SC1100 Calibration Option
Verification
7
90%
Rise time
measures
between
these two
points
10%
om033i.eps
Figure 7-9. Edge Rise Time
Table 7-12. Edge Rise Time Verification
Calibrator Output
Voltage
Frequency
DSO Vertical
Axis (mV/div)
A 11801
Measurement
B Corrected
Measurement
Tolerance
250 mV
1 kHz
20.0
< 300 ps
250 mV
1 MHz
20.0
< 300 ps
250 mV
10 MHz
20.0
< 350 ps
500 mV
1 kHz
50.0
< 300 ps
500 mV
1 MHz
50.0
< 300 ps
500 mV
10 MHz
50.0
< 350 ps
1V
1 kHz
100.0
< 300 ps
1V
1 MHz
100.0
< 300 ps
1V
10 MHz
100.0
< 350 ps
2.5 V
1 kHz
200.0
< 300 ps
2.5 V
1 MHz
200.0
< 300 ps
2.5 V
10 MHz
200.0
< 350 ps
Edged Aberration Verification
This procedure uses:
• Tektronix 11801 oscilloscope with SC22/26 sampling head
• Output cable supplied with the SC1100
To do edge aberration verification:
1. Make sure that the SC1100 is in the edge mode (the edge menu is shown in the
display), and set it to output 1 V p-p @ 1 MHz.
2. Push .
3. Connect the Calibrator to the oscilloscope as shown in Figure 7-8.
4. Set the oscilloscope vertical gain to 10 mV/div and horizontal time base to 1 ns/div.
5. Set the oscilloscope to show the 90 % point of the edge signal. Use this point as the
reference level.
7-37