Fume extraction
Depending on the bath liquid used and the operating method there is a possibility that
toxic vapours may be produced. In that case it is necessary to provide an appropriate
fume extraction. Pull out the mains plug before cleaning the bath with solvents. Provide
appropriate fume extraction. Before starting up the unit it is absolutely essential to
ensure that the bath contains no explosive mixtures. If necessary purge it with nitrogen!
4.4.5
Cooling water, steam production
When changing the bath liquid from water to heat transfer fluids for temperatures above
100°C any remaining water - including the one in the hoses and external system - has to
be removed completely. Otherwise there is a danger of burns because of delayed
boiling.