Download Avanti D-24 Specifications
Transcript
Troubleshooting Accessing the Rotor in Case of Power Failure Accessing the Rotor in Case of Power Failure WARNING Any maintenance procedure requiring removal of a panel exposes the operator to the possibility of electrical shock and/or mechanical injury. therefore, turn the power off (O) and disconnect the instrument from the main power source, and refer such maintenance to qualified service personnel. WARNING Before performing this procedure, verify that the rotor is not spinning by listening carefully for any noise coming from the chamber. Proceed only if he instrument is quiet. Never attempt to override the door interlock system while the rotor is spinning. If the facility power fails only momentarily, the centrifuge will resume operation when the power is restored and the rotor will return to the set speed. In the event of an extended power failure, it may be necessary to override the door-locking mechanism manually to remove the rotor and retrieve your sample. To access the door-locking mechanism, you must remove the centrifuge front panel. Two latches secure the front panel in place; these latches are accessible through two holes at the upper right and left of the panel (see Figure 4.1). Figure 4.1 Emergency Door Release Latch Access J326XP-IM-5AB 4-5 4