Download 5522A Service Manual
Transcript
SC1100 Calibration Option Verification 7 90% Rise time measures between these two points 10% om033i.eps Figure 7-9. Edge Rise Time Table 7-12. Edge Rise Time Verification Calibrator Output Voltage Frequency DSO Vertical Axis (mV/div) A 11801 Measurement B Corrected Measurement Tolerance 250 mV 1 kHz 20.0 < 300 ps 250 mV 1 MHz 20.0 < 300 ps 250 mV 10 MHz 20.0 < 350 ps 500 mV 1 kHz 50.0 < 300 ps 500 mV 1 MHz 50.0 < 300 ps 500 mV 10 MHz 50.0 < 350 ps 1V 1 kHz 100.0 < 300 ps 1V 1 MHz 100.0 < 300 ps 1V 10 MHz 100.0 < 350 ps 2.5 V 1 kHz 200.0 < 300 ps 2.5 V 1 MHz 200.0 < 300 ps 2.5 V 10 MHz 200.0 < 350 ps Edged Aberration Verification This procedure uses: • Tektronix 11801 oscilloscope with SC22/26 sampling head • Output cable supplied with the SC1100 To do edge aberration verification: 1. Make sure that the SC1100 is in the edge mode (the edge menu is shown in the display), and set it to output 1 V p-p @ 1 MHz. 2. Push . 3. Connect the Calibrator to the oscilloscope as shown in Figure 7-8. 4. Set the oscilloscope vertical gain to 10 mV/div and horizontal time base to 1 ns/div. 5. Set the oscilloscope to show the 90 % point of the edge signal. Use this point as the reference level. 7-37