NEC Display 20” LCD 20080UX+
N/A
418331-002
PRN50 printer
N/A
418708-001
Key Pad
N/A
CIC Pro Clinical Information Center
2001099-156
E-7
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
For your notes
E-8
CIC Pro Clinical Information Center
2001099-156
Revision C
0459
gemedical.com
World Headquarters
GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel: + 1 414 355 5000
1 800 558 5120 (US only)
Fax: + 1 414 355 3790
European Representative
GE Medical Systems
Information Technologies GmbH
Munzinger Straße 3-5
D-79111 Freiburg
Germany
Tel: + 49 761 45 43 - 0
Fax: + 49 761 45 43 - 233
Asia Headquarters
GE Medical Systems
Information Technologies Asia; GE (China) Co., Ltd.
24th Floor, Shanghai MAXDO Center,
8 Xing Yi Road, Hong Qiao Development Zone
Shanghai 200336, P.R. China
Tel: + 86 21 5257 4650
Fax: + 86 21 5208 2008