6.2.3 Taring
The weight of the hook and, if used, that of the sample holder must be tared before the first measurement and after each change of measurement mode.
Experiment
new
Main menu

New weight
exp.
Tare
Suspend hook freely and hook on sample holder.
Stdv
New weight exp. OK
Free500

Stdv 0.1 mg
Stdv 0.2 mg
Stdv 0.3 mg
Meas. parameter
Tare
Calibrate
Standard deviation of taring.
This value defines the required
taring stability.
The deviation must not be greater
than the set value during the
measurement.
MENU

HELP

The time period for the measurement extends from 2 s to a maximum of 10 s if the required deviation is not achieved.
Generally, the highest accuracy of 0.1 mg should be the target value. It is only when this
value cannot be obtained (e.g. in the case of external vibrations) that larger values should
be chosen for the standard deviation.
However, this reduces the accuracy and reproducibility of the following measurements.
New weight exp. OK
Free500

Taring in progress
The weight measurement for taring
is running.
Taring in progress

62
Do not disturb nor interrupt this process.
Avoid vibrations and mechanical shocks.
Measurement of density and weight
YAME0022 / 04.11.11