6 REMOVAL AND REPLACEMENT Adapta® Summit / Region™ 3 and 7 Physical Therapy Platforms 6.5 POSTURAL DRAIN ACTUATOR (CONTINUED) NOTE: The Shoulder Bolt from the Gas Spring and the Shoulder Bolt from the Actuator are end to end. It may be necessary, after loosening the Actuator Shoulder Bolt, to use a flat-head screwdriver to force the Shoulder Bolt away from the Gas Spring Shoulder Bolt for removal of the Nut. Refer to Figure 6.20. ACTUATOR FIGURE 6.20 2. Unplug the Power Cable from the Actuator. Refer to Figure 6.21. 3. Using a 5 mm Hex Wrench and a 13 mm Combination Wrench, remove the Shoulder Bolt and Nut from the other end of the Actuator. Refer to Figure 6.22. 4. Reverse for installation of the replacement. NOTE: If the length of the Actuator varies from the original, slightly lift the table and adjust the Postural Drain Elevation to accommodate the length of the replacement. FIGURE 6.21 NOTE: When installing the replacement, slip the end of the Actuator into the metal sleeve before connecting the Shoulder Bolt on the Rod end of the Actuator. Refer to Figure 6.22. FIGURE 6.22 48 GAS SPRING