Download IND780 Weighing Terminal User's Guide

Transcript
User's Guide
IND780
Weighing Terminal
IND780 Weighing Terminal
Essential Services for Dependable Performance of Your IND780 Weighing Terminal
Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper use of your
new equipment according to this Manual and regular calibration and maintenance by our factorytrained service team ensures dependable and accurate operation, protecting your investment.
Contact us about a service agreement tailored to your needs and budget. Further information is
available at www.mt.com/service.
There are several important ways to ensure you maximize the performance of your investment:
1.
Register your product: We invite you to register your product at
www.mt.com/productregistration so we can contact you about enhancements, updates and
important notifications concerning your product.
2.
Contact METTLER TOLEDO for service: The value of a measurement is proportional to its
accuracy – an out of specification scale can diminish quality, reduce profits and increase
liability. Timely service from METTLER TOLEDO will ensure accuracy and optimize uptime and
equipment life.
a. Installation, Configuration, Integration and Training: Our service representatives are factorytrained, weighing equipment experts. We make certain that your weighing equipment is
ready for production in a cost effective and timely fashion and that personnel are trained for
success.
b. Initial Calibration Documentation: The installation environment and application
requirements are unique for every industrial scale so performance must be tested and
certified. Our calibration services and certificates document accuracy to ensure production
quality and provide a quality system record of performance.
c. Periodic Calibration Maintenance: A Calibration Service Agreement provides on-going
confidence in your weighing process and documentation of compliance with requirements.
We offer a variety of service plans that are scheduled to meet your needs and designed to
fit your budget.
© METTLER TOLEDO 2014
No part of this manual may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying and recording, for any purpose without the express written
permission of METTLER TOLEDO.
U.S. Government Restricted Rights: This documentation is furnished with Restricted Rights.
Copyright 2014 METTLER TOLEDO. This documentation contains proprietary information of METTLER
TOLEDO. It may not be copied in whole or in part without the express written consent of METTLER
TOLEDO.
METTLER TOLEDO reserves the right to make refinements or changes to the product or manual
without notice.
COPYRIGHT
®
METTLER TOLEDO is a registered trademark of Mettler-Toledo, LLC. All other brand or product
names are trademarks or registered trademarks of their respective companies.
METTLER TOLEDO RESERVES THE RIGHT TO MAKE REFINEMENTS OR CHANGES
WITHOUT NOTICE.
FCC Notice
This device complies with Part 15 of the FCC Rules and the Radio Interference Requirements of the
Canadian Department of Communications. Operation is subject to the following conditions: (1) this
device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case
the user will be required to correct the interference at his or her expense.
Declaration of Conformity is located on the documentation CD.
NOTE ON FIRMWARE REVISIONS
This manual describes features and functions of the IND780 terminal with version 6.4.xx
firmware. Terminals with version 6.3.xx firmware or lower will differ in some areas. The
following lists indicate the key differences between versions:

New in version 5.1 – Continuous Extended Output; enhanced SICS scale interface to
support WM/WMH, 4-Series and Excellence bases; Ethernet Terminal Clustering; ID and
Prompt sequences; serial input for text entry; remote SQL Server Database Access;
Modbus TCP interface; DeviceNet™ interface; Chinese language support; support for
Axle-780 and Com-780.

New in version 6.1.01 – Support for SICS levels 2 and 3; revised CalFREE™
functionality; additional display and keyboard language options; DHCP option added for
Ethernet/IP and Modbus TCP interfaces; External Diagnostics Load Cell Symmetry page;
addition of Watchdog Timer function.

New in version 6.1.08 – Support for POWERCELL PDX load cells, MT Service View for
PDX Diagnostics, PDX Performance Log.

New in version 6.3.03 – Support for Dual-Channel Analog Output option board.

IND780 terminals now equipped with faster, new-generation ETX board. The IND780
main PCB is compatible with both types of board, but the terminal must use firmware
versions as follows:
Original ETX board – version 6.x or earlier
New ETX board – version 7.x or later

New in version 7.x – Support for Flow Meter boards
RoHS Compliance Statement.

The majority of our products fall within categories 8 and 9. Those categories currently do not fall
within the scope of the Directive 2002/95/EG (RoHS) of January 27, 2003. If our products are
intended for use in other products which themselves fall within the scope of the RoHS Directive,
compliance requirements have to be separately negotiated contractually.

Those products which fall within categories 1-7 and 10 will be in compliance with the EU RoHS
Directive from no later than July 1, 2006.

If it is not possible for technical reasons to replace any non-RoHS-compliant substances in any of
the above products as required, we plan to inform our customers in a timely manner
Statement regarding harmful substances
We do not make direct use of harmful materials such as asbestos, radioactive substances or
arsenic compounds. However, we purchase components from third party suppliers, which may
contain some of these substances in very small quantities.
Warnings and Cautions
•
READ this manual BEFORE operating or servicing this equipment and FOLLOW these
instructions carefully.
•
SAVE this manual for future reference.
WARNING
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT TO PROPERLY
GROUNDED OUTLET ONLY. DO NOT REMOVE THE GROUND PRONG.
WARNING
NOT ALL VERSIONS OF THE IND780 ARE DESIGNED FOR USE IN HAZARDOUS (EXPLOSIVE)
AREAS. REFER TO THE DATA PLATE OF THE IND780 TO DETERMINE IF A SPECIFIC TERMINAL
IS APPROVED FOR USE IN AN AREA CLASSIFIED AS HAZARDOUS BECAUSE OF
COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES
WARNING
IN ORDER TO INSTALL THE DIVISION 2 APPROVED IND780 TERMINAL UTILIZING THE U.S.
APPROVAL, METTLER TOLEDO CONTROL DRAWING 64069877 MUST BE FOLLOWED
WITHOUT EXCEPTION. IN ORDER TO INSTALL THE CATEGORY 3 MARKED IND780 UTILIZING
THE EUROPEAN APPROVAL, THE DEMKO APPROVAL CERTIFICATE 07ATEX0520819X AND
ALL LOCAL REGULATIONS MUST BE FOLLOWED WITHOUT EXCEPTION. FAILURE TO DO SO
COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE. REFER TO THE IND780
DIVISION 2 AND ZONE 2/22 INSTALLATION GUIDE 64063214 FOR ADDITIONAL
INFORMATION.
WARNING
IF THE KEYBOARD, DISPLAY LENS OR ENCLOSURE IS DAMAGED ON A DIVISION 2
APPROVED OR CATEGORY 3 MARKED IND780 TERMINAL THAT IS USED IN A DIVISION 2 OR
ZONE 2/22 AREA, THE DEFECTIVE COMPONENT MUST BE REPAIRED IMMEDIATELY. REMOVE
AC POWER IMMEDIATELY AND DO NOT REAPPLY AC POWER UNTIL THE DISPLAY LENS,
KEYBOARD OR ENCLOSURE HAS BEEN REPAIRED OR REPLACED BY QUALIFIED SERVICE
PERSONNEL. FAILURE TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY
DAMAGE.
WARNING
WHEN THIS EQUIPMENT IS INCLUDED AS A COMPONENT PART OF A SYSTEM, THE
RESULTING DESIGN MUST BE REVIEWED BY QUALIFIED PERSONNEL WHO ARE FAMILIAR
WITH THE CONSTRUCTION AND OPERATION OF ALL COMPONENTS IN THE SYSTEM AND THE
POTENTIAL HAZARDS INVOLVED. FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT
IN BODILY HARM AND/OR PROPERTY DAMAGE.
Warnings and Cautions
CAUTION
BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR
INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER
AND WAIT AT LEAST THIRTY (30) SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT IN DAMAGE
TO OR DESTRUCTION OF THE EQUIPMENT AND/OR BODILY HARM.
NOTICE
OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES.
Disposal of Electrical and Electronic Equipment
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies
to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from
which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
Thank you for your contribution to environmental protection.
Contents
64057247 | 13 | 08/2014
1
Introduction ................................................................................. 1-1
1.1.
IND780 Terminal Versions ............................................................. 1-1
1.2.
Warnings and Precautions ............................................................. 1-2
1.3.
Operating Environment .................................................................. 1-3
1.3.1.
1.3.2.
1.3.3.
Temperature and Humidity ................................................................................... 1-3
Environmental Protection...................................................................................... 1-3
Hazardous Areas ................................................................................................. 1-3
1.4.
Inspection and Contents Checklist ................................................... 1-3
1.5.
Model Identification ....................................................................... 1-5
1.6.
Physical Dimensions ..................................................................... 1-6
1.7.
Specifications ............................................................................... 1-7
1.8.
Main PCB .................................................................................... 1-9
1.8.1.
1.8.2.
Overview ............................................................................................................ 1-9
ETX Board, HMI and LCD Updates ......................................................................... 1-9
1.8.2.1.
Compatibility ................................................................................................................... 1-9
1.9.
Scale Bases ................................................................................. 1-9
1.9.1.
1.9.2.
1.9.3.
1.9.4.
1.9.5.
Analog Load Cell Scale Base ................................................................................ 1-9
IDNet Scale Base ............................................................................................... 1-9
SICS Scale Base ................................................................................................ 1-10
®
®
POWERCELL MTX Scale Base ......................................................................... 1-10
®
®
POWERCELL PDX Scale Base ......................................................................... 1-10
1.10.
Flow Meter Interface .................................................................... 1-10
1.11.
Options...................................................................................... 1-10
1.11.1.
1.11.2.
1.11.3.
Discrete I/O ....................................................................................................... 1-11
Serial Ports ....................................................................................................... 1-11
PLC Interfaces ................................................................................................... 1-12
1.11.3.1.
1.11.3.2.
1.11.3.3.
1.11.3.4.
1.11.3.5.
1.11.3.6.
1.11.3.7.
Analog Output ................................................................................................................ 1-12
A-B RIO......................................................................................................................... 1-12
ControlNet and EtherNet IP .............................................................................................. 1-12
DeviceNet ...................................................................................................................... 1-12
Modbus TCP .................................................................................................................. 1-12
Profibus DP ................................................................................................................... 1-13
Profinet ......................................................................................................................... 1-13
1.11.4.
Application Software .......................................................................................... 1-13
1.11.4.1.
1.11.4.2.
1.11.4.3.
Axle-780 ....................................................................................................................... 1-13
Drive-780 ..................................................................................................................... 1-13
COM-780...................................................................................................................... 1-13
1.11.5.
1.11.6.
TaskExpert™ .................................................................................................... 1-14
InSite SL Configuration Tool .............................................................................. 1-14
1.12.
Display and Keyboard ................................................................. 1-14
METTLER TOLEDO IND780 User's Guide
1
Contents
2
2
Operating Instructions .................................................................. 2-1
2.1.
Security ....................................................................................... 2-1
2.1.1.
Metrology Switch ................................................................................................. 2-2
2.2.
Display Operation ......................................................................... 2-2
2.2.1.
Softkeys and Icons .............................................................................................. 2-2
2.3.
Understanding the Navigational Interface.......................................... 2-7
2.3.1.
2.3.2.
2.3.3.
2.3.4.
2.3.5.
2.3.6.
Navigation Keys .................................................................................................. 2-8
Softkeys ............................................................................................................. 2-9
Application Keys ............................................................................................... 2-10
Scale Function Keys .......................................................................................... 2-10
Alpha Keys ....................................................................................................... 2-11
Numeric Keys ................................................................................................... 2-13
2.4.
Home Screen .............................................................................. 2-14
2.5.
Screen Backlight Timeout and Screen Saver Operation ..................... 2-14
2.6.
Basic Functionality ...................................................................... 2-14
2.6.1.
2.6.2.
2.6.3.
Select Scale ...................................................................................................... 2-15
Sum Scale ........................................................................................................ 2-16
Zero ................................................................................................................. 2-17
2.6.3.1.
2.6.3.2.
2.6.3.3.
Automatic Zero Maintenance ............................................................................................ 2-17
Power-Up Zero ............................................................................................................... 2-17
Pushbutton Zero ............................................................................................................. 2-17
2.6.4.
Tare ................................................................................................................. 2-18
2.6.4.1.
2.6.4.2.
2.6.4.3.
2.6.4.4.
2.6.4.5.
2.6.4.6.
Pushbutton Tare ............................................................................................................. 2-18
Keyboard Tare ............................................................................................................... 2-19
Tare Table ..................................................................................................................... 2-20
Net Sign Correction ......................................................................................................... 2-20
Auto Tare ...................................................................................................................... 2-21
Clearing Tare ................................................................................................................. 2-22
2.6.5.
2.6.6.
2.6.7.
2.6.8.
Unit Switching ................................................................................................... 2-22
IDNet Class II .................................................................................................... 2-24
Expand By 10 ................................................................................................... 2-24
Print ................................................................................................................. 2-24
2.6.8.1.
2.6.8.2.
2.6.8.3.
Print Interlock ................................................................................................................ 2-25
Automatic Print .............................................................................................................. 2-25
Repeat Print ................................................................................................................... 2-25
2.6.9.
2.6.10.
2.6.11.
MinWeigh ......................................................................................................... 2-25
Information Recall ............................................................................................. 2-26
Target Comparison ............................................................................................ 2-26
2.6.11.1.
2.6.11.2.
2.6.11.3.
Material Transfer Applications .......................................................................................... 2-27
Over/Under Applications .................................................................................................. 2-30
Target Comparison Parameter Entry .................................................................................. 2-30
2.6.12.
Comparators ..................................................................................................... 2-32
2.6.12.1.
Using Comparators......................................................................................................... 2-33
2.6.13.
2.6.14.
ID Mode ........................................................................................................... 2-33
SmartTrac™ ..................................................................................................... 2-35
2.6.14.1.
Display Sizes ................................................................................................................. 2-35
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
64057247 | 13 | 08/2014
2.6.14.2.
2.6.14.3.
2.6.14.4.
2.6.14.5.
Display Colors ............................................................................................................... 2-37
Bar Graph Mode............................................................................................................. 2-37
Over/Under Mode ........................................................................................................... 2-38
Cross Hairs Mode........................................................................................................... 2-41
2.6.15.
Clustering ......................................................................................................... 2-41
2.6.15.1.
2.6.15.2.
Overview ....................................................................................................................... 2-41
Viewing a Clustered Terminal Remotely ............................................................................. 2-42
2.6.16.
2.6.17.
2.6.18.
Time and Date .................................................................................................. 2-43
Reports ............................................................................................................ 2-44
Calibration Test ................................................................................................. 2-45
2.7.
Alibi Memory Direct Access .......................................................... 2-48
2.8.
Table Searches ........................................................................... 2-50
3
Service and Maintenance ............................................................. 3-1
3.1.
Precautions .................................................................................. 3-1
3.2.
List of Tools Required .................................................................... 3-2
3.3.
Cleaning and Maintenance ............................................................. 3-2
3.4.
Firmware Upgrade Procedure.......................................................... 3-3
3.4.1.
3.4.2.
3.4.3.
Upgrade Using InSite™ ........................................................................................ 3-3
Upgrade Using USB Memory ................................................................................. 3-5
Power Up and Screen Saver Graphics .................................................................... 3-7
3.4.3.1.
3.4.3.2.
3.4.3.3.
Power Up Graphic ............................................................................................................ 3-7
Screen Saver Graphic........................................................................................................ 3-7
Installing New Graphics .................................................................................................... 3-7
3.5.
System Backup and Restore Procedures .......................................... 3-7
3.5.1.
3.5.2.
Backup .............................................................................................................. 3-7
Restore............................................................................................................... 3-8
3.6.
Service ........................................................................................ 3-9
3.7.
Troubleshooting .......................................................................... 3-10
3.7.1.
3.7.2.
Troubleshooting Overview .................................................................................. 3-10
Board LED Interpretation..................................................................................... 3-12
3.7.2.1.
3.7.2.2.
Main PCB ...................................................................................................................... 3-12
Option Boards ............................................................................................................... 3-13
3.7.3.
3.7.4.
3.7.5.
3.7.6.
3.7.7.
AC Power Test ................................................................................................... 3-15
Power Supply Voltage Test ................................................................................. 3-16
RS-232 Serial Output Voltage Test ....................................................................... 3-16
Battery Test ....................................................................................................... 3-17
Internal Diagnostic Testing ................................................................................. 3-18
3.7.7.1.
3.7.7.2.
3.7.7.3.
3.7.7.4.
3.7.7.5.
3.7.7.6.
3.7.7.7.
3.7.7.8.
Memory Test .................................................................................................................. 3-19
Keyboard Test ................................................................................................................ 3-19
Scale ............................................................................................................................ 3-20
Load Cell Output ............................................................................................................ 3-20
Calibration Values .......................................................................................................... 3-21
Shift Values ................................................................................................................... 3-21
Diagnose Cell ................................................................................................................ 3-21
Statistics ....................................................................................................................... 3-22
METTLER TOLEDO IND780 User's Guide
3
Contents
4
3.7.7.9.
3.7.7.10.
3.7.7.11.
Serial Test ..................................................................................................................... 3-23
Discrete I/O Test ............................................................................................................. 3-24
Network Test .................................................................................................................. 3-25
3.7.8.
MT Service View ................................................................................................ 3-25
3.7.8.1.
3.7.8.2.
3.7.8.3.
3.7.8.4.
3.7.8.5.
Scale Load Cell Temperature ............................................................................................ 3-26
Scale Load Cell Supply Voltage ........................................................................................ 3-27
Scale Loadcell COM Voltage ............................................................................................ 3-27
Load Cell Information...................................................................................................... 3-28
PDX Terminal ................................................................................................................ 3-29
3.7.9.
External Diagnostics .......................................................................................... 3-30
3.7.9.1.
3.7.9.2.
3.7.9.3.
3.7.9.4.
3.7.9.5.
3.7.9.6.
3.7.9.7.
3.7.9.8.
3.7.9.9.
3.7.9.10.
3.7.9.11.
3.7.9.12.
Index Page .................................................................................................................... 3-30
Help File ....................................................................................................................... 3-31
System Information......................................................................................................... 3-32
Resource Utilization ........................................................................................................ 3-32
Scale Statistics ............................................................................................................... 3-33
Read Shared Data .......................................................................................................... 3-33
Scale Data .................................................................................................................... 3-35
PDX Scale Data (POWERCELL PDX bases only) ................................................................. 3-35
PDX Terminal (POWERCELL PDX bases only) .................................................................... 3-36
PDX Load Cell (POWERCELL PDX bases only) ................................................................... 3-36
Load Cell Statistics and Load Cell Data (POWERCELL and PDX bases only) .......................... 3-37
Load Cell Symmetry (POWERCELL and PDX bases only) .................................................... 3-38
3.7.10.
3.7.11.
Watchdog Monitoring ........................................................................................ 3-38
Master Reset ..................................................................................................... 3-38
3.7.11.1.
To initiate a master reset ................................................................................................. 3-39
3.8.
Main PCB .................................................................................. 3-39
3.8.1.
3.8.2.
To remove a main PCB from an IND780, follow these steps .................................. 3-39
Main PCB Installation ........................................................................................ 3-41
3.9.
ETX Board .................................................................................. 3-41
3.9.1.
3.9.2.
3.9.3.
ETX Board Versions ........................................................................................... 3-41
ETX Board Removal ........................................................................................... 3-42
ETX Board Installation ........................................................................................ 3-43
3.9.3.1.
To install an ETX board on the main PCB .......................................................................... 3-43
3.9.4.
CF Card Removal and Installation ....................................................................... 3-44
3.10.
Power Supply ............................................................................. 3-44
3.10.1.
Power Supply Removal ...................................................................................... 3-44
3.10.1.1.
To remove a power supply .............................................................................................. 3-44
3.10.2.
Power Supply Installation ................................................................................... 3-45
3.10.2.1.
To install a power supply ................................................................................................ 3-45
3.11.
Option Board Installation.............................................................. 3-46
3.11.1.
To install an option board on the main PCB ......................................................... 3-46
3.12.
PLC Interface Module Removal and Installation ............................... 3-46
3.12.1.
To install a PLC interface module on the main PCB ............................................... 3-46
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
1
Introduction
DIV 2 AND ZONE 2/22 INSTALLATION
IF YOU WISH TO INSTALL THE IND780 IN A DIVISION 2 OR ZONE 2/22 AREA, REFER TO THE DIVISION 2 AND
ZONE 2/22 INSTALLATION INSTRUCTIONS INCLUDED ON THE RESOURCE CD PROVIDED WITH THE TERMINAL.
FAILURE TO COMPLY WITH THE INSTRUCTIONS PROVIDED THERE COULD RESULT IN BODILY HARM AND/OR
PROPERTY DAMAGE.
This chapter covers
•
•
•
•
•
•
•
•
•
•
•
•
IND780 Terminal Versions
Warnings and Precautions
Operating Environment
Inspection and Contents Checklist
Model Identification
Physical Dimensions
Specifications
Main PCB
Thank you for purchasing the IND780 industrial terminal. The IND780 combines
the cumulative experience of over 100 years of industrial weighing experience
and the innovation of the original manufacturer of electronic weighing solutions,
with the latest in METTLER TOLEDO technology to provide you with a versatile
weighing terminal that can be configured to meet your specific requirements. The
IND780 terminal is a high-performance, single- or multiple-range weighing
terminal for use with analog load cells, high-precision IDNet, SICS,
POWERCELL® MTX® or POWERCELL® PDX® scale interfaces used in industrial
weighing applications. The IND780 is capable of supporting up to four
measurement channels and can provide a metrologically correct Sum Scale.
Scale Bases
Flow Meter
Options
Display and Keyboard
1.1.
