Download Bacharach Fyrite Tech 50 Specifications
Transcript
Fyrite Pro Appendix A – Formulas and Tables CO2 = CO2 max (20.9 – O2%) 20.9 qA = (TA – TL) x ( λ (Excess Air) = A2 +B) 20.9 – O2 20.9 20.9 – O2 Eta = (100% – qA) + [Kf, if TA < (T0 – 1.2 x O2)] Efficiency = Eta – Kf CO Air Free (Undilute) = CO x λ CO/CO2 = COppm x 10 –6 CO2% x 10 –2 Where: O2 = Measured Oxygen in percent TA = Measured stack temperature in °C TL = Measured combustion air temperature in °C CO = Measured carbon monoxide level in ppm CO2 = Calculated carbon dioxide level in percent qA = Calculated stack loss in percent λ = Calculated LAMBDA A2, B, CO2 max, T0, and Kf are constants (see table below): Fuel Code Fuel A2 B CO2 max T0 Kf F1 Natural Gas 0.66 0.009 11.8 58 °C 9.6% F2 Propane 0.63 0.008 13.8 54 °C 7.6% F3 Light Oil 0.68 0.007 15.4 51 °C 5.3% F4 City Gas 0.63 0.011 13.1 0 °C 0% Instruction 24-9388 51