Download Mode d'emploi

Transcript
E
Code
‘S’
SET ?
Fonction
SYS RST
STO 0 or OFF
STO 1 or ON
STO ?
‘U’
UNI ?
UNI 0 or OFF
UNI 1 or ON
‘V’
VER ?
Sends the configuration of the unit for the current channel.
1 or 2 (measuring unit, Resolution, Direction, Rel/Abs, locked
display and keyboard status).
Reset the unit (customer parameters)
Unlock the display
Lock the display
Print out the display status
Sends the current unit mode (mm or ‘’)
Lock the Unit function
Unlock the Unit function
Sends the version and date of software (Vx.y dd.mm.yy)
RS 232 messages of transmission errors
Code
ERR0
ERR1
ERR2
ERR4
ERR5
ERR6
ERR7
ERR8
ERRA
ERRB
Display
‘fct oFF’
‘rS Err’
‘rS codE’
‘rS FULL’
‘no Func’
‘rS orun’
‘rS Err’
‘rS Err’
‘MEMO Err’
‘MEM1 Err’
P1 ERR
P2 ERR
P12 ERR
‘no Prob1’
‘no Prob2’
‘no Prob’
Type of error
Command not executed, deactivated function
Parity error
Unknown format
Capacity Overflow, more than 200 characters without CR
Command not executed, unauthorized function.
Overrun error
Frame error
Stopped transmission error
Not critical error of Flash memory
Critical error of Flash memory, requires a keyboard
acknowledge.
Probe 1 not connected
Probe 2 not connected
Probes 1 and 2 not connected
RS 232 parameters of the unit
Baud rate
Parity
Data Bits
Stop bits
Flow Control
4800 (transmission speed)
even
7
2
None
54