Download Service manual
Transcript
1. GENERAL Specifications ECG input Input impedance Electrode offset tolerance Input unit protection Standard sensitivity Common mode rejection ratio Frequency response Waveform data processor Sample rate AC line filter High-cut filter EMG filter Time constant Waveform status detection Sensitivity selection 10 MΩ or more ±500 mV or more Isolated and defibrillator protected 10 mm /mV ±2% 100 dB or more 0.05 to 150 Hz (– 3 dB or more) 500 samples/s (input unit: 8,000 samples/s) 50/60 Hz, OFF 75, 100, 150 Hz 25/35 Hz 3.2 s or more Electrode detachment (polarization voltage), Noise (high frequency) 5, 10 , 20 mm/mV LCD (monochrome with CCFT backlight) (for ECG-9020K only) Size 5.6 inch Number of dots 320 × 240 ECG waveform 3 channel:Real time 12 ECG lead complexes 6 channel: 2.5 s Rhythm lead: 10 s Displayed data Waveform, patient information, recording settings, operation mode, heart rate, QRS sync mark, error message, electrode detachment, noise Recorder Printing method Printing density Scanning line density Recording width Number of recording channels Paper speed Printed data Mechanical noise External input/output External input Signal output Serial I/O Service Manual ECG-9010/9020 Rev A High resolution thermal printer head 200 dpi (8 dots/mm) 1 ms 104 mm 2, 3, 4, 6 10, 12.5, 25, 50 mm/s (10 and 12.5 mm/s available for ECG-9020K only) Program type, version, date and time, paper speed, sensitivity, lead name, filter, hospital name, patient information, timing mark, event mark, electrode detachment, noise 48 dB or less at paper speed 25 mm/s 10 mm/0.5 V ±5%, input impedance 100 kΩ or more 0.5 V/1 mV ±5%, output impedance 100 Ω or less Communication method: RS-232C Baud rate: 2400, 4800, 9600, 19200, 38400, 57600, 115200 1.3