IND780 Terminal Versions
The IND780 terminal is available with the following capabilities and versions:
• Basic weighing terminal used in safe areas and in certain hazardous locations
• Panel Mount or harsh desk/wall-mount enclosures
• Connection for up to four scale channels and a metrologically correct sum
• Support for up to sixteen 350Ω analog load cells per terminal, with up to eight 350Ω analog
load cells per scale channel
• 320 x 240 pixel backlit, active TFT, color LCD with the capability of displaying weight with
characters as large as 34 mm high; alternate multiple channel display
• Real-time clock with battery backup
• Two serial ports (RS232 and RS232/422/485) for asynchronous, bidirectional communication
and print output
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-1
• 10/100 Base-T Ethernet port
• USB Master
Introduction
• 100–240 VAC power input range
• Support for the following option boards:
 Analog Load Cell interface
 DeviceNet™ interface
®
®
 IDNet Scale interface
 POWERCELL PDX interface
®
®
 Serial communications
 Flow Meter Interface
 Allen Bradley RIO interface
 Discrete I/O interface
 ControlNet™ interface
 Profibus-DP interface
 EtherNet / IP (also for Modbus TCP)
 Profinet interface
 Analog Output interface
 POWERCELL MTX interface
®
®
• Basic weighing functions including scale selection, zero, tare, and printing
• Single or multiple range or interval weighing
• Rate calculation for each scale channel with selectable weight and time units
• Selectable over/under classifying mode of operation with graphics
• Selectable material transfer mode
• Comparators – simple targets for comparison of weight or rate with target values or ranges
• ID mode for prompted transaction sequencing
• SmartTrac™ graphical display
• Two memory tables for storage of Tare and Target values
• Unit switching, including custom units
• Alibi memory storage for up to 256,000 records
• Grand total and subtotal registers for accumulating weight
• Ten customizable print templates and report printing
• TraxDSP™ digital filtering for analog load cells
• TraxEMT™ performance monitoring and recording, including web-based tools
• Traditional Calibration with 5-point linearization
• CalFree™ calibration without test weights
• Step Calibration procedure
• Ethernet network clustering, up to 20 terminals for remote console, data and interface sharing
1.2.
Warnings and Precautions
Please read these instructions carefully before putting the new terminal into operation.
1-2
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Before plugging in the terminal, make sure that the voltage stated on the terminal's label matches
the local power supply voltage. If this is not the case, do not connect the terminal under any
circumstances.
Although the IND780 is ruggedly constructed, it is nevertheless a precision instrument. Use care in
handling and installing the terminal.
1.3.
Operating Environment
When selecting a location:
• Choose a stable, vibration-free surface
• Ensure there are no excessive fluctuations in temperature and no direct exposure to sunlight
• Avoid drafts (for example, from fans or air conditioning)
• Readjust (calibrate) the terminal after any major change of geographical position
1.3.1.
Temperature and Humidity
The IND780 can be operated at temperatures and relative humidity conditions as listed under
Operating Environment in Table 1-1. The terminal can be stored at temperatures ranging from –40°
to 60° C (–40° to 140° F) at 10 to 95% relative humidity, non-condensing.
1.3.2.
Environmental Protection
The harsh enclosure meets IP69K requirements. The Panel Mount front panel sealing provides type
4x and type 12 protection—comparable to IP65 rating.
1.3.3.
Hazardous Areas
WARNING!
THE STANDARD IND780 IS NOT INTRINSICALLY SAFE! DO NOT USE IN AREAS CLASSIFIED
AS HAZARDOUS BY THE NATIONAL ELECTRICAL CODE (NEC) BECAUSE OF COMBUSTIBLE
OR EXPLOSIVE ATMOSPHERES.
Not all versions of the IND780 can be operated in areas classified as Hazardous by the National
Electrical Code (NEC) because of the combustible or explosive atmospheres in those areas. Contact
an authorized METTLER TOLEDO representative for information about hazardous applications. When
an approved IND780 is installed in an area classified as Division 2 or Zone 2/22, special AC
wiring requirements must be met. See document 64063214, IND780 Division 2, Zone 2/22
Installation Guide.
1.4.
Inspection and Contents Checklist
Verify the contents and inspect the package immediately upon delivery. If the shipping container is
damaged, check for internal damage and file a freight claim with the carrier if necessary. If the
container is not damaged, remove the IND780 terminal from its protective package, noting how it
was packed, and inspect each component for damage.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-3
If shipping the terminal is required, it is best to use the original shipping container. The IND780
terminal must be packed correctly to ensure its safe transportation.
Introduction
The package should include:
1-4

IND780 Terminal

Documentation CD (includes all manuals)

Installation manual

Bag of parts including ferrites, grommets, etc.,
depending on terminal configuration
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
1.5.
Model Identification
The IND780 model number is located on the data plate on the back of the terminal along with the
serial number. Refer to Figure 1-1 to verify the IND780 that was ordered.
Figure 1-1: IND780 Model Identification Numbers
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-5
Introduction
1.6.
Physical Dimensions
The IND780 terminal physical dimensions for the Panel Mount enclosure are shown in Figure 1-2
in inches and [mm].
Figure 1-2: IND780 Panel Mount Enclosure Dimensions
The IND780 terminal physical dimensions for the harsh desk/wall-mount enclosure are shown in
Figure 1-3 and Figure 1-4 in inches and [mm].
Figure 1-3: Harsh Environment Enclosure Dimensions, Front
Figure 1-4: Harsh Environment Enclosure Dimensions, Side
1-6
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
1.7.
Specifications
The IND780 terminal conforms to the specifications listed in Table 1-1.
Table 1-1: IND780 Specifications
IND780 Specifications
Enclosure Type
Panel Mount stainless steel front panel
Harsh environment desk/wall/column-mount type 304 L stainless steel
enclosure
Dimensions (l × w × d)
Panel Mount: 320 mm × 220 mm × 110 mm
(12.6 in. × 8.7 in. × 4.3 in.)
Harsh Environment: 299 mm × 200 mm × 141 mm
(11.8 in. × 7.9 in. × 5.6 in.)
Shipping Weight
5 kg (11 lb)
Environmental Protection
Panel Mount front panel sealing provides Type 4x and Type 12 protection –
comparable to IP65 rating
Harsh Environment meets IP69K requirements
Operating Environment
The terminal (both enclosure types) can be operated at temperatures
ranging from −10° to 40° C (14° to 104°F) at 10% to 95% relative
humidity non-condensing
Hazardous Areas
Not all versions of the IND780 can be operated in areas classified as
Hazardous by the National Electrical Code (NEC) because of the
combustible or explosive atmospheres in those areas. Contact an
authorized METTLER TOLEDO representative for information about
hazardous applications
Power
Operates at 100–240 VAC, 49–61 Hz, 400 mA (both enclosure types)
Panel Mount version provides a terminal strip for AC power connections
Harsh environment version includes a power cord configured for the country
of use
Note: When an IND780 is installed in an area classified as Division 2 or
Zone 2/22, special AC wiring requirements must be met. See document
64063214, IND780 Division 2, Zone 2/22 Installation Guide.
Display
320 x 240 pixel backlit graphic, active, TFT color LCD with the capability of
displaying weight in 34-mm high characters; alternate multiple channel
display
Weight Display
Displayed resolution of 1,000,000 counts for analog load cell scales
Display resolution for high-precision IDNet bases is determined by the
specific base used
Scale Types
64057247 | 13 | 08/2014
Analog load cells or IDNet, High-Precision K Line, POWERCELL MTX,
POWERCELL PDX, SICS
METTLER TOLEDO IND780 User's Guide
1-7
Introduction
IND780 Specifications
Number of Cells
Eight 350-ohm load cells (2 or 3 mV/V) per analog channel; (16) 350Ω
per terminal
Number of Scales
Interface for up to four scale channels plus a sum
Flow Meters
Up to four flow meter channels, for a maximum of four devices (scales
or flowmeters)
Analog/Digital Update Rates
Internal Analog: >366 Hz
IDNet: determined by base
POWERCELL MTX, POWERCELL PDX: determined by cell
Target Comparison: 50 Hz
PLC Interface: 20 Hz
Load Cell Excitation Voltage
10 VDC
Minimum Sensitivity
0.1 microvolts
Keypad
30 keys; 1.22-mm thick polyester overlay (PET) with polycarbonate
display lens
Communications
Serial Interfaces
Standard: Two serial ports COM1 (RS-232) and COM2
(RS-232/RS-422/RS-485), 300 to 115,200 baud; Ethernet 10/100 Base-T
Protocol
Serial Inputs: ASCII characters, ASCII commands for CTPZ (Clear, Tare,
Print, Zero), SICS (most level 0 and level 1 commands)
Serial Outputs: Continuous or Demand with up to ten configurable print
templates or SICS host protocol, report printing, interfaces with external
ARM100 Input/Output modules, and DeviceNet Bridge
Approvals
Weights and Measures
USA: NTEP CoC # 06-017
Class II, 100,000d
Class III, IIIL, 10,000d
Canada: AM-5592
Class II 100,000d
Class III 10,000d and Class IIIHD 20,000d
Europe: TC6944
Class II, approved divisions determined by platform
Class III, IIII, 10,000e
Hazardous Areas (IECEx UL 10.0014X)
EX nA nL (nL) IIB T4
Ex tc IIIC T85°C Dc IP69K (Harsh) or IP65 (Panel)
-10°C ≤ T amb ≤+40°C
Product Safety
UL, cUL, CE
1-8
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
1.8.
Main PCB
1.8.1.
Overview
The IND780 terminal’s main printed circuit board (PCB) includes provisions for the microprocessor,
main memory, battery, application module key, Ethernet, USB and serial communications, and
mounting of option boards.
The main board contains the COM1 and COM2 serial ports. COM1 provides RS-232
communication, while COM2 supports RS-232, RS-422, or RS-485 communication. These ports
are bidirectional and can be configured for various functions such as demand output, SICS host
communications, continuous output, ASCII command input (C, T, P, Z), ASCII character input,
report printing, totals printing, or connection to a remote ARM100 module.
1.8.2.
ETX Board, HMI and LCD Updates
Newer IND780 terminals are equipped with an updated ETX processor, a new color LCD, and a
new HMI board. These updates provide an overall improvement in system performance of 25 to
30%, and a power-saving LED backlight that provides better display readability at all angles.
1.8.2.1.
1.9.
Compatibility
•
The new ETX board (PXA270) is compatible with the existing main PCB. However, in order to
use the updated ETX, the terminal must use firmware version 7.x or above. The existing ETX
board (PXA255) is not compatible with this firmware, and must be used with version 6.x or
below.
•
The new HMI board and LCD must be used together – neither is compatible with the previous
version of the other. If it is necessary to replace either component in a system currently using
the original HMI and LCD, both must be replaced.
Scale Bases
The IND780 supports Analog, IDNet, SICS, POWERCELL MTX and POWERCELL PDX bases.
1.9.1.
Analog Load Cell Scale Base
The IND780 supports this scale type through an analog load cell interface. The terminal can drive
up to sixteen 350-ohm analog load cells, with up to eight 350-ohm load cells on one channel.
1.9.2.
IDNet Scale Base
The IND780 supports both the newer T-brick style of high-precision base and the older “PIK-brick”
transducers, through the IDNet scale interface. For T-brick bases, the interface provides the +12
volts and communication required to operate this newer style of base. The port also provides +30
volts, to support PIK-brick high-precision bases. The base’s cable determines which voltage is
used.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-9
Introduction
1.9.3.
SICS Scale Base
The IND780 supports Mettler Toledo high precision scales and balances that utilize the SICS
communications protocol. These scales and balances are branded as the Mettler Toledo Excellence
balances, X-bases/platforms, WM/WMH and 4-series scales (BBx4xx, IND4xx). The SICS scales
are connected to the IND780 via the serial interfaces. Four SICS scales can be supported per
terminal, when optional Serial boards are installed. Depending on the type of SICS scale connected,
different configuration settings will be available in the IND780 terminal setup screens.
1.9.4.
POWERCELL® MTX® Scale Base
The IND780 supports scales that use the POWERCELL MTX communications network found in large
hopper / tank applications as well as vehicle scales that use the MTX load cell. This interface also
supports the use of the RAAD Box, which converts analog load cell signals into digital ones.
1.9.5.
POWERCELL® PDX® Scale Base
The IND780 supports scales that use the POWERCELL PDX communications network, typically
found in vehicle scale and large tank weighing applications that use the PDX load cell. When used
with an external power supply, the IND780 PDX interface can support as many as 24 cells. Up to
four independent scale bases can be logically addressed by the terminal. When used in
combination with a PDX network, the IND780 provides several diagnostic features such as
predictive failure, automated alerts and cell performance monitoring. These features help lower
maintenance costs and minimize downtime.
1.10.
Flow Meter Interface
The Flow Meter Interface Board is a two-channel isolated Counter/Flow Meter board for use in the
IND780batch terminal. It is intended to provide a flow-meter totalizer target comparison to directly
control on-board discrete outputs. The module is capable of counting input pulses at up to 50 kHz
on each of two isolated input channels, as well as measuring the frequency of the input signal. A
jumper-selectable switching threshold for each input channel is available, as well as a jumperselectable 15 kHz analog filter. The input level range for the AC mode is 50mV to 50Vrms. The
input level range for DC mode is 2.5 volts to 42 volts.
The outputs are 7407 open-collector drivers. Each module provides 150 mA of 5V power to drive
opto-22 or similar devices. A total of two flow meters may be connected to a single flow meter
card. Each terminal can connect to as many as 4 flow meters.
1.11.
Options
The following additional options are available for the IND780:
•
Discrete I/O
•
Internal, high-level discrete I/O (4 inputs and 4 outputs)
•
Remote discrete I/O via ARM100 module (4 inputs and 6 outputs)
• Serial Communications
1-10
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
• Programmable Logic Control (PLC) interfaces, including:
o
Allen-Bradley (A-B) RIO
o
Analog output
o
Allen-Bradley ControlNet™
o
DeviceNet™
o
EtherNet / IP
o
Modbus TCP
o
Profibus DP
o
Profinet
• Axle-780 (application software)
• Drive-780 (application software)
• COM-780 (communications module)
• Task Expert™
• InSite Configuration Tool
• Various brackets for wall and column mounting of the harsh enclosure
The scale measurement channel, serial and discrete I/O options are connected to the IND780
through six internal option slots. Various combinations of options may be ordered to match the
application solution requirements.
1.11.1.
Discrete I/O
The discrete I/O interface options include both internal and remote I/O.
• The internal version is available with dry-contact relay or solid state relay outputs. Both types
will switch up to 30 volts DC or AC and up to 1 amp of current. The inputs are switch-selectable
as either active (for simple pushbutton control) or passive (for connection to PLCs or other
devices that supply their own power for the I/O). Each internal board supports four inputs and
four outputs.
• The remote I/O is supported with the ARM100 remote module that provides dry-contact outputs.
The inputs are passive on the ARM100. Each ARM100 supports four inputs and six outputs. An
external 24-volt DC supply is required to operate the ARM100.
• A total of two internal Discrete I/O boards (8 inputs and 8 outputs) are supported and an
additional 32 inputs and 48 outputs through a maximum of eight remote I/O modules.
1.11.2.
Serial Ports
Additional communications cards provide RS-232, RS-422 or RS-485 communication at rates
from 300 to 115.2k baud. A maximum of two serial communications cards may be installed in the
IND780.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-11
1.11.3.
PLC Interfaces
Introduction
The IND780 PLC interface options include analog output, A-B RIO, ControlNet™, DeviceNet™,
EtherNet/IP®, Modbus TCP, Profibus DP and Profinet.
1.11.3.1.
Analog Output
The Analog Output option kit includes a dual-channel option board. This board provides two
channels of an isolated 4-20 mA analog signal output for displayed weight, gross weight, rate or
application variables (selected in setup). One or two boards can be installed in the IND780, for a
maximum of four output channels.
The analog option uses a 16-bit D/A converter for a very precise output. The output signals will be
at the lower limit (4 mA) when the value represented is at zero. When the value reaches its
maximum limit, the output signal will increase to the higher limit (20 mA). Any value between zero
and the maximum limit will be represented as a percentage of the output proportional to the
percentage of the value.
1.11.3.2.
A-B RIO
The A-B RIO option enables data exchange by bi-directional communications using the Discrete
Data Transfer or Block Transfer mode. The IND780 Terminal initiates a communication exchange
with the PLC approximately 20 times per second utilizing the Allen-Bradley Discrete Data Transfer
protocol. This communication is a high-speed, real-time message interface between the IND780
Terminal and the PLC for process control. Division, integer, and floating point values are supported.
The IND780 A-B RIO interface also supports Block Transfer mode for transmission of larger
amounts of data. Additional details about this interface can be found in the IND780 PLC Interface
Manual on the documentation CD.
1.11.3.3.
ControlNet and EtherNet IP
The IND780 supports ControlNet communications or EtherNet / IP interface options and the
appropriate driver software.
1.11.3.4.
DeviceNet
DeviceNet is an RS-485 based network using CAN chip technology. This network was created for
bit- and byte-level devices. The network can be configured to run up to 500kbits per second
depending on cabling and distances. Messages are limited to 8 unfragmented bytes. The network
can include up to 64 nodes including the master, which is commonly called the scanner.
1.11.3.5.
Modbus TCP
Modbus TCP is used to establish master-slave/client-server communication between intelligent
devices. It is an open standard network protocol, widely used in the industrial manufacturing
environment. The Modbus TCP protocol takes the Modbus instruction set and wraps TCP/IP around
it. The Modbus TCP protocol is supported by the Ethernet / IP interface board, version 1.32 or
higher.
1-12
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
1.11.3.6.
Profibus DP
The IND780 Terminal communicates to a Profibus-DP master according to DIN 19 245. The
Profibus option consists of a module, together with firmware that resides in the IND780 Terminal to
implement the data exchange.
1.11.3.7.
Profinet
The Profinet option allows the IND780 terminal to communicate with Profinet-enabled
Programmable Logic Controllers (PLCs) at 100 Mbps speed, via a direct connection to the Profinet
network. The option consists of an internal module and internal software to implement the data
exchange.
1.11.4.
Application Software
The following application software modules can be added to the IND780 terminal to provide
additional functionality for specific workplaces and industries.
1.11.4.1.
Axle-780
The Axle-780 application supports vehicle weighing on a single-platform axle scale. It can be
operated in automatic (unattended) or manual (attended) modes, and features the following:
1.11.4.2.
• Weighs vehicles with up to 12 axles
• Transaction ID enterable via keyboard or
RFID (badge) reader
• Configurable to flag overloads and control
ticket printing for overloaded trucks
• Integrates use of traffic lights to instruct
driver when to move
Drive-780
Drive-780 provides additional inbound / outbound vehicle weighing and control of traffic lights or
gates associated with a truck scale. It includes the following:
1.11.4.3.
• One-pass weighing using stored tare
weights
• Two-pass weighing using temporarily stored
tare weights
• Transient vehicle weighing
• Commodity conversion
• Traffic light and gate control
• Tare expiration
COM-780
The COM-780 option is a specialized software module focused on the needs of users utilizing
legacy communication protocols. The IND780com maintains all of the standard features and
functions of the IND780 in addition to the specific features and functions of the COM-780. COM780 allows the IND780 to communicate using the following protocols:
• 8142
• PT6S3
• 8530
• SMA
Additional information can be found in the COM-780 Technical Manual on the module’s document
CD.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-13
Introduction
1.11.5.
TaskExpert™
TaskExpert functionality provides a way to modify the standard capabilities of an IND780 so that it
more closely aligns with the application requirements. TaskExpert is a combination of a
programming visualization tool, an execution engine and the basic functionality of the terminal.
Modifications may be made to the sequence of operation and additional functionality added to the
basic operation of the terminal. In addition, pre-configured application software such as drive-780
can be used as the starting point.
1.11.6.
InSite SL Configuration Tool
The IND780 terminal can connect to a PC running InSite SL (version 2.01 or higher) via Ethernet
to:
• Save configuration information locally on the PC
• Load a saved configuration file into other devices
• Restoring to a known state for service purposes
1.12.
Display and Keyboard
The IND780 is available with a backlit, active TFT, graphic color LCD. Weight information may be
displayed in a variety of formats, including single or multiple channel displays and with or without
a tare or rate window.
The display layout is designed with a system line reserved at the top to show system messages
and any asynchronous errors. The middle portion of the display is reserved for the weight display
and/or SmartTrac display. Directly entered data is shown in the bottom of this area. The bottom of
the display is reserved for showing the graphic labels (icons) for the softkeys. Display positions are
provided for up to five softkey icons.
Three sets of five softkeys can be configured to activate a wide variety of built-in capabilities of the
IND780, ranging from setting time and date, to accessing specific memory tables, to controlling the
IND780’s operation. Your installation technician will work with you to determine the appropriate mix
of softkey functionality to suit your specific needs.
Four Application keys (A1-A4), located below the softkeys, can be defined to perform a variety of
functions, including display adjustments, calibration testing, triggering custom outputs, and
switching between units.
The terminal’s 12-key numeric keypad is used to enter data and commands. The numeric keys are
located on the right side of the terminal front panel. Alphanumeric data may be entered using the
softkeys, through the use of an external USB keyboard, or scanned in from a bar code or other
external device.
Five navigation keys are located below the numeric keypad. These keys enable the operator to
navigate through setup options in the menu tree and within setup and application screens .
1-14
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 1-5 shows the IND780’s display and keyboard layout.
