15 Sterilisation chip card Sterilisation Parameter: 160 °C / 4:00 h Unit-ID: ________________________ Sterilisation Process Control Card As protection against contamination the HCP is equipped as standard with a STERICard. The STERICard starts an automatic and pre-set sterilisation procedure. To activate, insert the STERICard into the slit in the control field, select the Start symbol turn control and the depressed SET key. with the push/ The auto-sterilisation sequence begins as soon as the sterilisation procedure is activated: Ramp 1: Ramp 2: Ramp 3: Ramp 4 heating to 160°C hold for 4 hours (after the sterilisation temperature has been reached) cooling to 70°C wait 10 minutes The automatic sterilisation programme can not be altered by the user. After successful sterilisation the control displays STERILISATION OK. For documentation purposes the sterilisation process can be reported on the STERICard and can be read using „Celsius 2007“. Important note: On incubators the sterilisation programme does not serve for sterilising the load but solely for sterilising the chamber. The incubator is not a steriliser within the scope of the medical product regulations. HCP page 32