Burner System Maintenance 1. Extinguish the flame, allow the burner head to cool, shut down the gases at source, and bleed the gas supply lines (procedure: page 3-7). 2. Switch off the spectrometer. 3. Flush the drain system (procedure: page 3-32). 4. Unscrew the retainer cap from the drain vessel. 5. Carefully lift out the drain trap assembly. Do not disconnect the electrical lead. 6. Place the drain trap assembly into a suitably large bucket or container so that it is vertical. If the drain tap assembly is laid on its side, liquid can run out onto the floor. 7. Clean the drain trap assembly carefully with mild laboratory detergent solution using a test tube brush. Make sure that the float moves freely. 8. Rinse the assembly well with water. 9. Empty the drain vessel (procedure: page 3-33). Note: Dispose of hazardous or corrosive solutions properly and refer to your local safety regulations for proper disposal procedures. 10. Place the drain trap assembly back into the drain vessel and screw on the retainer cap. 0993-5257 3-35