Download SciCan Statim 2000 Setup guide
Transcript
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx Cycle Fault Number Description of Fault Probable Cause of Fault Cycle Fault #1 The cassette temperature failed to reach 95ºC within a time-out period. Cycle Fault #2 Software Revisions R1xx or R2xx only Cycle Fault #3 The cassette temperature failed to increase from 95°C to 100°C within a time-out period. Cycle Fault #4 The cassette has failed to achieve sterilization conditions within a timeout period. Cycle Fault #5 Software Revision R1xx only The software causes the pump to activate at predetermined times. If a request to pump water occurs outside of the predetermined time Cycle Fault #5 occurs. This fault may be caused by a large cassette leak in conjunction with an extremely large load or a blown thermal fuse caused by weak pump delivery and/or contaminated steam generator or a faulty steam generator triac. If the triac has failed, the thermal fuse may have also failed. A faulty steam generator (steam generator resistance failed open) could also cause this fault. This may be cause by a major steam leak from the cassette, incorrect chamber temperature calibration or an extremely large load. This may be caused by a faulty cassette seal, a damaged cassette or a faulty solenoid valve (failed to close). Verify check valve and pressure relief valves for leaks. For PCB rev other than rev. 7.xx verify Vref to be 2.520V± 0.001V. This may be caused by a faulty cassette seal, a damaged cassette, improperly installed copper tubing, a failed or faulty solenoid valve (failed to close) , or improper alignment of the probe bracket. Verify check valve and pressure relief valves for leaks. For PCB rev other than rev. 7.xx verify Vref to be 2.520V± 0.001V This may be caused by a leaky cassette, a kink in the steam generator inlet tube, a leaking solenoid outlet valve, a weak pump and a very large load in the cassette. The cassette has failed to pressurize and achieve a temperature of 110ºC within a time-out period. 32