Transcript
This routine must be followed in order to operate the system correctly (see fig. 3 accordingly). 1. Make sure that the system to be tested (pos. 1) is depressurised. 2. Connect the Lloyds valve system to the 1/4" adapter mounted on the hose (pos. 2). 3. Start pumping out oil from the MGC calibrator system until the oil is coming out of the Lloyds valve system. 4. Close the Lloyds valve system (pos. 3). 5. Connect the Lloyds valve system to the valve with test connector located on the system, which is to be tested (pos. 4). 6. Open the valve on the system to be tested (pos. 5). 7. Open the Lloyds valve system (pos. 6). 8. Create pressure by operating the MGC-pump (pos. 7). 9. Read the pressure on the precision pressure gauge and on the object for calibration and repeat step 8 to proceed with the calibration. 10. Release the pressure and close the valve on the system to be tested (pos. 5). 11. Close the Lloyds valve system (pos. 3) and disconnect it from the pressure system with test connector (pos. 1). 124945 01 02-10-2007 21