3. Specifications 3. Specifications GB Specifications Voltage range Input protection Output Current range Uncertainty Resolution Drive capability 0.000 … 24.000 mA -25.00 … +125.00 % DC 0.000 … 28.000 V Fuseless, up to AC 250 V 0.000 … 24.000 mA -25.00 … +125.00 % ± 0.015 % of reading ± 2 µA 1 µA, 1 mV ■■ Without HART™-resistor 1200 Ω Loop supply voltage DC 24 V ■■ With HART™-resistor Range select Special features Permissible ■■ Operating temperature ■■ Storage temperature 950 Ω Decade; incremental to 0.001 mA steps Auto step/ramp, Auto “%-Error” function, built-in resistor for HART™ communications -10 … +55 °C Range -20 … +70 °C 5 1/2-digit with character size 9 mm and %-announcement mA, V, % Power supply DC 9 V battery Display Recording options ■■ Battery life ■■ Low battery indicator EMC / CE conformity Ingress protection Calibration Dimensions Weight MIN-/MAX-storage 12 hours (12 mA into 250 Ω continuous) displayed icon near the end of battery life 2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immunity (portable equipment) IP 52 3.1 calibration certificate per DIN EN 10204 (option: DKD/ DAkkS-calibration certificate) 144.7 x 80.0 x 36.3 mm approx. 340 g For further specifications see WIKA data sheet CT 81.01 and the order documentation. 8 WIKA operating instructions, model CEP1000 11570351.02 10/2010 GB/D/F/E/I Input Current range