Download DustTrak DRX Manual
Transcript
g. Open filter cassette using enclosed tool PN 7001303. Figure 4-9: Open Filter using Enclosed Tool h. Remove screen mesh from filter cassette and blow out using compressed air. Blow in reverse direction to remove captured particulate. i. Replace mesh in filter cassette and press halves together. Ensure filter has been fully closed. The filter tool PN 7001303 can be used to ensure the filter is fully closed. Figure 4-20: Replace Mesh in Filter Holder j. Place filter cassette back into position and close blue retaining clip. Make sure retaining clip snaps back into place. Notes Replacement filters were shipped with the new instrument. Additional filters (HEPA and 37-mm Filter Cassette with mesh filter) can be order from TSI under PN 801673. TSI does not supply any filter media for the filter cassette. Any commercially available 37-mm filter media may be used with the DUSTTRAK™ II or DRX desktop instruments to collect gravimetric reference samples. 3. It is important to reset the instruments filter counter after replacing filters. Resetting the counter will clear the filter error condition shown on the main screen. Reset the counters by the following: a. Turn on the instrument. b. Press the Setup button to go into the setup screen. 48 Chapter 4