PROGRAMMING INSTRUCTIONS
If you select a protocol « Paging » or « DECT » you can at the end determine how many
characters per information (criterion, floor, room, bed, spaces) have to be transmitted to
the paging/DECT system -> filter possibility.
The display 3 indicates (filtering of the criterion) :
CRITER.←
CRIT 07 + 0 SPACES
You can choose 0-11 characters, followed by 0-9 spaces.
☛ With buttons 4 or 5 you choose the number of characters, with buttons 6
or 7 you switch between « criterion » and « spaces » .
☛ Press button 1 to get to the filtering of floor/room/bed (or single number).
The display 3 indicates:
INFO. : fff rrr 00b
FL: 3+3 +RO: 3+1 +BE:3
☛ With buttons 4 or 5 you choose the number of characters, with buttons 6
or 7 you switch between « floor » , « room » , « bed » (or « single number » )
and « spaces » .
☛ Press button 1 to get to the filtering of the actual position (only if
« locator » function is activated). The display 3 indicates :
BE POS xyz
SPACES : 1
☛ With buttons 4 or 5 you choose the number of characters.
☛ Press Confirm button 2 to return to Ready State.
Example :
CR_←
INFO : frrb
BE__POS xyz
This means 2 characters for the criterion (CR), 1 space, 1 character for floor (f),
0 spaces, 2 characters for room (rr), 0 spaces, 1 characters for bed (b),
2 spaces, position indication.
In case of a call for help from floor 008, room 023, bed 1 with the actual
position 248 the following sequence is sent out : « HE_8231__POS_248 ».
45