Children CarotidAngio/ CarotidAngio042s (optional) Indications: CT angiography of the carotid arteries, e. g. carotid stenosis or occlusion, vascular abnormalities of the carotids or vertebral arteries, etc. A typical range of 17 cm covered in 10 sec. CarotidAngio kV 80 Effective mAs 100 Rotation time 0.5/0.42 sec.** Slice collimation 0.75 mm Slice width 5.0 mm Feed/Rotation 9.0 mm Increment 5.0 mm Kernel B30f CTDIVol 2.6 mGy Effective dose (mSv) Male: 0.6* Female: 0.7* 2ndreconstr. 21.0 mm 0.7 mm B20f Male: 0.6* Female: 0.7* * The conversion factor for an 8-week-old, and a scan range of 80 mm was used. ** optional For the 2nd reconstruction the Autoload into MIPthin Range on the 3D Card is activated. The images will be automatically loaded into 3D, MIPthin, and a coronal MIPthin Range will pop up. Please notice, if you are not satisfied with the Range preset, adapt the parameters to your needs and link them to the series. Contrast medium IV injection Start delay exam dependent Flow rate dependent upon needle size/Access site Total amount 1 – 2 ml per kg of body weight 228