Download Teledyne 2103 Network Router User Manual
Transcript
APPENDIX A Firmware Commands Teledyne API Ultrafine Particle Monitor - Model 651 RIE – Read Instrument Errors RIE reads the instrument errors (displayed as a 16-bit integer in hexadecimal format). The number may be a combination of the values of more than one flag. RIE Command Response RIE Bit 0 x XXXX Example Command Response RIE C00 Bit = Hexadecimal character A-F. When the bit is set, the parameter is in error. XXXX = 4-digit number 0x0001 = Conditioner Temperature 0x0002 = Growth Tube Temperature 0x0004 = Optics Temperature 0x0008 = Vacuum Level 0x0020 = Laser Status 0x0040 = Water Level 0x0080 = Concentration Over-range 0x0100 = Pulse Height Fault 0x0200 = Absolute Pressure 0x0400 = Nozzle Pressure 0x0800 = Water Separator Temperature 0x1000 = Warmup 0x2000 = Reserved 0x4000 = Service Reminder 0x8000 = Reserved Read Instrument Errors Water Separator Temperature and Nozzle Pressure faults (Nozzle Pressure = hexadecimal 4. Water Separator = hexadecimal 8. Added together they make hexadecimal C.) Note: Hexadecimal is a numerical system using a base of 16. The symbols 0-9 represent the values zero to nine, and the letters A-F represent the values ten to sixteen. It is a useful “shorthand” for computer engineering because each hexadecimal digit represents four binary digits. RIF – Read Aerosol Flow Rate RIF reads the inlet flow rate in liters per minute (L/min). A-6 RIF Command Response RIF X Example Command Response RIF 0.3 X = Floating point number either 0.12, 0.6, or 3.0 Read Inlet Flow Rate 3.0 L/min 07506C DCN6727