STARTUP AND OPERATION 3 The self-test can be repeated at any time by activating the TEST softkey in the Self Test Results menu. 8. Set the desired oven setpoint in the range of 93°C to 204°C (200°F to 400°F). 9. Wait for the Purge Air green LED to illuminate. 10. Introduce the remaining supply gases. Perform leak check. (See Specifications page(s) in the Preface section of this manual) 11. Set and verify the internal gas pressures. INTERNAL PRESSURE REGULATOR BURNER AIR TYPICAL OPERATING PRESSURES 965 to 1103 hPa-gauge (14 to 16 psig) FUEL 1516 to 1723 hPa-gauge (22 to 25 psig) SAMPLE (NON-ADJUSTABLE) 206 to 290 hPa-gauge (3.0 to 4.0 psig) Purge air of the following specifications must be present: FLOW: 16 to 18 L/min. SUPPLY PRESSURE: 689 to 1378 hPa-gauge (10 to 20 psig) Noncompliance could cause damage to the module. At the very least, the module's safety system, which requires a certain volume of purge air flowing through the case before allowing burner ignition, will not allow the instrument to operate. The lowest purge air flow/pressure setting possible during burner operation is preferable. Thus, the user should set the external purge air pressure initially at 689 hPa-gauge (10 psig). Check the Miscellaneous Control Parameters screen under Technical Diagnostics, and note whether the Purge Gas (switch) variable is "ON." If it is "OFF," increase purge air supply by 69 hPa-gauge (1 psig), and recheck the Purge Gas variable until it reads "ON." DO NOT EXCEED 1378 hPa-GAUGE (20 PSIG). If the maximum setting is reached, and the Purge Gas variable does not read "ON," contact factory. If the safety system is initiated successfully (Purge Gas variable is "ON"), continue with the remainder of the startup procedure. NOTE Do not restrict the PURGE OUT port and the pressure relief valve. They must be vented to atmospheric pressure. 748414-A August 1999 Heated Flame Ionization Detector Analyzer Module Rosemount Analytical NGA 2000 19