10. Instructions how to check the sensor
Cable connection
S2 (reference Pt100)
(1) red
(3) blue
S1 (heater Pt100)
(6) brown
(8) white
(2) pink (5) yellow
(4) grey (7) green
S1 (heater)
S2 (reference)
Cable test
(1) red
(3) blue
<1
at 25°C ca. 110
(2) pink
(4) grey
<1
sheeld
(5) yellow
(7) green
<1
at 25°C ca. 110
(6) brown
(8) white
<1
Sensor check
-switch off power supply
-disconnect sensor
-check cables following above diagram using a Ω-meter
Important
no contact from shield to any line
INSTALLATION AND OPERATING INSTRUCTIONS Type SF-586b
24/01/2011 Subject to technical changes