Download Quick Operating Tips for Versaflo L000
Transcript
QuickOperatingTips for VersafloL000 ALARMS To acknowledge and silencean alarm,pressthe l=FI',t=| button. I CLEAR I WARNING!Priorto restartingthe pump, doublecheckthat the conditionthat triggeredthe alarm has been correctedto prevent harm to your patient. pressthe ml startReminder Alarm- if an infusion is not startedwithin3 minutes, button I STOP I to beginyour infusion Door OpenAlarm- checkdoor to ensureit's closedproperly;the pump will not startthe infusionuntil fixed Air in line Alarm* - an air bubbleis detectedin the lVtubing,removethe air bubbleand restart y o u ri n f u s i o n . Ccclasiciii:t line*lerry1- pressurebuild up hasoccurredin thE"Milrte;releesethe pi'cssu:"e from the lV lineand restartthe infusion. EmptyAlarm* - this alarmcan be triggeredby two issues,if you are usingthe drop sensor,the sensoris not detectingthe dropsasthey passthroughthe sensor.Doublecheckthat your sensoris placedat the top of your chamber,the fluid in the chamberdoesnot go abovethe sensor;your lV bagis not at an angle;the sensoris pluggedinto the backof the pump;andthat the sensorwas not removedfrom the drip chamber.Alsodoublecheckyour fluidshavenot run out. CompleteAlarm* - the correctamountof fluidshavebeendeliveredto your patient. Change your lV bagand enterthe new amountto be deliveredto your patient. Low BatteryAlarm- the batterypower is almostdepleted,plugpump into an electricaloutlet t o c h a r g et h e p u m p . your DriveMechanismError- a mechanismerror hasoccurredwithinthe pump. Disconnect patientfrom the pump,turn the pump off. lf the batteryis almostdepletedon the pump,plug the pump into an electricaloutlet and retry. NOTE:Pumpwill chaneeto KVOmod QuickOperatingTips for Versaflo1000 WARNING!ThislV pumpshouldbe usedunderdirectsupervision of trainedpersonnel. WARNING!Pleaseread the user'smanual prior to operationof this pump NOTE:To assureaccuracy, we recommend that you usethe lV linesthat werepackaged with the pump. AnyotherlV lineusedwill requireadjustments andyou shouldreferto the usermanualfor propercalibration of the lV lines. 1 . P r e s st h e O N / O F F s w i t c ho n t h e b a c ko f t h e p u m pa n dt u r n t h e p u m po n . 2 . C l o s er o l l e rc l a m po n s e t ,s p i k ea n d h a n gf l u i dc o n t a i n e rF. i l ld r i p c h a m b e tro f i l l l i n e . 3 . Openrollerclampto fill all the set with fluid.Closerollerclamp. 4 . Openthe door of the pump by pressingdown on the leveron the front. I n s t a lllV l i n ei n s i d et h e p u m p ,m a k i n gs u r et h a t i t i s p r o p e r l yi n s t a l l e d throughthe guidesinsidethe pump includingthe Infrared Air Detector, Up tube guide,fleristaltic tube guid{ and tiie oicfusionsens6i,and that y o u r l i n eh a sn o s l a c ki n s i d et h e p u m p . C l o s et h e d o o r . O p e nr o l l e r clamp. - ftr{Ttral 5 . Press l:ffi1 | ^ i-l pressor hold onceto enterthe infusion ratesettingscreen, l v l| or | - | to setdesired infusionrate.press l#lonce confirmthe desiredrateandjumpto volumesetting. 6. Press or troldr ll o..Fl to to setdesired volume to beinfused. Press|llfE8lto confirm lsErl the settingandreturnto the mainscreen. 7. Attachthe infusion setto vein-puncture device. 8. HoldlM9Elto primethe infusion set. I BOLUS I 9. Connect thevein-puncture device to thepatient. nress lffilto pressm-l 10.Tostopinfusion, I sroP I be8ininfusion. 11.To restartor clearthe total infusedvolume,pressON/OFFswitchon backof pumpto turn the pump off, wait minimum3 seconds,and turn it on again.