Download Cane Crono User guide
Transcript
FrEE SECTION 11 WITHDRAWING THE Pusher 1. Withdrawal of the pusher before the end of the infusion This function allows the interruption of an active infusion, withdrawing the pusher to the start position of the infusion. To stop an active infusion, do the following: • switch the pump off by pressing the button; • press the and buttons together: the display shows End for 10 seconds and then begins to withdraw the pusher; • during the 10 seconds that the display shows End you can button. cancel the withdrawal request by pressing the 2. Withdrawal of the pusher at the end of the infusion Ten minutes before the end of the infusion, the device gives an intermittent beep for 2 seconds. This signal is repeated twice at 5 minutes from the end of the infusion. At the end of the infusion an audible signal is given and the display shows End. After a few seconds, the pusher starts withdrawing until it reaches the start position of the infusion. NOTES • While the display shows End, the withdrawal function can be cancelled by pressing the button. The display shows StoP. • The withdrawal function can be interrupted by pressing button. The display then alternates between End the and OFF. At this point the only active button is the button. When you press again the pump recommences the withdrawal of the pusher. 51 MAN 01/EN/03 CRONO PAR 50 06/13