Scale function keys
Navigation keys
Softkeys
Application keys
Enter key
Figure 1-5: The IND780 Front Panel Layout
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
1-15
2
Operating Instructions
This chapter covers
• Security
• Display Operation
• Understanding the Navigational
Interface
• Home Screen
• Screen Backlight Timeout and
Screen Saver Operation
• Basic Functionality
• Alibi Memory Direct Access
• Table Searches
2.1.
The IND780 Terminal is simple to use yet sophisticated, with flexibility of
configuration to meet a variety of operating requirements. The setup menu system
provides the power of configuration in a user-friendly operational environment.
While reading this manual and operating the terminal, keep in mind that various
functions may or may not have been enabled for your installation, and that the
screens shown in this manual may vary from terminal to terminal, depending on
how it is set up and configured.
This document provides instructions for performing typical weighing operations on
the IND780 terminal.
Security
The IND780 supports the use of username/password for setup security at four levels.
• Administrator—An Administrator account has unlimited access to all areas of the operating and
setup system. There can be multiple Administrator accounts. There is a Primary Administrator
account, which can be changed but never deleted. The terminal is pre-configured at the factory
with the Primary Administrator account with no password. The unit as configured at the factory
requires no login or password entry to enter the setup mode. All functions of the terminal are
available to all users until a password for the Primary Administrator account is set up.
When the Metrology switch is turned “on” (refer to the next section of this chapter), all users with
Administrator rights are reduced to the Maintenance level. This is done to protect metrologically
significant parameters that cannot be changed when the terminal is “approved.” Refer to the
IND780 Technical Manual Appendix B, Default Settings for the security level of each setup
parameter.
Once a password is set up, be sure to remember it. If the password is changed or forgotten,
access to the setup menu will not be available. Be sure to protect the password from access by
unauthorized personnel. The password provides access to the entire setup menu. However, if
the metrology switch is placed in the approved position, metrologically significant settings
cannot be modified.
• Maintenance—Access is generally the same as the Administrator level with the exception of
access to metrologically significant areas of the setup.
• Supervisor—Access is generally limited to editing tables and setting time and date.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-1
Operating Instructions
• Operator—One default operator account is provided. Sites with validation requirements might
create many operator accounts, each with a username and password entry requirement. The
Operator-class of security is the most restrictive, allowing the user to use and view, but not
change records within tables.
If a password has been programmed for the default Administrator username in Setup, and all other
users have a password assigned, a login screen is presented whenever the Setup softkey is
pressed. A valid username and password must be entered. Depending on the access level of the
user logged in, setup screens may be visible only, or visible and available for modification.
If a login fails, the display exits the login page and returns to the home screen.
2.1.1.
Metrology Switch
If the metrology switch (S1) is placed in the approved position (On), changes to the Scale branch
of setup and other metrologically significant areas are not permitted. Access to the metrology switch
may be sealed in conformity with local regulations in “legal-for-trade” applications. Figure 2-1
shows the location of the metrology switch, immediately adjacent to the Compact Flash card and
PLC module location.
S1
Figure 2-1: Metrology Switch, Location (left), shown in its OFF position (right)
2.2.
Display Operation
Key names and commands are identified in this manual by upper- and lower-case letters. Key
names, such as ENTER, are in all upper-case letters, and commands, such as “select,” are in
lower-case (unless they begin a sentence, in which case the first initial is upper-case). For
example:
• “Press START...” means to press the START softkey
.
• “Select an option...” means to use the UP or DOWN navigation keys to select a setting, then
press ENTER.
2.2.1.
Softkeys and Icons
Softkeys and application displays use graphic icons for identification. Table 2-1 shows graphic
icons and their functions, categorized by where they are used. An asterisk (*) in the Softkey column
indicates that this icon can be assigned to a softkey position. The Setup
and Information Recall
softkeys are always present – they can be moved, but not deleted.
2-2
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Table 2-1: Icons and Functions
Icon
Softkey
Function
Explanation
System Recall
Information
Recall
*
Provides access to the Recall Screens: Weight, System
Info, Metrology, Totals
Weight Recall
Shows the current gross, tare, net values
System
Information
Recall
Shows the model, serial number, IDs, software versions,
and installed hardware
Metrology Recall
Shows the firmware version number and the time and date
of the last calibration
Provides access to the subtotal and grand total
accumulations
Totals Recall
Print
Prints the selected memory to an attached peripheral
Clear a Subtotal
Clears the subtotal and grand total registers from the
Totals Report
Clears both the subtotal and grand total registers from the
totals memory
Clear All
Icon to Enter Setup
*
Provides access to the setup parameters and procedures
of the instrument
Setup
Softkey Menu Icons
Up and Down
View the first or last five softkeys
More Down
View the next lower row of softkeys
More Up
View the next higher set of softkeys
Calibration Test Icons
*
Calibration Test
Provides access to the defined technician to perform the
calibration test
Calibration Test
Weight
Sequence
Provides access to the test weight sequence for the
calibration test
Start
Begins the defined sequence
Skip
Skips a failed step of the Calibration Test
Internal
Calibration Test
64057247 | 13 | 08/2014
Initiates an internal calibration test of a SICS scale
METTLER TOLEDO IND780 User's Guide
2-3
Operating Instructions
Calibration Icons
Capture Zero
Resets the zero condition of the scale
Capture Span
Resets the span value with known test weights
Calibrates the scale using a substitution method with a known
weight and a substitution mass
Step Calibration
CalFREE
Adjusts span to precalibrate a scale without test weights
Service Mode
Service Mode (access the IDNet service mode)
Internal
Calibration
Initiates an internal calibration of a SICS scale
Manual
Calibration
Initiates a manual calibration of a SICS scale
Initial Adjustment
Starts an initial adjustment of a SICS scale
Start
Begins the defined calibration sequence
Stop / Abort
Stops or aborts the calibration sequence
Reset
Resets the SICS scale to factory defaults
Table and Memory Icons
*
Reports
Provides access to the enabled tables – Alibi, Tare or
Target
Table Search
Provides search capabilities for the selected tables –
Alibi, Tare or Target
Search/ View
Locates and displays a record specified by the user’s
selected parameters
Capture Tare
Captures the current weight as a Tare weight
Clear All
Clear Totals
Print
2-4
*
Repeat Print
*
Tare Table
METTLER TOLEDO IND780 User's Guide
Clears both the subtotal and grand total registers from
the totals memory
Clears all totals from the Tare table
Prints the selected memory to an attached peripheral –
Tare or Target
Reprints the most recent transaction or Custom Print
output, with heading “DUPLICATE”
Provides access to stored tare values
64057247 | 13 | 08/2014
Icon
Softkey
Function
Explanation
System Recall
*
Alibi Memory
Provides access to stored transactional data
*
Target Table
Provides access to stored target values
*
Custom Trigger
1
*
Custom Trigger
2
*
Custom Trigger
3
*
Custom Trigger
4
*
Custom Trigger
5
*
Counter Reset
Reset
When configured in Setup at Communication >
Connections and assigned as a softkey, triggers
associated output.
Recalls and/or resets the next scale sequential number
and transaction counter number
Resets (clears) the currently selected log – Change,
Maintenance or Error
Task Selection Icons (only when TaskExpert™ is installed)
*
Task List
Displays list of assigned TaskExpert applications
*
Task 1
Starts TaskExpert application designated as Task 1
*
Task 2
Starts TaskExpert application designated as Task 2
*
Task 3
Starts TaskExpert application designated as Task 3
ID Mode Icons
*
ID1 Sequence
Initiates ID1 programmed transactional prompt sequence
*
ID2 Sequence
Initiates ID2 programmed transactional prompt sequence
Target Action Icons
64057247 | 13 | 08/2014
*
Comparators
Access Comparators table to permit selection of simple
setpoint
*
Target
Defines the current target, spill, fine feed, tolerance, and
description
METTLER TOLEDO IND780 User's Guide
2-5
Icon
Softkey
Function
Explanation
Operating Instructions
System Recall
Provides start, stop, pause, abort control of the current
target
*
Target Control
*
Start
Begins the defined sequence
Pause
Pauses the defined sequence
Stop / Abort
Stops or aborts the defined sequence
Display Icons
MinWeigh
Defines the Minimum weight value, below which a sign
appears beside the displayed weight, and the weight
display appears in red
*
Times Ten
(x10 Display)
Expands the displayed weight by a power of ten
*
Unit Switching
Toggles between the defined units of weighment
*
SmartTrac
toggle
*
Time and Date
Provides access to edit the hour, minutes, day, month,
year
*
Select Terminal
Switches between clustered IND780 terminals
Toggles the SmartTrac display on or off
Editing Icons
Exit
Exits a screen or parameter saving the values
Edit
Modifies the parameters of the selected object
New
Creates or inserts a new object containing information that can be
presented to the user
Delete
Print
OK / Accept
Prints the selected memory to an attached peripheral
Accepts or stores the new object parameter
Cancel
Skips or ignores a setting or parameter
Escape
Leaves a screen or parameter without saving
Copy
2-6
Clears an object
METTLER TOLEDO IND780 User's Guide
Makes a copy of the currently selected item
64057247 | 13 | 08/2014
Icon
Softkey
Function
Explanation
System Recall
Email Alert Test
When pressed, sends test message to recipient currently in focus.
Discrete I/O Icons
Output Off
Toggles the output off in the discrete I/O test
Output On
Toggles the output on in the discrete I/O test
PDX Service Icons
Sort And
Address
Unlock
Lock
*
2.3.
PDX
Performance
Log
Sort and readdress cells in an ascending order
corresponding to the cells’ S/N
Initiates MT Service Security unlock process
Locks MT Service Security and inhibits MT Service View
Trigger a log record in the PDX Performance Log
Understanding the Navigational Interface
Navigate in the applications and configure the IND780 using
• Navigation keys
• Scale function keys
• Softkeys
• Alpha keys (that appear on-screen when a field requiring
alphanumeric data entry is selected)
• Application Keys
• Numeric keys
• Optional external keyboard
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-7
The locations of these keys and the display screen area are shown in Figure 2-2.
Scale Function Keys
Operating Instructions
Display
Numeric Keypad
Softkey Icon
and Alpha
Keys Location
Softkeys
Navigation
Keys
Application keys
Clear Key
Enter Key
Figure 2-2: Front Panel Components and Key Locations
2.3.1.
Navigation Keys
Navigation keys (see Figure 2-2) enable navigation within the setup menu tree, setup screens, and
application screens.
These keys are used to move the focus to different setup options within the menu tree (focus is
indicated by highlighted text); to select different fields within a setup page; and to switch to another
page of available softkeys in the Home screen. The UP and DOWN and LEFT and RIGHT navigation
keys are also used to move the view up, down, left or right when a scroll bar indicates that more
information is available than can be shown in one screen. The example in Figure 2-3 shows a view
of a table with scroll bars visible. The bars indicate the position of this view relative to all available
information.
Figure 2-3: Tare Table Screen Showing Scroll Bars
Press the DOWN navigation key to move the focus down the rows. When the focus reaches the
second to last row the screen moves up to reveal the next row. Press the RIGHT navigation key to
move the highlight sideways and view additional columns.
2-8
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
• LEFT and RIGHT navigation keys
•
Expand (RIGHT) the setup options in the menu tree
•
Collapse (LEFT) the setup options in the menu tree
•
Move the cursor position to a specific character in text areas
•
Enable left and right scrolling to view all information available on a screen
• UP and DOWN navigation keys
2.3.2.
•
Move up and down through menu tree branches
•
Move field focus up and down on menu screens
•
Scroll up and down through tables to display additional rows-
•
ENTER key
o
Opens the selected setup page for viewing and editing setup parameters
o
Moves the focus from a field label to a setup value for that field
o
Accepts new values entered in a field and moves the focus to the next field label
Softkeys
A softkey setup screen (Figure 2-4), accessed at Setup > Terminal > Softkeys, is used to add and
remove softkey assignments and change softkey positions. Softkeys assignments can be modified
only by users with Maintenance or Administrator level access.
Figure 2-4: Softkey Setup Screen
Icons that can be assigned as softkeys are indicated in Table 2-1.
Five softkeys are located along the bottom of the display screen (see Figure 2-2). Some
installations might have up to three rows of softkey icons for a total of 15 possible functions. A
MORE DOWN or MORE UP symbol ( or ) displayed at the lower-right corner of the Home
screen (to the far right of the softkey icons) indicates that more softkey selections are available.
Press the DOWN navigation key to display additional softkey icons. Press the UP navigation key to
display the previous softkey set of icons.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-9
The IND780 Technical Manual Appendix E, Softkey Mapping and Application Key Configuration,
explains in detail how the softkeys can be assigned and edited in Setup.
Operating Instructions
2.3.3.
Application Keys
The four application keys, A1 through A4, are indicated in Figure 2-2. These may be assigned (via
Setup > Terminal > Application Keys) to give direct access to features such as the Tare Table or
SmartTrac, or to functions in specialized applications such as Task Expert modules. Each
application key can be assigned a custom label.
2.3.4.
Scale Function Keys
The four scale function keys (see Figure 2-2) are explained in Table 2-2.
Table 2-2: Scale Function Keys
Icon
Function
Select
Scale
Zero
Explanation
When multiple scales are connected to the indicator, this button permits the user
to switch between them, including the Sum Scale if one is configured.
To select a specific scale, enter the scale number using the numeric keys and
press the Select Scale function key.
When the scale platform or weighbridge is empty, the terminal should indicate
zero. The gross zero reference is recorded during calibration. Press the ZERO key
to capture a new gross zero reference point if pushbutton zero is enabled in
configuration and the weight is within the zero range. Pressing ZERO when a
Tare has been taken and the scale is in Net mode will cause an error message
to display – “Zero Failed – Scale in Net Mode. Press ENTER to continue.”
Tare is the weight of an empty container. Tare is normally used to determine the
net weight of the contents of a container. Press the TARE key when an empty
container is on the scale. The terminal then captures the tare value and displays
a net weight of zero. The B/G display changes to read NET, and a box appears
at upper right in the display (Figure 2-14), showing the tare value and units. As
the container is loaded, the terminal will display the net weight of the contents.
Pushbutton tare must be enabled to use the key in this manner.
Tare
When the empty weight of the container is a known value enter the tare directly
using the numeric keys, then press the TARE scale function key. The terminal will
display the net weight of the contents of the container. Keyboard tare must be
enabled in order to use this key in this manner.
To return to B/G (gross) mode after the TARE key has been pressed, press
(clear) on the keypad.
2-10
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Icon
2.3.5.
Function
Explanation
Print
Press the PRINT key to initiate a demand print of an assigned print template. A
printer must be connected to a serial port and the terminal must be configured to
match the serial port setting of the printer. Configuration is necessary to connect
a template or report to the selected serial or Ethernet port, and to define the
selected template or report. When a demand print command is issued, “Printing”
appears in the system line for 3 seconds.
Alpha Keys
On some screens permitting data entry, when field requiring non-numeric input (such as a
password) is selected, softkeys and application keys function as alpha keys used to enter
alphabetic characters. Figure 2-5 shows an example of the initial alpha key screen, with the
softkeys shown to indicate their relationship to the display.
Figure 2-5: Alpha Keys and Softkeys
The alpha keys are automatically displayed when the cursor is moved into a data entry box that
supports alpha entry. The first set of softkeys shown is groups of upper-case letters (Figure 2-5).
Press the MORE DOWN key to view lower-case groups (Figure 2-6).
Figure 2-6: Lower Case Alpha Key Groups
In addition to those accessed via softkeys, additional sets of characters are accessible using the A3
and A4 keys. A1 functions as an ESCAPE key, which takes the display back one level, either to the
previous set of alpha keys or, if the display is currently on the first set of alpha keys, out of alpha
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-11
Operating Instructions
key mode, returning focus to the currently selected input field. Other assignments include a variety
of symbols. For each of these sets of characters, a group of four to six letters is shown per softkey.
To enter a character, press the softkey under the group that includes the required character. If the set
of keys shown does not include the required character, press the DOWN navigation key to view the
next set. Press the DOWN navigation key and UP navigation key as required until the set of keys
shows the required character. Note the MORE UP and MORE DOWN symbol to the right of the
softkeys (visible at right in Figure 2-5 and Figure 2-6), indicating that additional sets of characters
are available.
After pressing the softkey associated with the group of characters that contains the required
character, the characters above the softkey will change to show one character per softkey (in the
upper row) and application key (in the lower row). Examples of the possible displays are shown in
Figure 2-7, with the lower-case letter screen associated with the first softkey at the top, and all the
symbol screens below it. Further rows of characters may be accessed using the MORE UP and
MORE DOWN (arrow) keys. In the case of the A-Z/a-z characters, these arrows toggle between
upper and lower case.
Row of
characters
associated with
application
keys
Row of
characters
associated with
softkeys
Figure 2-7: One Character per Key
Now, press the softkey or application key that corresponds to the required character; the character
will appear in the selected alpha input field.
Each group includes an ESC softkey
, assigned to A1, that steps back in the entry sequence in
case the wrong group of characters was selected.
Repeat this process until all alpha characters have been entered (Figure 2-8). The CLEAR key
on the keypad can be used to backspace and delete unwanted characters in the entry box.
2-12
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
To save the entry once all alpha characters have been entered, press the ENTER key on the keypad
to accept the entry, exit the text field and return to the initial entry screen (Figure 2-8). Then press
the OK softkey
to save the changes and exit the screen.
Figure 2-8: Entry Complete—Escape and OK Softkeys Showing
To void the entry, press A1 (the Alpha ESCAPE key
) to exit alpha key mode, and then press
the ESCAPE softkey
to return to the initial entry screen. The entry field will revert to its original
state, containing the data displayed in the text field before the entry was initiated. Press ESCAPE
again to leave the editing screen without making any changes.
2.3.6.
Numeric Keys
Use the terminal’s 12-key numeric keypad (see Figure 2-2) to enter data and commands.
To use numeric keys, position the cursor in the field (see Navigation Keys) and press the numeric
keys to enter the appropriate data. Press the DECIMAL key
to enter decimal points where
necessary.
The CLEAR key functions like a backspace key. Position the cursor at the end of data to be deleted
and press the CLEAR key once for each character to be deleted. When a data entry box is first
selected by pressing ENTER, the previous entry will be in focus (white text on a black background).
Pressing CLEAR at this point will clear the entire entry.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-13
2.4.
Home Screen
Operating Instructions
The Home screen displays when the terminal is idle (an example is shown in Figure 2-9). The
Home screen is the only screen that provides operator access to the programmable softkeys. Except
when the setup menu tree is accessed, or one of the programmable softkeys or application keys
has been pressed, the Home screen displays.
System
Line
Application
Area
Message
Area
MORE
DOWN
Icon
Softkey
Icons
Figure 2-9: Weighing Operation Home Screen
The Home Screen includes:
• System Line—System messages and application data
• Application Area—Weight, units, tare and other application-specific weighing data
• Message Area—Messages and SmartTrac display
• Softkey Icons—Icons for the active softkey functions. A MORE DOWN
symbol indicates that more softkey selections are available.
2.5.
symbol or a MORE UP
Screen Backlight Timeout and Screen Saver
Operation
The screen backlight will turn off and a screen saver will appear after independently set periods of
inactivity. These delays are configured in Setup at Terminal > Display.
To exit the Screen Saver and/or restore the Backlight, press any key on the terminal or on an
optional external keyboard. The key press will not perform the function normally associated with
that key.
2.6.
Basic Functionality
This section provides information about the IND780’s basic functionality. Configuration of these
areas of functionality can be accessed through the different areas of the setup menu tree – an
example view is shown in Figure 2-10. Additional areas of functionality specific to application
2-14
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
software available for the IND780 are addressed in the specific application manuals. Basic
functions addressed in this section include:
• Select Scale
• MinWeigh
• Bar Graph Mode
• Sum Scale
• Information Recall
• Over/Under Mode
• Zero
• Target Comparison
• Cross Hairs Mode
• Tare
• Comparators
• Clustering
• Unit Switching
• ID Mode
• Time and Date
• IDNet Class II
• SmartTrac
• Reports
• Expand By 10
• Display Sizes
• Calibration Test
• Print
• Display Colors
Figure 2-10: Setup Menu Tree Example
2.6.1.
Select Scale
The Select Scale function button
is used to switch between scale bases when multiple
platforms are configured for the IND780, including the Sum Scale. It determines which scale is
shown on the display and controlled from the keypad. It is possible to have multiple scales
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-15
displayed on the screen simultaneously, in which case front keypad functions (tare, zero, etc.)
affect the selected scale.
Operating Instructions
Pressing the Select Scale button switches between the scale bases in sequence. To select a scale
directly, entering its number using the numeric keys, then press the Select Scale button. Figure 2-11
shows a home screen with three scales plus a Sum Scale, with a highlight indicating that Scale 3 is
selected.
Figure 2-11: Scale Selection
2.6.2.
Sum Scale
When configured in Setup (at Scale > Sum Scale > Type), the Sum Scale allows for a logical
measurement channel that works from the arithmetic sum of the weight values from two or more
scales. Each scale may have a different capacity and increment size, but they must all be
configured with the same units.
In general, the Sum Scale display behaves like any other independent physical scale channel.
However, any zero command issued to the Sum Scale is redirected to each of its component scale
channels. With regard to the weight display, if one scale included in the Sum Scale is overcapacity, both its display and the Sum Scale will show “^ ^ ^ ^ ^“. If a component scale is under
range, the display will also show “v v v v v“. These conditions are shown in the screen images in
Figure 2-12. In both cases, the Sum Scale is selected.
