Automatic Tensiometer TD 3 6.2.3 Taring The weight of the hook and, if used, that of the sample holder must be tared before the first measurement and after each change of measurement mode. Experiment new Main menu New weight exp. Tare Suspend hook freely and hook on sample holder. Stdv New weight exp. OK Free500 Stdv 0.1 mg Stdv 0.2 mg Stdv 0.3 mg Meas. parameter Tare Calibrate Standard deviation of taring. This value defines the required taring stability. The deviation must not be greater than the set value during the measurement. MENU HELP The time period for the measurement extends from 2 s to a maximum of 10 s if the required deviation is not achieved. Generally, the highest accuracy of 0.1 mg should be the target value. It is only when this value cannot be obtained (e.g. in the case of external vibrations) that larger values should be chosen for the standard deviation. However, this reduces the accuracy and reproducibility of the following measurements. New weight exp. OK Free500 Taring in progress The weight measurement for taring is running. Taring in progress 62 Do not disturb nor interrupt this process. Avoid vibrations and mechanical shocks. Measurement of density and weight YAME0022 / 04.11.11