Download Operating Instructions - VEGAPULS SR 68 - 4 ? 20 mA
Transcript
3 Product description - Safety Manual (SIL) VEGAPULS SR 68 (optional) Test certificate measuring accuracy, depending on the instrument VEGAPULS SR 68 (optional) Operating instructions manual "Indicating and adjustment module" (optional) Ex-specific "Safety instructions" (with Ex versions) if necessary, further certificates 3.2 Principle of operation Application area The instrument is designed for level measurement of bulk solids even under difficult process conditions. Its mechanical construction and electronics are specially optimised for such applications. Many application possibilities can be found in the food processing industry, in plastics processing and steel production as well as in the building industry. The instrument is particularly suitable for average-size silos and vessels. The instrument is also suitable for applications in liquids. The use of the instrument is possible with products with an εr value ≥1.5. The reachable value depends on the measurement conditions, the antenna system, the standpipe or bypass. Functional principle The antenna of the radar sensor emits short radar pulses with a duration of approx. 1 ns. These pulses are reflected by the product and received by the antenna as echoes. The running time of the radar pulses from emission to reception is proportional to the distance and hence to the level. The determined level is converted into an appropriate output signal and outputted as measured value. 3.3 Packaging, transport and storage Packaging Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180. The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies. Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device. Transport inspection The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with. 8 VEGAPULS SR 68 • 4 … 20 mA/HART - four-wire 38295-EN-110209 Transport