Figure 2-12: Summed Scales Over Capacity (left) and Under Range (right)
2-16
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
If one scale included in the Sum Scale is over-capacity and another is under range, the respective
scales will indicate their conditions accordingly, and the Sum Scale display will show “----------“,
indicating an invalid weight value. This condition is shown in Figure 2-13.
Figure 2-13: Summed Scales Simultaneously Over Capacity and Under Range
2.6.3.
Zero
The Zero function is used to set or reset the initial zero reference point of the IND780. There are
three types of zero setting modes:
• Automatic Zero Maintenance (AZM)
• Power-Up
• Pushbutton
2.6.3.1.
Automatic Zero Maintenance
Automatic Zero Maintenance (AZM) enables the IND780 to compensate for the buildup of small
amounts of weight and track itself back to the center of zero. Within the AZM operating range
(programmable from 0.0 to 9.9 divisions), when the terminal is in a no motion condition it makes
small adjustments to the current zero reading to drive the weight reading toward the true center-ofzero. When the weight is outside of the programmed AZM range, this feature is not functional.
2.6.3.2.
Power-Up Zero
Power-Up Zero enables the IND780 terminal to capture a new zero reference point after power is
applied. If there is motion during a power-up zero capture function, the terminal will continue to
check for a no-motion condition until zero is captured.
Power-up zero can be disabled (RESTART on power up) or enabled (Reset on power up), and a
range above and below calibrated zero can be configured. The range is programmable from 0% to
100% of capacity and can include a positive range and also a range below calibrated zero.
2.6.3.3.
Pushbutton Zero
The pushbutton (semi-automatic) zero function can be accomplished by pressing the ZERO
scale function key, by programming a discrete input, by a PLC or serial command, or by an
application.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-17
Operating Instructions
The range for all types of semi-automatic zero is selectable (0% to 100%) plus or minus from
either the calibrated zero point (if power-up zero is disabled) or from the initial zero setting point (if
power-up zero is enabled).
Remote initiation of the semi-automatic Zero command is possible via a discrete input, an ASCII ‘Z’
command sent serially (CPTZ and SICS), a command initiated by the PLC interface, or from an
application.
2.6.4.
Tare
For detailed information about tare functionality when using SICS balances, please refer to the
SICS section in the IND780 Technical Manual Appendix D, Communication.
Tare is the weight of an empty container. A tare value subtracts from the gross weight
measurement, providing the computation of the net weight (material without the container). The tare
function can also be used to track the net amount of material being added to or removed from a
vessel or container. In this second case, the weight of the material in the container is included with
the tare weight of the container as tare. The display then reflects the net amount being added to or
removed from the vessel.
The tare value can be displayed with the net weight. The operation of this secondary display is
determined in Setup at Terminal > Display.
Tare
Net
weight
display
Tare type –
T, PT or M
Tare units
Figure 2-14 Screen Displaying Captured Tare
Tare types and associated operations available on the IND780 include:
2.6.4.1.
• Pushbutton Tare
• Auto Tare
• Keyboard Tare (Preset Tare)
• Clearing Tare
• Table Tare
•
Manual Clear
• Net Sign Correction
•
Auto Clear
Pushbutton Tare
Pushbutton tare can be configured in Setup as enabled or disabled. When this feature is disabled,
the TARE scale function key
cannot be used to obtain a tare.
2-18
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
If pushbutton tare is enabled, pressing the pushbutton TARE scale function key
initiates a
semi-automatic tare. The IND780 will attempt to perform a tare process. If it is successful, the
display changes to a zero net weight indication and the previous weight on the scale is stored as
the tare value. The net mode will be indicated on the display.
Several conditions could inhibit the pushbutton tare function:
• Motion—Pushbutton tare cannot be taken when the scale is in motion. If motion is detected
when a pushbutton tare command is received, the IND780 will wait up to three seconds for a
no-motion condition. If a stable (no motion) weight condition occurs before the three seconds
expire, the pushbutton tare command is executed.
If there is still motion at the end of three seconds, the command is aborted and a “Tare Failure
– In Motion” error displays.
• Pushbutton Tare Disabled—If pushbutton tare is configured as disabled, the TARE scale
function key will not initiate a semi-automatic tare.
• Negative Gross Weight—Any pushbutton tare attempted when the gross weight is at or below
zero is ignored and a “Tare Failed–Too Small” error displays. Ensure that the gross weight is
above zero.
2.6.4.2.
Keyboard Tare
A keyboard (preset) tare is a numeric tare that is entered manually through the numeric keypad,
received serially from a peripheral, or retrieved from the Tare Table memory. The preset tare value
cannot exceed the capacity of the scale. Data entered is interpreted to have the same units as the
current displayed value. Motion does not impact the entry of preset tare values.
Keyboard tare can be configured in Setup as enabled or disabled. When disabled, the numeric
keypad and the TARE scale function key
cannot be used to obtain a tare.
To manually enter a preset tare value, use the numeric keypad to enter the tare value (the entry will
display above the softkey icons) and press the TARE scale function key
. The tare display box
will include PT to indicate that a preset tare is in use.
If configured in Setup, remote equipment can enter a preset tare value using a serial command or
PLC command. (For further information, refer to the IND780 Technical Manual, Communication,
Configuration, the Serial and PLC Interface sections.)
If the preset tare is successful, the display changes to a net weight indication, and the entered
preset tare value is stored as the tare value in the Tare Table.
Several conditions could inhibit the preset tare function:
• Keyboard Tare Disabled—If keyboard tare is configured in Setup as disabled, the numeric
keypad and the TARE scale function key
cannot be used to obtain a tare.
• Over-Capacity or Under-Zero Condition—Preset tare is not allowed when the weight display
indicates over capacity or under zero conditions. Any preset tare attempted when the scale is
over capacity is ignored and a “Tare Failed – Over Cap” error displays. Any preset tare attempted
when the weight display indicates an under zero condition is ignored and a “Tare Failure –
Below Zero” error displays.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-19
Operating Instructions
Preset tare can be entered in free format. If the entered value does not match the displayed
weight decimal point location or display interval, the entered tare value is rounded to the nearest
display interval and the decimal point adjusted to match the gross weight. The rounding method
is that 0.5 or more of a display interval (d) is increased to the next display interval and 0.49 or
less of a display interval is decreased to the next lower display interval.
When entering a preset tare value less than 1.0, the operator can enter the data without the
leading zero (left of the decimal point), but all subsequent display, storage, or printing of this
value will include the leading zero. For example, a preset tare entry of .05 will display as 0.05.
If a preset tare has already been established and another preset tare is entered, the second
preset tare replaces the previous value (it does not add to the previous value). The replacement
tare can be larger or smaller than the original tare value.
2.6.4.3.
Tare Table
The IND780 terminal contains a tare table for storing tare weights that can be recalled by the
operator for use instead of manually entering them for each transaction. This is especially useful
when certain tare values are used repeatedly.
A description up to 40 characters long can be included with each record. This can be used to help
distinguish one tare record from another. Each tare record in the tare table also contains a
totalization field. When totalization is enabled for the tare table, each time a transaction is
completed using a specific tare ID, the selected weight value (gross or net weight) will be added to
the total and the corresponding counter will be incremented by one.
A tare memory can be recalled by picking from a list of all available records (accessed using the
Tare Table softkey). If the ID for the tare value is known it can be recalled directly by entering the ID
(note that ID entry is not case-sensitive) and pressing the Tare Table softkey. Unless the tare ID is
set as a number, an external keyboard is required to access a tare memory directly.
A printed report of the records in the Tare Table is also available. Additional details about the tare
table are explained in the IND780 Technical Manual Appendix C, Table and Log File Structure.
2.6.4.4.
Net Sign Correction
Net sign correction enables the terminal to be used for both shipping (inbound empty) and
receiving (inbound loaded) operations. Net sign correction can be either disabled or enabled on the
IND780. Refer to the IND780 Technical Manual, Chapter 3, Configuration, the Scale section, for
further information about disabling and enabling net sign correction.
If net sign correction is disabled in Setup, any stored weight value in the tare register is assumed to
be a tare regardless of the gross weight present on the scale at the time of the final transaction and
net values can be negative.
If net sign correction is enabled, the terminal will switch the gross weight and tare weight fields
when necessary, so that the larger weight is the gross weight, the smaller weight is the tare weight,
and the difference is always a positive net weight. Net sign correction affects the display, stored
data, weight recall, and printed data.
2-20
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Net sign correction will operate with pushbutton tare, preset tare, or tare memories. An example of
weight values with and without net sign correction is shown in Table 2-3. In this example, the tare
register value is 53 kg and the live weight on the scale is 16 kg.
Table 2-3: Weight Values With and Without Net Sign Correction
Net Sign Correction
Printed and Displayed
Gross
Disabled
Enabled
16 kg
53 kg
Tare
53 kg
16 kg
Net
−37 kg
37 kg
When net sign correction has been enabled, the tare weight field in the recall display will be
labeled with the letter “M” to indicate “Memory” instead of “T” or “PT.” Performing a transaction
switches the label to “PT.”
2.6.4.5.
Auto Tare
The IND780 can be configured so that tare is automatically taken (auto tare) after the weight on the
scale exceeds a programmed tare threshold weight. Auto tare can be configured in Setup as
enabled or disabled. When auto tare is enabled, the display changes to a zero net weight indication
after the weight exceeds the threshold value. The previous weight on the scale is stored in the Alibi
Table as the tare value. Auto tare operations involve:
• Tare Threshold Weight—When weight on the scale platform exceeds the tare threshold value,
and there is no motion, the terminal automatically tares.
• Reset Threshold Weight—The reset threshold weight must be less than the tare threshold
weight. When the weight on the scale platform falls below the reset threshold value, such as
when a load has been removed, the terminal automatically resets the auto tare trigger.
• Motion Check—A motion check is provided to control the re-arming of the auto tare function. If
disabled, the auto tare trigger will be reset as soon as the weight falls below the reset value. If
enabled, the weight must settle to no-motion below the reset threshold before the next auto tare
can be initiated.
Several conditions could inhibit the auto tare function:
• Motion—Auto tare cannot be taken when the scale is in motion. If motion is detected after the
weight on the scale exceeds a preset tare threshold weight, the IND780 will wait for a no-motion
condition. If a stable (no motion) weight condition occurs within three seconds, the auto tare
command is executed.
• Auto Tare Disabled—Auto tare can be configured in Setup as enabled or disabled.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-21
2.6.4.6.
Clearing Tare
Operating Instructions
Tare values can be cleared manually or automatically.
2.6.4.6.1.
Manual Clear
Manually clear tare values by pressing the CLEAR key
on the numeric keypad when the
IND780 is in the net mode and has completed the weighing operation. Motion on the scale will not
impact a manual clear.
If configured in Setup, pressing the ZERO scale function key
will first clear the tare before
issuing a zero command (refer to the IND780 Technical Manual, Chapter 3, Configuration, the
Scale section, Auto Clear).
2.6.4.6.2.
Auto Clear
The IND780 can be configured to automatically clear tare when the weight returns to a value below
a programmable threshold or by the print command. Once the tare is cleared, the display returns to
the gross weighing mode.
Auto clear is disabled or enabled in Setup. If auto clear is enabled, the following parameters as
configured in Setup affect the auto clear operation:
• Clear Threshold Weight—The clear threshold weight is the gross weight value below which the
IND780 will automatically clear a tare after settling to a value.
• Motion Check—A motion check is provided to control the automatic clearing of tare. If the
motion check is disabled, the tare value is cleared as soon as the weight drops below the
threshold weight (auto clear threshold), regardless of the motion status.
If Auto Clear and Motion Check are enabled in the Tare Auto Clear screen, once the Auto Clear
Threshold Weight value is passed and the then weight falls below the threshold (i.e. a load to be
weighed is placed on the scale and then removed), the IND780 will wait for a no-motion
condition then automatically clear the tare.
• Clear After Print—If enabled, tare is automatically cleared and the scale returned to the gross
mode after data has been transmitted by pressing the PRINT scale function key
or from a
remote source.
Refer to the IND780 Technical Manual, Chapter 3, Configuration, the Scale section, Auto Clear for
further information about configuring auto clear.
2.6.5.
Unit Switching
In order to support locations and applications that utilize multiple measurement units, the IND780
supports unit switching. The UNIT SWITCHING softkey
enables switching between primary units
(the main unit of measure) and alternate (primary or secondary) units. (For further information,
refer to the IND780 Technical Manual, Communication, Configuration, the Serial and PLC Interface
sections.)
When the UNIT SWITCHING softkey
is pressed, the display changes from the primary unit to the
secondary unit. The secondary unit may be a standard unit of measure or a custom unit. Custom
conversions support a division factor, a name and an increment size. An international standard unit
of measure should not be used as a conversion unit.
2-22
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
When units are switched, the units value changes to the correspondingly selected units, and the
conversion of the display value occurs. The display division changes to an equivalent weight value
in the switched unit (for example, from 0.02 lb to 0.01 kg) and the decimal location switches to
accommodate the conversion.
Figure 2-15 shows the Home screen with primary units (kilograms) displayed.
Figure 2-15: Home Screen With the Primary Units Displayed
Figure 2-16 shows the Home screen after the Unit Switching softkey
displaying the gross weight in pounds.
has been pressed,
Figure 2-16: Home Screen After the Unit Switching Softkey has Been Pressed
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-23
2.6.6.
IDNet Class II
Operating Instructions
If an IDNet scale base is set up as a Class II W&M approved scale with the appropriate capacity
and increment size, the IND780 will show a weight display with the last digit in a smaller size
(Figure 2-17).
Figure 2-17: Class II-Approved IDNet Base Weight Display
A demand print of weight data in this case will include parentheses around the last digit of each
weight field, to indicate the Class II approval condition.
Example:
Gross:
Tare:
Net:
2.6.7.
2.767(8) kg
1.719(3) kg T
1.048(5) kg
Expand By 10
The EXPAND BY 10 softkey
is used to increase the selected weight display resolution by one
additional digit. For example, a weight display of 40.96 could increase by one additional digit to
display as 40.958. When the EXPAND BY TEN softkey
is pressed again, the display reverts
to the normal display of weight.
If programmed as approved with the metrology switch ON, the Expand By Ten mode is displayed
for five seconds then automatically returns to normal resolution. When the weight is expanded and
the terminal is programmed as approved printing is disabled.
If an IDNet scale base is set up as a Class II W&M approved scale via the IDNet Service Mode, the
EXPAND BY 10 mode is unavailable, and will not display when the softkey
is pressed.
2.6.8.
Print
The print function (demand print) can be initiated by pressing the PRINT function key
or by
automatic print. Demand printing of data may also be initiated as part of a particular sequence of
operation or special application software. A “Printing” message appears in the system line for 3
seconds when the terminal is carrying out a demand print command.
2-24
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2.6.8.1.
Print Interlock
Print Interlock is designed to enforce a single print output per transaction. Print Interlock can be
disabled or enabled. If enabled, the print command is ignored until the measured gross weight
exceeds the print interlock threshold. After the first print command is executed, subsequent print
commands are ignored until the gross weight indication falls below the print interlock reset
threshold. If a print command is blocked by print Interlock, a synchronous “2nd Print Prohibited”
error is generated.
2.6.8.2.
Automatic Print
A demand print is automatically initiated when gross weight exceeds the minimum threshold and
there is no motion on the scale. After initiation, gross weight must return below the reset threshold
before a subsequent automatic print can occur.
If Automatic Print is enabled, it can be triggered and reset by weight exceeding thresholds, or by
weight deviation from a previously stable reading.
2.6.8.3.
Repeat Print
The Repeat Print softkey
permits the output of the most recent transaction, or of a Custom
Print from an application, to be printed again with a DUPLICATE header or footer to distinguish it
from the original print.
Output templates can be flagged with a DUPLICATE header or footer to indicate that the data in the
output template was generated as a repeat of a previous print transaction.
In firmware versions 6.5xx and above, the last printed data used for the Repeat Print function is not
retained during a power loss so the Repeat Print function will not be available after power is cycled
until after a print is generated.
2.6.9.
MinWeigh
Certain industries such as pharmaceuticals and food require a guarantee that the weighing
equipment selected for a particular measurement is adequate for the task. One way to ensure that
appropriate weighing equipment is selected is by creation and use of a minimum weighment value
(MinWeigh), below which a particular piece of weighing equipment cannot be used.
When the MinWeigh function is enabled, the MINWEIGH softkey
allows modification of the
MinWeigh value without entering Setup. If security has been enabled, a valid user must log in
before a change can be made.
The IND780 compares the current net weight with the programmed MinWeigh value. If the net
weight is greater than or equal to MinWeigh, all equipment functions behave normally. If the
absolute value of the net weight is less than MinWeigh, the weight display includes an icon ( ) to
the left of the weight, and the weight display appears in red. If the user attempts to record the weight
while in this condition, the printout includes an asterisk (∗).
For specific step-by-step instructions on how to setup MinWeigh, see the Scale | MinWeigh section
of Chapter 3, Configuration, in the IND780 Technical Manual.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-25
2.6.10.
Information Recall
Operating Instructions
Press the INFORMATION RECALL softkey
in Table 2-4 appear:
to access the Recall screen where the softkeys shown
Table 2-4: Recall Screen Softkeys
Icon
Explanation
Weight Recall—Displays ID, Gross, Tare, and Net active weight readings for each scale.
System Information Recall—Displays system information, including the model and
serial number, Terminal ID #1, Terminal ID #2, Terminal ID #3, and software and
hardware information. This ID information is entered in the Setup mode. The IND780
Technical Manual, Chapter 3, Configuration, Terminal, Device, gives details on entering
Terminal ID information.
Metrology Recall—Displays the terminal metrology controlled firmware version number,
whether the terminal is programmed as approved or not, and the last calibration time
and date for each scale. On the high-precision IDNet version, the Ident Code (calibration
tracking) is also displayed.
Totals Recall—Displays the sub-total and grand-total transaction count and total weight
for the terminal. Press the CLEAR SUBTOTAL softkey
to clear the sub-total amounts.
Press the CLEAR softkey
to clear both the grand-total and sub-total amounts. Press
the PRINT softkey
to print a report of the totals. Note that, if Clear Total or Clear
Subtotal After Print are enabled in Setup, these values will clear after printing.
2.6.11.
Target Comparison
Target comparison is used to compare either the gross or net weight on the scale to a
predetermined target value. This feature can be useful in both automatic and manual processes. As
an example, an automatic filling system using target comparison could provide a Start signal to the
IND780 and the IND780 could control the feeder system to fill a container to a desired target.
An example of a manual process could be a checkweighing station where an operator is checking
net weight of packages. The IND780 can use its target comparison feature to provide a graphical
Over - OK - Under indication to help the operator determine if the weight of each package is
acceptable or not compared to an ideal weight.
Target (setpoint) comparison is most often used in two types of applications:
• Material transfer applications—Requires that a control device deactivate when a target value is
achieved
• Over/Under applications—Classify a load placed on the scale platform as above or below the
target value or within the target tolerance range
The IND780 target comparison rate (for analog load cell scales) is 50 comparisons per second.
The comparison rate for high precision IDNet bases depends on the module in the base.
2-26
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Refer to the IND780 Technical Manual, Chapter 3, Configuration, the Application section, Memory,
Target Table and Operation, Target for further information about configuring parameters for target
comparison operations.
2.6.11.1.
Material Transfer Applications
If the target comparison of the IND780 will be used to control the flow of material, the application
can be classified as a Material Transfer application. These types of applications are usually
automated but can also be manual. A single or two-speed feeding system is used to either add
weight to or remove weight from the scale. The terminal monitors the change in weight and
compares it to a previously entered target and other control parameters.
Table 2-5 defines some terms used in Material Transfer applications.
Table 2-5: Terms used in Material Transfer applications
Term
64057247 | 13 | 08/2014
Explanation
Abort
After a target comparison process has been paused, it can be totally stopped by
pressing the Abort softkey
under the Target Control softkey
or by triggering a
Target Pause discrete input. If abort is selected, the target comparison process is
aborted.
Coincidence
Outputs
A coincidence output is always active and does not require a start or stop signal. If the
weight on the scale is below the target minus the spill value, the outputs are “on”. If
the weight is above the target minus the spill value, the outputs are “off”. This type of
output typically requires external logic to provide the required control for feeding
systems.
Concurrent
Outputs
This describes the operation of the feed output in a 2-speed feeding system. If the
feeder type is programmed as concurrent, both the fast feed and feed outputs are
active at the beginning of a target comparison. When the weight reaches the point at
which a slower feed is to occur, the fast feed output is turned off.
Fast Feed
Output
This refers to the physical output connection that is used for the faster feed in a 2speed feeding system. This output is not used in a single speed feeding system.
Feed Output
This refers to the physical output connection that is used for the slower feed in a 2speed feeding system or the only feed output in a single speed feeding system.
Fine Feed
The value entered for the amount of material that will be fed in the slower rate of feed
in a 2-speed feeding system. This value and the spill value are subtracted from the
target to determine the point that the fast feed output turns off.
Independent
Outputs
This describes the operation of the feed output in a 2-speed feeding system. If the
feeder type is programmed as independent, the feed output does not turn on until the
fast feed output turns off. Only one feeder output is active at a time. It is the opposite of
concurrent outputs.
Latched
Outputs
Latched outputs turn off at the target weight minus the spill value and remain off
(regardless of additional plus or minus weight changes) until a “start” signal is
received. These types of outputs do not usually require external logic to perform
standard weigh-in or weigh-out sequences.
