Download Reocor S - BIOTRONIK USA
Transcript
368703--G_GA_Reocor_S_multi.book Page 56 Tuesday, April 24, 2012 4:16 PM 56 English Behavior before use Lead connection General Description • Keep an external defibrillator, oxygen, intubation equipment and emergency drugs on hand. • Before use, Reocor S should be visually inspected for damages and dirt. • Never use a device that is damaged or shows abnormal behavior. Replace any cable that shows even slight damage. • Before using Reocor S, the patient cable or leads, the user should touch the patient to equalize electrical potentials. • It is strongly recommended to examine all set parameters before the leads are connected to Reocor S. • Even though Reocor S is protected from dripping water, the device and all plugs should be kept clean and dry. • Reocor S cannot be sterilized. • The connections of Reocor S and the temporary pacing leads must be secured and checked regularly. • The patient cable must first be connected to Reocor S and then to the leads. • The temporary leads, to which the Reocor S is connected, represent a low-impedance conductor to the myocardium for electric current. Therefore line-powered devices that are operated in the patient's vicinity must be grounded in accordance with established guidelines. • When handling already implanted leads, their connector pins and metal contact surfaces must not touch or come into contact with electrically conductive or wet surfaces. • If the cable is disconnected from the Reocor S, it must be reconnected immediately and the security of the connection has to be examined. • When using unipolar leads, two unipolar leads must be used for effective pacing.