Download User's Guide / LogXR Software / XR5 Data Logger
Transcript
XR5 Data Loggers Battery Voltage / Ext. Power: Start Logging: The total voltage of the two internal batteries, or the external power voltage (Q terminal) if it is higher than the battery voltage. The reading is updated every time a setup is sent to the XR5, and at one hour intervals from the time that the setup is initially sent (it is not updated when Status | Extract is selected). The batteries should be replaced when the Battery Voltage reads below 6.0 V. See page 65 for details. The Battery Voltage is also written to the header section of all XR5 data files. The time the first data value was logged (or will be logged, if the Log mode has a Preset Start Time which has not yet been reached). Exception for Wrap modes: For Manual/Wrap or Preset/Wrap logging modes, if the memory is full and the XR5 is currently logging, then although "Start Logging" indicates when logging began, the stored dataset begins at a later time, equal to the current time minus the Maximum Log Time (the earliest readings are overwritten in Wrap mode once the memory is full). Stop Logging: The calculated time that logging will end (when the memory becomes full). This time cannot be calculated if the Log Mode is ‘Log on Ext. Trigger’ or ‘Log on Alarms’. Alarm summary: Displays user selected alarm limits. The ‘Tripped’ column displays ‘Yes’ if an alarm condition has occurred anytime since the start of the logging session. Serial #: Displays the unit’s serial number. The serial number is factory set and cannot be changed. A label on the side of the battery holder inside the XR5 also displays the serial number. HwRev: Displays the hardware revision number (2.0 for all XR5-SE models). 19