METTLER TOLEDO IND780 User's Guide
2-27
Operating Instructions
Term
Explanation
Material
Transfer
Mode
The target comparison mode that provides control for delivering a measured amount of
material from one container or vessel to another. The transfer can apply to material
entering or leaving a container or vessel. It covers applications traditionally using
terms such as weigh-in, weigh-out, filling, and dosing.
Pause
A pause function is provided in target comparison control in case a target comparison
must be temporarily paused. This can be done by pressing the Pause softkey
under the Target Control softkey
or by triggering a discrete input programmed as
Target Pause. When paused, power is removed from the Feed and Fast Feed (if used)
discrete outputs. After a target comparison has been paused, the process can either be
resumed or aborted.
Resume
After a target comparison process has been paused, it may be resumed by pressing
the Resume softkey
or by triggering a discrete input programmed as Target
Resume. When a target comparison is resumed, it continues to use the original target
values.
Spill
The amount of material that will be added (on a weigh-in) or removed (on a weighout) from the scale after the final feed is turned off. In a weigh-in process, this is the
material in suspension that will still fall onto the scale when the feed is turned off. This
value is subtracted from the target value to determine when the feed output turns off.
Start
When using latched targets, the default condition is with the latch set or in the “off”
condition. To turn any outputs “on”, a start signal is required. This could be the Start
softkey
softkey
2.6.11.1.1.
, a discrete input programmed as Target Start, or via the Target Control
.
Target
The target is the weight value that is the end goal of the material transfer process. If a
container should be filled with 10 kg of material, the target value is 10 kg.
Tolerance
The weight range above and below the target value that will be acceptable as an “in
tolerance” target comparison. The tolerance can either be entered as a weight
deviation from the target or a percentage deviation from the target, depending upon
setup.
Target Control
Two different levels of target control are provided by the terminal for the Material Transfer mode.
Latching outputs must be enabled to utilize either type of automated control. For simple
applications, no control is provided. The comparison runs at all times. Whenever the scale weight
is less than the target value, the output is True/enabled.
For those applications where latching outputs are used or more control is needed for an occasional
pause or to abort a comparison process, the TARGET CONTROL softkey
is used. Pressing the
TARGET CONTROL softkey provides a dedicated display for the target comparison with the
comparison status and appropriate softkeys shown. Possible status modes are “Ready”, “Running”
or “Paused”. The appropriate control softkeys include START
, PAUSE
, and STOP
. Target
control functionality does not work when coincidence targets are configured.
When the weight data is invalid (e.g. under zero, over capacity), the Target Control screen will
show a status with dashes (- - - - -) and only the BACK softkey
will be available.
2-28
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
An example of each target control display is shown in Figure 2-18, Figure 2-19 and Figure 2-20.
Figure 2-18: Example of Ready State in Target Control
Figure 2-19: Example of Running State in Target Control
Figure 2-20: Example of Paused State in Target Control
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-29
2.6.11.2.
Over/Under Applications
Operating Instructions
If the target comparison of the IND780 will be used to checkweigh the weight of a product, the
application can be classified as an Over/Under application. These types of applications are usually
manual processes but can also be automated. The SmartTrac display indicates the classification of
the weight placed on the scale and discrete I/O can be used to trigger external lights or other
controls. The terminal compares the current weight to a previously entered target with tolerance
values, or to zone limits; it then indicates the results. In Table 2-6 definitions will explain some
terms used in Over/Under applications, and are specific to those applications:
Table 2-6: Over/Under Applications Terms
Term
2.6.11.3.
Explanation
Over Limit
If a tolerance mode of “Weight value” is selected in Setup, the IND780 does not require
the entry of a target value. In this mode, only the upper and lower limits for the OK
zone are entered. When programmed for this mode, the over limit value is the
maximum weight that will be classified as OK. Any weight above this value will be
classified as Over.
Over/Under
Mode
This target comparison mode provides three zone classifications for weight placed on
the scale. The classification can be Under if the weight is below the target minus the tolerance, OK if within the tolerance range or Over if the weight is above the target plus
the +tolerance.
Target
The target is the ideal weight value for the product being tested. If a container should
weigh exactly 10 kg, the target value is 10 kg.
Tolerance
The weight range above and below the target value that will be accepted as “OK” in the
target comparison. The tolerance can either be entered as a weight deviation from the
target, a percentage deviation from the target or as absolute weight (zone edges)
depending upon setup.
Under Limit
If a tolerance mode of “Weight value” is selected in Setup, the IND780 does not require
the entry of a target value. In this mode, only the upper and lower limits for the OK
zone are entered. When programmed for this mode, the under limit value is the
minimum weight that will be classified as OK. Any weight below than this value will be
classified as Under.
Zones
The IND780 provides 3 different classifications or “zones” in the Over/Under mode. The
zones are Under, OK and Over.
Target Comparison Parameter Entry
The active record is the record currently in use by the terminal.
2.6.11.3.1.
Direct Editing of Active Records
To directly edit an active record:
1. Press the TARGET softkey
. (Refer to the IND780 Technical Manual Appendix E, Softkey
Mapping and Application Key Configuration, for information about how to set up softkeys.) The
Edit Target screen displays. The example shown in Figure 2-21 shows the fields available in
Material Transfer mode.
2-30
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 2-21: Edit Target Screen
2. Use the UP and DOWN navigation keys to scroll through the fields available for editing.
3. Press the ENTER key to select a field to edit.
Depending whether material transfer or over/under mode was selected and the type of tolerance
selected, the active target record will show different fields to edit. Use the navigation and
numeric keys to enter values for all of the fields as needed (Table 2-7).
Table 2-7: Target Record Fields by Mode and Tolerance
Material Transfer Mode
Tolerance – Either Mode
Over/Under Mode Tolerance –
Weight Deviation or % of Target
Over/Under Mode Tolerance –
Weight Value
Target
Target
Over Limit
+ Tolerance
+ Tolerance
Under Limit
- Tolerance
- Tolerance
Spill
Fine Feed
4. Use the alpha keys to enter a description. Refer to the Alpha Keys section of Understanding the
Navigational Interface.
5. Press the OK softkey
to accept the edits to the active record, or Press the ESCAPE softkey
to return to the weighing operation screen without saving the active record edits.
2.6.11.3.2.
Loading Records from the Target Table
The Target Memory softkey
can be used in two ways to load records from the Target Table:
Quick Access and List Selection.
Quick Access
Use the Quick Access mode when the ID of the Target Table record to be loaded is known. Use the
numeric keypad to enter the ID and then press the Target Memory softkey
to load the record. If
the record is available, the data is loaded. If the record is not found, an “ID not found” error
displays and the currently loaded target is retained.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-31
List Selection
Operating Instructions
Use the List Selection mode when the ID of the Target Table record is unknown. To use the List
Selection mode:
1. Press the Target Memory softkey
displays.
without any preceding data entry. The Target Search screen
2. Enter any search restrictions required or leave selections as they are to retrieve all records.
3. Press the SEARCH softkey
to view the selected records in the table.
4. Use the UP and DOWN navigation keys to scroll through the list until the desired record is
highlighted.
5. Press the OK softkey
to load the selected record from the list and return to the weighing
operation screen, or Press the EXIT softkey
to return without loading the record.
2.6.12.
Comparators
Comparators are simple targets, twenty of which can be configured in setup. They are controlled
either by coincidence or by comparison with a target or range. The source for comparison can be
the Gross Weight, Displayed Weight, Rate or assigned by a custom TaskExpert application. When
assigned to the home screen, the Comparator softkey
permits direct access to the Comparators
for the current selected scale. As seen in Figure 2-22, this screen displays each Comparator’s ID,
Description, Limit, Units, High Limit (if applicable) and Active operator, provided that they have
been configured in setup.
To make changes to a Comparator directly from the home screen, if the Comparator ID is known,
press the ID number and then the Comparator softkey
. If the ID is not known, press the
Comparator softkey
, use the UP/DOWN keys to select the desired Comparator from the list, and
press the Edit softkey
. Only the limit value/s can be edited. To edit parameters other than limits,
the Comparator configuration screen must be accessed in setup.
Figure 2-22: Comparators List for Scale 2
When the Active operator value is <, <=, =, < >, >= or >, the selected source is compared to a
target weight. When the Active value is within (>_ _<) or outside (_< >_) a range, the selected
source is compared to a range defined by the Limit and High Limit values.
2-32
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2.6.12.1.
Using Comparators
To use a Comparator, it must be assigned to a discrete output. Comparators are not associated
with SmartTrac.
2.6.12.1.1.
Target Value Comparators
In this case, the output will be triggered depending on the Active setting. For example, if the Active
operator is < (less than) and the Limit is 1000 kg, the output will turn OFF when the measured
value reaches 1000 kg.
2.6.12.1.2.
Range Comparators
In the case of a Range mode Comparator, the Limit value sets the lower of the two limits that define
the range, within or outside (depending on the operator) which the output is active. For example, if
the Active operator is Within (>_ _<), the Limit is 3.0 kg/sec and the High Limit is 8 kg/sec, the
output will be active provided the rate remains between 3 and 8 kg/sec. If the rate falls below 3 or
exceeds 8 kg/sec, the output will turn OFF.
Note that comparisons using a Rate source require the rate calculation to be enabled and the
Rate display to be turned on. The Rate value configured in setup at Scale > Rate permits weight
and time units to be defined, together with the period between individual measurements and the
period over which the measurements are averaged to generate the output value.
A Range Comparator can also use Displayed or Gross weight as its source. For example, if
displayed weight (possibly a net value, allowing for the weight of a hopper) is the source, the
Active operator is Outside, and the Limit and High Limit are 1000 kg and 1500 kg respectively, the
output will remain active as long as the source value is below 1000 kg or above 1500 kg. In this
case a large enough value for the High Limit should be set to ensure that, once the Limit is reached
and the output turned off, there is not sufficient spill to reach the High Limit and reactivate the
output.
2.6.13.
ID Mode
The ID function provides a convenient and simple way to ensure that a specific sequence of
operation is carried out the same way each time. Each step can include an on-screen message that
instructs the operator to perform an action, or a prompt asking the operator to enter some
information for a transaction printout. These steps may include data entry (using the numeric
keypad, alphanumeric keys, or some other input such as a barcode scanner) and actions (such as
placing a package on the scale, taking a tare, or printing a ticket for the transaction). When a step
is completed, depending on the type of step it will either automatically advance or pressing the
ENTER key moves it to the next step.
The following list shows the types of steps that are programmable in a sequence.
64057247 | 13 | 08/2014
• Alphanumeric prompt input
• Select Tare ID
• Clear Tare
• Select Target ID
• Numeric prompt input
• Start Sequence
• Print
• Tare Automatically
• Select Scale
• Tare Preset
METTLER TOLEDO IND780 User's Guide
2-33
Operating Instructions
The ID function can be used in combination with the Material Transfer Control function and latched
outputs to perform a simple semi-automatic filling sequence. To allow for this operation, the use of
Target -Start, Target-Pause and Target-Resume discrete inputs is also necessary. An example of an
operator guided container fill sequence is shown in Table 2-8.
Table 2-8: Container Fill Sequence ID Example
Sequence Steps
Type
Enter Batch No.
Alphanumeric Prompt
Load container
Alphanumeric Prompt
Scale is tared automatically
Tare-Auto
Select Material ID
Select Target ID
Press target start. Feed finished?
(Target-Start discrete input is triggered, and wait for feed to complete,
then acknowledge prompt).
Alphanumeric Prompt
Information is printed automatically
Print
Remove container
Alphanumeric Prompt
Tare is cleared automatically
Clear Tare
The IND780 supports two separate ID sequences, ID1 and ID2. As many as 20 steps may be
programmed in each sequence, and the sequence can be triggered in one of two ways:
•
Manually, using the ID softkeys
or
, application keys A1-A4, discrete input or
shared data triggers xc0149 (ID1) and xc0150 (ID2).
•
Automatically, by threshold and reset weight readings from the pre-defined scale.
If the sequence is triggered manually, it may be programmed to continuously loop using the Start
is pressed or an ESC softkey
Sequence step, repeating all steps until either the EXIT softkey
is pressed while an alphanumeric entry field is selected.
When ID mode is set to function automatically, placing on the pre-defined scale a weight that
exceeds a pre-configured threshold value initiates the respective sequence. When all the steps have
been performed or the sequence is exited and the weight is removed, the reading from the scale
falls below a user-defined reset value. At this point the terminal is ready to begin the next ID
sequence.
Only one ID sequence can be triggered to run at a time. Starting an ID sequence while one is
currently running will generate a system line message indicating that the ID sequence is busy. If a
step fails to execute due to an invalid operation (eg. print connection not found, or tare too small),
the ID sequence will be exited automatically and will need to be restarted.
Typically the step’s number, prompt and operator data entry box (if applicable) appear together for
each step in the Message Area below the weight and SmartTrac display (if enabled). Considering
space restrictions, the terminal will attempt to resize resize its on-screen objects to fit the prompts
and data entry box, as shown in Figure 2-23 below. If an automatic resize is not possible, the
2-34
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
prompts and entry box will be superimposed on the existing screen objects (see Figure 2-24).
When the sequence is completed or exited, the display objects will return to their original sizes.
Figure 2-23: Single Scale with SmartTrac and ID Sequence Display
Figure 2-24: Two Scales plus Sum with SmartTrac and ID Sequence Display
For details on ID mode configuration, refer to the IND780 Technical Manual, Chapter 3,
Configuration.
2.6.14.
SmartTrac™
SmartTrac is a graphical display visualization of either the displayed weight or gross weight as
selected in Setup. The visualization may be a bar graph, cross hairs, or an over/under display. The
following conditions allow SmartTrac to display:
• A Target must be defined and selected.
• A SmartTrac size other than none must be selected at Setup > Terminal > Display.
• A SmartTrac graphic display type must be selected for the scale at Setup > Application >
Operation > Target > Scale #.
Refer to the IND780 Technical Manual, Chapter 3, Configuration, for details about these settings.
2.6.14.1.
Display Sizes
Figure 2-25 shows a small size SmartTrac bar graph. Figure 2-26 shows an over/under display in
medium size and Figure 2-27 a cross hairs display in large size. Note the effect on the main weight
display area. In each of these examples, the display shows the weight on target. The small bar
graph does not include target weight, description and tolerance information; with a medium or large
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-35
Operating Instructions
bar graph, the target description appears above the bar and target and tolerance below it (see
Figure 2-28).
Figure 2-25: Small Size SmartTrac, Bar Graph
Figure 2-26: Medium Size SmartTrac, Over/Under
Figure 2-27: Large Size SmartTrac, Cross Hairs
2-36
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2.6.14.2.
Display Colors
• Red—Over tolerance
• Green—Within tolerance
• Blue—Under tolerance
In cross hairs mode, the colors listed above are used as background fill. In this case, the cross
hairs appear in white when within tolerance (see Figure 2-27), and yellow when over or under
tolerance.
2.6.14.3.
Bar Graph Mode
The operator adds material until the measured value is within an acceptable tolerance limit.
Normally an operator will add material quickly when the container is not nearly full and more
slowly as the target value approaches. Figure 2-28 shows a SmartTrac bar graph with a displayed
value below the lower tolerance limit.
Figure 2-28: Bar Graph, Large SmartTrac Display
While the display appears to be one continuous bar graph, it can actually be broken into four
separate sections or ranges. Depending upon the target and tolerance values used, there could be
one, two, or three speeds at which these graphical sections are filled in. Figure 2-29 illustrates the
ranges, and includes the numerical display of the target and tolerance values. Note the white
triangle, indicating that in this case the measured weight is precisely on target.
Figure 2-29: Medium Size SmartTrac Bar Graph With Display Ranges
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-37
Operating Instructions
2.6.14.3.1.
Under Tolerance
The Under Tolerance range is labeled “A” in Figure 2-29, and represents the amount of material
between 10% of the target value and the target minus the negative tolerance value. The bar graph
begins to populate the “A” range when more than 10% of the target has been added. Until the 10%
point is reached, no fill is shown.
2.6.14.3.2.
Acceptable Tolerance Under Target
Region B in Figure 2-29 represents the zone of acceptable tolerance below the target value.
2.6.14.3.3.
On Target
When the measured value exactly equals the target value, an unfilled triangle indication displays,
as shown in Figure 2-29.
2.6.14.3.4.
Acceptable Tolerance Over Target
Region C in Figure 2-29 represents the zone of acceptable tolerance above the target value.
2.6.14.3.5.
Over Tolerance
Region D in Figure 2-29 represents the zone of unacceptable tolerance above the target value. In
Region D, the fill rate reverts to the same rate as used in Region A. If the measured value is large
enough that the number of dot columns required for display exceeds the number available, the
display saturates. Further increases in the measured value do not affect the graphical display.
2.6.14.4.
Over/Under Mode
The Over/Under SmartTrac visualization may be displayed in medium or large sizes. If the Target
Table’s Tolerance Type is Target Deviation or % of Target, the target value and the amount of
deviation from it are shown, together with a graphic indicating the direction of deviation (Figure
2-31 and Figure 2-32). If the Target Table’s Tolerance Type is Weight Value, only the graphic
display appears indicating whether the current value is above, within or below the acceptable range
(Figure 2-34). In either case, a black oblong indicates that the current weight is within the
acceptable range.
Figure 2-30: Example of Over/Under Display
2-38
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 2-31 shows the Over/Under portion of the display, indicating weights above the high
tolerance limit (top) and below the low tolerance limit (bottom).
Over tolerance
arrow
Amount over
tolerance
Target
description
High
tolerance
Target
weight
Low
tolerance
Amount under
tolerance
Under tolerance
arrow
Figure 2-31: Elements of the Over/Under SmartTrac Display
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-39
Operating Instructions
In the Over/Under Mode, when Tolerance Type is set to Weight Value, the graphic displays as
shown in Figure 2-32. When Tolerance Type is set to either of the other two values, the table
requires a target value to be entered. From top to bottom are graphics for over tolerance, under
tolerance and in tolerance.
Figure 2-32: Over/Under SmartTrac Displays, No Target Weight
2-40
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2.6.14.5.
Cross Hairs Mode
The cross hairs SmartTrac visualization may be displayed in medium or large sizes. Figure 2-33
shows a large size display indicating that the measured weight has been reached. The cross hairs
become an expanded “eye” to indicate that the weight is precisely on target. The target weight and
tolerance values display to the right of the screen.
Figure 2-33: Cross Hairs SmartTrac Display, On Target
Figure 2-34 shows three cross hairs displays. From top to bottom, they are above high tolerance;
in tolerance range but below target; and below low tolerance. The horizontal bar moves vertically as
the measured weight approaches the tolerance range, giving a coarse indication of the relationship
between current weight and target weight. The vertical bar begins to move when measured weight
approaches the lower tolerance line, giving a fine indication of the relationship between current
weight and target weight. The intersection of these bars indicates the current weight.
Coarse target
weight line
Fine target weight
line
Weight
within
tolerance
Vertical
(fine) bar
Low
tolerance
line
High
tolerance
line
Horizontal
(coarse)
bar
Figure 2-34: Cross Hairs SmartTrac Displays, Off Target
2.6.15.
Clustering
2.6.15.1.
Overview
Clustering is a means of networking up to 20 individual IND780 terminals together in a peer-topeer Ethernet network. This is accomplished by assigning each IND780 terminal a unique IP
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-41
Operating Instructions
address and a terminal number. Within a cluster, terminals can share Shared Data, operator
consoles, printers, and PLC interfaces. The IND780 also extends its clustering services to host PCs
or file servers on an Ethernet LAN for data exchange.
An IND780 terminal can operate as a remote operator console to any clustered IND780. A
dedicated SELECT TERMINAL softkey
, assigned to the home screen, can be pressed to open a
list of names of clustered terminals. Once a terminal is selected, pressing OK
allows the local
terminal to access the display, keypad and setup configuration of the remote terminal. The local
terminal’s display is replaced with that of the selected remote terminal. All front panel keys now
function as if they belonged to the remote terminal, and softkeys assigned on the remote terminal
appear on the home screen. The local terminal will not lock out the controls for the remote terminal
during this remote connection.
2.6.15.2.
Viewing a Clustered Terminal Remotely
To select a clustered terminal to view from the home screen, press the SELECT TERMINAL softkey
. A screen like the one in Figure 2-35 appears, with a drop down list including all clustered
terminals.
Figure 2-35: List of Clustered Terminals
2-42
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Select the desired terminal and press ENTER, then press the OK softkey to confirm the choice. If the
selected terminal is already being viewed by another terminal, a message will appear:
Figure 2-36: Remote Viewer Disconnect Screen
Press OK to confirm the disconnection, or ESC to return to the home screen without disconnecting.
Once the remote view connection has been closed, the clustered terminal is again available to be
selected for remote viewing.
If the connection to the remote terminal is successful, after a brief delay the current home screen
display from that terminal will appear. If the connection is unsuccessful, an error message might
appear (Figure 2-37) indicating that the remote terminal’s network/view server is disabled in setup.
Figure 2-37: Remote Viewer Not Enabled Error Message
To disconnect from the remote viewer, press the SELECT TERMINAL softkey and follow the screen
instructions as shown in Figure 2-36.
2.6.16.
