Transcript
• If option <LASt> is choosen in the scale program displays estimated unit mass of the last piece (about 3 seconds) and then goes to Counting pieces automatically setting the previously displayed value as valid for the procedure. • If the <FrEE> option is selected you will see: • Using where: and enter the required sample quantity, - selection of digit position, • Confirm the value by pressing - setting the digit, , • You will see <LoAd> on the display and then: • If weighing is performed in a container put the container on the pan first and then tare it. Then put the declared quantity of pieces on the pan and confirm it when stable (signalled by ): • The program will automatically calculate the mass of a single piece and go on to the Piece Counting mode (pcs). You will see the following display: - 46 -