Download Betriebsanleitung Operating instructions Mode d'emploi Manual de
Transcript
8. Communication Software GB Shifting the zero point and the span We recommend you not to change the span. The in-factory adjustment should be changed by you only if you have available sufficient calibration equipment (at least 3-times better precision than the accuracy given). For instruments with absolute pressure measuring ranges or +/- measuring ranges, a sufficient calibration equipment as well as a vacuum pump are also necessary for the zero point adjustment. Shifting the zero point and setting the range can be done under the „Adjustment“ tab. There you can select the corresponding button „Zero point“ or „Span“. The zero point and the span can be shifted in steps using the arrow keys and then confirmed by pressing the „Apply“ key. Note: Adjusting the span using an unsuitable calibrating standard will no longer guarantee the accuracy class. The zero point and the span can be reset to the original factory settings by using the „Reset factory settings“ button. BD_BE_BF_BSP 721 a (Recommended recalibration cycle: yearly) Betriebsanleitung/Operating instructions/Mode d'emploi/Instrucciones de servicio 37