Time and Date
Time and date are used for reporting, error and transaction log timestamps, and triggering service
events. The Time & Date softkey
accesses the Set Time & Date screen where the user can set
the time and date, including the hours, minutes, day, month, and year. When the time is set,
seconds are set to 0. The portion of the system line which displays date and time can be enabled
or disabled (the default setting) from the Format Time & Date screen.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-43
Although the format for the time and date can be selected according to local preferences, the use of
a timestamp in log files is not selectable. Timestamp formats are always fixed as:
Operating Instructions
• Date: YYYY/MM/DD (for example, July 23, 2005 becomes the fixed format date 2005/07/23)
• Time: HR:MM:SS stored in 24 hour format (for example, 10:01:22 PM becomes the fixed
format time 22:01:22). Seconds are not displayed on the screen.
2.6.17.
Reports
Reports may be generated, viewed, and printed from database tables including:
• Alibi Memory
• Tare Table
• Target Table
The REPORTS softkey
must display as a softkey or be assigned to an Application key (A1—A4)
to generate table reports. Otherwise, table reports can only be generated from within Setup.
To generate a report:
1. Press the REPORTS softkey
. The Reports Run screen appears.
2. While the report is printing, the CANCEL softkey
operation.
displays. Pressing it will cancel the print
3. To select a sub-set of records from the table, press the TABLE SEARCH softkey
Screen for the selected report type displays (see Figure 2-38).
. The Search
Search Field
selection box
Comparison
Data entry
fields
Data Sort
selection box
SEARCH
softkey
Figure 2-38: Tare Table Search Screen
When Alibi is selected in the Report Type field and the Table Search softkey is pressed, allow
some time for the Alibi Search View table to populate.
4. When the table Search View is on screen, the PRINT softkey
displays again. Pressing
PRINT will print the entire contents of the table, and a “Printing” system line message appears
for 5 seconds.
2-44
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2.6.18.
Calibration Test
The IND780 terminal provides a programmable Calibration Test sequence that can contain up to
25 individual steps. The Calibration Test is designed to lead the tester through a set of
predetermined steps and compare the terminal calibration with known test weights.
The test is accessed by pressing the CALIBRATION TEST softkey
on the home page or as an
Application key (A1—A4). A screen (Figure 2-39) displays, including a field for the name of the
tester.
Figure 2-39: Calibration Test Name-Entry Screen
Once the technician’s name has been entered, press the Test Weight Information softkey
screen shown in Figure 2-40 opens.
. The
Figure 2-40: Calibration Test Weight View
Use the NEW
and EDIT
softkeys to enter the weight and serial number of each test weight to
be used. The weights and weight units should match those programmed for the Calibration Test.
This information provides a traceable record for the calibration test.
Once the test weights are entered, return to the Calibration Test screen by pressing the EXIT softkey
.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-45
Operating Instructions
The calibration test is started by pressing the START softkey
. A display is shown (Figure 2-41)
that provides the active scale weight, the target and tolerance weight values and, at the bottom,
instructions (as to where to place the weights).
Figure 2-41: Calibration Test Step
When the tester has carried out these instructions, he or she presses the OK softkey
, and the
terminal compares the actual weight on the scale to the programmed target weight for this step. If
the comparison passes, the test continues to the next step. If it fails, a message (Figure 2-42)
displays indicating that the result is out of tolerance.
Figure 2-42: Tolerance Error Message
After the error is acknowledged, the screen shown in Figure 2-43 displays, and the tester can abort
the test (by pressing the EXIT softkey
), retest this step (by pressing the OK softkey
), or skip
this step (by pressing the SKIP softkey
) knowing that it failed.
2-46
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 2-43: Skip Calibration Test Step
After progressing through all the steps in the calibration test procedure, a Test Complete message
displays, along with a status message of either pass or fail (Figure 2-44). Pressing the PRINT
softkey
generates a printout of the calibration test report.
Figure 2-44: Calibration Test Complete Screen
A full description of how to program the calibration test can be found in Chapter 3 of the IND780
Technical Manual, Configuration, in the Maintenance | Configure/View, Calibration Test section.
If a SICS scale is present, an internal calibration test function is also available, provided the base
supports the ‘TST3’ SICS function. The tester can initiate the internal calibration test by pressing the
INTERNAL CALIBRATION TEST softkey
from the Calibration Test screen. The Internal Calibration
Test screen displays with a tolerance weight value and a status message directing the tester to
empty the scale and to press the START softkey (Figure 2-45).
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
2-47
Operating Instructions
Figure 2-45: Internal Calibration Test Screen
The tolerance value indicated can be programmed only in the setup mode. It is used as the
pass/fail criterion of the calibration test.
The test process is started by first emptying the scale and then pressing the START softkey
“Testing” message displays, to indicate the status of the internal calibration operation.
.A
When the test operation is completed, a Test Complete message displays along with a Status
message of either Pass or Fail (Figure 2-46). The weight deviation since the last calibration is also
indicated. The calibration test report can be printed using the PRINT softkey
.
Figure 2-46: Internal Calibration Test Screen, Complete
2.7.
Alibi Memory Direct Access
The Alibi Memory stores transaction data that can be retrieved in order to verify transaction
information.
The information stored in the Alibi Memory includes:
• Transaction counter value
• Date and time of transaction
2-48
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
• Gross, net, and tare weights including units of measure
To view the Alibi Memory:
1. Depending which is programmed as one of the home page softkeys, press the Alibi softkey
, or press the REPORTS softkey
and select Alibi Memory from the report selection
box.
2. Press the VIEW TABLE softkey
. The Alibi Search View screen displays, with two Search
Fields and associated Data limiting boxes.
3. To limit the data:
A
Use the Search Field selection box to select a desired search field. The Alibi Search screen
has two sets of limiting fields, allowing the search to be further limited. Table 2-9 lists the
options by Table, with default values indicated with an asterisk.
Table 2-9: Table Search Field Options
Table
Search Field Options
Alibi
None*, Date (2005/08/21), Time (18:27:44), Transaction Counter
Tare
ID*, Description, Tare
Target
B
ID*, Description, Target, +Tol, -Tol
Use the Comparison selection box to select how the data will be limited. Table 2-10 shows
the options offered in this selection box.
Table 2-10: Table Comparison Field Options
Symbol
64057247 | 13 | 08/2014
Comparison
Symbol
Comparison
<
Is less than
<>
Is not equal to
<=
Is less than or equal to
>=
Is greater than or equal to
=*
Is equal to (default)
>
Is greater than
C
Use the numeric keypad to enter the limiting factor in the text box (the * character is the
“wild-card” character and returns all results).
D
For the Tare and Target Tables, select the Sort By method. Sort By offers the same choices
(Table 2-10) as the Search Field.
E
Press the SEARCH softkey
to view the limited data in the table. Unless specified
otherwise in the Sort By field of the Search screen, Tare and Target records are always
listed in order by ID, lowest to highest, Alibi records in order by date and time, oldest first.
Figure 2-47 shows an example of search results for a Tare Table. The UP, DOWN, LEFT
and RIGHT arrows can be used to scroll around this view in order to see more rows and
columns of data.
METTLER TOLEDO IND780 User's Guide
2-49
Operating Instructions
Figure 2-47: Tare Table Search View
F
The PRINT softkey
can be used to print a report of the selected data.
4. Press the SEARCH softkey
. The Alibi Search View screen displays the search results.
Records are ordered by date and time, with the most recent record shown last.
Figure 2-48: Alibi Search View
2.8.
Table Searches
Table searches can be accessed by various methods including the TARE TABLE softkey , TARGET
TABLE softkey , ALIBI softkey
, REPORTS softkey
, and from the setup menu tree.
To search a table:
1. Access the table to be viewed using one of the methods noted above. Refer to the IND780
Technical Manual Appendix C, Table and Log File Structure for more information about how to
access specific tables.
2. If accessing the table from the Reports Run screen, press the TABLE SEARCH softkey
. The
Search View screen displays. If a table is accessed from its dedicated softkey, the Search View
screen displays directly.
2-50
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3. Use the Search Field, Data entry and Sort By boxes to enter specific search information to limit
the search, or accept the default * (the “wild-card” character) to view all table information.
Options in the search screens are as described in Table 2-9 and Table 2-10.
4. Press the SEARCH softkey
64057247 | 13 | 08/2014
. The Search View screen displays with the search results.
METTLER TOLEDO IND780 User's Guide
2-51
3
Service and Maintenance
This chapter covers
•
•
•
•
•
•
•
•
•
•
•
List of Tools Required
Cleaning and Maintenance
Upgrading Firmware
Backup and Restore of system configuration
Service
Troubleshooting
The IND780 terminal is designed to provide years of dependable
operation. However, METTLER TOLEDO recommends that – as with any
industrial measurement equipment – the IND780 terminal and the
connected scale system be serviced periodically. Timely, factory-specified
maintenance and calibration by a METTLER TOLEDO service technician
will ensure and document accurate and dependable performance to
specifications.
Main PCB
ETX Board
Power Supply
Option Board Installation
PLC Interface Module Removal and
Installation
3.1.
Precautions
•
READ this manual BEFORE operating or servicing this equipment and FOLLOW these
instructions carefully.
•
SAVE this manual for future reference.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL. EXERCISE CARE WHEN
MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING
TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY
DAMAGE.
WARNING
IF THE KEYBOARD, DISPLAY LENS OR ENCLOSURE IS DAMAGED ON A DIVISION 2 APPROVED
OR CATEGORY 3 MARKED IND780 TERMINAL THAT IS USED IN A DIVISION 2 OR ZONE 2/22
AREA, THE DEFECTIVE COMPONENT MUST BE REPAIRED IMMEDIATELY. REMOVE AC POWER
IMMEDIATELY AND DO NOT REAPPLY AC POWER UNTIL THE DISPLAY LENS, KEYBOARD OR
ENCLOSURE HAS BEEN REPAIRED OR REPLACED BY QUALIFIED SERVICE PERSONNEL. FAILURE
TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-1
WARNING
Service and Maintenance
WHEN THIS EQUIPMENT IS INCLUDED AS A COMPONENT PART OF A SYSTEM, THE
RESULTING DESIGN MUST BE REVIEWED BY QUALIFIED PERSONNEL WHO ARE FAMILIAR
WITH THE CONSTRUCTION AND OPERATION OF ALL COMPONENTS IN THE SYSTEM AND THE
POTENTIAL HAZARDS INVOLVED. FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT IN
BODILY HARM AND/OR PROPERTY DAMAGE.
CAUTION
RISK OF EXPLOSION IF BATTERY IS REPLACED WITH WRONG TYPE OR CONNECTED
IMPROPERLY. DISPOSE OF BATTERY ACCORDING TO LOCAL LAWS AND REGULATIONS.
CAUTION
BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR INTERCONNECTING
WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY (30)
SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE
PRECAUTIONS COULD RESULT IN DAMAGE TO OR DESTRUCTION OF THE EQUIPMENT AND/OR BODILY HARM.
NOTICE
OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES.
3.2.
List of Tools Required
The following tools are required to perform these procedures:
• Soft cloth and mild glass cleaning solution
• Voltmeter
• Anti-static mat and wrist strap
• Two sizes of Philips head screwdrivers
• Flat blade screwdriver
• Wrenches to fit 33 mm and 19 mm, for use with cable glands
• Nut driver with 7 and 8 mm sockets
3.3.
Cleaning and Maintenance
Clean the IND780 terminal’s keypad and cover with a clean, soft cloth that has been dampened
with a mild glass cleaner. Do not use any type of industrial solvent such as toluene or isopropanol
(IPA) that could damage the terminal’s finish. Do not spray cleaner directly on the terminal.
Regular maintenance inspections and calibration by a qualified service technician are
recommended. The IND780 is a rugged stainless steel enclosed instrument; however, the front
3-2
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
panel is a polyester covering over sensitive electronic switches and a lighted display. Care should
be taken to avoid any punctures to this surface or any vibrations or shocks to the instrument.
Should the front panel become punctured, ensure that steps are taken to prevent dust and moisture
from entering the unit until the terminal can be repaired.
3.4.
Firmware Upgrade Procedure
There are two ways to install an upgraded version of the IND780’s firmware. The preferred method
uses the PC-based InSite™ software tool, but it is also possible to do so without a computer, using
a USB flash memory drive (thumb drive) connected to the IND780’s USB port.
3.4.1.
Upgrade Using InSite™
To upgrade existing installed field units, two upgrade files (Upgrade.L78 and Upgrade.Z78) are
required. These files and the method described below can only be used if you are upgrading from
IND780 revision 3.xx or higher and the terminal has at least 256MB of CF card memory.
If the terminal has a 2.xx or lower revision firmware or a smaller CF card memory, the upgrade
operation requires a re-format of the CF memory card and a complete flash of the new firmware files
and folders structure. For this condition, contact a local METTLER TOLEDO representative for
assistance in upgrading your terminal.
With the IND780 connected to a network:
1. Run the InSite Configuration Tool (only version 5 or higher).
2. In InSite, access Home | Connection I Settings, or click the Settings icon
address the terminal.
, to set up InSite to
3. Access Options | Flash Download, or click on the Flash Download icon
. From the
Firmware Download window (Figure 3-1), select the files to copy to the IND780’s Upgrade
folder. Refer to the InSite documentation or help system for details on performing this step.
The files selected for download must not be read-only. If necessary, right-click the files in
Windows Explorer, select Properties, and modify their attributes – clear the Read-Only box at
lower left, and then click OK to confirm the change.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-3
Service and Maintenance
Figure 3-1: InSite Firmware Upgrade File Selection Screen
4. Press the SETUP softkey
to access the IND780’s setup tree. When prompted, enter a valid
user name and password at an access level that permits setup parameters to be changed.
5. Use the arrow keys to move focus to Maintenance > Run > Install Software Upgrade. When the
node is in focus, press ENTER. The screen shown in Figure 3-2 will appear.
Figure 3-2: Firmware Upgrade Screen
If the terminal is sealed this screen will appear without a START button, and with a Status
message that software upgrade will not be permitted when the metrological security switch S-1
is ON (locked). The terminal must be unsealed and the switch set to OFF, in order to permit a
software upgrade.
6. By default, the Internal File option is selected in the Install from list. This is the correct setting
for upgrade from files downloaded using InSite.
7. Press the START softkey
to perform the upgrade. If the files are not present in the Upgrade
folder, the screen shown in Figure 3-3 appears. Performing the copy operation from InSite will
3-4
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
continue the process. Pressing the EXIT softkey
setup tree.
abandons the process and returns to the
Figure 3-3: Firmware Upgrade – Waiting for Files
8. If the IND780 finds the files in the UPGRADE directory, the screen shown in Figure 3-4 appears
briefly, and the terminal restarts.
Figure 3-4: Firmware Upgrade in Progress
9. On-screen messages will indicate the progress of the upgrade. Do not disconnect power from
the IND780 during this procedure. The terminal will back up existing files, unpack the upgrade
files and perform clean-up operations. Once this process is complete, start-up continues as
normal.
10. If prompted to do so, manually cycle power to the terminal to complete the process.
3.4.2.
Upgrade Using USB Memory
To upgrade existing installed field units, two upgrade files (Upgrade.L78 and Upgrade.Z78) are
required. These files and the method described below can only be used if you are upgrading from
IND780 revision 3.xx or higher and the terminal has at least 256MB of CF card memory.
If the terminal has a 2.xx or lower revision firmware or a smaller CF card memory, the upgrade
operation requires a re-format of the CF memory card and a complete flash of the new firmware files
and folders structure. For this condition, contact a local METTLER TOLEDO representative for
assistance in upgrading your terminal.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-5
In general, USB drives that are USB 2.0 certified and 1.1 compatible are supported by the IND780.
To perform an upgrade:
Service and Maintenance
1. Create a folder called UPGRADE in the root of a USB Memory drive (“thumb drive”).
2. Copy the two UPGRADE files into the UPGRADE folder.
3. Connect the USB Memory to the IND780’s USB port, either by plugging it in directly or via a
USB extension cable connected to the port.
4. Access Setup > Maintenance > Run > Install Software Upgrade, as above.
5. With focus on the Install from box, press ENTER, and use the DOWN arrow key to highlight USB
Memory, as shown in Figure 3-5, then press ENTER to make the selection.
Figure 3-5: Firmware Upgrade From USB Memory
If the terminal is sealed this screen will appear without a START button, and with a Status
message that software upgrade will not be permitted if the metrological security switch S-1 is
ON (locked). The terminal must be un-sealed and the switch set to OFF in order to permit a
software upgrade.
6. Press the START softkey
. If the USB drive is not attached or the folder isn’t present, a Status
message will appear – “USB Memory Error.” Otherwise, the terminal will display a message
“Copying USB to CF.” Wait while the process is complete, and another message appears –
“Unplug USB Memory.” The terminal automatically re-starts. Status messages will indicate that
files are being backed up, unpacked and upgraded. Do not turn off power to the terminal during
this procedure.
7. If prompted to do so, manually cycle power to the terminal to complete the process.
3-6
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.4.3.
Power Up and Screen Saver Graphics
The graphic used by the IND780 as a splash screen during start up and the image used as a
screen saver (which displays when the Screen Saver setting at Terminal > Display has a value
other than zero) can both be changed by downloading replacement files to the terminal’s UPGRADE
folder.
3.4.3.1.
Power Up Graphic
The graphic used at power up must be a bitmap file measuring no more than 320 pixels wide and
140 pixels high. It must be named pwrup.bmp.
3.4.3.2.
Screen Saver Graphic
The screen saver graphic is also a bitmap file, which may be in 256 color format. The image
should measure no more than 320 pixels wide and 200 pixels high, and must be named
saver.bmp.
3.4.3.3.
Installing New Graphics
Either or both of the power up and screen saver graphics can be installed on the IND780 by
making an ftp connection to the terminal via EtherNet, logging on as an administrator (with write
privileges), and copying the file/s to the UPGRADE folder. When the terminal is restarted, the files
are automatically moved to their correct location, where they replace the existing images.
3.5.
System Backup and Restore Procedures
The configuration of an IND780 may be exported (backed up) to and restored from, USB memory.
Files can be restored to the original terminal or copied to another IND780, which acquires a
configuration identical to that of the original. This permits the restoration of setup should a main
PCB require replacement, for example, and allows the creation of functionally identical terminals.
Once the backup dataset is copied to a USB drive, it can be saved to a PC. However, for
restoration to an IND780 the file must always reside in the directory indicated below.
3.5.1.
Backup
When the IND780 is backed up, the dataset includes all log files in .csv format. These can be
viewed or printed from a PC.
To back up files from an IND780:
1. Connect a USB (flash) memory device to the IND780’s USB port, either directly or via a USB
extension cord.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-7
Service and Maintenance
2. Access Setup > Maintenance > Run > Backup to USB. The screen shown in Figure 3-6 will
appear.
Figure 3-6: Backup to USB
3. By default, the Dataset Name is BK1; press ENTER to select the field and enter another name, if
required.
4. Press the START softkey
. The IND780 will check for the presence of a valid USB memory
device. On finding one, it checks for available space. If a device is not found, or a device is
present that has insufficient space for the operation, a prompt (“USB Memory Error”) informs
the operator and the backup does not proceed. If space is available, the IND780 creates a
folder titled Mettler Toledo/Backup/IND780/ <dataset name > (where “dataset name” is the
name entered in step 3), then begins the transfer of data. An on-screen message (“Working”)
indicates that the procedure is in progress.
5. Once the terminal indicates that the backup is complete, the USB device can be unplugged and
the EXIT softkey
pressed to return to the menu tree.
3.5.2.
Restore
System Restore cannot be performed if the terminal is locked.
An Administrator level login must be in effect to carry out this procedure.
A system restore operation does not copy log files from the dataset, and the terminal’s existing log
files are not over-written on restore.
To restore files to an IND780:
1. Connect a USB (flash) memory device to the IND780’s USB port, either directly or via a USB
extension cord.
3-8
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2. Access Setup > Maintenance > Run > Restore from USB. A screen like the one shown in Figure
3-7 will appear.
Figure 3-7: Restore from USB
3. Press ENTER to select the Dataset Name field, then use the alphanumeric keys or an external
keyboard to enter the name of the dataset to be uploaded to the terminal.
4. If scale calibration parameters are to be included in the restore procedure, select Enabled in the
Restore Scale Calibration box.
5. Press the START softkey
to initiate the transfer of files. The IND780 will check for the
presence of a valid USB memory device and for the existence, in a folder named /Mettler
Toledo/Backup/IND780/<dataset name>, of the file named in step 3. It then begins the transfer,
with an on-screen message (“Working”) showing that the procedure is in progress. If a USB
device is not found, or the named dataset is not present in the correct folder, a message
(“Could not find Dataset Name, please re-enter.”) appears.
6. Once the terminal indicates that the transfer is complete, the USB device can be unplugged and
the EXIT softkey
pressed to return to the menu tree.
7. It may be necessary to restart the terminal in order to complete the restore operation.
3.6.
Service
Only qualified personnel should perform installation, programming, and service. Please contact a
local METTLER TOLEDO representative for assistance.
In general, once the IND780 is installed, programmed, and calibrated for a given application, only
routine calibration service is required.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL. EXERCISE CARE WHEN
MAKING CHECKS, TESTS, AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON.
FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY
DAMAGE.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-9
3.7.
Troubleshooting
Service and Maintenance
Troubleshooting activities for the IND780 include:
3.7.1.
• Board LED interpretation
• Internal diagnostic testing
• AC power test
• MT Service View
• Power supply voltage test
• External diagnostics
• Battery test
• Watchdog monitoring
• RS-232 serial output voltage test
• Master reset
Troubleshooting Overview
Each of the procedures listed in Table 3-1 is described in detail in the following sections. If the
result of any of these procedures indicates a problem, refer to this table to isolate the possible
cause/s and to identify the appropriate response.
