CAUTION Glass door and inner chamber become hot during sterilization. Danger of burning. ∅ Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting sterilization. Aborting sterilization after less than 6 hours When aborting the sterilization prematurely, it may be that the cells/pathogens inside the unit have not all been killed. You should repeat the sterilization. CAUTION Interruption of temperature reaction time. Ineffective sterilization. Repeat the sterilization. CAUTION Glass door and inner chamber become hot during sterilization. Danger of burning. ∅ Do NOT touch the glass door and inner surfaces for approx. 7 hours after aborting sterilization. • When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in the CO2 sensor. The CO2 sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F. • Turn on the unit (chap. 5). Three events which can abort the automatic sterilization cycle prematurely: • A manual abortion (chap. 15.5.1) • Opening the outer door (chap. 15.5.2) • Turning off the power to the C 150 or a brief power failure (chap. 15.5.3) 15.5.1 Aborting the hot-air sterilization manually You can abort the sterilization cycle manually. • Press the controller button “180 °C” for three seconds during the sterilization cycle. The lower display will show the operating function “StE”, and the value “ON” will appear in the upper display. • Press “▲”. The upper display toggles from “ON” to “OFF“. • Press “MODE” to end the sterilization cycle. • Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C / 98.6°F. C 150 (E2) 12/2012 page 69/90