ECH2O-TE/EC-TM Probe Operator’s Manual 4. Probe VWC Calibration 4. Probe VWC Calibration The ECH2O-TE and EC-TM have been calibrated in media types including mineral, organic, and potting soil and rock wool. The goal of these calibrations was to create a generic calibration equation that will work in all types of each media, with an accuracy of better than ± 3% volumetric water content (VWC). If you need more accuracy, you can perform a media-specific calibration to get the accuracy down to ±1-2%. For more information on how to perform your own soil-specific calibration, or to have Decagon’s calibration service perform one for you, visit http://www.decagon.com/echo/ calibration.html. Mineral Calibration The probes were tested in four types of soil (sand, sandy loam, silt loam, and clay) at solution electrical conductivities of 0.1 to > 5 dS/m and water contents from 0 to 35% volumetric water content (VWC). The calibration equation generated from these tests is: θ = 1.087 * 10-3* Raw - 0.629 where θ is vol. water content and Raw is the output of the probe sensors (see appendix). It is worth noting here that this equation is linear. In fact, the linear fit provided as good a fit in the mineral soil range as did a higher order polynomial, with the added benefit that it is relatively easy to work with in datalogger programs, etc. However, one drawback with this 14