Table 3-1: Troubleshooting
Diagnostic
Procedure
Expected Outcome
Response/s to Incorrect Outcome
AC Power Test
Power source within
range -5% to +10% of
rated VAC (100240 VAC, depending on
terminal configuration)
If display operates incorrectly, or there are intermittent
problems with Terminal function:
Check whether power source is within correct range
If power source is not OK, find a reliable power source
If problem persists, perform Power Supply Voltage Test
Power Supply
Voltage Test
12 VDC: In range 11.712.3 VDC
5 VDC: In range 4.95.1 VDC
If voltages are absent or outside specified ranges, replace the
Power Supply
Battery Test
Not less than 2.5 VDC
If voltage is below 2.5 VDC, replace battery
BRAM Test
BRAM PASS
If BRAM FAIL displays, replace Main PCB
Front panel keypad:
If no response, or incorrect character displays:
•
Keyboard Test
On-screen display of
character associated
with pressed key
Check connection between HMI board and Main PCB
If problem persists, replace HMI board, or HMI harness, or
Main PCB, or ETX board
External keyboard:
If no response, or incorrect character displays:
•
•
Check keyboard connection to USB port
Attach another external keyboard and re-test
If problem persists, replace Main PCB
3-10
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Diagnostic
Procedure
Load Cell Output
Test
Expected Outcome
Raw counts displayed
vary as expected,
indicating proper
function of load cells
Response/s to Incorrect Outcome
If no value displays for load cell, or values are
inappropriate:
•
Check connection to load cell
If revisiting Load Cell Output screen shows same result,
replace load cell or affected scale option board
If test fails:
Serial Port Test
Sent characters are
received
•
Check physical connection to port (or correct installation
of self-test jumper wire)
•
•
Check port configuration
Check that receiving/sending device is functioning
properly
If problem persists, replace Serial option board, or Main
PCB, or ETX board
If bit state does not change:
Discrete I/O Test
Appropriate changes to
bit state
•
Check that correct setup screen is in use (i.e. installed
I/O device is addressed – DIO board, ARM100 module)
•
•
Check physical connection to external device
Check function of external device (powered, correctly
configured, etc.)
If problem persists, replace DIO board or remote module
If Timeout displays for test:
Network Test
PASS on all four tests
•
•
Check physical connection to network
•
Check that Gateway and Email server are available/online
Check terminal configuration – IP, Email and Gateway
addresses set up correct
If TCP/IP Stack or This Terminal test times out, replace Main
PCB or ETX board
RS-232 Serial
Output Voltage
Test
64057247 | 13 | 08/2014
Demand: Stable within
range -5 to +15 VDC
Continuous: Fluctuation
within range -10 to
+10 VDC
If either mode fails, replace Serial Option board, or ETX
board, or Main PCB
METTLER TOLEDO IND780 User's Guide
3-11
Service and Maintenance
3.7.2.
Board LED Interpretation
The following sections describe the function of LEDs visible in the IND780. The functions of the LED
arrays found on ControlNet and Ethernet / IP boards are described in the pertinent chapters of the
IND780 PLC Interface Manual.
3.7.2.1.
Main PCB
D30: USB functional
D73: Ethernet link
D13: 5V supply
D74: Ethernet active
D12: 12V
l
D75: 100MB
D72: Software flag
Figure 3-8: Main PCB LEDs
Table 3-2: Main PCB LEDs
3.7.2.1.1.
LED
Color
Function
D12
Yellow
Indicates 12V supply is present.
D13
Yellow
Indicates 5V supply is present.
D30
Green
When lit, indicates USB port has been identified and is functional.
Note that this LED remains lit whether or not a USB device is connected.
D72
Amber
Used for software development purposes; blinking during normal system operation.
D73
Yellow
Indicates that an Ethernet connection exists.
D74
Green
When blinking, indicates that the Ethernet connection is actively receiving or transmitting
data.
D75
Red
When lit, indicates that Ethernet connection is 100 MB.
D73 lit, D75 off, indicates that Ethernet connection is 10 MB.
ETX Board
The ETX board features a single, green LED, mounted beneath the lower card guide. Its approximate
location is indicated in Figure 3-8. When power is supplied to the board, this LED is lit steadily,
indicating that the core power supply is on. This in turn indicates that both on-board regulators are
functioning correctly.
3-12
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.2.2.
Option Boards
3.7.2.2.1.
Analog Load Cell and POWERCELL MTX Boards
Figure 3-9 indicates the locations of LEDs on the Analog Load Cell and POWERCELL boards.
POWERCELL
MTX Network
USB
USB
Figure 3-9: Analog Load Cell (left) and POWERCELL MTX (right) LEDs
Table 3-3: Analog Load Cell and POWERCELL MTX LEDs
3.7.2.2.2.
LED
Color
Analog Load Cell: USB
Amber
POWERCELL: USB
Amber
POWERCELL: PC
Amber
Function
Slow blink (~1 Hz) indicates no USB communication with Main PCB
Fast blink (~4 Hz) indicates USB communication with Main PCB is
operating properly
Slow blink (~1 Hz) indicates POWERCELL MTX not connected
Fast blink (~4 Hz) indicates POWERCELL MTX connected
IDNet, Serial and Discrete I/O Boards
All other option boards – IDNet, Serial, Discrete I/O – feature a pair of green LEDs, only one of
which is functional in the IND780.
D1
D3
D1
D1
D3
D3
Figure 3-10: LED Locations: IDNet (left), Serial (center) and DIO (right) Cards
These LEDs function as follows:
64057247 | 13 | 08/2014
LED
Color
D1
N/A
D3
Green
Function
N/A
Slow blink (~1 Hz) indicates there is no USB communication with main PCB.
Fast blink (~4 Hz) indicates USB communication with main PCB is operating properly.
METTLER TOLEDO IND780 User's Guide
3-13
3.7.2.2.3.
POWERCELL PDX Board
Service and Maintenance
Figure 3-11 indicates the locations of LEDs on the POWERCELL PDX board. Table 3-4 describes
the function of the LEDs.
3-14
I1
I2
I3
I4
I5
Figure 3-11: PDX Board LEDs
Table 3-4: PDX Board LEDs Function
LED
Color
Function
I1: SW
Amber
I2: USB
Amber
ON – the USB connection to the main board is active
OFF – the USB connection to the main board is inactive
I3: A1
I4: A2
Amber
I3 solid ON, I4 OFF – indicates that the CAN communication to the load cells
is active
I3 and/ or I4 blinking – indicates that the CAN communication to the load
cells is interrupted
I5: IPC
Amber
Slow blink (~1 Hz) indicates that the board’s processor is not working
correctly
Fast blink (~ 3 Hz) indicates that the board’s processor is working correctly
ON – the board’s discrete output is turned on
OFF – the board’s discrete output is turned off
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.2.2.4.
Analog Output Board
Figure 3-11 indicates the locations of LEDs on the Analog Output board. Table 3-4 describes the
function of the LEDs.
LED 3
LED 2
LED 1
LED 4
Figure 3-12: Analog Output Board LEDs
Table 3-5: Analog Output Board LEDs Function
3.7.3.
LED
Color
Function
LED 1: USB
Green
Solid ON: USB communicating with IND780 main board
LED 2: Channel 1
Green
Solid ON: Channel 1 is functioning
LED 3: Channel 2
Green
Solid ON: Channel 2 is functioning
LED 4: Board OK
Green
Slow blink: checking board communication and function
Fast blink: board communicating and functioning
AC Power Test
If the display is blank or if intermittent problems are occurring, the condition of the AC power source
should be checked. Use a multi-meter to check the AC input power. Input power must be within 15% to +10% of the standard 100–240 voltage range.
If no AC power is present, have a qualified electrician restore power at the source. After power is
restored, test the IND780 for correct operation.
There is no fuse in the IND780. The external AC power is connected directly to the internal
power supply.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-15
3.7.4.
Power Supply Voltage Test
Service and Maintenance
The power supply provides 5 VDC to both the Main PCB and the Display PCB, and 12 VDC to the
Main PCB. Verify the 12 VDC voltage is between 11.7 and 12.3 volts when measured from Pin 5
(+V) to Pin 3 or 4 (Ground). Verify the 5 VDC voltage is between 4.9 and 5.1 volts when
measured from Pin 1 (+5V) to Pin 3 or 4 (Ground), and Pin 2 to Pin 3 or 4 (Ground). If the
voltages are absent or deviate from the specified range, replace the Power Supply PCB. See Figure
3-13 for the location and description of Pins 1, 2, 3, 4 and 5 on the Power Supply harness.
USB
CONNECTOR
ETHERNET
CONNECTOR
POWER
CONNECTOR
COM1
CONNECTOR
VIDEO CABLE
COM2
CONNECTOR
Figure 3-13: Main PCB Connections and Mounting Screws
Red
1
Red
2 +5
+5
Black
3 GROUND
Black
4 GROUND
1
5 +12V
Yellow
Figure 3-14: Pins on the Power Supply Harness
3.7.5.
RS-232 Serial Output Voltage Test
If the send and receive functions of the serial (COM) ports test fail, use the following procedure to
determine whether the RS-232 serial port is operational:
1. Disconnect AC power from both the IND780 terminal and the printer.
2. Disconnect the serial cable from the COM port on the IND780 main PCB.
3. Set the voltmeter to read 20 VDC.
4. Connect the red lead to the transmit terminal of the COM port and connect the black lead to the
ground terminal of the COM port.
3-16
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
5. Apply power to the IND780. The meter should read as follows:
•
Demand mode—The meter should read a stable value (without fluctuation) between −5
and +15 VDC.
•
Continuous mode—The meter should fluctuate continuously within the range −10 to +10
VDC. The actual values and degree of fluctuation observed will depend on the type and
sensitivity of meter used. The constant fluctuation on the meter display indicates that the
scale/indicator is transmitting information.
To test Demand baud rates, press the PRINT key
. The display should fluctuate as for
continuous mode for the duration of the transmission, then become stable again. This fluctuation
indicates the terminal has transmitted data.
When measuring the higher baud rates in the Demand mode, the meter display will fluctuate for
a shorter period of time.
3.7.6.
Battery Test
CAUTION
RISK OF EXPLOSION IF BATTERY IS REPLACED WITH WRONG TYPE OR CONNECTED
IMPROPERLY. DISPOSE OF BATTERY ACCORDING TO LOCAL LAWS AND REGULATIONS.
If setup parameters change uncontrollably or programming is lost, check the BRAM battery voltage.
Battery voltage is tested at the battery assembly on the Main PCB. The battery is shown in Figure
3-15.
Battery
Figure 3-15: Main PCB, Battery Location Indicated
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-17
Service and Maintenance
Use a Volt-Ohm meter to measure voltage between terminal 1 (positive) and terminal 2 (negative)
to the left and right of the battery with the AC power disconnected. This measurement should be
approximately 3.0 VDC. Replace the battery (Panasonic CR2032 or comparable) if the measured
voltage is below 2.5 VDC.
Positive
Negative
Figure 3-16: Battery Installed
The battery replacement procedure is detailed in the Maintenance > Run section of Chapter 3,
Configuration in the IND780 Technical Manual.
3.7.7.
Internal Diagnostic Testing
The IND780 provides several internal diagnostic tests that are accessible in setup mode. Press the
SETUP softkey
to view the setup menu tree. Use the DOWN navigation key to scroll down the
menu tree to Maintenance. Press the RIGHT navigation key to expand the menu tree selections for
Maintenance. Scroll down and expand Run. Scroll down and expand Diagnostics. Available
diagnostic setup screens are shown in the setup tree view in Figure 3-17.
Figure 3-17: Maintenance Diagnostic Options Tree
3-18
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.7.1.
Memory Test
To test the BRAM, press the START
to return to the menu tree.
softkey. When testing is complete, press the EXIT softkey
Figure 3-18: Memory Test Screen
3.7.7.2.
Keyboard Test
Enables testing of the keyboard. Each key that is pressed will be shown on the display. When
testing is complete, press the EXIT softkey
to return to the menu tree.
Figure 3-19: Keyboard Test Screen
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-19
3.7.7.3.
Scale
Service and Maintenance
Shows the diagnostics options for a scale:
Figure 3-20: Scale Diagnostics Menu Tree Options
3.7.7.4.
Load Cell Output
Displays the current load cell output (active weight) for the scale. The numbers in the scale below
feature load cell addresses for each output. For digital load cells, individual cell counts will be
displayed. Individual cell counts are the raw output of the load cells, which the indicator translates
into weight by calculation, using information it acquires through the calibration process.
Figure 3-21: Load Cell Output Screen
3-20
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.7.5.
Calibration Values
Displays the current calibration values for each scale. If these values are recorded after a scale
calibration, and the scale channel PCB is replaced in the future, the calibration values can be
manually entered here to “transfer” the previous calibration to the new scale channel PCB. This is
not available with IDNet bases.
Figure 3-22: Calibration Values Screen, Zero in Focus
3.7.7.6.
Shift Values
Used for POWERCELL and PDX scales only. The following screen shows values used to adjust gain
for each load cell in a multiple load cell system. The values compensate for difference in response
between cells. The load cell numbers represent the node cell addresses. Where more than 12 load
cells are present this screen displays a scroll box to reveal further rows.
The diagnostic purpose of the Load Cell Output and Shift Values screens is to permit a technician to
confirm valid calibration values were obtained when the calibration routine (Setup > Scale > Scale
n > Calibration) was carried out.
Figure 3-23: Scale Shift Values Screen
3.7.7.7.
Diagnose Cell
Used for POWERCELL scales only (but not POWERCELL PDX). The purpose of the Diagnose Cell
screens is to allow the technician to verify a load cell’s node address and raw counts. If the cell that
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-21
Service and Maintenance
is being diagnosed has a known address, select that address from the Load Cell Address selection
box. If the cell address is unknown, then select Connected Cell. If the cell is faulty or it has an
invalid address, the terminal will fail to communicate with the cell. If the cell responds correctly, the
node address and the cell counts will be displayed on the screen (Figure 3-24).
Figure 3-24: Diagnose Cell Screen
3.7.7.8.
Statistics
Displays statistical information for each scale such as the total number of weighments, the number
of scale overloads, the peak weight weighed on the scale, the total number of zero commands, and
failed zero commands (Figure 3-25). These are very helpful when diagnosing scale problems.
These statistics accumulate for each scale, and are saved until a master reset is performed. It is not
considered likely that a master reset will be a frequent occurrence, so these statistics are effectively
permanent.
Statistics cannot be deleted without losing all other setup values.
Figure 3-25: Scale Statistics Screen
3-22
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.7.9.
Serial Test
Enables testing of the transmit and receive functions on the serial (COM) ports. Select the COM port
to be tested using the selection box for Com Port (Figure 3-26). Only those ports installed are
available for testing.
Figure 3-26: Serial Test Setup Screen
Press the START softkey
and a data string is output repeatedly approximately once every 3
seconds Figure 3-27). The data is: [Testing COMx: nn] where “x” is the COM port and “nn” is an
incrementing value beginning at 00 and continuing to 99. Each transmission increments this
number by one.
Figure 3-27: Serial Test Screen
If a jumper wire is placed between the transmit and the receive terminals (Figure 3-28) on the port
being tested, the same data string that is transmitted displays as the receiving field.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-23
Service and Maintenance
Jumper wire
Figure 3-28: Jumper Wire Placed Between the Transmit and Receive Terminals
If another device is connected to the receiving port, any ASCII data received displays in the receiving
field.
To stop the serial port test, press the STOP softkey
3.7.7.10.
.
Discrete I/O Test
Provides a view of the status of the discrete inputs and allows enabling or disabling of the discrete
outputs for diagnostic purposes. With Discrete I/O highlighted (menu tree), expand its branch by
pressing the RIGHT navigation key. Choose whether the test will be performed on the local (internal)
discrete I/O option or one of the remote ARM100 modules. Press ENTER when that branch is
highlighted.
A warning (Figure 3-29) will appear to remind the tester that the outputs can be turned on
manually during this test, so any control power to the discrete outputs should be removed.
Figure 3-29: Discrete I/O Test Screen
WARNING
THE DISCRETE OUTPUTS OF THE IND780 TERMINAL WILL BE MANUALLY ENABLED DURING
THIS TEST. REMOVE OUTPUT CONTROL POWER SO EXTERNAL EQUIPMENT WILL NOT BE
ENERGIZED BY MISTAKE. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND
ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO OBSERVE THESE
PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
3-24
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
When viewing the test screen, the input status will be shown at the top of the display and the output
status will be shown at the bottom of the display. Focus is shown on Output #1 first.
Focus can be moved to any of the other outputs by using the RIGHT and LEFT navigation keys on
the front panel, or pressing ENTER.
Note that when the Discrete I/O Test screen is exited, all outputs will be turned off again.
3.7.7.11.
Network Test
Enables testing of the Ethernet programming/hardware/firmware in the terminal. Figure 3-30 shows
the result of running a network test.
Figure 3-30: Network Test Screen
The following tests are run during this diagnostic process:
•
TCP/IP Stack
•
This Terminal
•
Gateway Address
•
Email Server
Each step will indicate “Testing” during the test then indicate a Pass or Timeout status. If the status
is shown as Timeout, it indicates that the IP address or gateway address has not been programmed
in the terminal. Refer to the Communication and Network section of Chapter 3, Configuration, in the
IND780 Technical Manual for information on entering the IP and gateway addresses.
3.7.8.
MT Service View
The MT Service View functionality provides a METTLER TOLEDO authorized service representative
access to various diagnostics data collected from a POWERCELL PDX scale. This functionality is
accessed in setup mode but the MT Service Security must be first unlocked. MT Service View
screens include:
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-25
Service and Maintenance
• Scale Loadcell Temperature
• Scale Loadcell Com Voltage
• PDX Terminal
• Scale Loadcell Supply Voltage
• Load Cell Information
Figure 3-31: MT Service View Screen
3.7.8.1.
Scale Load Cell Temperature
From the MT Service View screen, choose Scale in the Device selection box and Temperature in the
View selection box. Select the PDX scale channel to view and press the VIEW softkey
to display
the Scale Load Cell Temperature screen. This screen provides a view of the temperature sensor
readings measured in every PDX cell for the selected scale. For each cell node address shown is
the corresponding temperature value.
Figure 3-32: Scale Load Cell Temperature Screen
The PDX cells rely on their temperature sensors to compensate for temperature variations and
continuously provide accurate weight measurements. To ensure that the sensors are operating
correctly, it is important to compare the temperature readings between the load cells. Readings
should be reasonably close between the cells if there are no obvious differences in external
influences (e.g. sunshade, heating pipes).
3-26
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.8.2.
Scale Load Cell Supply Voltage
From the MT Service View screen, choose Scale in the Device selection box and Supply Voltage in
the View selection box. Select the PDX scale channel to view and press the VIEW softkey
to
display the Scale Load Cell Supply Voltage screen. This screen provides a view of the input supply
voltage readings measured in every PDX cell for the selected scale. For each cell node address, the
corresponding supply voltage value is shown in volts. Measurements are automatically updated
hourly, and after power-up.
Figure 3-33: Scale Load Cell Supply Voltage Screen
The nominal supply voltage is based on whether the IND780 uses its internal 12 VDC or the
external 24 VDC supply to power the cells. It is normal to have the voltage readings from one
cell to another drop in value. This is the result of the additional power consumption attributed to
each cell on the network and the voltage dissipation along the length of cabling. The expected
load cell supply voltage ranges are:
• At 12 VDC: 12.5 – 7.9 VDC
• At 24 VDC: 24.5 – 15 VDC
Most importantly the supply voltage measured at any cell must not be below 7.8 VDC for the PDX
cell to operate reliably. Wiring shorts or an overloaded supply will cause these readings to be out of
range. A system that has supply voltages that deteriorate over time may also indicate a potential
problem.
3.7.8.3.
Scale Loadcell COM Voltage
From the MT Service View screen, choose Scale in the Device selection box and COM Voltage in the
View selection box. Next, select the PDX scale channel to view and press the VIEW softkey
.A
warning and confirmation screen will be displayed after the VIEW softkey is pressed. Press the OK
softkey
to proceed to display the Scale Load Cell COM Voltage screen. This operation will take
several minutes as each load cell is required to interrupt its communication lines temporarily in
order to take CAN voltage measurements.
This screen provides a view of the CAN-High and CAN-Low signal voltages measured in every PDX
cell when they are placed into the Dominant and Recessive communication mode. For each cell
node address, the corresponding CAN signal voltage values are shown in volts. Measurements are
static and are only updated each time this screen is viewed.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-27
Service and Maintenance
Figure 3-34: Scale Load Cell COM Voltage Screen
The CAN-High and CAN-Low Dominant states combine to represent a logical bit 0 whereas CANHigh and CAN-Low Recessive states represent a logical bit 1. When checking for proper operation,
verify not only the voltage levels of each signal but also the voltage differential between the signals.
The expected voltage ranges are:
CAN Signals
Dominant Mode Voltage
Recessive Mode Voltage
CAN-High
3 – 5 VDC
2 – 3 VDC
CAN-Low
0 – 2 VDC
2 – 3 VDC
CAN-High and CAN-Low Differential
≈ 2 VDC
≈ 0 VDC
Wiring shorts or poor network termination may cause these readings to be out of range.
3.7.8.4.
Load Cell Information
From the MT Service View screen, choose Load Cell in the Device selection box then select the PDX
scale channel and the specific load cell node to view. Press the VIEW softkey
to display the
Load Cell Information screen. This screen provides a collective view of various diagnostic
parameters for a selected POWERCELL PDX cell. Each data field will have both a current value and
a value recorded at the time of calibration. This enables the user to track changes and make data
comparisons to a known working condition of the cell.
