Download Agilent Technologies 8753ES Technical data
Transcript
Performance Tests 12. Receiver Magnitude Dynamic Accuracy 11.Take a power meter calibration sweep. • Cal • ONE SWEEP PWRMTR CAL −30 x1 TAKE CAL SWEEP 12.Verify that the power meter reads approximately −30 dBm. Full 2-Port Calibration 13.Connect the equipment as shown in the Figure 2-23. Be sure to reconnect the cable to the B channel connectors. Figure 2-23 Full 2-Port Calibration 14.Perform a full 2-port error correction with isolation. 15.Save the results of the new cal set. Measure Channel B Magnitude Dynamic Accuracy 16.Connect the equipment as shown in Figure 2-24. Confirm that the step attenuator is set to 20 dB. Chapter 2 2-53
Related documents
8753D Option 011 Network Analyzer Service Guide
Agilent Technologies 8752C User's Manual
Agilent Technologies 8753ES User`s guide
8720ES Service Guide
HP 8719D/20D/22D Network Analyzer Service Guide
What is a Solar 8000 Patient Monitor?
User Guide - Performance Audio
SERVICE MANUAL
ENH900EXT User Manual - EnGenius Technologies, Inc.
ENS1200 User Manual - EnGenius Technologies, Inc.