9. Troubleshooting
9. Troubleshooting
Problem
Cause/Remedy
Stand
The microscope does not respond upon being
switched on.
Ensure that the AC outlet has power.
Ensure that the electronics box is connected
to an AC outlet.
Check the cable connections.
Inform Service and have the supply unit fuse
checked.
Illumination
The image is completely dark.
Open the shutter (→ p. 59).
Check the connections of the lamp housings
on the microscope (transmitted/incident light)
Ensure that the lamps are connected to the
power supply and are not defective.
Inform Service and have the ebq 100 supply
unit fuse checked.
The image is unevenly or not uniformly illuminated.
Remove all unneeded filters from the light
path.
Center the lamp (→ p. 62ff).
Replace the old lamp (→ p. 23ff, 42f).
The illumination flickers.
Be sure that there is no loose connection at
the power supply.
Replace the old lamp (→ p. 23ff, 42f).
The fluorescence lamp does not illuminate immediately upon being switched on.
Hot Hg lamps should cool down before
switching on again.
The ebq 100 must be switched-on repeatedly.
Replace the old lamp (→ p. 25ff).
87