Figure 3-35: Load Cell Information Screen
3-28
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
The recorded load cell parameters include:
Data Field
Cell S/N
Description
Represents the unique factory serial number embedded in the cell.
Install Date
Indicates the date the load cell was last addressed.
Temperature
Temperature sensor reading in the cell.
LC Supply Voltage
Input supply voltage measured in the cell in volts.
CAN-High Dominant X Cell CAN-High voltage recorded for Dominant mode in volts.
1
1
CAN-Low Dominant X
Cell CAN-Low voltage recorded for Dominant mode in volts.
CAN-High Recessive
Cell CAN-High voltage recorded for Recessive mode in volts.
CAN-Low Recessive
Cell CAN-Low voltage recorded for Recessive mode in volts.
Gas Concentration
Level (%) of inert gas concentration enclosed in the cell as a measure of the
hermetic sealing. Small losses of gas levels are expected over the years. A
sudden rapid loss of gas levels is an apparent indication of a cell enclosure
breach condition.
1
1
Note
1. These values are stored from the last time the Load Cell COM Voltage Screen was accessed.
3.7.8.5.
PDX Terminal
From the MT Service View screen, choose Terminal in the Device selection box and press the VIEW
softkey
to display the PDX Terminal screen. This screen provides a view of the minimum and
maximum voltages and current detected on the IND780 PDX option board. These values are
updated dynamically while the screen is viewed.
Figure 3-36: PDX Terminal Screen
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-29
The displayed information includes:
Service and Maintenance
Data Field
3.7.9.
Description
Minimum CAN-High Voltage
Typically associated with the CAN-High voltage value (volts) measured
during a Recessive communication state. Normal range: 2 – 3 V.
Maximum CAN-High Voltage
Typically associated with the CAN-High voltage value (volts) measured
during a Dominant communication state. Normal range: 3 – 5 V.
Minimum CAN-Low Voltage
Typically associated with the CAN-Low voltage value (volts) measured
during a Dominant communication state. Normal range: 0 – 2 V.
Maximum CAN-Low Voltage
Typically associated with the CAN-Low voltage value (volts) measured
during a Recessive communication state. Normal range: 2 – 3 V.
Minimum CAN Voltage Difference
Minimum difference between CAN-High and CAN Low voltages. Normal
range: ≈ 0 V.
Maximum CAN Voltage Difference
Maximum difference between CAN-High and CAN Low voltages. Normal
range: ≈ 2 V.
Minimum Supply Voltage
Minimum supply voltage measured at the present voltage level as driven
by the internal (12 VDC) or external (24 VDC) power supply. Normal
range: ≈ 5 % tolerance of nominal values.
Maximum Supply Voltage
Maximum supply voltage measured at the present voltage level as driven
by the internal (12 VDC) or external (24 VDC) power supply. Normal
range: ≈ 5 % tolerance of nominal values.
Minimum Supply Current
Minimum supply current consumption measured with the cells connected.
Normal range: ≈ 10 mA – 38 mA per cell.
Maximum Supply Current
Maximum supply current consumption measured with the cells
connected. Normal range: ≈ 10 mA – 40 mA per cell.
External Diagnostics
The IND780 includes a utility that provides valuable tools for examining the operation of the
terminal, its installed options and software, and (for POWERCELL bases only) of associated load
cells. Information available in Web Help includes:
• System hardware and software configuration
• Resource utilization
• Share data variables, viewed either statically or dynamically
• Scale and (if digital load cells are used) load cell status and statistics
Web Help is accessed with a web browser via an Ethernet connection. Connect to the Terminal
using its IP address.
Depending on how it is configured, the terminal’s IP address may appear at upper left in the home
screen. It may also be viewed and modified in Setup at Communication > Network > Ethernet.
At the bottom of each page, a HOME button returns the view to the index page seen in Figure 3-37.
3.7.9.1.
Index Page
When Web Help is first accessed, the index page shown in Figure 3-37 appears in the browser
window.
3-30
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 3-37: Web Help Home Screen
Click the links on this page to visit the respective views, each of which is detailed in the following
sections.
3.7.9.2.
Help File
A help file, accessible from each page via the help button
features. Figure 3-38 shows a portion of the help file.
explains the full range of the utility’s
Figure 3-38: Help File
At the bottom of the page, a CLOSE PAGE button allows the user to close this browser window
without exiting the Web Help utility.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-31
3.7.9.3.
System Information
Service and Maintenance
The system information page (Figure 3-39) provides information about asset information, installed
component part numbers, software versions and serial numbers, and installed application software.
Figure 3-39: System Information Page
3.7.9.4.
Resource Utilization
The resource utilization page (Figure 3-40) gives a snapshot of the current status of the terminal,
including CPU usage and information about the various types of memory used by the IND780.
Figure 3-40: Resource Utilization Page
3-32
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.9.5.
Scale Statistics
The scale statistics page (Figure 3-41) shows a variety of data for each installed scale.
Figure 3-41: Scale Statistics Page
3.7.9.6.
Read Shared Data
The shared data page (Figure 3-42) allows the user to interact with the terminal, requesting a
display of current values for up to ten shared data variables at a time. Values are accessed by
entering the shared data name in one of the fields at left, and then clicking on the Save Changes
button at bottom. The button is in focus when a new entry has been made in a name field.
Figure 3-42: Shared Data Page
To refresh the displayed values without changing any variable names, click on the Manual Refresh
button.
Values may also be refreshed automatically, at one second intervals, by clicking on the Start
button. Once Start is selected, the Stop button is in focus and clickable. The current status of
dynamic display (Stopped or Running) is shown in a field between the two buttons. Note that, as
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-33
Service and Maintenance
indicated on the page, Auto Refresh must be stopped before new shared data names can be
entered.
3-34
For a complete list of available shared data names, refer to the IND780 Shared Data Manual. In the
example shown in Figure 3-42, the shared data information displayed is as follows:
Table 3-6: Shared Data Information
Name
Data Displayed
Interpretation
wt0200
96980^ …etc.
Composite of entire data block for Scale 2
xp0117
10163
Cumulative power on time, in minutes
ca0111
0
SmartTrac Appearance. 0 = Bar Graph
ce0204
1
Number of ranges defined for Scale 2
ce0223
1950012
Encoder counts at the middle calibration value for Scale 2
ce0225
16
GEO code set in the Scale 2 calibration page
ce0245
3
Analog Load Cell gain jumper setting. 3 = 3 mV/V
pl0102
2
PLC Interface Module type. 2 = PROFIBUS
nt0102
192.168.000.001
Ethernet IP Address
wt0203
kg
Primary units set in the Scale 2 calibration page
When a string is longer than the data display width of 40 characters (e.g. wt0200 in the example
above), the left/right scroll buttons
may be used to move through the string to view all of it.
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.9.7.
Scale Data
The scale data page (Figure 3-43) shows information, including current configuration and
calibration values, for each installed scale. The signal quality display permits troubleshooting of the
connection between the IND780 and load cells. When quality is poor, a red bar is displayed. A
yellow bar indicates adequate quality, and green bar means the quality is good, as in the example
below.
Figure 3-43: Scale Data Page
3.7.9.8.
PDX Scale Data (POWERCELL PDX bases only)
The PDX Scale Data page (Figure 3-44) shows individual load cell information, for a PDX type
scale. This page provides access to several views:
• Adjusted Counts
• Temperature (MT Service only)
• Gross Weight
• Input Voltage (MT Service only)
• Communication Errors
• Gas Concentration (MT Service Only)
As indicated in the list above, some views are accessible only if the terminal’s MT Service Security
has been unlocked.
In the data pages each cell node is represented by a circle, with different colors to indicate if certain
fault conditions exists. A circle with a red border (as in cells 2 and 7 in the example below)
indicates that a fault condition has been detected, but in a different view. A completely red circle
indicates that a fault is present in the current view.
Indicated fault conditions include cell communication errors, cell zero drift, cell overloads and
enclosure breach. Click on each individual cell to go to a specific detailed PDX Load Cell page
(Figure 3-46).
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-35
Service and Maintenance
Figure 3-44: PDX Scale Data Page
3.7.9.9.
PDX Terminal (POWERCELL PDX bases only)
The PDX Terminal page (Figure 3-45) shows the minimum and maximum voltages and current
detected on the IND780 PDX option board. This page can be used to determine if the terminal’s
PDX option board is providing the proper electrical parameters to support the CAN Bus
communication and to power the cells. It also helps to indicate if there are issues with poor wiring,
a short condition or an overloaded power supply. Most of the information on this page is only
accessible if the terminal’s MT Service Security has been unlocked.
Figure 3-45: PDX Terminal Page
3.7.9.10.
PDX Load Cell (POWERCELL PDX bases only)
The PDX Load Cell page (Figure 3-46) provides a collective view of various diagnostic parameters
for a selected POWERCELL PDX cell. Most data fields will have both a current value and a value
recorded at the time of calibration. The data is arranged into five main categories – weight,
puncture, tilt, temperature, voltages and other errors. When a category or value is highlighted in
red, it indicates that a value is out of its normal range, which may mean that an error condition is
present. Most of the information on this page is only accessible if the terminal’s MT Service Security
has been unlocked.
3-36
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 3-46: PDX Load Cell Page
3.7.9.11.
Load Cell Statistics and Load Cell Data (POWERCELL and PDX bases only)
Load cell statistics (Figure 3-47) and data (Figure 3-48) are collected for POWERCELL bases. In
the examples shown, a POWERCELL option board is installed, but no load cell is connected.
Figure 3-47: Load Cell Statistics Page
Figure 3-48: Load Cell Data Page
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-37
Service and Maintenance
3.7.9.12.
Load Cell Symmetry (POWERCELL and PDX bases only)
When symmetry monitoring is enabled, the View/Load Cell Symmetry page (Figure 3-49) shows
live load cell symmetry difference percentages. The symmetry difference values are displayed only
when the start threshold for the monitoring is exceeded. This page can be used in troubleshooting
load cell failure based on symmetry difference, or it can be used to determine a suitable difference
threshold value for SETUP.
Figure 3-49: Load Cell Symmetry Page
3.7.10.
Watchdog Monitoring
The main hardware microprocessor of the IND780 has a built-in hardware watchdog timer that is
used to constantly monitor the performance of the measurement channels, weight display,
peripheral communications and the discrete I/O functions. If any one of these functions fails to
respond to the watchdog within a fixed time interval, the watchdog timer will timeout and the entire
terminal will automatically reboot within 40 seconds. Given the seriousness of the system failure, if
feasible the terminal will record the watchdog update failure event in the error log. Watchdog
monitoring allows the terminal to detect certain system failures and to automatically recover from
them.
3.7.11.
Master Reset
A hardware master reset switch is provided, that sets all IND780 terminal settings to the factory
default settings (see Appendix B, Default Settings). This includes scale setup and calibration
information, log files (except for the error log) and counters. The master reset typically is performed
under these circumstances:
• When a hardware configuration change is made, such as adding a new scale or Discrete I/O
board to the terminal.
• When a software configuration problem arises that cannot be resolved without starting from the
factory default settings.
• When security is enabled to protect setup, and the password is lost.
• After a firmware upgrade is performed.
• After a hardware key is installed or removed.
Two types of master reset are possible, depending on the position of switch S2 (shown in its
OFF position in Figure 3-50): A full reset including metrologically significant scale configuration
data, and a reset that preserves scale data.
3-38
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
3.7.11.1.
To initiate a master reset
1. Remove AC power.
2. If the reset is to include scale configuration data, set S2 to its ON position. If scale configuration
data are to be retained, leave S2 OFF.
3. Press and hold the MASTER RESET button (Figure 3-50).
S2
Master
reset
button
Figure 3-50: Main PCB, Master Reset Button and S2 Locations Indicated
4. Apply AC power. Hold the MASTER RESET button until the unit beeps and the “BRAM reset to
Factory” message pops up.
5. Press ENTER to acknowledge each message box as it appears.
6. If S2 was set to ON to include metrologically significant data in the reset, return it to its OFF
position.
3.8.
Main PCB
3.8.1.
To remove a main PCB from an IND780, follow these steps
1. Remove AC power.
2. Use a grounding strap to prevent risk of electrostatic discharge.
3. At the back of the terminal, identify and label all option board cables.
4. Disconnect the option board cables and power cable from the back of the terminal.
5. It may not be necessary to remove the panel mount from its location. If removal is required:
6. Use an 8 mm nut driver to remove the eight nuts securing the terminal to its mounting location.
7. Remove the backing plate and lift the terminal out of the opening in which it is installed.
8. Open the terminal by removing its rear cover (in the case of a panel mount terminal) or its front
panel (in the case of a harsh environment terminal).
9. Remove the two screws attaching the upper card guide, shown in Figure 3-51, and lift the
guide from its stand-offs.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-39
Service and Maintenance
Figure 3-51: Upper Card Guide
The option board connector pins are delicate. Take care to avoid damaging them when
removing the boards.
10. Note the locations of the option boards then remove them and set them aside on a static-free
mat.
11. Remove all peripheral connectors from their sockets – COM1, COM2, Ethernet, USB, power and
video. Connector locations are indicated in Figure 3-52.
USB
connector
Ethernet
connector
ETX board with CF
card mounted
COM1
connector
Power
Connector
PLC module
COM2
connector
Video cable
Figure 3-52: Main PCB Connections and Mounting Screws
12. Disconnect any PLC interface module cables or connectors from their sockets.
13. Unscrew and remove the six screws holding the main PCB to the enclosure. Their locations are
circled in Figure 3-52. In the case of a panel mount terminal, the four outer fasteners will be
hex post stand-offs, as shown in Figure 3-53.
3-40
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 3-53: Stand-Off on Main PCB
14. Carefully lift the PCB out of the enclosure and set it on a static-free mat.
3.8.2.
Main PCB Installation
All mounting screws should be installed with 23 inch pounds (2.6 Nm) of torque.
1. Position the main PCB board over the six stand-offs in the terminal enclosure.
2. Install the hex post stand-offs:
For a panel mount terminal, install screws in the center holes and in the four hex post standoffs in the outer holes (see Figure 3-53).
3. For a harsh enclosure terminal, install six screws.
4. Reconnect all peripherals – COM1, COM2, Ethernet, USB, power and video.
5. Install option boards as necessary.
6. Reconnect the PLC interface module cable or connectors if appropriate.
7. Replace the rear cover or front panel.
3.9.
ETX Board
3.9.1.
ETX Board Versions
Newer IND780 terminals are equipped with a faster, new-generation ETX board. The IND780 main
PCB is compatible with both types of board, but the terminal must use firmware versions as
follows:
Original ETX board
Firmware version 6.x or earlier
New ETX board
Firmware version 7.x or later
To determine which type of ETX board is installed in an IND780 terminal, watch the boot-up
display. The processor type will appear in the upper left corner of the screen:
MSC ET(e) – PXA255 = original board
MSC ET(e) – PXA270 = new board
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-41
To identify the two types of ETX board, refer to Figure 3-54. Note the change in orientation of the CF
card.
Service and Maintenance
Original / PXA255
New / PXA270
Figure 3-54: Original and New ETX Boards
3.9.2.
ETX Board Removal
To remove the ETX board from the main PCB, first remove the main board from the IND780
chassis. Then:
1. Remove the two screws holding the lower card guide and the ETX board to the main PCB
(Figure 3-52). The locations of these screws are indicated in Figure 3-55.
Figure 3-55: Lower Card Guide Screw Locations
2. Remove the other two screws securing the ETX board in place, shown in Figure 3-56.
3-42
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Figure 3-56: ETX Board Mounting Screws
3. Taking precautions to avoid static discharge, lever the ETX board up from its connectors,
starting at the CF card end indicated in Figure 3-56.
4. Lift the ETX board off the main PCB and set it down on a static-free mat.
3.9.3.
ETX Board Installation
This procedure must be performed with the main PCB removed from the IND780 enclosure:
All mounting screws should be installed with 23 inch pounds (2.6 Nm) of torque.
3.9.3.1.
To install an ETX board on the main PCB
1. Note the orientation of the ETX board indicated in Figure 3-52. The ETX connectors are arranged
so that it cannot be installed backwards.
2. Position the ETX board over its connectors. Figure 3-57 shows two of the connectors, and
Figure 3-58 shows a side view of the board resting on its connectors.
Figure 3-57: Close-Up View of ETX Connectors on Main PCB
Figure 3-58: ETX Board Resting on Connectors
3. With the main PCB resting flat on a static-free mat, press down firmly over each ETX board
connector to seat it. The connectors seat with an audible snap.
Note that simply installing the ETX board mounting screws will not seat the connectors properly
– see Figure 3-59. Each connector must be pressed into place.
Figure 3-59: ETX Board Connectors Improperly Seated
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-43
Service and Maintenance
4. Check to ensure that all connectors are fully seated, as shown in Figure 3-60.
Figure 3-60: ETX Board Connectors Fully Seated
5. Install screws to secure the board to the main PCB. Two of these screws also attach the lower
card guide. This bar is asymmetrical, and must be mounted in the orientation indicated in
Figure 3-55. Note that the longer end is oriented toward the main PCB battery.
3.9.4.
CF Card Removal and Installation
To remove a CF card from the ETX board, hold the sides of the CF card and carefully slide it out of
its socket, in the direction indicated in Figure 3-61. It may be necessary to move the card slightly
side to side to disengage it. If any PLC interface module option is installed, it may need to be
removed first.
Note that Figure 3-61 shows the CF card in the same orientation as in Figure 3-52.
Figure 3-61: CF Card Removal – Original (left) and New (right) ETX
To install a CF card in the ETX board, slide the CF card into its socket. Keep the card straight, to
avoid damaging the socket’s pins. The card will snap into place. There are two guides on the CF
Card. One is thinner than the other. Do not force the CF Card into the carrier.
If the CF card has been completely reformatted on a PC, it is important to select the FAT file
system as the formatting option.
3.10.
Power Supply
3.10.1.
Power Supply Removal
Components in the power supply may be hot to the touch. Allow them to cool before performing the
procedure described below.
3.10.1.1.
To remove a power supply
1. Disconnect the incoming power cord and the cord from the main PCB from their connectors on
the power supply board. These connectors are keyed and can only be attached in the correct
orientation. They are shown attached in Figure 3-62, and disconnected in Figure 3-63.
3-44
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
Main PCB power
connector
Incoming
power
connector
Figure 3-62: Power Supply, Connected
Figure 3-63: Power Supply, Disconnected
2. Use a 7 mm nut driver to remove the two nuts and washers (indicated in Figure 3-63) from
each the two posts holding the power supply bracket to the enclosure. In a panel mount
terminal (shown here) this bracket is mounted beside the main PCB. In a harsh enclosure
terminal, the bracket is mounted to one end wall of the enclosure.
3. Lift the bracket and power supply out of the enclosure.
4. Remove the power supply PCB from the bracket by removing the four screws, one at each
corner, visible in Figure 3-63.
3.10.2.
Power Supply Installation
All mounting screws should be installed with 23 inch pounds (2.6 Nm) of torque.
3.10.2.1.
To install a power supply
1. Mount the power supply PCB to the mounting bracket using four screws, one at each corner.
Note the orientation of the power supply relative to the bracket in Figure 3-63.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-45
2. Position the bracket over the two mounting posts in the terminal enclosure, and install two nuts
with star washers to secure it in place.
Service and Maintenance
3. Connect the incoming power cord and power cord to the main PCB as shown in Figure 3-62.
3.11.
Option Board Installation
All mounting screws should be installed with 23 inch pounds (2.6 Nm) of torque.
The option board connector pins are delicate. Take care not to bend them when aligning them
with their socket on the main PCB.
3.11.1.
To install an option board on the main PCB
1. Loosen and remove the screws holding the upper card guide (Figure 3-51) in place on its
stand-offs.
2. Remove the upper card guide.
3. Position the option board over its connector.
4. Ensure that the connector pins are correctly aligned with the socket, and press down gently to
seat the board and fit the board into the slot in the lower card guide.
5. Replace the upper card guide, taking care to position the upper edge of the option board/s in
the slot/s in the upper guide (see Figure 3-51).
3.12.
PLC Interface Module Removal and Installation
To remove a PLC interface module from the main PCB, loosen and remove the three screws
(indicated in Figure 3-64) holding it in place, and carefully lift the module off its connector.
PLC Interface
Connector
Location
Figure 3-64: PLC Interface Module Installed on Main PCB
3.12.1.
To install a PLC interface module on the main PCB
All mounting screws should be installed with 23 inch pounds (2.6 Nm) of torque.
1. Position the module over its connector.
3-46
METTLER TOLEDO IND780 User's Guide
64057247 | 13 | 08/2014
2. Ensure that the module’s pins are correctly located over the connector, and press down gently
to engage them.
Figure 3-65: PLC Interface Module Socket on Main PCB (left) and Connector Pins on the Underside of a
Module (right)
3. Install the three screws indicated in Figure 3-64.
64057247 | 13 | 08/2014
METTLER TOLEDO IND780 User's Guide
3-47
To protect your METTLER TOLEDO product’s future:
Congratulations on choosing the quality and
precision of METTLER TOLEDO. Proper use
according to these instructions and regular
calibration and maintenance by our factory-trained
service team ensure dependable and accurate
operation, protecting your investment. Contact us
about a METTLER TOLEDO service agreement
tailored to your needs and budget.
We invite you to register your product at
www.mt.com/productregistration so we can contact