NIV – Non-invasive ventilation
Mask ventilation
NIV can only be activated as a supplementary function in the
pressure-controlled ventilation modes BIPAP (PCV+),
BIPAP/ASB (PCV+/PS), CPAP, CPAP/ASB (CPAP/PS).
Mask leakages are detected by the device, compensated and
included in the displayed flow curve and measured values for
V T and MV.
The measured values for V T and MV will be incorrect
during ventilation with leakages without active NIV
function.
Use of NIV
The dead space increases when using masks.
● Note the mask manufacturer's directions!
● Application mode »NIV« must not be activated with
intubated patients!
● Check alarm limits after deactivating »NIV« mode!
● Avoid high airway pressure – risk of aspiration!
To switch on NIV
● Press »Set. ∆∆« key until screen page
2/2
appears.
● Activate line »NIV off«
● Select »on« and confirm
The supplement NIV appears in the top line of the screen.
12837171
Oxylog 3000 automatically adjusts to the requirements of
mask ventilation. Leakage flows are compensated automatically and the leakage alarm is inactive.
● The minimum ventilation required for the patient must be
assured by setting the lower alarm limit »MV
«.
12937171
Apnoea ventilation is not permitted by the ventilator when NIV
is active.