Changes for Software Version 3
In addition to one- and two-point calibration, the meter can perform a threepoint calibration.
Three-point calibration must be activated in a new configuration step
(3P-CAL On/OFF, see page 20). During calibration, you can stop the procedure
after each completed buffer evaluation by pressing the meas key.
Depending on the number of buffer evaluations made, the meter performs a
one-, two-, or three-point calibration.
With three-point calibration the zero and slope are calculated using a mean
straight line (to DIN 19268). Therefore, there is still only one zero point and
one slope for Sensoface® evaluation.
The calibration record has been supplemented by the third buffer.