Download Agilent Technologies U1251B Technical data
Transcript
Maintenance 5 3 To test Fuse 1, place the tip of the test probe on the right half of input terminal . Ensure that the probe tip touches the metal inside the input terminal, as shown in the figure above. 4 To test Fuse 2, place and touch the tip of the test probe on the right half of input terminal . Ensure that the probe tip touches the metal inside the input terminal. 5 Observe the reading on the instrument's display. Refer to Table 5- 3 below for the possible readings that could appear). 6 Replace the fuse when OL is displayed. Table 5-3 Measurement readings for fuse checking Fuse OK (approximately) Current input terminal Fuse Replace fuse Fuse rating Displayed readings 1 440 mA/1000 V 6.2 MΩ OL 2 11 A/1000 V 0.06 Ω OL Agilent U1251B/U1252B User’s and Service Guide 113