368703--G_GA_Reocor_S_multi.book Page 56 Tuesday, April 24, 2012 4:16 PM

56
English
Behavior
before use
Lead connection
General Description
•
Keep an external defibrillator, oxygen, intubation
equipment and emergency drugs on hand.
•
Before use, Reocor S should be visually inspected
for damages and dirt.
•
Never use a device that is damaged or shows
abnormal behavior. Replace any cable that shows
even slight damage.
•
Before using Reocor S, the patient cable or leads,
the user should touch the patient to equalize electrical potentials.
•
It is strongly recommended to examine all set
parameters before the leads are connected to
Reocor S.
•
Even though Reocor S is protected from dripping
water, the device and all plugs should be kept
clean and dry.
•
Reocor S cannot be sterilized.
•
The connections of Reocor S and the temporary
pacing leads must be secured and checked regularly.
•
The patient cable must first be connected to
Reocor S and then to the leads.
•
The temporary leads, to which the Reocor S is connected, represent a low-impedance conductor to
the myocardium for electric current. Therefore
line-powered devices that are operated in the
patient's vicinity must be grounded in accordance
with established guidelines.
•
When handling already implanted leads, their connector pins and metal contact surfaces must not
touch or come into contact with electrically conductive or wet surfaces.
•
If the cable is disconnected from the Reocor S, it
must be reconnected immediately and the security
of the connection has to be examined.
•
When using unipolar leads, two unipolar leads
must be used for effective pacing.