Configuring the RS485 Network
During manufacture, Hydronix assigns each sensor a unique ID and a common network
address of 16. When more than one sensor is connected on a network, each must be
assigned a different network address otherwise the network will fail to operate correctly.
There is also the facility for a user-defined name for each sensor.
Since new sensors will always appear at network address 16, they should be added to the
network individually and their network addresses reassigned using the following procedure:
2.2
1
Temporarily disconnect any sensor having an existing network address of 16
by unplugging the connector.
2
Connect the new sensor to the network by plugging in its 10-way connector. Press
the ‘Search’ button to locate this new sensor at address 16.
3
Select an unused network address
from the address drop-down list.
4
Rename the sensor if appropriate
using the sensor name box.
5
Press the ‘Update’ button. The new
sensor will be renamed internally and
will reappear at the address specified.
User defined sensor name
6
Attach a label to the sensor to indicate its new network address to prevent later
confusion.
7
Repeat steps 2 through 6 for any further sensors being installed at this time.
Troubleshooting the network
If a new sensor does not appear in the Active sensors list.
2.3
•
Check power and all connections to the sensor.
•
If you have other sensors connected, temporarily disconnect these – the ‘new’ sensor
may have already been configured to a conflicting network address and will need to
be changed.
•
Try a slower search (selected on the Search Speed menu) – this allows the sensor
more time to respond to Hydro-Com (for example if it is on a long cable run)
Sensor Measurements
The two drop-down lists can be set to display any of the variables generated within a given
sensor. The items are identical in each menu. Note that not all the variables in the list are
available in each sensor model. (See ‘Hardware’ in Appendix D for more information.) If a