8 Sources of error It is unnecessary to go through all the troubleshooting, but it is important to distinguish some of the most common sources of error. Most of the trouble is caused by contamination. Impurities in the column or in the liner are visible in noise and/or drifting baseline. Even ghost peaks may appear when running a sample or a standard. Ghost peaks mean peaks appearing at positions where peaks are not expected. Briefly, be aware of avoiding contamination. When running samples or standards, every other run should be plain water, because it avoids samples to contaminate each other. To maintain Picture 12. In the back, there is a liner, septums and o-rings from the column clean, the helium flow must left to right. In front of them are tools for replacing them in the inlet and for maintaining the column. to be on all the time. It also helps to “bake out” the impurities from the column increasing the oven temperature to 200-230 °C for ten minutes every day. The column won’t last very long if it’s baked continuously in its maximum temperature, but short baking at times purifies it. Every time when there are no samples to run, the time can be spent to purify the column by baking the oven and/or running water. Organic acids consume the stationary phase so the retention times change slightly within time, but that can be corrected with calibration. If the ghost peaks keep continuously appearing, it may be necessary to cut a little piece from the beginning of the column (a cutter and explanation in the Picture 13), because most of the contamination piles up there. Regular maintenance will help avoiding chromatographic issues. Other supplies that have to be replaced from time to time are the liner, septum and oring (Picture 12) especially if problems occur. It is Picture 13. Used for cutting a piece from the column tubing. Use the side with black stripes to scratch it a couple of times and bend the column tubing off. recommended to replace the o-ring and septum after a hundred injections. A new liner is needed when the old one is dirty (glass wool turns black) and it depends on how much impurities there are in samples. Remember 15