Chapter 4: Performance Features and Safety Precautions Motor/Blower The motor/blower assembly pulls room air into the front of the cabinet, and re-circulates it internally. During its recirculation, the air is split into two separate streams. One path leads through the exhaust HEPA filter and out of the unit. The second path flows through the supply HEPA filter, which then flows down through the work area, as shown in Figure 4-4. The motor in the Purifier Logic Series Cabinet is an electronically commutated motor (ECM). The ECM is a brushless DC motor that includes its own power supply to convert the incoming alternating current to direct current, as well as its own microprocessor to control and measure the motor’s operation. The motor is preprogrammed to deliver a consistent volume of air, even as the HEPA filters load over time. Figure 4-4 18 Product Service 1-800-522-7658