Download Agilent Technologies U1252A Technical data
Transcript
Performance Tests and Calibration 7 Now you can use 1234 as the security code. If you want to enter a new security code, see “To change the instrument calibration security code from the front panel” on page 129. Make sure you record the new security code. Calibration process To complete a full instrument calibration use following general procedures: 1 Read “Test Considerations” on page 118. 2 Perform the verification tests to characterize the instrument (incoming data). 3 Unsecure the instrument for calibration (see “Calibration Security” on page 119). 4 Perform the adjustment procedures (see “Adjustments Consideration” on page 133). 5 Secure the instrument against calibration. 6 Note the new security code and calibration count in the instrument's maintenance records. NOTE Make sure to quit the Adjustment Mode before turning off the instrument. Agilent U1251A/U1252A User’s and Service Guide 131