MAINTENANCE MAINTENANCE WORK 8.4.3 Checking the test space tightness The test space must be sealed up tightly. Check the tightness as follows: • Place a paper strip between test space door and seal • Pull it out - there must be a noticeable resistance • Repeat this procedure all around the door If the sealing is not tight, contact our service organisation. 8.4.4 Cleaning the air-cooled condenser The air-cooled condenser is located in the mechanical section → Fig. 2-1Test system (page 11) Cleaning the fins DANGER The fins of the condenser may cause hand injuries • Be sure to wear protective gloves Dust deposits on the fins of the air-cooled condenser will cause nonpermissible pressure increase in the refrigerating unit. Open the front flap • Check the air-cooled condenser regularly for dust deposits • Clean it every three months, more often in dusty environments. • Use a vacuum cleaner, compressed air or a brush Betriebsanleitung für Klimaprüfschränke kap8.fm 63837023 GB 03.2003 • 64 – 68 1) option 2) not assigned 3) only for test systems with 600 ltr and more