5.Front Panel Operation
(1) Press Utility and choose Disp Setup, press
(2) Press
F2
F2
to select Sep.
to switch between Comma, Space, Off.
Take the Frequency parameter for instance:
Comma
Space
Off
To Set the Screen Saver
The screen saver will run automatically if no operation is taken for any key within the set
time. Press any button to resume.
(1) Press Utility and choose Disp Setup, press
(2) Press
F3
F3
to select Scrn Svr.
to switch between On/Off.
(3) If On is selected, you can set the screen saver time. Turn the knob to change the value;
or press the number keys to input the desired time in minutes, press F4 to select
the unit. The screen saver time range is 1~999 minutes.
To Set Output Parameter
To Set the Output Load
For the Main output terminal on the Front panel, the Generator has a built-in 50Ωseries
impendence. If the actual load does not match the set one, the displayed amplitude and
offset are incorrect. This function is used to match the displayed voltage with the expected
one.
Steps for setting the Load:
(1) Press Utility and choose Output Setup, press
( “*” represents a value).
F1
to select HighZ or *Ω
(2) To change the load value, after selecting *Ω, turn the knob to change it directly; or
press the number keys to input the desired value. Press F3 or F4 to select the
unit. The load range is 1Ω~10KΩ.
Note:
AG4151 has a fixed 50Ω Series Impendence. No matter what Value the set parameter is, if
the real load is different from the set one, the displayed voltage will not equal the real
voltage.